Dominic J.
Hare
*ab,
Fred
Fryer
c,
Bence
Paul
d,
David P.
Bishop
a and
Philip A.
Doble
*a
aElemental Bio-imaging Facility, University of Technology Sydney, PO Box 123, Broadway, New South Wales, Australia. E-mail: dominic.hare@uts.edu.au; philip.doble@uts.edu.au; Tel: +61 3 9035 9549 Tel: +61 2 9514 1792
bThe Florey Institute of Neuroscience and Mental Health, The University of Melbourne, Parkville, Victoria, Australia
cAgilent Technologies Australia, Mulgrave, Victoria, Australia
dSchool of Earth Sciences, The University of Melbourne, Parkville, Victoria 3052, Australia
First published on 3rd October 2016
Dried micro-droplets were used to characterise the formation of polyatomic interferences in a commercial laser ablation-inductively coupled plasma-mass spectrometer (LA-ICP-MS). Droplets containing 4 ng each of Al, As, Ca, Cd, Co, Cu, Fe, Mg, Mn, P, Se and Zn in the presence of potentially interfering isotopes were deposited on silicate microscope slides. Comparisons of the total signal recorded for each dried droplet showed no detectable influence of polyatomic matrix-based interferences. Visualisation of acquired signal was achieved by the generation of two-dimensional maps. The natural abundance pattern of elements with two measurable isotopes was confirmed for each droplet. The lack of interferences was due to the absence of major gas molecules (e.g. N2, O2, CO2) from the laser cell and minimal total matrix load on the plasma when compared to standard solution nebulisation (approx. ng s−1versus 10–20 μL s−1) ICP-MS conditions.
It is often assumed that LA-ICP-MS suffers from significant isobaric interferences from polyatomic species3 in line with traditional solution nebulisation (SN) ICP-MS. While it has been suggested that interferences, particularly those arising from water in the sample matrix, are considerably lower in LA-ICP-MS,4,5 reduced interference formation in the context of bio-imaging has not been confirmed. ICP-MS instruments with quadrupole mass filters typically have mass resolution of 0.1 atomic mass units. Therefore, polyatomic species with the same nominal mass as the ion of interest cause significant interfering signal that precludes accurate measurement. For SN-ICP-MS either a higher resolution mass filter (such as a sector field design6) or collision/reaction gas7,8 is required to negate the effect of, or physically remove interferences. Sector field instruments command a considerably higher price than quadrupole instruments with minimal or even no increase in sensitivity and a more limited range of sequentially scanned isotopes.9 Reaction/collision cell equipped quadrupole ICP-MS exhibit a significant drop in signal abundance (often by a factor of 10 or more), which is unfavourable when introducing comparatively smaller amounts of material to the plasma via LA.
Ablated analytes transported to the ICP-MS reach the plasma in a very different state when compared to SN. Most significantly, solvent-free particles of a much smaller diameter10 are produced by commercially available UV lasers. The ablation process involves desorption, vapourisation and ionisation of the analyte and subsequent condensation and aggregation from the laser plasma to form laser aerosols.11,12 Therefore, it is reasonable to assume that applying operational parameters similar to solution nebulisation is unnecessary in LA-ICP-MS to achieve adequate determination.
MO+, MN+ and MH+ based interferences primarily arise from air (N2, O2etc.) and the solvent in use (mainly H2O based matrices, along with Cl+ and N+ species in certain acidic matrices). These highly abundant ions are generated within the plasma,13 and are somewhat controlled in solution nebulisation ICP-MS by manipulation of parameters that influence plasma temperature such as carrier gas flow rate, radio frequency (RF) power; or residence time such as sampling depth, and to a lesser extent sampling cone orifice dimesions.14 The degree to which these parameters can be controlled is limited, as adequate desolvation and ionisation is dependent on the duration the particle resides within the plasma and the applied power. The ‘dry’ nature of the plasma in LA-ICP-MS is advantageous, as the confounding effects of H2O and N on polyatomic interference from the typical acidic matrices used in solution nebulisation ICP-MS are significantly reduced. Reducing the potential for interference formation can sufficiently negate the need for post-plasma treatment, such as collision gases or high resolution mass filters.
