Andrea
Bazzano
ab,
Kris
Latruwe
b,
Marco
Grotti
a and
Frank
Vanhaecke
*b
aDepartment of Chemistry and Industrial Chemistry, University of Genoa, Via Dodecaneso 31, 16146 Genoa, Italy
bDepartment of Analytical Chemistry, Ghent University, Krijgslaan 281-S12, 9000 Ghent, Belgium. E-mail: Frank.Vanhaecke@UGent.be
First published on 23rd February 2015
Reliable determination of Pb isotope ratios in Antarctic snow is challenging because of the low analyte concentration and the low volume of sample typically available. In this work, a combination of a total sample consumption introduction system (the torch-integrated sample introduction system, TISIS) with multi-collector ICP-mass spectrometry (MC-ICP-MS) was used for this purpose. With this instrumental setup, accurate and precise determination of Pb isotope ratios was possible at concentrations as low as 0.5 ng mL−1, while using 0.2 mL of solution only (total amount of Pb: 100 pg). At 10 ng mL−1, the repeatability for the 207Pb/206Pb ratio was 0.16‰ RSD. The concentration range was further extended downwards by using 100-fold analyte element preconcentration via freeze-drying of 20 g of snow. The Pb concentration in procedural blanks was 0.5 ± 0.3 pg g−1, enabling the determination of Pb isotope ratios in snow samples containing down to 5 pg g−1 of Pb. After development and validation, the procedure was applied to snow samples collected at Dome C (East Antarctic Plateau) on a monthly basis during the 2006 and 2010 campaigns. The method developed was able to reveal a seasonal variation in the Pb isotope ratios occurring during 2006 and strong inter-annual variation between the two campaigns.
Isotope dilution-thermal ionisation mass spectrometry (ID-TIMS) is capable of the simultaneous determination of the concentration of Pb and its isotopic composition and therefore appears to be the method of choice for the analysis of snow and ice samples of limited size.5,6 Multi-collector inductively coupled plasma-mass spectrometry (MC-ICP-MS) can be considered a strong competitor of TIMS for many applications,7 but so far, it has not been employed for the determination of Pb isotope ratios in snow samples from polar areas. For achieving an adequate isotope ratio precision, a relatively high target element concentration is required due to the low analyte transport efficiency from the sample solution to the detectors and the modest sensitivity of the Faraday cups used for ion monitoring. For many sample types, this problem can be circumvented via an analyte pre-concentration step, but this approach requires relatively large sample amounts, thus reducing the spatial/temporal resolution of the information provided in the specific case of Pb isotopic analysis of snow samples. A way to increase the instrumental sensitivity, while adhering to a low sample consumption, is the use of a so-called “total consumption” sample introduction system. Among these total consumption devices, the torch-integrated sample introduction system (TISIS)8,9 was already proven capable of efficient introduction of microsamples into an ICP source, in both atomic emission (ICP-AES)10,11 and mass spectrometry (ICP-MS).12,13 In its final design, the TISIS consists of a single-pass high-temperature evaporation chamber, with a lateral port to introduce a sheathing gas stream, in a location close to the aerosol production point. In this way, the aerosol solvent evaporation was enhanced and the impact of droplets against the cavity walls reduced.
Paredes et al. investigated the performance of a TISIS coupled with MC-ICP-MS for the determination of Pb14 and Sr15,16 isotope ratios. Accurate data were obtained at a very low sample uptake rate (5–30 μL min−1) and at a total sample consumption of less than 0.3 mL of solution. However, in these works, relatively high element concentrations were considered, resulting in a signal strength of about 1 V for the most abundant isotope monitored. The low concentration levels of Pb in snow from remote areas brings us far away from these ideal conditions.
In this work, snow samples were first submitted to analyte pre-concentration, based on sample volume reduction by freeze-drying.17 The solutions thus obtained were introduced into an MC-ICP-MS instrument for Pb isotopic analysis using a TISIS. This analytical protocol was optimised, characterised in terms of precision and accuracy at ultra-trace concentration level, and, finally, applied to the analysis of recent snow collected from Antarctica during two year-round expeditions.
