Improved plutonium concentration analysis in specimens originating from the nuclear fuel cycle using high resolution ICP-OES

Michael Krachler * and Rafael Alvarez-Sarandes
European Commission – Joint Research Centre, Institute for Transuranium Elements, P.O. Box 2340, D-76125 Karlsruhe, Germany. E-mail: michael.krachler@ec.europa.eu; Web: http://www.itu.jrc.ec.europa.eu Fax: +49 7247 951 588; Tel: +49 7247 951 884

Received 22nd May 2015 , Accepted 3rd June 2015

First published on 4th June 2015


Abstract

An accurate and robust ICP-OES procedure for the reliable determination of the Pu concentration in radioactive samples was developed and subsequently cross-validated using ICP-MS. Method optimisation focused on identifying Pu emission wavelengths that are spectrally not interfered by the occurrence of emission signals of concomitant elements (e.g. Am, Np, Th, and U) in the analyte solution, thereby avoiding an otherwise necessary chemical separation of Pu from the matrix elements. To this end, a total of 43 Pu emission wavelengths were tested for their selectivity and sensitivity. The signal to background ratio of the 27 most prominent Pu emission lines ranged from ∼2 to 20, while relative sensitivities varied only within a factor of two. Peak widths of ICP-OES Pu signals extended from ∼5 pm to ∼11 pm, with the majority remaining <6 pm. Using a desolvating nebuliser, instrumental detection limits (LOD) were lowered by an order of magnitude compared to all previous studies. Among the finally selected, seven worthwhile Pu emission wavelengths, LODs of 2.3 μg L−1, 3.1 μg L−1, and 3.2 μg L−1 were obtained at λ = 299.649 nm, λ = 299.409 nm, and λ = 297.251 nm. Applying external calibration, several radioactive samples originating from pyrochemical separation experiments were analysed for their Pu concentration. Typical relative standard deviations of replicate Pu measurements amounted to ∼1% to 2%. The complementary employment of the standard addition approach confirmed the accuracy of this data as well did comparative in-house sector field ICP-MS analysis. Preliminary experiments with spent fuel solutions revealed promising Pu data highlighting the potential of ICP-OES for reliable Pu analysis in the nuclear field.


Introduction

Even though plutonium (Pu) has been present in the universe ever since the first supernova occurred about 100[thin space (1/6-em)]000 years after the Big Bang,1 Pu is principally considered a man-made element. During the last few decades substantial amounts (∼1300 tonnes) of Pu have also been produced as a by-product during the operation of uranium-driven nuclear reactors all over the globe.2 Of course, Pu is also a major component of various nuclear weapons. Because Pu is a highly toxic, radioactive, fissile element, its use is strictly controlled by national and international safeguard programmes.3

Mass spectrometric techniques such as ICP-MS, MC-ICP-MS, accelerator mass spectrometry (AMS) and TIMS, but also alpha and gamma spectrometry, are commonly used to determine both Pu concentrations and the abundance of various Pu isotopes.4–7 These instrumental techniques normally require a chemical separation of Pu from U (e.g. via liquid–liquid extraction using TBP (tri-n-butyl phosphate) or using commercial pre-packed columns such as TEVA) to reduce detrimental interferences that would otherwise hamper the reliable determination of specific Pu isotopes. The quantification of 238Pu using ICP-MS, for example, suffers from the isobaric spectral interference of 238U (99.275% natural abundance), while the formation of 238UH deteriorates the reliable analysis of 239Pu by mass spectrometry.4–7 Similarly, the presence of 241Am in a sample complicates the mass spectrometric measurement of the 241Pu. In alpha spectrometry, in turn, the signals of 239Pu and 240Pu occur as a single peak because both isotopes have similar energies which often cannot be discriminated sufficiently.4,6 All these constraints hamper reliable Pu concentration analyses, which comprises the sum of the contributions of all individual Pu isotopes.

The disadvantageous analytical effects described above may be overcome using ICP-OES if sufficiently sensitive Pu emission wavelengths are identified that do not overlap with spectral lines arising from other concomitant elements present in the sample to be analysed. In fact, being aware of the limitations of α- and γ-spectrometry, Claudon et al.8 investigated the impact of the presence of Am and U in the analyte solution on various prominent Pu emission wavelengths employing ICP-OES already almost three decades ago. Due to the low optical resolution of 20 pm of their employed ICP-OES, only the Pu emission wavelength at λ = 453.62 nm was recommended for Pu analysis.8 Later, Kulkarni et al.9 studied the influence of the presence of U and Th on two prominent Pu ICP-emission wavelengths. Already at a 10-fold excess of either U or Th, the Pu emission signal at λ = 300.06 nm yielded distinctly elevated Pu concentrations.9 The Pu emission signal at λ = 453.62 nm, in turn, was found to be free from these spectral interferences.9 More recently, the spectral performance of the two above mentioned Pu emission wavelengths was tested using a high resolution (HR-)ICP-OES instrument.10 Employing experimentally established inter-elemental correction factors, the potential impact of a 10-fold excess of U, Th and selected rare earth elements was limited to an increase of 8% with respect to the target Pu concentration.10

