Balamurugan Thirumalraj‡
a,
Selvakumar Palanisamy‡a,
Shen-Ming Chen*a,
Cheng-Yu Yanga,
Prakash Periakaruppan*b and
Bih-Show Lou*c
aElectroanalysis and Bioelectrochemistry Lab, Department of Chemical Engineering and Biotechnology, National Taipei University of Technology, No. 1, Section 3, Chung-Hsiao East Road, Taipei 106, Taiwan, Republic of China. E-mail: smchen78@ms15.hinet.net; Fax: +886-2-27025238; Tel: +886-2-27017147
bDepartment of Chemistry, Thiagarajar College, Madurai-625009, Tamilnadu, India
cChemistry Division, Center for Education, Chang Gung University, 259, Wen-Hwa 1st Road, Kwei-Shan, Tao-Yuan 333, Taiwan, Republic of China
First published on 7th September 2015
In the present work, a glucose biosensor was fabricated based on the direct electrochemistry of glucose oxidase at glassy carbon modified with a reduced graphene oxide (RGO) and fullerene-C60 (C60) composite. The reduced graphene oxide/fullerene (RGO–C60) composite was prepared by electrochemical reduction of a graphene oxide (GO) and C60 composite at −1.4 V for 200 s in pH 5 solution; while the GO–C60 composite was prepared by a simple sonication of C60 in GO solution for 6 hours at 45 °C. A well-defined and enhanced reversible redox peak of GOx was observed at RGO–C60 composite compared with other modified electrodes. The heterogeneous electron transfer rate constant (Ks) and the surface coverage concentration of GOx at RGO–C60/GOx modified electrode were calculated to be 2.92 s−1 and 1.19 × 10−10 mol cm−2, respectively. Under optimum conditions, the amperometry response of the biosensor was linear against the concentration of glucose from 0.1 to 12.5 mM with a response time of 3 s. The limit of detection was estimated to be 35 μM based on S/N = 3 with a high sensitivity of 55.97 μA mM−1 cm−2. In addition, the fabricated biosensor showed a good practical ability for the detection of glucose in human blood serum samples.
Reduced graphene oxide (RGO) is a carbon allotrope and the single atomic layer of carbon atoms that are arranged in a honeycomb lattice.2 Since the discovery of RGO in 2004, it has been regularly used for different potential applications including energy conversion and storage, electronic devices, electrocatalysis, sensors and biosensors.12,13 However, the DET of GOx at pristine RGO surface is difficult due to the van der Waals force of attraction between each RGO sheets. Hence, the different approaches or micro or nanomaterials have been used with RGO to further enhance the electron transfer properties of GOx thus leading to high sensitivity for the detection of glucose.7 Recently, the composites of RGO with carbon nanomaterials, metal nanoparticles and metal oxides have been used as an electrode material for immobilization of GOx, and they have shown an enhanced DET with better catalytic performance of GOx towards glucose compared to RGO.14–16
Fullerene C60 (C60) is one of the carbon allotropes and is made up of 60 closely packed carbon atoms into fused hexagons and pentagons without any edges.17 Furthermore, it has been used as a promising electron transfer mediator for biomolecules and oxidase based amperometric biosensors and electrochemiluminescence (ECL) biosensors.19–21 In addition, it is reported earlier that C60 can be easily dispersed into the graphene oxide (GO) aqueous solution via covalent and non-covalent interactions.20,21 In addition, GO has been widely used for the preparation of hybrid carbon materials including carbon nanotubes and C60 owing to the presence of abundant oxygen functional groups.21 In the present study, we have dispersed the C60 into the GO aqueous solution by simple sonication and the resulting GO–C60 composite has been electrochemically reduced to RGO–C60 composite by electrochemical method. Compared with other reduction methods, electrochemical method is more helpful to retain some of the oxygen functionalities on the carbon basal plane and the properties of RGO and C60. The reasons why we have chosen RGO–C60 composite for biosensor are: (i) the electrochemical behaviour of C60 is greatly improved in the presence of RGO, since pristine C60 has poor electrochemical behavior in aqueous solutions; (ii) the electron transfer rate of enzyme towards the electrode surface is much improved at RGO–C60 composite when compared with pristine C60 and RGO, because the amino terminal group of enzyme can be easily attracted by oxygen functional group of RGO and electron-releasing nature of enzymes and therefore proteins can be easily attracted by conjugate π electron structure of C60 molecules.