All work, except for the pre-concentration procedure, was carried out in a metal-free class-10 clean lab facility at Ghent University. The pre-concentration procedure was accomplished in a class-100 laminar flow area at the University of Genoa.
A 10 ng mL−1 in-house isotopic standard was prepared by serial dilution of a commercially available 1 mg mL−1 stock solution (Inorganic Ventures, Christiansburg, VD, USA; lot G2-PB03044) and is further referred to as “A&MS-Pb”. Lead isotope ratios for this solution are 207Pb/206Pb = 0.90413 ± 0.00002, 208Pb/206Pb = 2.15331 ± 0.00003 and 208Pb/207Pb = 2.38167 ± 0.00003 (average ± 95% confidence interval; n = 15; determined at 100 ng Pb mL−1 using a conventional sample introduction system working at sample uptake rate of 200 μL min−1).
Instrumental parameters | |
---|---|
Sample uptake rate (μL min−1) | 20 |
Plasma gas flow rate (L min−1) | 15 |
Auxiliary gas flow rate (L min−1) | 0.90 |
Nebuliser gas flow rate (L min−1) | 0.72 |
Sheating gas flow rate (L min−1)a | 0.35 |
Temperature of the chamber (°C)a | 150 |
RF power (W) | 1200 |
Sampling cone | Ni; 1.1 mm Ø orifice |
Skimmer | Ni; H-type; 0.8 mm Ø orifice |
Lens settings | Optimised for maximum Pb signal intensity |
Resolution mode | Low |
Data acquisition parameters | |
---|---|
a Parameters of the TISIS chamber. b As a result of the low isotopic abundance of 204Pb and the spectral overlap from 204Hg, Pb isotope ratios including this Pb nuclide were not considered. | |
Scan type | Static; multi-collection |
Number of blocks | 7 |
Number of cycles/block | 6 |
Integration time (s) | 8.4 |
Magnet settling time (s) | 0 |
Idle time (s) | 3 |
Cup configurationb | L3: 202Hg; L2: 203Tl; L1: 204Pb, 204Hg; C: 205Tl; H1: 206Pb; H2: 207Pb; H3: 208Pb |
Resistivity of the pre-amplifier | C, L1, L2, L3: 1011 Ω; H1, H2, H3: 1012 Ω |
All sample and standard solutions were prepared in 0.05% ultrapure HNO3 doped with 10 ng mL−1 of Tl (NIST SRM 997) and the correction for the bias introduced by instrumental mass discrimination was performed offline using the method proposed by Woodhead et al.18 and further refined by Baxter et al.19 With this method, a regression line in ln–ln space is determined by measuring both the isotope ratio for Tl (NIST SRM 997) and that selected for Pb (NIST SRM 981) in a series of standard solutions. Based on the regression line thus obtained and the Tl isotope ratio measured for NIST SRM 997 Tl, admixed to the sample, the mass bias is corrected for.
The A&MS-Pb solution was analysed as a quality control sample every five samples, whereas a full procedural blank was analysed for every analytical batch.
The signal produced by the reagents blank (0.05% of ultrapure HNO3) was about 10−5 V for 208Pb and was stable during approximately six hours of analysis. After this period of time, signal spikes usually started to appear, necessitating the analysis to be stopped. In such an event, the TISIS chamber was thoroughly cleaned by immersion in a bath of hot HNO3 (about 10% v/v) for several hours. This phenomenon was already described by Paredes et al.14 and is likely due to the re-nebulisation of dry analyte particles from the nebuliser tip and the spray chamber walls.
