Michael
Krachler
* and
Rafael
Alvarez-Sarandes
European Commission – Joint Research Centre, Institute for Transuranium Elements, P.O. Box 2340, D-76125 Karlsruhe, Germany. E-mail: michael.krachler@ec.europa.eu; Web: http://www.itu.jrc.ec.europa.eu Fax: +49 7247 951 588; Tel: +49 7247 951 884
First published on 4th June 2015
An accurate and robust ICP-OES procedure for the reliable determination of the Pu concentration in radioactive samples was developed and subsequently cross-validated using ICP-MS. Method optimisation focused on identifying Pu emission wavelengths that are spectrally not interfered by the occurrence of emission signals of concomitant elements (e.g. Am, Np, Th, and U) in the analyte solution, thereby avoiding an otherwise necessary chemical separation of Pu from the matrix elements. To this end, a total of 43 Pu emission wavelengths were tested for their selectivity and sensitivity. The signal to background ratio of the 27 most prominent Pu emission lines ranged from ∼2 to 20, while relative sensitivities varied only within a factor of two. Peak widths of ICP-OES Pu signals extended from ∼5 pm to ∼11 pm, with the majority remaining <6 pm. Using a desolvating nebuliser, instrumental detection limits (LOD) were lowered by an order of magnitude compared to all previous studies. Among the finally selected, seven worthwhile Pu emission wavelengths, LODs of 2.3 μg L−1, 3.1 μg L−1, and 3.2 μg L−1 were obtained at λ = 299.649 nm, λ = 299.409 nm, and λ = 297.251 nm. Applying external calibration, several radioactive samples originating from pyrochemical separation experiments were analysed for their Pu concentration. Typical relative standard deviations of replicate Pu measurements amounted to ∼1% to 2%. The complementary employment of the standard addition approach confirmed the accuracy of this data as well did comparative in-house sector field ICP-MS analysis. Preliminary experiments with spent fuel solutions revealed promising Pu data highlighting the potential of ICP-OES for reliable Pu analysis in the nuclear field.
Mass spectrometric techniques such as ICP-MS, MC-ICP-MS, accelerator mass spectrometry (AMS) and TIMS, but also alpha and gamma spectrometry, are commonly used to determine both Pu concentrations and the abundance of various Pu isotopes.4–7 These instrumental techniques normally require a chemical separation of Pu from U (e.g. via liquid–liquid extraction using TBP (tri-n-butyl phosphate) or using commercial pre-packed columns such as TEVA) to reduce detrimental interferences that would otherwise hamper the reliable determination of specific Pu isotopes. The quantification of 238Pu using ICP-MS, for example, suffers from the isobaric spectral interference of 238U (99.275% natural abundance), while the formation of 238UH deteriorates the reliable analysis of 239Pu by mass spectrometry.4–7 Similarly, the presence of 241Am in a sample complicates the mass spectrometric measurement of the 241Pu. In alpha spectrometry, in turn, the signals of 239Pu and 240Pu occur as a single peak because both isotopes have similar energies which often cannot be discriminated sufficiently.4,6 All these constraints hamper reliable Pu concentration analyses, which comprises the sum of the contributions of all individual Pu isotopes.
The disadvantageous analytical effects described above may be overcome using ICP-OES if sufficiently sensitive Pu emission wavelengths are identified that do not overlap with spectral lines arising from other concomitant elements present in the sample to be analysed. In fact, being aware of the limitations of α- and γ-spectrometry, Claudon et al.8 investigated the impact of the presence of Am and U in the analyte solution on various prominent Pu emission wavelengths employing ICP-OES already almost three decades ago. Due to the low optical resolution of 20 pm of their employed ICP-OES, only the Pu emission wavelength at λ = 453.62 nm was recommended for Pu analysis.8 Later, Kulkarni et al.9 studied the influence of the presence of U and Th on two prominent Pu ICP-emission wavelengths. Already at a 10-fold excess of either U or Th, the Pu emission signal at λ = 300.06 nm yielded distinctly elevated Pu concentrations.9 The Pu emission signal at λ = 453.62 nm, in turn, was found to be free from these spectral interferences.9 More recently, the spectral performance of the two above mentioned Pu emission wavelengths was tested using a high resolution (HR-)ICP-OES instrument.10 Employing experimentally established inter-elemental correction factors, the potential impact of a 10-fold excess of U, Th and selected rare earth elements was limited to an increase of 8% with respect to the target Pu concentration.10
Despite several earlier investigations,8–10 the potential of HR-ICP-OES for the reliable determination of Pu in nuclear specimens is not yet fully exploited. While most previous studies focussed on only a few Pu emission wavelengths, the current study investigates as many as 43 of them aiming at arriving at a list of sensitive Pu emission wavelengths that works reliably with genuine samples. This endeavour also includes setting up measures to lower Pu detection limits (LOD) compared to values reported in the current scientific literature. Specific emphasis is targeted on the identification of wavelengths that do not suffer from spectral interferences and thus do neither need correction factors to be applied nor a chemical separation prior to Pu analysis. Altogether, this study attempts to develop an analytical procedure allowing the reliable, simple and fast quantification of Pu in samples related to the nuclear fuel cycle using HR-ICP-OES.