Herein, we report the electrochemical preparation of the RGO–C60 composite by electrochemical reduction of GO–C60 composite at −1.2 V for 200 s in pH 5 solution. The GO–C60 composite was prepared by a simple sonication of C60 in GO aqueous solution for 6 hours at 45 °C. The fabricated RGO–C60 composite was further used as an immobilization matrix for GOx. The DET of GOx is greatly improved in the presence of C60 with RGO. The GOx immobilized RGO–C60 composite electrode was subsequently used for the construction of glucose biosensor. The practicality of the biosensor in human blood serum samples have also been studied in detail.
Cyclic voltammetry (CV) and amperometry i–t measurements were performed using the CHI 1205b and CHI 750a electrochemical stations. The surface morphology studies were carried out JSM-6500F scanning electron microscope. Raman spectrum was recorded using a Raman spectrometer (Dong Woo 500i, Korea) equipped with a 50× objective and a charge-coupled detector. Ultraviolet-visible (UV-Vis) absorption spectra were obtained using a Hitachi U-3300 UV spectrophotometer. A conventional three-electrode system consisting of modified glassy carbon electrode (electrochemically active surface area = 0.12 cm2) as a working electrode, an external saturated Ag/AgCl as a reference electrode and a platinum wire as the auxiliary electrode was used. Amperometric i–t measurements were performed using a PRDE-3A (as distributed by BAS Inc. Japan) rotating ring disc electrode apparatus with an electrochemically working surface area of 0.033 cm2. The electrochemically active surface area (EASA) of the GCE and RDE was calculated by CV according to our previous reported method.22
:
2, v/v%) and ultrasonicated for 6 hours at 45 °C. After 6 hours of ultrasonication at 45 °C, the colour of GO solution changes from brown to dark brown indicates the formation of GO–C60 composite. Ultrasonicator DC150H (Taiwan Delta New Instrument Co. Ltd) with operating frequency of 40 kHz and ultrasonic power output of 150 W has been used for sonication.
The effect of loading amount of GO and C60 on the biosensor response was optimized and the optimized conditions were used for the fabrication of the biosensor. The prepared GO–C60 composite was centrifuged at 2000 rpm to remove the unattached C60. About 8 μL of GO–C60 composite (optimum) was drop casted onto the pre-cleaned GCE and dried at room temperature. The GO–C60 composite modified GCE was transferred to N2 saturated pH 5 solution and was electrochemically reduced to RGO–C60 composite by applying −1.4 V for 200. For comparison, the RGO modified GCE was prepared by electrochemical reduction of GO modified electrode without C60 by applying −1.4 V for 200 s. A schematic representation for the preparation of RGO–C60 composite is shown in Fig. 1.
![]() | ||
| Fig. 2 SEM images of (A) GO and GO sonicated for 6 h at 45 °C (inset), GO/C60 (B), RGO (C) and RGO/C60 (D). The inset of B shows the SEM image of C60. Scale bar = 1.5 μm. | ||
On the other hand, the C60 reveals the typical rod-shape morphology of C60. The SEM image of GO–C60 composite clearly shows that the C60 rods are enfolded by GO sheets which indicate the formation of GO–C60 composite. While, the C60 rods are clearly seen on the RGO nanosheets and the morphology of RGO does not affect upon the electrochemical reduction of GO–C60 composite. The findings confirm the formation of RGO–C60 composite.
The formation of RGO–C60 composite was further confirmed by Raman spectroscopy and UV-Vis spectroscopy. Raman spectroscopy is widely used for the confirmation of the electronic and lattice structures of carbon materials. Fig. 3A shows the Raman spectra of GO (a), C60 (b) and RGO–C60. It can be seen that C60 exhibits two distinct peaks at 1460 and 1680 cm−1, which ascribed to the Ag and Hg mode of C60. While, the Raman spectrum of RGO–C60 composite shows D and G band at 1342 and 1596 cm−1, which is related to the vibrations of sp3 carbon atoms of disordered graphene nanosheets and vibrations of sp2 carbon atom domains of graphite.18 It should be noted that the Hg mode of C60 is not obviously visible on RGO–C60 composite, which maybe overlapping of Hg mode of C60 with the G-band of RGO. Similar results has been reported before for the presence of C60 with carbon nanomaterials.18 In addition, the ID/IG ratio increase (ID/IG = 1.12) in RGO–C60 composite compared to GO (ID/IG = 0.91) which is due to the removal of oxygen functional groups in GO and formation of conjugated graphene network (sp2 carbon) by electrochemical reduction process. The results confirm the formation of RGO–C60 composite.