Despite several earlier investigations,8–10 the potential of HR-ICP-OES for the reliable determination of Pu in nuclear specimens is not yet fully exploited. While most previous studies focussed on only a few Pu emission wavelengths, the current study investigates as many as 43 of them aiming at arriving at a list of sensitive Pu emission wavelengths that works reliably with genuine samples. This endeavour also includes setting up measures to lower Pu detection limits (LOD) compared to values reported in the current scientific literature. Specific emphasis is targeted on the identification of wavelengths that do not suffer from spectral interferences and thus do neither need correction factors to be applied nor a chemical separation prior to Pu analysis. Altogether, this study attempts to develop an analytical procedure allowing the reliable, simple and fast quantification of Pu in samples related to the nuclear fuel cycle using HR-ICP-OES.

Experimental

Instrumentation

All measurements were carried out with a commercial high resolution (HR-)ICP-OES instrument (Ultima2, HORIBA Jobin Yvon, Longjumeau, France). The entire sample introduction system including the autosampler (AS500, HORIBA) of this sequentially working optical spectrometer was installed in a glove box enabling the analysis of the radioactive samples. This demanding adaptation of the original ICP-OES set-up protected the analyst from α and β radiation of the sample solutions to be analysed. A PolyPro ST nebuliser (Elemental Scientific, Inc., Omaha, NE, USA) was attached to a high efficiency sample introduction system (Apex E, Elemental Scientific) enhancing ICP-OES performance. The ICP was operated at 1000 W forward power. More detailed operating and data acquisition parameters of the HR-ICP-OES are reported elsewhere.11,12

For quality assurance selected samples and the Pu stock solution used for calibrating the ICP-OES response were additionally analysed for their Pu concentration in-house using sector field (SF-)ICP-MS (Element2, Thermo Scientific). Similar to the ICP-OES set-up, a glove box hosted the sample introduction system of the SF-ICP-MS that was run under routine conditions according to accredited standard operating procedures.

Reagents and standards

High purity water (18.2 MΩ cm) from a Milli-Q Element (Millipore) water purification system and sub-boiled nitric acid were used for the preparation of all solutions.

Because of their radioactive nature, all investigated samples were handled in appropriate glove boxes following strict safety and security procedures. As such, all analysed samples had to be bagged-in into a dedicated glove box through a lock system prior to any measurement. To minimise the radioactive dose to the operator, samples were fed to the sample introduction system of the ICP-OES with an autosampler.

A concentrated Pu solution was prepared by dissolving a piece of Pu metal (∼0.55 g; Pu content: 99.90 ± 0.04%) in hydrochloric acid as recommended by the supplier of this reference material (Plutonium Metal “MP2”, CETAMA, CEA VALHRO, Marcoule, France). A 100 mg kg−1 working stock solution was obtained by diluting an aliquot of the concentrated Pu solution with 8 M nitric acid using an analytical balance. The purity and actual Pu concentration of this working stock standard solution was confirmed by in-house SF-ICP-MS analysis. Seven Pu solutions with nominal Pu concentrations of 0, 150, 350, 500, 650, 850, and 1000 μg kg−1 in 0.14 M HNO3 were prepared for setting up calibration curves at all investigated Pu emission wavelengths.

Additional standard addition experiments involved the analysis of the actual sample plus three spiked sample aliquots with increasing Pu concentrations. All solutions were prepared on a weight basis to avoid inaccuracies related to the delicate handling of small volumes with pipets inside the glove box.

For the interference study, single element solutions of natural U and Th (prepared from 1000 mg kg−1 stock solutions, Specpure®, Alfa Aesar) in 1 M nitric acid as well as thoroughly characterised 241Am and 237Np solutions (prepared and characterised in-house)12–14 were employed.

Radioactive samples

Pyrochemical actinide partitioning. The analysed samples originated from in-house pyrochemical experiments focussing on the homogeneous recovery of all actinides from metallic nuclear fuels.15 Briefly, the fuel was dissolved anodically in a molten salt consisting of a LiCl–KCl eutectic mixture. Salt samples generated throughout this process were subsequently dissolved in a mixture of 1 M HNO3, 0.1 M HF and 1 M HCl. For analysis in the present study, all solutions were diluted with 0.14 M HNO3. Details of the experimental procedure as well as the preparation and nature of the investigated nuclear samples can be found elsewhere.12,15
Spent fuel. Two different kinds of spent fuel were considered for this study. All corresponding solutions were diluted gravimetrically. First, a two mm slice of an irradiated (Th, Pu)O2 fuel rod was dissolved by boiling under reflux in a so-called “Thorex solution” containing Al(NO3)3·9H2O, NaF and 14.4 M HNO3 in a chemical hot-cell at JRC-ITU using telemanipulators within the framework of a Th irradiation project called LWR-Deputy.16 The dissolved fuel solution was further diluted ∼200 times with 0.14 M nitric acid for subsequent ICP-OES analysis.

