Santiago
Medina-Rodríguez
a,
Sergey A.
Denisov
b,
Yanouk
Cudré
c,
Louise
Male
c,
Marta
Marín-Suárez
a,
Alberto
Fernández-Gutiérrez
a,
Jorge F.
Fernández-Sánchez
*a,
Arnaud
Tron
b,
Gediminas
Jonusauskas
d,
Nathan D.
McClenaghan
*b and
Etienne
Baranoff
*c
aDepartment of Analytical Chemistry, Faculty of Sciences, University of Granada, Avenida Fuentenueva s/n, 18071 Granada, Spain. E-mail: jffernan@ugr.es
bUniversité Bordeaux/CNRS, ISM, 351 cours de la Libération, 33405 Talence Cedex, France. E-mail: nathan.mcclenaghan@u-bordeaux.fr
cSchool of Chemistry, University of Birmingham, Edgbaston B15 2TT, UK. E-mail: e.baranoff@bham.ac.uk
dUniversité Bordeaux/CNRS, LOMA, 351 cours de la Libération, 33405 Talence Cedex, France
First published on 14th April 2016
A doubly pyrene-grafted bis-cyclometallated iridium complex with engineered electronically excited states demonstrates reversible electronic energy transfer between adjacent chromophores giving rise to extremely long-lived red luminescence in solution (τ = 480 μs). Time-resolved spectroscopic studies afforded determination of pertinent photophysical parameters including rates of energy transfer and energy distribution between constituent chromophores in the equilibrated excited molecule (ca. 98% on the organic chromophores). Incorporation into a nanostructured metal–oxide matrix (AP200/19) gave highly sensitive O2 sensing films, as the detection sensitivity was 200–300% higher than with the commonly used PtTFPP and approaches the sensitivity of the best O2-sensing dyes reported to date.
The oxygen-sensitive luminescent dye is a central component to the optical sensor. As the excited dye has to be quenched by the triplet ground state of molecular oxygen, phosphorescent complexes based on heavy metals such as Ir(III), Pt(II), Pd(II), Re(I) and Ru(II) complexes have been particularly studied, which correspond to kinetic and energetic prerequisites.6,7,10,21–31 The lifetime of the emissive excited state, τ0, and hence likelihood to interact with diffusing O2, is a key parameter influencing the performance of the sensor, which is evaluated by measuring the Stern–Volmer constant KSV = kq·τ0 with kq being the quenching rate constant. Consequently, the most commonly used highly sensitive dyes to date are Pd(II)/Pt(II) porphyrins such as Pt/PdTFPP (TFPP = meso-tetra(pentafluorophenyl)porphin) that have excited triplet state lifetime >50 μs, possibly reaching milliseconds for Pd porphyrins.21,22,32 Increasing the lifetime of the excited triplet state of dyes is therefore an appealing strategy to improve the performance of the oxygen sensor.33–37 To this end, the triplet state of organic and metalloid-containing organic molecules has been successfully harnessed for highly sensitive O2-sensing systems due to lifetime of excited states easily reaching values >milliseconds. For example, C70 (τ0 > 20 ms) reached KSV ≈ 230 kPa−1 using ethyl cellulose as the matrix.34 Aluminum complexes of 6-hydroxy-7H-benzo[de]anthracen-7-one (τ0 = 350 ms) reach KSV ≈ 330 kPa−1 when highly O2-permeable Teflon AF is used as the supporting matrix.36 However, in these cases, kq is low (kq(C70) ≈ 11.5 Pa−1 s−1 and kq(Al complex) ≈ 0.94 Pa−1 s−1) compared to iridium complexes. For example, N969 (KSV1(0–10% O2) = 4.79 kPa−1 and τ0 = 3.8 μs)32 and EB146 (KSV1(0–10% O2) = 1.70 kPa−1 and τ0 = 1.9 μs)32 have kq values around 1000 Pa−1 s−1. Such compounds are therefore very promising as highly sensitive O2-sensing dyes for ultra-low O2 concentrations upon greatly extending the lifetime of the excited state (i.e. >100 μs).