![]() | ||
| Fig. 3 (A) Raman spectra of GO (a), C60 (b) and RGO–C60 (c). (B) UV-Vis spectra of C60 (a), GO (b), GO–C60 (c), RGO–C60 (d) and RGO (e). | ||
The formation of GO–C60 composite is confirmed by UV-Vis spectroscopy and the results are shown in Fig. 3B. The C60 (curve a) shows a shoulder peak at 332 nm, which is assigned to π–π* electronic transitions of C60 molecules. The GO (curve b) exhibits an absorption maximum at 310 nm corresponding to n → π* transitions of C
O bonds.25 On the other hand, RGO did not show the absorption peak at 310 nm, is due to the deoxygenation of the GO by electrochemical reduction processes. The GO–C60 composite shows absorption peaks at 331 nm and is related to the π–π stacking of C60 and GO. Since, the peak of C60 at 335 nm is a vital indicator for the π–π interaction between the C60 and other organic compounds.20 The RGO–C60 exhibits absorption peak at 332 nm and the peak of C60 at 310 nm disappear, which indicates that C60 molecule is assembled on graphene sheets through π–π interaction. The observed results are much similar to the previously reported RGO–C60 composite.20,26
![]() | ||
| Fig. 4 (A) CV response of C60/GOx (a), GO–C60/GOx (b), RGO (c) and RGO–C60/GOx modified electrode in N2 saturated PBS at a scan rate of 50 mV s−1. | ||
On the other hand, RGO–C60/GOx modified electrode shows a well-defined redox peak for GOx and Epa and Epc of redox couple is observed at −0.439 and −0.405 V with a peak to peak separation (ΔEp) of 34 mV. Furthermore, the Ipa and Ipc of redox couple is 5.6 folds higher than those observed at C60 and GO–C60 modified GCEs. RGO/GOx modified GCE shows less intense redox couple for GOx compared with other GOx immobilized modified GCEs. The results clearly indicate that RGO–C60 composite is favourable for DET between the electrode surface and the redox centers of GOx compared to other modified electrodes. The enhanced DET of GOx at RGO–C60 modified electrode is due to the combined unique properties of RGO and C60. Hence, RGO–C60/GOx modified electrode can be used as a sensitive biosensor for the detection of glucose.
Fig. S1A† shows the effect of scan rate on the DET of GOx immobilized RGO–C60 modified electrode in PBS. It can be seen that the redox peak current of GOx increases with increasing the scan rates; while the Epa and Epc of redox couple is slightly shifted towards the positive and negative direction upon increasing the scan rates from 10 to 200 mV s−1. As shown in Fig. S1B,† the anodic (Ipa) and cathodic peak current (Ipc) of redox couple have a linear dependence over the scan rates from 10 to 120 mV s−1 with the correlation coefficient of 0.9978 and 0.9983, respectively. The result suggests that the redox electrochemical behavior of GOx at RGO–C60 modified electrode is controlled by a surface-confined electrochemical process. The heterogeneous electron transfer rate constant (Ks) of GOx at RGO–C60/GOx modified electrode is calculated when the ΔEp > 100 mV; since the number of electrons transferred in electrochemical reaction of GOx is 2. The value of Ks is calculated as 2.92 s−1 by using the following Laviron equation.28
| Γ = Q/nFA |
The pH is an essential parameter to evaluate the DET of GOx at RGO–C60 modified electrode, since it greatly affects the DET behavior of GOx. Fig. S2A† shows the CV response of GOx immobilized RGO–C60 modified electrode in N2 saturated different pH solutions from 3 to 9 at a scan rate of 50 mV s−1. It can be seen that a well-defined reversible redox peaks of GOx observed in each pH solution; the Epa and Epc of redox couple of GOx shift towards the positive and negative direction upon decreasing and increasing the pH of the solution. As shown in Fig. S2B,† the formal potential (E0′,
) of GOx has a linear dependence over the pH range from 3 to 9 with a slope of −51.9 mV per pH and a correlation coefficient of 0.9939. The observed slope value (−51.9 mV per pH) is close to the theoretical value of the Nernst equation for an equal number of proton and electron transfer reversible electrochemical process and it confirms that the DET of GOx involves the equal number of protons (H+) and electrons (e−) transferred reaction. The DET of GOx can be written by the following eqn. as reported previously.31
| GOx − (FAD) + 2e− + 2H+ ↔ GOx − (FADH2) |
| GOx − (FAD) + 2e− + 2H+ ↔ GOx − (FADH2) (DET of GOx) |
| GOx − (FADH2) + O2 → GOx − (FAD) + H2O2 (O2 reduction) |
Fig. S3B† shows the CV response of RGO–C60/GOx modified electrode in the absence (a) and presence of different concentrations of glucose (b)–(g) in oxygen saturated PBS at a scan rate of 50 mV s−1. It can be seen that the cathodic peak current decreases with increasing the concentration of glucose into the oxygen saturated PBS. The decrease in the cathodic peak current is due to the consumption of molecular oxygen and the formation of H2O2. The high activity of GOx with molecular oxygen is resulting to the decrease of oxidized form of GOx on electrode surface. This is the reason why the addition of glucose restrained the electrocatalytic reaction and lead to the decrease of reduction current in the presence of oxygen. The similar phenomenon has been reported early in the literatures for oxidation of glucose by GOx in the presence of oxygen.31–33 The mechanism of glucose oxidation by GOx in the presence of oxygen can be expressed by following equations, and the schematic representation of electro-oxidation of glucose in the presence of oxygen at RGO–C60/GOx composite is shown in Fig. 5A.