Similarly, a metallic fuel, irradiated in the frame of the METAPHIX-1 Project,17 was dissolved and investigated for its Pu concentration. The experimental research fuel has been doped with 2% rare earth elements and 2% minor actinides in order to study the impact of irradiation on these elements. The dissolution of the fuel samples was carried out by boiling under reflux in a mixture of 3 M HNO3 and 0.2 M HF in a chemical hot-cell at JRC-ITU. From these dissolved fuel solutions, further sample dilutions (∼4000 times) have been prepared using 0.14 M HNO3 for ICP-OES analysis.

Results and discussion

Selection of wavelengths

Because the recent scientific literature is lacking a compilation of suitable, sensitive ICP-OES emission wavelengths for Pu, all 43 wavelengths available in the instrument software were tested for potential Pu analysis. To this end, emission spectra centred around the reported Pu peak maxima were recorded within a distance of ±50 pm in 1 pm increments aspirating a 1 mg kg−1 Pu solution. In addition, spectra of a blank solution containing 0.14 M HNO3 were investigated at the above mentioned wavelength regions aiming at identifying non-Pu-specific emission signals. After this initial screening, the use of 16 Pu emission wavelengths was no longer pursued because they were either of low sensitivity or consisted of unfavourable peak shapes and/or a noisy spectral background.

Subsequently the remaining 27 Pu wavelengths were investigated in more detail, specifically for their sensitivity, selectivity and detection limits.

Spectroscopic characterisation of Pu emission lines

These 27 most promising Pu emission wavelengths are listed in chronological order in Table 1, together with some relevant spectroscopic characteristics. The first ten wavelengths from λ = 292.555 nm to λ = 300.057 nm provide the best signal-to-background (S/B) ratio, while starting from λ = 319.844 nm S/B ratios clearly deteriorate. It should be noted, however, that these values reveal a day-to-day variability of approximately ±10%. A similar trend holds true for the slope of the calibration curves (Table 1), as exemplified for three selected Pu emission wavelengths in Fig. 1. However, as indicated by the relative sensitivity of the tested 27 emission wavelengths, the spread between the most and the least sensitive emission line is only a factor of about two (Table 1). In other words, sensitivity issues are not the main selection criteria for a specific wavelength because the instrumental response of all investigated Pu wavelengths is comparable within a factor of two.
Table 1 Spectroscopic characteristics of the 27 most prominent Pu emission lines as determined using HR-ICP-OES
Wavelength [nm] Signal/background Slopea Relative sensitivityb Peak width [pm]
a Net intensity (counts) per mg kg−1 Pu. b Normalised to the emission intensity obtained at λ = 437.990 nm.
292[thin space (1/6-em)]555 10.6 84[thin space (1/6-em)]240 1.7 5.8
293[thin space (1/6-em)]912 11.8 95[thin space (1/6-em)]590 2.0 5.1
295[thin space (1/6-em)]005 14.9 95[thin space (1/6-em)]970 2.0 5.0
295[thin space (1/6-em)]455 11.1 90[thin space (1/6-em)]460 1.9 6.0
296[thin space (1/6-em)]464 11.6 85[thin space (1/6-em)]290 1.8 5.8
297[thin space (1/6-em)]251 15.9 88[thin space (1/6-em)]450 1.8 5.3
298[thin space (1/6-em)]023 10.2 94[thin space (1/6-em)]160 2.0 5.3
299[thin space (1/6-em)]409 15.1 94[thin space (1/6-em)]590 2.0 6.8
299[thin space (1/6-em)]649 19.6 92[thin space (1/6-em)]570 1.9 5.2
300[thin space (1/6-em)]057 18.5 103[thin space (1/6-em)]440 2.1 5.6
319[thin space (1/6-em)]844 3.3 68[thin space (1/6-em)]430 1.4 5.7
340[thin space (1/6-em)]110 6.6 81[thin space (1/6-em)]410 1.7 5.2
346[thin space (1/6-em)]510 5.2 75[thin space (1/6-em)]070 1.6 4.8
358[thin space (1/6-em)]587 5.7 74[thin space (1/6-em)]160 1.5 5.1
363[thin space (1/6-em)]221 6.5 76[thin space (1/6-em)]940 1.6 9.3
390[thin space (1/6-em)]721 4.4 81[thin space (1/6-em)]740 1.7 5.7
397[thin space (1/6-em)]220 2.7 57[thin space (1/6-em)]570 1.2 5.3
397[thin space (1/6-em)]543 3.1 60[thin space (1/6-em)]610 1.3 5.7
402[thin space (1/6-em)]154 3.3 48[thin space (1/6-em)]500 1.0 4.6
435[thin space (1/6-em)]271 2.6 66[thin space (1/6-em)]220 1.4 8.2
437[thin space (1/6-em)]990 1.7 48[thin space (1/6-em)]280 1.0 10.0
439[thin space (1/6-em)]645 2.6 63[thin space (1/6-em)]560 1.3 8.4
447[thin space (1/6-em)]270 2.7 67[thin space (1/6-em)]090 1.4 10.5
449[thin space (1/6-em)]378 3.0 71[thin space (1/6-em)]920 1.5 8.3
450[thin space (1/6-em)]492 3.1 76[thin space (1/6-em)]870 1.6 8.3
453[thin space (1/6-em)]615 6.8 93[thin space (1/6-em)]420 1.9 9.0
476[thin space (1/6-em)]717 2.0 52[thin space (1/6-em)]110 1.1 7.9



image file: c5ja00187k-f1.tif
Fig. 1 Representative Pu calibration curves up to a concentration of 1 mg kg−1 at three selected prominent Pu emission wavelengths, i.e. λ = 299.409 nm, λ = 390.721 nm, and λ = 439.645 nm. Correlation coefficients of these regression curves are always >0.999, but absolute values of their slopes might vary up to 10% on a day-to-days basis. Relative standard deviations of replicate measurements (N = 3) of single calibration points are <2%.