Much effort has been devoted to increasing the excited state lifetime of emissive transition metal complexes. The strategies used rely upon modification of the ligands to modify the energetics of the 3MLCT and 3MC states38–41 or to modify the geometry around the metal centre.42 Another strategy is to design systems containing additional weakly interacting chromophores resulting in reversible electronic energy transfer (REET) between the core complex and the appended organic chromophore(s).43–47 The latter strategy is very effective at significantly increasing the excited state lifetime without changing other parameters such as energetics (emission wavelength, redox potentials) and photoluminescence quantum yields and has been successfully applied to cyclometallated iridium complexes.47–49 Surprisingly, REET has not been widely adopted for oxygen sensing beyond gaining insights into the photophysical processes occurring in the systems50,51 or simply measuring the emission intensity ratio between air-equilibrated and degassed solutions. We have found only one report of a bichromophoric ruthenium-pyrene system exhibiting REET actually used for O2-sensing related application, a pressure sensitive paint.52 Yet this system has very low sensitivity to oxygen, 0.11 kPa−1 for the main component, which is unsuitable for measuring low concentrations of oxygen.
In this work we demonstrate with a series of dyes, Ir0, Ir1 and Ir2, that REET can be used to improve the sensitivity of low concentration oxygen sensing dyes far beyond commonly used platinum(II) porphyrins and approaches the best sensitivities reported to date. The dyes are based on a bis-cyclometallated iridium core complex of the type Ir(C^N)2(bpy)+ with C^N = 2-(naphthalen-1-yl)pyridine and bpy = 2,2′-bipyridine substituted with 0, 1 and 2 pyrene moieties (Scheme 1). It was anticipated that the two auxiliary pyrene groups would substantially increase the luminescence lifetime of Ir2 compared to that of complex Ir1 with a single appended pyrene (due to entropy-induced modification of the effective driving force for energy transfer, which favours location of excitation energy on organic chromophores in the former case exalting the energy reservoir effect), and hence greatly benefit the oxygen sensing performance, without compromising the emission profile and quantum yield of emission.44
Complex |
E
ox![]() |
E
red![]() |
λ em. max (nm) |
Φ
air![]() |
Φ
degas![]() |
τ (μs)e | K eq | τ eq (ns) |
---|---|---|---|---|---|---|---|---|
a E = (Ec + Ea)/2, vs. Fc+/0. b Only the cathodic peak is observed, Ec is given. c Luminescence quantum yield (Φlum) in air-equilibrated CH3CN solution cf. [Ru(bpy)3]2+ in H2O (bpy = 2,2′-bipyridine), Φlum = 0.028.54 . d Luminescence quantum yield in degassed CH3CN solution cf. [Ru(bpy)3]2+. e MLCT luminescence lifetime in dilute degassed CH3CN. | ||||||||
Ir0 | 0.75 | −1.85 | 590, 625 | 3.8 × 10−3 | 10 × 10−2 | 8.3 ± 0.3 | — | — |
Ir1 | 0.76 | −1.83 | 590, 625 | 5 × 10−4 | 9.5 × 10−2 | 225 ± 15 | 27.8 ± 2 | 5 |
0.92b | −1.76b | |||||||
Ir2 | 0.76 | −1.81 | 590, 625 | 1 × 10−4 | 9.6 × 10−2 | 480 ± 15 | 61 ± 3 | 1.8 |
0.89b | −1.73b |
![]() | ||
Fig. 2 Steady-state luminescence spectra of Ir0, Ir1 and Ir2 in degassed (solid) and air-equilibrated (dashed lines) acetonitrile at room temperature (λexc = 413 nm). |
A small energy gap value can be estimated at ∼700 cm−1 between high energy emission features (see ESI†), corresponding to the energy gap between lowest lying electronically excited states on adjacent chromophores assuming similar 3MLCT levels in all complexes (at room temperature similar absorption and emission spectra are observed for the emitting moiety). Note: More accurate measurements of the energy gap can be obtained through time-resolved spectroscopies, vide infra.