For quality assurance selected samples and the Pu stock solution used for calibrating the ICP-OES response were additionally analysed for their Pu concentration in-house using sector field (SF-)ICP-MS (Element2, Thermo Scientific). Similar to the ICP-OES set-up, a glove box hosted the sample introduction system of the SF-ICP-MS that was run under routine conditions according to accredited standard operating procedures.
Because of their radioactive nature, all investigated samples were handled in appropriate glove boxes following strict safety and security procedures. As such, all analysed samples had to be bagged-in into a dedicated glove box through a lock system prior to any measurement. To minimise the radioactive dose to the operator, samples were fed to the sample introduction system of the ICP-OES with an autosampler.
A concentrated Pu solution was prepared by dissolving a piece of Pu metal (∼0.55 g; Pu content: 99.90 ± 0.04%) in hydrochloric acid as recommended by the supplier of this reference material (Plutonium Metal “MP2”, CETAMA, CEA VALHRO, Marcoule, France). A 100 mg kg−1 working stock solution was obtained by diluting an aliquot of the concentrated Pu solution with 8 M nitric acid using an analytical balance. The purity and actual Pu concentration of this working stock standard solution was confirmed by in-house SF-ICP-MS analysis. Seven Pu solutions with nominal Pu concentrations of 0, 150, 350, 500, 650, 850, and 1000 μg kg−1 in 0.14 M HNO3 were prepared for setting up calibration curves at all investigated Pu emission wavelengths.
Additional standard addition experiments involved the analysis of the actual sample plus three spiked sample aliquots with increasing Pu concentrations. All solutions were prepared on a weight basis to avoid inaccuracies related to the delicate handling of small volumes with pipets inside the glove box.
For the interference study, single element solutions of natural U and Th (prepared from 1000 mg kg−1 stock solutions, Specpure®, Alfa Aesar) in 1 M nitric acid as well as thoroughly characterised 241Am and 237Np solutions (prepared and characterised in-house)12–14 were employed.
Similarly, a metallic fuel, irradiated in the frame of the METAPHIX-1 Project,17 was dissolved and investigated for its Pu concentration. The experimental research fuel has been doped with 2% rare earth elements and 2% minor actinides in order to study the impact of irradiation on these elements. The dissolution of the fuel samples was carried out by boiling under reflux in a mixture of 3 M HNO3 and 0.2 M HF in a chemical hot-cell at JRC-ITU. From these dissolved fuel solutions, further sample dilutions (∼4000 times) have been prepared using 0.14 M HNO3 for ICP-OES analysis.
Subsequently the remaining 27 Pu wavelengths were investigated in more detail, specifically for their sensitivity, selectivity and detection limits.
