Correction: Achieving high resolution and optimizing sensitivity in spatial frequency encoding NMR spectroscopy: from theory to practice

Bertrand Plainchont , Daisy Pitoux , Ghanem Hamdoun , Jean-Michel Ouvrard , Denis Merlet , Jonathan Farjon and Nicolas Giraud *
Equipe de RMN en milieu orienté, ICMMO, UMR 8182 (CNRS-UPS), Université Paris-Saclay, 91405 Orsay cedex, France. E-mail: nicolas.giraud@u-psud.fr

Received 28th October 2016 , Accepted 28th October 2016

First published on 7th November 2016


Abstract

Correction for ‘Achieving high resolution and optimizing sensitivity in spatial frequency encoding NMR spectroscopy: from theory to practice’ by Bertrand Plainchont et al., Phys. Chem. Chem. Phys., 2016, 18, 22827–22839.


The authors would like to correct an error in the published results: the E-Snob pulse whose amplitude profile is presented in Fig. 8 in the main text, and Fig. S4 in ESI, was calibrated for 90° (instead of 270°).

The authors note that this calibration to 90° does not correspond to the optimal setting for an E-Snob shape. E-Snob pulses calibrated for 270° are actually as suitable as E-Burp2 and Gaussian pulses for performing a Spatial Frequency Encoding excitation step (see for instance eMagRes, 2016, 5, 1377–1382, DOI: 10.1002/9780470034590.emrstm1528). The authors apologize for this error.

The Royal Society of Chemistry apologises for these errors and any consequent inconvenience to authors and readers.