In this respect, the ICP-OES performance of Pu is similar to that of Am, but it is in large contrast to the spectroscopic behaviour of other actinides such as Np and U.12,13,18 While relative sensitivities of up to ∼30 were established recently for Np at the most sensitive emission wavelengths,12 for Am corresponding maximum values amounted to only ∼1.8.13 Similarly, S/B ratios of up to ∼30 were achievable for U,18 numbers that are slightly superior to those of Pu (Table 1).

The peak width (full width at half maximum, FWHM) of most of the tested Pu emission signals range from ∼5 to ∼6 pm (Table 1). However, also broad peaks having widths of up to ∼11 pm are found, particularly at higher emission wavelengths. In general, narrower signals are preferred for data evaluation because they often provide a superior S/B ratio as well as a higher relative sensitivity (Table 1). In addition, the narrower the peak, the less likely is the occurrence of spectral interferences that might hamper the reliable determination of the analyte, i.e. Pu. One exception in this context, however, is the Pu emission signal at λ = 453.615 nm being broad and also offering high sensitivity at the same time. If peak area is considered for data evaluation, this constellation could lead to substantial performance improvement at this particular Pu emission wavelength. The software of the employed ICP-OES instrument, however, supports only peak height measurements, leaving this option untouched.

While the typical peak width of Pu emission signals acquired with the employed ICP-OES instrument is <6 pm, it can be as large as ∼11 pm. For U representative peak widths are ∼4 pm,18 while those of other actinides such as Am and Np range from ∼3–12 pm (ref. 13) and ∼4–30 pm (ref. 12), respectively. In this context, the spectroscopic behaviour of Pu is similar to that of the minor actinides Am and Np.

Analytical performance

Prior to the assessment of instrumental detection limits (LOD), the detector voltage (final range: 774–955 V) for each Pu emission wavelength was optimised using a 1 mg kg−1 Pu solution. This experimental approach established the optimum compromise between sensitivity, spectral baseline and the corresponding S/B ratio. Detection limits for Pu were computed based on the results of 25 measurements of a blank solution containing 0.14 M HNO3. To this end, the signal intensity of the 25 blank measurements was averaged for each wavelength. Subsequently, the LOD was calculated as the concentration corresponding to three times the standard deviation of this blank (3σ criterion).

An exhaustive list comprising the LOD obtained at each of the 27 investigated Pu emission wavelengths is given in Table 2. This record represents by far the most comprehensive collection of Pu emission wavelengths and corresponding Pu LOD in the recent scientific literature. To put the data of the current study into perspective and for a sound comparison, previously reported Pu LOD are additionally summarised in Table 2. Depending on both the selected Pu emission wavelength and study, best Pu LOD reported formerly range from 15 to 100 μg L−1 (Table 2). Interestingly, achievable LOD at a particular wavelength vary considerably. At the Pu emission wavelength λ = 300.057 nm (one of the most frequently selected Pu emission wavelengths), for example, Mainka et al.19 achieved the lowest Pu LOD of 20 μg L−1, while higher values of 28 μg L−1, 30 μg L−1, 54 μg L−1, 100 μg L−1 and 200 μg L−1 have been reported by other authors.8–10,19–21 This diversity in Pu LOD reflects, beside other issues, the different sensitivity of the employed detectors across the studied wavelength region.

Table 2 Comparison of Pu limits of detection [μg L−1] obtained at various ICP-OES emission lines as reported by several authorsa
Limit of detection, LOD
Wavelength [nm] This study Ref. 21 Ref. 8 Ref. 19 Ref. 20
a Limit of detection, based on the 3σ criterion. See text for details.
292[thin space (1/6-em)]555 2.9
293[thin space (1/6-em)]912 3.2
295[thin space (1/6-em)]005 2.5 46 250 104 76
295[thin space (1/6-em)]455 3.0
296[thin space (1/6-em)]464 6.2 61 200 108 61
297[thin space (1/6-em)]251 3.2 29 100 19 68
298[thin space (1/6-em)]023 2.1 52 51 122
299[thin space (1/6-em)]409 3.1 48 57
299[thin space (1/6-em)]649 2.3 30 150 35 81
300[thin space (1/6-em)]057 2.2 28 100 20 54
319[thin space (1/6-em)]844 11 167 456 81
340[thin space (1/6-em)]110 9.0 78 403 50
346[thin space (1/6-em)]510 20
358[thin space (1/6-em)]587 35
363[thin space (1/6-em)]221 7.3 49 300 51 25
390[thin space (1/6-em)]721 5.5
397[thin space (1/6-em)]220 15
397[thin space (1/6-em)]543 8.2
402[thin space (1/6-em)]154 12
435[thin space (1/6-em)]271 9.7
437[thin space (1/6-em)]990 20
439[thin space (1/6-em)]645 14
447[thin space (1/6-em)]270 9.4
449[thin space (1/6-em)]378 11
450[thin space (1/6-em)]492 10
453[thin space (1/6-em)]615 10 89 220 15
476[thin space (1/6-em)]717 18