A small energy gap value is conducive with reversible interchromophore electronic energy transfer taking place at room temperature. This would result in an excited state equilibrium being reached, which would consequently result in prolonged luminescence lifetime and oxygen sensitivity.44
![]() | ||
Fig. 3 Luminescence decays in the 580–620 nm range of dilute Ir0 (10 μM, λexc = 465 nm), Ir1 (2 μM, λexc = 355 nm) and Ir2 (2 μM, λexc = 355 nm) in degassed acetonitrile. |
Complementary transient absorption spectroscopy studies allow elucidation of the management of energy by the excited molecule prior to emission, and transient absorption signatures for both pyrene and iridium complexes have been described.48
Excitation into the MLCT absorption band of Ir2 at 430 nm rapidly led to the population of the pyrene triplet, denoted by a characteristic Tn ← T1 absorption at 410 nm, see Fig. 4. The kinetics of deexcitation of this absorption band exactly parallels those of the emission, clearly showing that while energy is principally located on this triplet (see ESI†), it is quantitatively transferred to the metal centre where it is subsequently emitted. This observation implies the presence of quasi-isoenergetic excited states on the adjacent chromophores permitting rapid and reversible electronic energy transfer, leading to a dynamic excited-state equilibrium.
The establishment of a dynamic excited-state equilibrium from an initial non-equilibrated excited state can be observed in real time by transient absorption spectroscopy, and a 5 ns and 1.8 ns rise time of the pyrene triplet absorption signature was measured for Ir1 and Ir2, respectively. This value gives the rate of establishment of equilibrium (keq = 2 × 108 s−1 and keq = 5.6 × 108 s−1), and is equal to the sum of forward (kf) and back (kb) energy transfer processes.44 The increased rate of attaining equilibrium Ir2 compared to Ir1 can be ascribed to the presence of two degenerate pyrene triplet manifolds in the former case and one in the latter complex. Due to an overlap between Ir-centre ground state bleaching signal with the transient MLCT absorption signal, the relative kf and kb values cannot be determined by direct observation of transient absorption signatures.
The variations of the luminescence intensity with the oxygen concentration as well as the Stern–Volmer plots for Ir2-AP200/19 and Ir2-PS are shown in Fig. 5, which also demonstrate the reversibility of the sensing process. The fitting parameters for all five studied sensing films are reported in Table 2. Ir0 immobilised in AP200/19 provided very irreproducible sensing films and therefore was not studied further.
![]() | ||
Fig. 5 Time trace curves (variation of the luminescence intensity with the oxygen concentration) for: (a) Ir2-PS and (b) Ir2-AP200/19, and calibration curves in the range 0–10 kPa O2 and Stern–Volmer plots (I0/I vs. pO2) for (c) Ir2-PS and (d) Ir2-AP200/19 obtained by intensity measurements. Circles in (c) and (d) correspond to experimental data and lines are the mathematical fitting by using a two-site model; the parameters of the fitting have been obtained according with the ref. 57 using the criteria “a” (minimization of the square error in the analytical signal). |
Range | Parameters | Intensity measurements (I0/I) | ||||
---|---|---|---|---|---|---|
Ir0-PS | Ir1-PS | Ir1-AP200/19 | Ir2-PS | Ir2-AP200/19 | ||
0–10 kPa pO2 | K SV1 (kPa−1) | 0.87 ± 0.06b | 2.94 ± 0.09c | 11.03 ± 0.35b | 4.499 ± 0.796d | 26.96 ± 0.25b |
x 1 | 0.9 ± 0.0 | 1.0 ± 0.0 | 0.9 ± 0.1 | 0.9684 ± 0.006 | 0.99 ± 0.03 | |
K SV2 (kPa−1) | 0.00 ± 0.00 | — | 0.00 ± 0.00 | — | 0.00 ± 0.00 | |
x 2 | 0.1 ± 0.1 | — | 0.1 ± 0.1 | — | 0.01 ± 0.01 | |
R 2 | 0.999 ± 0.007 | 0.9922 ± 0.0250 | 0.9999 ± 0.0004 | 0.9904 ± 0.0011 | 0.9998 ± 0.0009 |
Apparent lifetime measurements (τϕ) | |||
---|---|---|---|
Ir2-PS | Ir2-AP200/19 | ||
a [Dye concentration] = 1.5 mg mL−1; the results are the average of 3 replicas ±st/√n (n = 3, t = 4.303 (2P = 0.05), s = standard deviation), the fitting parameters have been obtaining according with ref. 57 using criteria “a”. b Data fitted with the Demas two-site model. c Data fitted with the Stern–Volmer model. d Data fitted with the Lehrer two-site model. | |||