Dried droplet ablation has been suggested as an alternative method to electrothermal vapourisation (ETV) ICP-MS for the discrete sampling of small volume aqueous samples.15 Günther et al.16 showed that direct ablation of dried solutions yielded no significant elemental fractionation. Dried droplet analysis was also shown to be suitable for the simultaneous determination of 21 isotopes with absolute detection limits in the sub ng g−1 range.5 Volumes as low as 20 μL have been used to quantify μg g−1 concentrations of selenomethionine in yeast via droplet analysis of HPLC fractions.17 It has also been shown that dried droplet analysis yielded a 100-fold decrease in hydride formation of radionucliotides when compared to both solution nebulisation and ETV-ICP-MS.18 Ablation of dried blood droplets on filter paper has been suggested as a suitable technique for the measurement of Pb toxicity.19 Recently, 5–20 μL droplets of acidified standard solutions placed on filter paper supports has been successfully used for multi-point external calibration in imaging experiments.20
The object of this study was to characterise interference formation in a commercial LA-ICP-MS instrument using dried sub-μL volume droplets. The experimental design was based on the theoretical formation of polyatomic isobaric interferences to determine the influence of matrix components on true analyte signal.
![]() | ||
Fig. 1 (a) Schematic representation of imaging dried droplets on a standard microscope slide by LA-ICP-MS. (b and c) Relative intensity maps of iron (m/z 56, 57) and copper (m/z = 63, 65) in deposited droplets with corresponding composite images, showing negligible interferences on any measured mass due to theoretical polyatomic species (see Table 2). The area left of the dashed white line indicates the gas blank signal, showing no discernible increase in signal arising from potential 40Ar16O+ or 23Na40Ar+ interferences on 56Fe or 63Cu, respectively. |
Five solutions were prepared (see Table 1). These consisted of a blank solution containing only amido black and solutions containing potential interfering elements (a list of theoretical polyatomic interferences are listed in Table 2), analyte elements and a mixture of all elements used (Fig. 1a). The amido black was added to locate the dried spots using the CCD camera on the laser unit. The amido black was also a source of carbon to evaluate potential 12C and 13C-based interferences. Nitrate from the 1% HNO3 matrix remaining on the slide surface after drying provided a potential source of 14N-based interferences. Approximate total amounts of N, O and C from the nitrate anion and 10 μg mL−1 amido black dye are 124 ng, 429 ng and 1.72 pg, respectively.
Analyte | Interfering species |
---|---|
27Al+ | 12C15N+, 13C14N+, 14N2+ spread |
75As+ | 40Ar35Cl+, 59Co16O+, 36Ar39K+, 43Ca16O2+, 23Na12C40Ar+ |
43Ca+ | 27Al16O+ |
44Ca+ | 12C16O2+, 14N216O+, 28Si16O+ |
111Cd+ | 95Mo16O+ |
112Cd+ | 40Ca216O2+, 40Ar216O2+ |
59Co+ | 43Ca16O+, 24Mg35Cl+ |
63Cu+ | 40Ar23Na+, 40Ca23Na+, 31P16O2+ |
65Cu+ | 40Ar25Mg+ |
56Fe+ | 40Ar16O+, 40Ca16O+ |
57Fe+ | 40Ar16OH+, 40Ca16OH+ |
24Mg+ | 12C2+ |
25Mg+ | 12C2H+ |
26Mg+ | 12C14N+, 12C2H2+, 12C13CH+ |
55Mn+ | 40Ar14NH+, 39K16O+ |
31P+ | 14N16OH+ |
78Se+ | 40Ar38Ar+ |
80Se+ | 40Ar2+ |
64Zn+ | 32S16O2+ |
66Zn+ | 34S16O2+, 40Ar14N2+ |
Laser parameters | |
Laser type | Nd:YAG |
Wavelength | 213 nm |
Spot diameter | 80 μm |
Scan speed | 80 μm s−1 |