It is important to stress that the vast majority of Pu LOD determined in the current study at any emission wavelength is lower than the best LOD reported before for optimum Pu wavelengths (Table 2). In general, LOD presented here were improved by about an order of magnitude compared to earlier work, including other studies employing the same ICP-OES model used in the current study.10 In fact, the two emission wavelengths λ = 300.057 nm and λ = 453.615 nm considered in this latter study10 yielded LOD for Pu of 30 μg L−1 (vs. 2.2 μg L−1 in the current investigation) and 70 μg L−1 (10 μg L−1), respectively. This difference in LOD can be partly explained by the different nebulisers used in the two studies, i.e. pneumatic nebuliser (Meinhard TR 50-C1 made from glass)10 operated at 1 mL min−1vs. a low flow nebuliser (PolyPro ST made from polypropylene) in combination with a desolvator aspirating 0.3 mL of sample solution min−1 in this study.

Compared to previous work the use of the low flow nebuliser allowed for a 3-fold reduction of the sample volume required for analysis in the current work, producing less radioactive waste that is expensive to dispose of.

Spectral interferences

One of the potential benefits of ICP-OES over ICP-MS for Pu analysis of nuclear samples is the fact that no chemical separation of the actinide from matrix elements might be required when using ICP-OES. If doable, this advantage obviously speeds up the analysis and thereby reduces the waste amounts and also lowers the radiation dose to the analyst. Excess amounts of U frequently present in spent nuclear fuel solutions, for example, create a huge ICP-MS response in the mass spectrum at m/z 238 that spills over into the much smaller Pu signal at m/z 239. In addition, the formation of the polyatomic 238UH species always adds to the 239Pu signal in mass spectrometry and therefore needs to be corrected for.5,6 Consequently, Pu determinations employing mass spectrometry require utmost attention and in addition, frequently a chemical separation of U from Pu prior to analysis.6,7 For ICP-OES, in turn, emission wavelengths of U and Pu might be far away from each other, allowing for an accurate quantification of Pu in samples containing high amounts of U without the need of a chemical separation of Pu prior to analysis.

Previous ICP-OES studies on the determination of Pu investigated either only two9,10 or just a few Pu emission wavelengths.8,19–21 This restriction limited the chances of identifying suitable Pu emission wavelengths that are free from spectral interferences originating from concomitant elements such as Am, Np, U, and Th that are possibly present at elevated concentrations in the analyte solution. Using an ICP-optical emission spectrometer having an spectral resolution of 20 pm, Claudon et al.,8 for example, excluded the four prominent Pu emission wavelengths λ = 293.912 nm, λ = 297.251 nm, λ = 298.023 nm, and λ = 299.409 nm because the Pu signals overlapped with intensive emission wavelengths of Am. Among the remaining Pu emission wavelengths, λ = 453.615 nm was found to be the most suitable one, tolerating even a ten-fold excess of U relative to Pu.8 In the same vein, another more recent study10 also favoured the Pu emission wavelength at λ = 453.615 nm over that at λ = 300.057 nm being less prone to spectral interferences.

To set the above mentioned facts into perspective, one should mention that typical UO2-based spent fuels contain ∼1 mass% of Pu and 96 mass% of U. In mixed oxide (MOX) fuels the amount of PuO2 can reach up to 30%. While metallic fuels consist of ∼60 mass% U and ∼20 mass% Pu, thoria-based (Th, Pu)O2 fuels may comprise 1.7 mass% U, 0.6 mass% Pu and 83 mass% Th.22

To shed more light on the potential impact of concomitant elements such as U and Th on Pu ICP-OES analysis, an extensive interference study was carried out. Because of its relevance to reprocessing experiments, the impact of excess amounts of Al, K, Li, Zr and Nd (as a typical representative for fission products) on the Pu ICP-OES signal was also tested. To this end emission spectra of both synthetic standard solutions and actual sample solutions were recorded at the 27 Pu emission wavelengths listed in Table 2.

Part 1 of the interference study, employing standard solutions containing either 0.5 mg kg−1 241Am or 1 mg kg−1 of each Pu, U, Th, and 237Np, excluded further potential Pu emission wavelengths. Among them was the Pu emission line at λ = 293.912 nm, as showcased in Fig. 2. While the presence of Th and 237Np did not spectrally interfere with the Pu emission signal, small amounts of U and particularly Am hampered the reliable determination of Pu at this sensitive Pu emission wavelength (Fig. 2). The presence of Th gave rise to a peak overlapping with the Pu ICP-OES signal at λ = 319.844 nm, for example, also excluding this emission wavelength for Pu analysis (data not shown).


image file: c5ja00187k-f2.tif
Fig. 2 High resolution ICP-OES spectrum of 1 mg kg−1 Pu centred around λ = 293.912 nm showcasing the impact of 0.5 mg kg−1 241Am, and 1 mg kg−1 of each 237Np, Th, and U. Each displayed spectrum comprises 101 data points recorded in 1 pm increments.