Wavelength [nm] | Signal/background | Slopea | Relative sensitivityb | Peak width [pm] |
---|---|---|---|---|
a Net intensity (counts) per mg kg−1 Pu. b Normalised to the emission intensity obtained at λ = 437.990 nm. | ||||
292![]() |
10.6 | 84![]() |
1.7 | 5.8 |
293![]() |
11.8 | 95![]() |
2.0 | 5.1 |
295![]() |
14.9 | 95![]() |
2.0 | 5.0 |
295![]() |
11.1 | 90![]() |
1.9 | 6.0 |
296![]() |
11.6 | 85![]() |
1.8 | 5.8 |
297![]() |
15.9 | 88![]() |
1.8 | 5.3 |
298![]() |
10.2 | 94![]() |
2.0 | 5.3 |
299![]() |
15.1 | 94![]() |
2.0 | 6.8 |
299![]() |
19.6 | 92![]() |
1.9 | 5.2 |
300![]() |
18.5 | 103![]() |
2.1 | 5.6 |
319![]() |
3.3 | 68![]() |
1.4 | 5.7 |
340![]() |
6.6 | 81![]() |
1.7 | 5.2 |
346![]() |
5.2 | 75![]() |
1.6 | 4.8 |
358![]() |
5.7 | 74![]() |
1.5 | 5.1 |
363![]() |
6.5 | 76![]() |
1.6 | 9.3 |
390![]() |
4.4 | 81![]() |
1.7 | 5.7 |
397![]() |
2.7 | 57![]() |
1.2 | 5.3 |
397![]() |
3.1 | 60![]() |
1.3 | 5.7 |
402![]() |
3.3 | 48![]() |
1.0 | 4.6 |
435![]() |
2.6 | 66![]() |
1.4 | 8.2 |
437![]() |
1.7 | 48![]() |
1.0 | 10.0 |
439![]() |
2.6 | 63![]() |
1.3 | 8.4 |
447![]() |
2.7 | 67![]() |
1.4 | 10.5 |
449![]() |
3.0 | 71![]() |
1.5 | 8.3 |
450![]() |
3.1 | 76![]() |
1.6 | 8.3 |
453![]() |
6.8 | 93![]() |
1.9 | 9.0 |
476![]() |
2.0 | 52![]() |
1.1 | 7.9 |
In this respect, the ICP-OES performance of Pu is similar to that of Am, but it is in large contrast to the spectroscopic behaviour of other actinides such as Np and U.12,13,18 While relative sensitivities of up to ∼30 were established recently for Np at the most sensitive emission wavelengths,12 for Am corresponding maximum values amounted to only ∼1.8.13 Similarly, S/B ratios of up to ∼30 were achievable for U,18 numbers that are slightly superior to those of Pu (Table 1).
The peak width (full width at half maximum, FWHM) of most of the tested Pu emission signals range from ∼5 to ∼6 pm (Table 1). However, also broad peaks having widths of up to ∼11 pm are found, particularly at higher emission wavelengths. In general, narrower signals are preferred for data evaluation because they often provide a superior S/B ratio as well as a higher relative sensitivity (Table 1). In addition, the narrower the peak, the less likely is the occurrence of spectral interferences that might hamper the reliable determination of the analyte, i.e. Pu. One exception in this context, however, is the Pu emission signal at λ = 453.615 nm being broad and also offering high sensitivity at the same time. If peak area is considered for data evaluation, this constellation could lead to substantial performance improvement at this particular Pu emission wavelength. The software of the employed ICP-OES instrument, however, supports only peak height measurements, leaving this option untouched.
While the typical peak width of Pu emission signals acquired with the employed ICP-OES instrument is <6 pm, it can be as large as ∼11 pm. For U representative peak widths are ∼4 pm,18 while those of other actinides such as Am and Np range from ∼3–12 pm (ref. 13) and ∼4–30 pm (ref. 12), respectively. In this context, the spectroscopic behaviour of Pu is similar to that of the minor actinides Am and Np.
An exhaustive list comprising the LOD obtained at each of the 27 investigated Pu emission wavelengths is given in Table 2. This record represents by far the most comprehensive collection of Pu emission wavelengths and corresponding Pu LOD in the recent scientific literature. To put the data of the current study into perspective and for a sound comparison, previously reported Pu LOD are additionally summarised in Table 2. Depending on both the selected Pu emission wavelength and study, best Pu LOD reported formerly range from 15 to 100 μg L−1 (Table 2). Interestingly, achievable LOD at a particular wavelength vary considerably. At the Pu emission wavelength λ = 300.057 nm (one of the most frequently selected Pu emission wavelengths), for example, Mainka et al.19 achieved the lowest Pu LOD of 20 μg L−1, while higher values of 28 μg L−1, 30 μg L−1, 54 μg L−1, 100 μg L−1 and 200 μg L−1 have been reported by other authors.8–10,19–21 This diversity in Pu LOD reflects, beside other issues, the different sensitivity of the employed detectors across the studied wavelength region.