| GOx − (FAD) + glucose → gluconolactone + GOx − (FADH2) |
| GOx − (FADH2) + O2 → GOX − (FAD) + H2O2 |
Amperometric i–t method was used for the determination of glucose using the proposed biosensor, since it is more sensitive method than that of other voltammetric methods. This is the reason why amperometric i–t method has been widely used for enzymatic biosensors. Fig. 5B shows the amperometric i–t response of RGO–C60/GOx modified electrode for the addition of different concentrations of glucose into the constantly stirred oxygen saturated PBS with a working potential of −0.42 V. A well-defined amperometric response is obtained for the each addition of 500 μM glucose into the PBS (inset). The amperometric reduction current decreases upon the addition of further concentrations of glucose in to the PBS and shows the response up to 14.6 mM. The biosensor reaches its steady state current in 3 s. As shown in Fig. 5C, the amperometric i–t response current of the biosensor is linear against the concentration of glucose in the range of 0.1–12.5 mM. The limit of detection (LOD) was calculated as 35 μM based on S/N = 3. The sensitivity is calculated from the slope of the calibration plot as 55.97 μA mM−1 cm−2. The obtained linear response range of glucose is more suitable for the detection of glucose in human blood samples, since the normal glucose range in human blood in the range of 4.4–6.6 mM. The Michaelis–Menten constant (Km) is calculated as 4.44 mM using a following Lineweaver–Burk equation.34 The obtained Km value is lower than those previously reported glucose biosensors,33–36 which indicates the higher affinity of glucose towards biosensor electrode surface.
The selectivity of the proposed biosensor is always an important one especially for amperometric enzymatic biosensors; hence, the selectivity of the biosensor was investigated in the presence of potentially interfering compounds such as dopamine, ascorbic acid and uric acid. The selectivity of the biosensor was investigated using amperometry and shown in Fig. S4.† The experimental conditions are similar as of Fig. 5B. It can be seen that the 5 mM addition of interfering compounds causes less interference effect on the modified electrode; while a sharp response for the addition of 1 mM glucose. The negative working potential is more beneficial for the less interference effect on the detection of glucose. The result confirms the appreciable selectivity of the biosensor towards the detection of glucose and can be used for the real time sensing of glucose in human serum samples without any interference effect.
The practical ability of the biosensor was evaluated for the detection of glucose in human blood serum samples using amperometry. The concentration of glucose was predetermined by a commercial detector and the appropriate concentration of blood serum was diluted with PBS for the real sample analysis. The experimental conditions are similar as of Fig. 5B. The recovery of glucose in human serum samples was calculated using the standard addition method,31 and the recovery of glucose is listed in Table ST2.† It can be seen that the proposed sensor shows a satisfactory recovery towards glucose in human serum samples and the recovery was in the range of 97.5%. The satisfactory recovery of glucose in human serum samples authenticates the good practicality of the proposed biosensor.
The storage stability of the biosensor was evaluated by CV and the cathodic response current of biosensor towards 1 mM glucose containing PBS was measured periodically and the results are shown in Fig. S5;† the biosensor was stored in the refrigerator when not in use. The biosensor retains 90.9% of its initial response towards glucose after stored in six days, which indicates the good storage stability of the proposed biosensor. The strong immobilization of GOx in the composite matrix thus leads to the high activity of the biosensor. The repeatability and reproducibility of the biosensor was evaluated using CV for the detection of 1 mM glucose. The three biosensors were prepared and used for the detection of glucose in the PBS and the relative standard deviation (RSD) of 3.9% was observed. The RSD of 4.4% was observed for the 8 measurements of the detection of 1 mM glucose containing PBS using a single biosensor. These results indicate a good precision and accuracy of the proposed biosensor.