For part 2 of the interference study, four samples representing a variety of matrix compositions from in-house pyrochemical experiments were selected. Based on the emission spectra recorded for these samples at the 27 wavelengths above, the list of “useful” Pu emission wavelengths was further cut down to the seven wavelengths listed in Table 3. This decision was based on the fact that additional severe spectral interferences occurring at the majority of these 27 emission wavelengths – that were not obvious in the preliminary interferences check using standard solutions (see above) – hampered the reliable determination of Pu. Fig. 3 exemplifies these findings for a set of four samples originating from different steps of pyrochemical separation experiments as observed around one of the most popular Pu emission wavelength, i.e. λ = 300.057 nm. While the reason for the deteriorating peak shape for samples A18 and A32 have not been investigated in detail, the elevated (up to 20×) U concentration relative to their Pu amount might be the respective cause. For samples A16 and A30, in turn, Pu emission spectra revealed no spectral interferences around the peak maximum (Fig. 3). Even though being one of the most popular Pu emission lines, λ = 300.057 nm was not considered for quantification of the element in the investigated nuclear samples because of its incongruous spectral behaviour.

Table 3 Concentration of Pu [N = 12, mg kg−1] in nine sample solutions from pyrochemical experiments obtained at nine ICP-OES emission wavelengths on four consecutive days using external calibrationa
Sample Emission wavelength, nm
297.251 299.409 299.649 390.721 439.645 449.378 453.615
a Numbers in italics indicate results that indicate either diminished or elevated values compared to Pu concentrations obtained at λ = 297.251 nm, λ = 449.378 nm and λ = 453.615 nm. See text for details.
1 0.45 ± 0.01 0.46 ± 0.01 0.45 ± 0.01 0.42 ± 0.01 0.43 ± 0.01 0.47 ± 0.01 0.46 ± 0.02
2 0.48 ± 0.01 0.48 ± 0.01 0.48 ± 0.01 0.44 ± 0.00 0.45 ± 0.01 0.48 ± 0.02 0.47 ± 0.02
3 0.43 ± 0.01 0.42 ± 0.01 0.43 ± 0.01 0.39 ± 0.01 0.42 ± 0.00 0.46 ± 0.01 0.42 ± 0.01
4 0.46 ± 0.01 0.48 ± 0.01 0.48 ± 0.01 0.44 ± 0.00 0.43 ± 0.01 0.47 ± 0.01 0.47 ± 0.02
5 0.49 ± 0.01 0.50 ± 0.00 0.49 ± 0.01 0.46 ± 0.00 0.46 ± 0.01 0.49 ± 0.01 0.50 ± 0.02
6 0.55 ± 0.01 0.80 ± 0.00 0.95 ± 0.03 0.68 ± 0.01 0.51 ± 0.01 0.54 ± 0.01 0.54 ± 0.01
7 0.47 ± 0.01 0.58 ± 0.00 0.63 ± 0.02 0.53 ± 0.01 0.46 ± 0.01 0.47 ± 0.02 0.48 ± 0.02
8 0.41 ± 0.01 0.44 ± 0.00 0.44 ± 0.01 0.41 ± 0.01 0.41 ± 0.01 0.40 ± 0.01 0.42 ± 0.01
9 0.51 ± 0.02 0.57 ± 0.01 0.59 ± 0.01 0.53 ± 0.01 0.51 ± 0.01 0.51 ± 0.01 0.53 ± 0.02



image file: c5ja00187k-f3.tif
Fig. 3 High resolution ICP-OES spectra of four nuclear samples related to pyrochemical experiments highlighting the extent of potential spectral interferences deteriorating the reliable determination of Pu at one of the most popular Pu emission wavelengths, i.e. λ = 300.057 nm. Each displayed spectrum comprises 101 data points recorded in 1 pm increments. Denomination of samples refers to internal laboratory numbers.

Pu concentration analysis of nuclear samples

External calibration was used for quantification of Pu with ICP-OES in nine nuclear samples (Table 3). The results of 12 replicate analyses (3 replicates each on 4 days) revealed a high reproducibility of the measurements at each of the seven employed Pu emission wavelengths, i.e. relative standard deviations (RSD) of ∼2–4%. While the majority of the Pu concentrations obtained at the seven emission wavelengths were very similar, however, some discrepancies were also evident. Compared to the other five Pu emission wavelengths calculated Pu concentrations were systematically lower at λ = 390.721 nm and λ = 439.645 nm in samples 1 to 5 (Table 3). Similarly, signal intensities recorded at both λ = 299.409 nm and λ = 299.649 nm resulted in elevated Pu concentrations for samples 8 and 9. For samples 6 and 7, in turn, high Pu concentrations were obtained at λ = 299.409 nm, λ = 299.649 nm and λ = 390.721 nm as well as marginally diminished values at λ = 439.645 nm (Table 3). Spectral interferences of concomitant elements either overlapped partly with the Pu emission signal or slightly increased the spectral background leading to these deteriorating results discussed above.