Limit of detection, LOD | |||||
---|---|---|---|---|---|
Wavelength [nm] | This study | Ref. 21 | Ref. 8 | Ref. 19 | Ref. 20 |
a Limit of detection, based on the 3σ criterion. See text for details. | |||||
292![]() |
2.9 | — | — | — | — |
293![]() |
3.2 | — | — | — | — |
295![]() |
2.5 | 46 | 250 | 104 | 76 |
295![]() |
3.0 | — | — | — | — |
296![]() |
6.2 | 61 | 200 | 108 | 61 |
297![]() |
3.2 | 29 | 100 | 19 | 68 |
298![]() |
2.1 | 52 | — | 51 | 122 |
299![]() |
3.1 | 48 | — | — | 57 |
299![]() |
2.3 | 30 | 150 | 35 | 81 |
300![]() |
2.2 | 28 | 100 | 20 | 54 |
319![]() |
11 | 167 | — | 456 | 81 |
340![]() |
9.0 | 78 | — | 403 | 50 |
346![]() |
20 | — | — | — | — |
358![]() |
35 | — | — | — | — |
363![]() |
7.3 | 49 | 300 | 51 | 25 |
390![]() |
5.5 | — | — | — | — |
397![]() |
15 | — | — | — | — |
397![]() |
8.2 | — | — | — | — |
402![]() |
12 | — | — | — | — |
435![]() |
9.7 | — | — | — | — |
437![]() |
20 | — | — | — | — |
439![]() |
14 | — | — | — | — |
447![]() |
9.4 | — | — | — | — |
449![]() |
11 | — | — | — | — |
450![]() |
10 | — | — | — | — |
453![]() |
10 | 89 | 220 | — | 15 |
476![]() |
18 | — | — | — | — |
It is important to stress that the vast majority of Pu LOD determined in the current study at any emission wavelength is lower than the best LOD reported before for optimum Pu wavelengths (Table 2). In general, LOD presented here were improved by about an order of magnitude compared to earlier work, including other studies employing the same ICP-OES model used in the current study.10 In fact, the two emission wavelengths λ = 300.057 nm and λ = 453.615 nm considered in this latter study10 yielded LOD for Pu of 30 μg L−1 (vs. 2.2 μg L−1 in the current investigation) and 70 μg L−1 (10 μg L−1), respectively. This difference in LOD can be partly explained by the different nebulisers used in the two studies, i.e. pneumatic nebuliser (Meinhard TR 50-C1 made from glass)10 operated at 1 mL min−1vs. a low flow nebuliser (PolyPro ST made from polypropylene) in combination with a desolvator aspirating 0.3 mL of sample solution min−1 in this study.
Compared to previous work the use of the low flow nebuliser allowed for a 3-fold reduction of the sample volume required for analysis in the current work, producing less radioactive waste that is expensive to dispose of.
Previous ICP-OES studies on the determination of Pu investigated either only two9,10 or just a few Pu emission wavelengths.8,19–21 This restriction limited the chances of identifying suitable Pu emission wavelengths that are free from spectral interferences originating from concomitant elements such as Am, Np, U, and Th that are possibly present at elevated concentrations in the analyte solution. Using an ICP-optical emission spectrometer having an spectral resolution of 20 pm, Claudon et al.,8 for example, excluded the four prominent Pu emission wavelengths λ = 293.912 nm, λ = 297.251 nm, λ = 298.023 nm, and λ = 299.409 nm because the Pu signals overlapped with intensive emission wavelengths of Am. Among the remaining Pu emission wavelengths, λ = 453.615 nm was found to be the most suitable one, tolerating even a ten-fold excess of U relative to Pu.8 In the same vein, another more recent study10 also favoured the Pu emission wavelength at λ = 453.615 nm over that at λ = 300.057 nm being less prone to spectral interferences.
To set the above mentioned facts into perspective, one should mention that typical UO2-based spent fuels contain ∼1 mass% of Pu and 96 mass% of U. In mixed oxide (MOX) fuels the amount of PuO2 can reach up to 30%. While metallic fuels consist of ∼60 mass% U and ∼20 mass% Pu, thoria-based (Th, Pu)O2 fuels may comprise 1.7 mass% U, 0.6 mass% Pu and 83 mass% Th.22
To shed more light on the potential impact of concomitant elements such as U and Th on Pu ICP-OES analysis, an extensive interference study was carried out. Because of its relevance to reprocessing experiments, the impact of excess amounts of Al, K, Li, Zr and Nd (as a typical representative for fission products) on the Pu ICP-OES signal was also tested. To this end emission spectra of both synthetic standard solutions and actual sample solutions were recorded at the 27 Pu emission wavelengths listed in Table 2.