0–1 kPa pO2 | K SV1 (kPa−1) | 7.99 ± 0.72d | 94.81 ± 0.19b |
x 1 | 0.96 ± 0.01 | 0.76 ± 0.04 | |
K SV2(kPa−1) | — | 2.28 ± 0.10 | |
x 2 | — | 0.24 ± 0.03 | |
R 2 | 0.9986 ± 0.0009 | 0.9996 ± 0.0008 | |
0–10 kPa pO2 | K SV1 (kPa−1) | 8.79 ± 0.61d | 45.64 ± 0.16a |
x 1 | 0.941 ± 0.002 | 0.90 ± 0.03 | |
K SV2 (kPa−1) | — | 0.30 ± 0.08 | |
x 2 | — | 0.10 ± 0.02 | |
R 2 | 0.9972 ± 0.0012 | 0.9985 ± 0.0011 |
The increasing number of pyrene moieties on the complex clearly improves the oxygen sensitivity; the sensitivity of Ir2-PS is 1.5 and 5 times higher than the sensitivities of Ir1-PS and Ir0-PS, respectively, following the increase of excited state lifetime. The same effect is observed with the AP200/19 matrix, as the sensitivity of Ir2-AP200/19 is 2.4 times higher than with Ir1-AP200/19. The latter also highlights the effectiveness of the nanostructured support AP200/19 to increase the oxygen sensitivity, as previously observed for other oxygen sensitive dyes.23–25,32
Based on intensity measurements, the most sensitive sensing film is Ir2-AP200/19 (KSV of 26.96 kPa−1). Comparing these results with classical Pt(II), Ru(II), and Ir(III) sensing films using the same nanostructured matrix, it is clear that this sensing film is very promising for O2-sensing applications, showing a higher Stern–Volmer constant than one of the most sensitive complexes published in the literature so far (Ksv(PtTFPP-AP200/19) = 25.68 kPa−1).32
In order to determine if Ir2 can be used for analysing low oxygen concentration, lifetime measurements were carried out in the frequency domain (phase-resolved method)32,59,60 because intensity measurements are not sensitive enough to characterize the sensing films in the ranges 0–1 and 0–0.21 kPa pO2. The description of the methodology as well as the protocol for carrying out the measurements are described by Medina-Rodríguez et al.59,60 The modulation frequencies used for the measurement in these ranges (see ESI; Fig. S11†) were 1130 Hz for the Ir2-PS sensing film (average phase difference of 51.90 degrees) and 2260 Hz for the Ir2-AP200/19 sensing film (average phase difference of 32.61 degrees). The variation of the apparent luminescence lifetime of Ir2 incorporated into PS and into AP200/19 when exposed to different oxygen concentrations between 0–1 kPa pO2 are shown in Fig. 6 and the results are summarized in Table 2, demonstrating the high sensitivity of the film (KSV1 of Ir2-AP200/19 is 94.81 kPa−1 and Δτ0.05 of Ir2-AP200/19 = 68.51 ± 0.21%); the Δτ0.05 of PtTFPP-AP200/19 measured in the same conditions is 62.53 ± 3.66%.32
![]() | ||
Fig. 6 Calibration curves (τφ, expressed in microseconds) in the range 0–1 kPa O2 and Stern–Volmer plots (τ0/τ vs. pO2) for (a) Ir2-PS and (b) Ir2-AP200/19 at 21 °C. Circles corresponds with experimental data and lines are the mathematical fitting by using a two-site model; the parameters of the fitting have been obtained according with the ref. 57 using the criteria “a”. |
Finally, the measurement capability of Ir2-PS and Ir2-AP200/19 was also determined by comparing the measured oxygen concentrations with the real O2 concentrations (see ESI, Table S2†), which shows the reliability of the sensing films.
A key advantage of this strategy is that it provides a rational approach to further enhancing the performance of the dyes: simply increasing the number of energy reservoir units will increase the excited state lifetime linearly, as previously demonstrated with ruthenium dyes,61 which will result in an increase of sensitivity. Equally, reversible energy transfer could a priori be instilled in Pd/Pt porphyrins in an analogous fashion. Clearly ultrasensitive oxygen sensing dyes are now accessible.
Footnote |
† Electronic supplementary information (ESI) available: Synthetic details, crystallographic data, cyclic voltammograms, experimental details for spectroscopic and O2 sensing studies. CCDC 1452045 and 1452046. See DOI: 10.1039/c6an00497k |