Line spacing | 80 μm |
Laser power | ∼0.15 J cm−2 |
Frequency | 20 Hz |
![]() |
|
ICP | |
RF power | 1200 W |
Plasma gas flow rate | 15 L min−1 |
Carrier gas flow rate | 1.15 L min−1 |
Sampling depth | 4 mm |
Scanning mode | Peak hoping, time resolved analysis |
Sampling cone | Pt, Ni base |
Skimmer cone | Pt, Ni base |
A 5 × 30 mm area encompassing all deposited droplets was ablated to ensure all material was transported to the ICP. Analysis time was approximately four hours per sample set. Images for m/z 56 and 57 are shown in Fig. 1b. Correlation analysis of m/z 56 and 57 images showed a Pearson correlation coefficient of 0.979, indicating no additional signal above the background was recorded from one of the most prevalent polyatomic interference normally encountered, 40Ar16O+ on 56Fe+. A background signal was present for m/z 56 when the laser was not operational, produced from trace O2 impurities present in the argon gas. Furthermore, this background was stable throughout the entire experiment, indicating that the formation of matrix-based 40Ar16O+ was negligible.
Images for m/z 63 and 65, corresponding to the two natural isotopes of Cu are shown in Fig. 1c. 63Cu+ was potentially interfered by 40Ar23Na+, 40Ca23Na+ and 31P16O2+, and 65Cu+ by 40Ar25Mg+. Droplet ‘Int’ contained 10 mg kg−1 of both 31P and 25Mg, and droplet A1 contained 10 mg kg−1 40Ca. No signal was recorded for m/z 63 or 65 in these ablated droplets. This indicated that Ar- or O-based polyatomic interferences were below the instrument's limit of detection, and the Pearson's correlation coefficient was 0.954 for the resulting images.
Further investigation of the influence of polyatomic interferences was carried out by examining the abundance ratios of elements with two measurable isotopes (Table 4). Close correlation between each element's natural isotopic abundance and measured uncorrected abundance ratios were confirmed. While the sequential mass analyser in a quadrupole ICP-MS is more prone to intrinsic uncertainty due to the mode of detection when compared to multicollector designs as heavier isotopes are more efficiently transmitted and detected,28 our data was either within or close to the measured uncertainty expected for a quadrupole analyser. This comparison clearly demonstrated polyatomic interferences on the abundance ratio were negligible.
Isotopes | Natural abundance ratio | Uncorrected abundance ratio |
---|---|---|
24Mg+/26Mg+ | 7.17 | 6.89 ± 0.07 |
43Ca+/44Ca+ | 0.06 | 0.06 ± 0.01 |
56Fe+/57Fe+ | 43.3 | 41.5 ± 0.22 |
63Cu+/65Cu+ | 2.24 | 2.09 ± 0.01 |
64Zn+/66Zn+ | 1.74 | 1.67 ± 0.01 |
95Mo+/98Mo+ | 0.66 | 0.62 ± 0.01 |
111Cd+/114Cd+ | 0.45 | 0.43 ± 0.01 |
These data support the use of continued use of LA-ICP-MS for interference-free imaging applications in the biological sciences, showing that the most commonly expected matrix-based species arising from highly abundant elements such as Na, K, P and O are absent using typical LA-ICP-MS operating procedures, in contrast to that which was previously claimed but not experimentally demonstrated.29 We have previously discussed the utility of reaction gases, such as hydrogen (in an octopole-equipped ICP-MS)30 and oxygen (for mass-shifting using triple quadrupole-design ICP-MS systems),31 though in both cases gases were used for either improving signal to noise ratios for extremely small sample volumes or low-abundance elements such as Se, respectively.