Part 1 of the interference study, employing standard solutions containing either 0.5 mg kg−1 241Am or 1 mg kg−1 of each Pu, U, Th, and 237Np, excluded further potential Pu emission wavelengths. Among them was the Pu emission line at λ = 293.912 nm, as showcased in Fig. 2. While the presence of Th and 237Np did not spectrally interfere with the Pu emission signal, small amounts of U and particularly Am hampered the reliable determination of Pu at this sensitive Pu emission wavelength (Fig. 2). The presence of Th gave rise to a peak overlapping with the Pu ICP-OES signal at λ = 319.844 nm, for example, also excluding this emission wavelength for Pu analysis (data not shown).
For part 2 of the interference study, four samples representing a variety of matrix compositions from in-house pyrochemical experiments were selected. Based on the emission spectra recorded for these samples at the 27 wavelengths above, the list of “useful” Pu emission wavelengths was further cut down to the seven wavelengths listed in Table 3. This decision was based on the fact that additional severe spectral interferences occurring at the majority of these 27 emission wavelengths – that were not obvious in the preliminary interferences check using standard solutions (see above) – hampered the reliable determination of Pu. Fig. 3 exemplifies these findings for a set of four samples originating from different steps of pyrochemical separation experiments as observed around one of the most popular Pu emission wavelength, i.e. λ = 300.057 nm. While the reason for the deteriorating peak shape for samples A18 and A32 have not been investigated in detail, the elevated (up to 20×) U concentration relative to their Pu amount might be the respective cause. For samples A16 and A30, in turn, Pu emission spectra revealed no spectral interferences around the peak maximum (Fig. 3). Even though being one of the most popular Pu emission lines, λ = 300.057 nm was not considered for quantification of the element in the investigated nuclear samples because of its incongruous spectral behaviour.
Sample | Emission wavelength, nm | ||||||
---|---|---|---|---|---|---|---|
297.251 | 299.409 | 299.649 | 390.721 | 439.645 | 449.378 | 453.615 | |
a Numbers in italics indicate results that indicate either diminished or elevated values compared to Pu concentrations obtained at λ = 297.251 nm, λ = 449.378 nm and λ = 453.615 nm. See text for details. | |||||||
1 | 0.45 ± 0.01 | 0.46 ± 0.01 | 0.45 ± 0.01 | 0.42 ± 0.01 | 0.43 ± 0.01 | 0.47 ± 0.01 | 0.46 ± 0.02 |
2 | 0.48 ± 0.01 | 0.48 ± 0.01 | 0.48 ± 0.01 | 0.44 ± 0.00 | 0.45 ± 0.01 | 0.48 ± 0.02 | 0.47 ± 0.02 |
3 | 0.43 ± 0.01 | 0.42 ± 0.01 | 0.43 ± 0.01 | 0.39 ± 0.01 | 0.42 ± 0.00 | 0.46 ± 0.01 | 0.42 ± 0.01 |
4 | 0.46 ± 0.01 | 0.48 ± 0.01 | 0.48 ± 0.01 | 0.44 ± 0.00 | 0.43 ± 0.01 | 0.47 ± 0.01 | 0.47 ± 0.02 |
5 | 0.49 ± 0.01 | 0.50 ± 0.00 | 0.49 ± 0.01 | 0.46 ± 0.00 | 0.46 ± 0.01 | 0.49 ± 0.01 | 0.50 ± 0.02 |
6 | 0.55 ± 0.01 | 0.80 ± 0.00 | 0.95 ± 0.03 | 0.68 ± 0.01 | 0.51 ± 0.01 | 0.54 ± 0.01 | 0.54 ± 0.01 |
7 | 0.47 ± 0.01 | 0.58 ± 0.00 | 0.63 ± 0.02 | 0.53 ± 0.01 | 0.46 ± 0.01 | 0.47 ± 0.02 | 0.48 ± 0.02 |
8 | 0.41 ± 0.01 | 0.44 ± 0.00 | 0.44 ± 0.01 | 0.41 ± 0.01 | 0.41 ± 0.01 | 0.40 ± 0.01 | 0.42 ± 0.01 |
9 | 0.51 ± 0.02 | 0.57 ± 0.01 | 0.59 ± 0.01 | 0.53 ± 0.01 | 0.51 ± 0.01 | 0.51 ± 0.01 | 0.53 ± 0.02 |
As a consequence of these experimental facts and to have a single robust set of identical acquisition parameters, only the three Pu emission wavelengths at λ = 297.251 nm, λ = 449.378 nm and λ = 453.615 nm (Table 3) were considered for a reliable quantification of Pu. The selection of these three Pu emission wavelengths for calculating Pu concentrations was additionally supported by the results obtained through replicate standard addition experiments (N = 12) carried out with all nine investigated nuclear samples (Table 4).