As a consequence of these experimental facts and to have a single robust set of identical acquisition parameters, only the three Pu emission wavelengths at λ = 297.251 nm, λ = 449.378 nm and λ = 453.615 nm (Table 3) were considered for a reliable quantification of Pu. The selection of these three Pu emission wavelengths for calculating Pu concentrations was additionally supported by the results obtained through replicate standard addition experiments (N = 12) carried out with all nine investigated nuclear samples (Table 4).

Table 4 Comparison of Pu concentrations [mg kg−1] in nine sample solutions from pyrochemical experiments obtained using ICP-OES and SF-ICP-MSd
Sample ICP-OES SF-ICP-MS biasc
Standard additiona RSD, % External calibrationb RSD, % Absolute, mg kg−1 Relative, %
a Standard addition experiments using three aliquots of the actual sample and three spiked solutions for each sample aliquot on three consecutive days (N = 9). b External calibration experiments carried out on four consecutive days (N = 12). c Difference between the ICP-OES results (based on external calibration) in column 4 and comparative SF-ICP-MS data acquired in-house. Measurement uncertainties of SF-ICP-MS data are <1%. d Reported ICP-OES values represent the grand average (and the corresponding standard deviation) calculated from the average Pu concentrations obtained at the three Pu emission wavelengths λ = 297.251 nm, λ = 449.378 nm and λ = 453.615 nm (see Table 3).
1 0.45 ± 0.02 3.9 0.46 ± 0.01 2.1 −0.02 −4.8
2 0.48 ± 0.01 1.8 0.47 ± 0.01 1.3 −0.04 −8.6
3 0.45 ± 0.02 3.9 0.43 ± 0.02 4.7 −0.04 −8.4
4 0.45 ± 0.00 0.7 0.47 ± 0.01 1.4 0.00 0.0
5 0.47 ± 0.00 0.9 0.50 ± 0.01 1.7 −0.00 −0.1
6 0.54 ± 0.01 2.2 0.54 ± 0.00 0.7 0.00 0.9
7 0.47 ± 0.00 0.4 0.47 ± 0.01 1.2 0.00 0.4
8 0.39 ± 0.01 1.3 0.41 ± 0.01 2.1 −0.01 −2.0
9 0.50 ± 0.01 1.8 0.52 ± 0.01 1.8 0.00 0.4


As no certified matrix-matched reference materials are available for quality assurance, the use of a complementary analysis technique whose detection is based on a different physical principle is desirable.14 To further underpin the accuracy of the selected evaluation regime of the ICP-OES measurements, all samples were also analysed using SF-ICP-MS. To this end, an aliquot from the identical solution that has been used for ICP-OES analysis was also subjected to SF-ICP-MS in-house. In general, ICP-MS results agreed well with ICP-OES data within 2% (Table 4). The results of the samples 1–3, however, did not satisfy these criteria as highlighted in Table 4. In particular, ICP-MS results were up to 8.6% lower than corresponding ICP-OES data. The high matrix load of these three samples (Np 2×; Nd 20×; Li 25×; K 100×; Cl 220×) and the inferior capability of the ICP-MS to compensate for the resulting signal suppression is the most likely explanation for this obvious discrepancy between the Pu concentrations established with both analytical techniques. Both the standard addition and the external calibration ICP-OES approach, in turn, yielded equivalent results suggesting sufficiently robust ICP-OES measurement conditions. It is worth mentioning that RSDs of replicate measurements also deteriorated slightly for these three samples (Table 4). Taken together, however, the developed ICP-OES approach generates accurate Pu concentration data for the above mentioned nuclear samples, avoiding a laborious and time-consuming separation of the analyte from the matrix.

Within this study also promising preliminary data for two different spent fuels were obtained. The current experiments were limited by the small volume (a few millilitres) of dissolved spent fuel that was available for testing. As such, high resolution spectra were recorded at the seven Pu emission wavelengths selected previously for the above described nuclear samples. Among them, the two Pu lines at λ = 299.649 nm (Fig. 4A) and λ = 453.615 nm (Fig. 4B) gave the impression to work best with this type of samples. Even though the metallic spent fuel contained ∼60% U, the corresponding emission spectra emerged straightforward (Fig. 4A and B). The spectra of the spent (Th, Pu)O2 fuel, in turn, suffered from the signals originating from the ∼83% Th that were present in the non-separated spent fuel solution (Fig. 4A and B). Due to the high resolution capabilities of the employed ICP-OES instrument, however, the Pu emission signals of both spent fuel solutions were well separated from potential spectral interferences. Further in-depth studies will be needed to give more evidence to the potential of HR-ICP-OES for the reliable determination of Pu in such spent fuel solutions.


image file: c5ja00187k-f4.tif
Fig. 4 Smoothed high resolution ICP-OES spectra of solutions of spent, non-separated metallic and (Th, Pu)O2 fuels. (A) Emission spectrum centred around λ = 299.649 nm. Insert shows a close-up of the central part of the spectrum. (B) Emission spectrum centred around λ = 453.615 nm. See text for details.

Acknowledgements

The supply of the employed Pu certified reference material, the preparation of the investigated samples and comparative Pu concentration analysis using SF-ICP-MS by our colleagues B. Lynch, M. Vargas Zunica, P. Souček, and G. Rasmussen at EC-JRC-ITU is greatly appreciated.

References

  1. J. Silk, Origin of the heavy metals, in The Big Bang, W. H. Freeman and company, New York, USA, 1989, pp. 319–340 Search PubMed.
  2. http://www.world-nuclear.org/info/Nuclear-Fuel-Cycle/Fuel-Recycling/Plutonium/ .
  3. International safeguards in nuclear facility design and construction, IAEA nuclear energy series, ISSN 1995–7807, NP-T-2.8, Vienna, International Atomic Energy Agency, 2013.
  4. X. Hou and P. Roos, Anal. Chim. Acta, 2008, 608, 105–139 CrossRef CAS PubMed.
  5. J. Zheng, K. Tagami, S. Homma-Takeda and W. Bu, J. Anal. At. Spectrom., 2013, 28, 1676–1699 RSC.
  6. F. Luisier, J. Antonio Corcho Alvarado, P. Steinmann, M. Krachler and P. Froidevaux, J. Radioanal. Nucl. Chem., 2009, 281, 425–432 CrossRef CAS.
  7. F. Quinto, E. Hrnecek, M. Krachler, W. Shotyk, P. Steier and S. R. Winkler, Environ. Sci.: Processes Impacts, 2013, 15, 839–847 CAS.
  8. X. Claudon, J. C. Birolleau, M. Lavergne, B. Miche and C. Bergey, Spectrochim. Acta, Part B, 1987, 42, 407–411 CrossRef.
  9. J. Kulkarni, S. K. Thulasidas, N. Goyal, A. G. Page and M. D. Sastry, Anal. Lett., 1990, 23, 2095–2106 CrossRef PubMed.
  10. S. K. Thulasidas, V. C. Adya, M. Kumar, T. K. Seshagiri and S. V. Godbole, At. Spectrosc., 2011, 32, 228–233 CAS.
  11. M. Krachler, R. Alvarez-Sarandes and S. Van Winckel, J. Anal. At. Spectrom., 2013, 28, 114–120 RSC.
  12. M. Krachler, R. Alvarez-Sarandes, P. Souček and P. Carbol, Microchem. J., 2014, 110, 425–434 CrossRef PubMed.
  13. M. Krachler, R. Alvarez-Sarandes, P. Carbol, R. Malmbeck and S. Van Winckel, Microchem. J., 2013, 110, 425–434 CrossRef CAS PubMed.
  14. M. Krachler, R. Alvarez-Sarandes and S. Van Winckel, J. Radioanal. Nucl. Chem., 2015, 304, 1201–1209 CrossRef CAS.
  15. P. Souček, R. Malmbeck, C. Nourry and J.-P. Glatz, Energy Procedia, 2011, 7, 396–404 CrossRef PubMed.
  16. S. Van Winckel, L. Aldave de las Heras, P. Carbol, E. Hrnecek, R. Malmbeck, P. Van Belle, E. Zuleger and J.-P. Glatz, LWR-Deputy Project: Radiochemical analysis of a high burn-up (Th,Pu)O2 sample, irradiated in KWO Obrigheim, JRC Scientific and Technical Reports, JRC-ITU-TPW-2010/03, confidential, European Commission, 2010 Search PubMed.
  17. D. Papaioannou, M. Ougier, V. V. Rondinella, J.-P. Glatz and H. Ohta, Post-Irradiation Examinations (PIE) on the U-Pu-Zr Metal Alloy Fuel Rods from the METAPHIX-1 Experiment, JRC Scientific and Technical Reports, JRC-ITU-TPW-2010/04, confidential, European Commission, 2010 Search PubMed.
  18. M. Krachler and D. H. Wegen, J. Anal. At. Spectrom., 2012, 27, 335–339 RSC.
  19. E. Mainka, H. G. Müller and F. Geyer, Report KfK 3578 (in German), "Versuche zur Pu-Bestimmung mit Hilfe der ICP- Methode", Kernforschungszentrum Karlsruhe, Institut für Radiochemie, Karlsruhe, Germany, 1983 Search PubMed.
  20. M. C. Edelson, E. L. DeKalb, R. K. Winge and V. A. Fassel, Spectrochim. Acta, Part B, 1986, 41, 475–486 CrossRef.
  21. E. A. Huff and D. L. Bowers, Appl. Spectrosc., 1990, 44, 728–729 CrossRef CAS.
  22. M. Krachler, R. Alvarez-Sarandes and S. Van Winckel, J. Anal. At. Spectrom., 2014, 29, 817–824 RSC.

This journal is © The Royal Society of Chemistry 2015
Click here to see how this site uses Cookies. View our privacy policy here.