Sample | ICP-OES | SF-ICP-MS biasc | ||||
---|---|---|---|---|---|---|
Standard additiona | RSD, % | External calibrationb | RSD, % | Absolute, mg kg−1 | Relative, % | |
a Standard addition experiments using three aliquots of the actual sample and three spiked solutions for each sample aliquot on three consecutive days (N = 9). b External calibration experiments carried out on four consecutive days (N = 12). c Difference between the ICP-OES results (based on external calibration) in column 4 and comparative SF-ICP-MS data acquired in-house. Measurement uncertainties of SF-ICP-MS data are <1%. d Reported ICP-OES values represent the grand average (and the corresponding standard deviation) calculated from the average Pu concentrations obtained at the three Pu emission wavelengths λ = 297.251 nm, λ = 449.378 nm and λ = 453.615 nm (see Table 3). | ||||||
1 | 0.45 ± 0.02 | 3.9 | 0.46 ± 0.01 | 2.1 | −0.02 | −4.8 |
2 | 0.48 ± 0.01 | 1.8 | 0.47 ± 0.01 | 1.3 | −0.04 | −8.6 |
3 | 0.45 ± 0.02 | 3.9 | 0.43 ± 0.02 | 4.7 | −0.04 | −8.4 |
4 | 0.45 ± 0.00 | 0.7 | 0.47 ± 0.01 | 1.4 | 0.00 | 0.0 |
5 | 0.47 ± 0.00 | 0.9 | 0.50 ± 0.01 | 1.7 | −0.00 | −0.1 |
6 | 0.54 ± 0.01 | 2.2 | 0.54 ± 0.00 | 0.7 | 0.00 | 0.9 |
7 | 0.47 ± 0.00 | 0.4 | 0.47 ± 0.01 | 1.2 | 0.00 | 0.4 |
8 | 0.39 ± 0.01 | 1.3 | 0.41 ± 0.01 | 2.1 | −0.01 | −2.0 |
9 | 0.50 ± 0.01 | 1.8 | 0.52 ± 0.01 | 1.8 | 0.00 | 0.4 |
As no certified matrix-matched reference materials are available for quality assurance, the use of a complementary analysis technique whose detection is based on a different physical principle is desirable.14 To further underpin the accuracy of the selected evaluation regime of the ICP-OES measurements, all samples were also analysed using SF-ICP-MS. To this end, an aliquot from the identical solution that has been used for ICP-OES analysis was also subjected to SF-ICP-MS in-house. In general, ICP-MS results agreed well with ICP-OES data within 2% (Table 4). The results of the samples 1–3, however, did not satisfy these criteria as highlighted in Table 4. In particular, ICP-MS results were up to 8.6% lower than corresponding ICP-OES data. The high matrix load of these three samples (Np 2×; Nd 20×; Li 25×; K 100×; Cl 220×) and the inferior capability of the ICP-MS to compensate for the resulting signal suppression is the most likely explanation for this obvious discrepancy between the Pu concentrations established with both analytical techniques. Both the standard addition and the external calibration ICP-OES approach, in turn, yielded equivalent results suggesting sufficiently robust ICP-OES measurement conditions. It is worth mentioning that RSDs of replicate measurements also deteriorated slightly for these three samples (Table 4). Taken together, however, the developed ICP-OES approach generates accurate Pu concentration data for the above mentioned nuclear samples, avoiding a laborious and time-consuming separation of the analyte from the matrix.
Within this study also promising preliminary data for two different spent fuels were obtained. The current experiments were limited by the small volume (a few millilitres) of dissolved spent fuel that was available for testing. As such, high resolution spectra were recorded at the seven Pu emission wavelengths selected previously for the above described nuclear samples. Among them, the two Pu lines at λ = 299.649 nm (Fig. 4A) and λ = 453.615 nm (Fig. 4B) gave the impression to work best with this type of samples. Even though the metallic spent fuel contained ∼60% U, the corresponding emission spectra emerged straightforward (Fig. 4A and B). The spectra of the spent (Th, Pu)O2 fuel, in turn, suffered from the signals originating from the ∼83% Th that were present in the non-separated spent fuel solution (Fig. 4A and B). Due to the high resolution capabilities of the employed ICP-OES instrument, however, the Pu emission signals of both spent fuel solutions were well separated from potential spectral interferences. Further in-depth studies will be needed to give more evidence to the potential of HR-ICP-OES for the reliable determination of Pu in such spent fuel solutions.