The repeatability or internal precision (RSD for 42 successive measurements) was studied in a concentration range from 0.5 to 10 ng mL−1 of Pb (NIST SRM 981), equivalent to a total Pb amount from 0.1 to 2 ng. In Fig. 1, the RSD for the 207Pb/206Pb ratio is reported, showing values below 1.5‰ in the entire concentration range studied. At 10 ng mL−1 of Pb, the RSD was 0.16 ± 0.03‰ (average ± 95% confidence interval). This value is more than an order of magnitude better than that reported using the same sample introduction system, but using a quadrupole-based ICP-MS instrument with ammonia as a damping gas for smoothing the fluctuations of the ion beam intensity for Pb isotope ratio measurements.17
The intermediate or external precision was assessed by measuring the 207Pb/206Pb ratio in a solution with 10 ng mL−1 of Pb (NIST SRM 981) over a period of more than three months and was 0.11‰ (0.91465 ± 0.00005, n = 73) when expressed as twice the relative standard deviation of the ratios (not the standard error of the mean). This result is very similar to that obtained by Cocherie et al.,21 who reported a value of 0.1‰ using a similar MC-ICP-MS instrument and a sample uptake rate of 50 μL min−1. Similar values have also been reported by Kuritani et al.22 (0.13‰) and Amelin et al.23 (0.07‰) for isotopic analysis of absolute amounts of Pb of 1 and 2 ng by TIMS and double-spike TIMS, respectively. On the other hand, Makishima et al.24 reported a reproducibility down to 0.03‰ in the isotopic analysis of 1 ng of Pb at a sample uptake rate of 50 μL min−1 (0.2 mL total sample consumption) when using an Aridus II aerosol desolvation system and a Neptune MC-ICP-MS instrument equipped with a Jet-type skimmer cone (also termed X-type), characterised by enhanced transport efficiency. Finally, Thirlwall25 reported an external precision of 0.06–0.18 ‰ (2 SD) for isotopic analysis of ∼7 ng of Pb using double-spike MC-ICP-MS and the Aridus desolvation system.
When reducing the Pb concentration to 0.5 ng mL−1, the intermediate precision deteriorated to 0.31‰ (0.91438 ± 0.00028, n = 3) for the 207Pb/206Pb ratio.
The Pb concentration was estimated by comparing the Pb intensity observed for the blanks to that of the closest standards (NIST SRM 981 at 10 ng mL−1), resulting in a concentration of 0.5 ± 0.3 pg g−1 (or an absolute amount of 10 ± 6 pg Pb) and a Pb isotopic composition characterized by 207Pb/206Pb = 0.868 ± 0.008 and 208Pb/206Pb = 2.103 ± 0.020. Taking these results into account, the blank can induce variations in the isotopic composition ranging from 0.1 to 0.2% in snow samples with a Pb concentration of 5 pg g−1, the lowest Pb concentration encountered in the snow samples analyzed. A bias of this magnitude is relatively small compared to the natural variation in the isotopic composition of Pb as measured in Antarctic snow and ice.2,26
The relative contributions for the first three effects were estimated and propagated according to eqn (1), previously reported by Baxter et al.,19 and based on the assumption that the input quantities are not correlated
![]() | (1) |
The relative contribution from the blank was estimated as the difference in the combined uncertainty calculated before and after the subtraction of the intensity of the blank.
Finally, the expanded uncertainty was calculated by multiplying the combined uncertainty, uc(Rx,corr) with a coverage factor k = 2.
In Fig. 2, the different contributions to the combined uncertainty are visualized at concentrations of 5, 23 and 700 pg g−1, corresponding to the minimum, median and maximum concentration found in the Antarctic snow samples analysed (see next section). In the first case, the major contributions came from the instrumental precision (57%) and the blank (37%), whereas the contribution from the uncertainty on the NIST SRM 981 certified value accounted for 6% only. On the other hand, at higher concentrations, 55–77% of the uncertainty is attributed to the uncertainty of the certified value, 20–41% is stemming from the Pb isotope ratio measurement precision, while the contribution to the uncertainty coming from the blank is lower than 2%.
Despite the increased influence of the blank at low concentrations, it did not appear to be the main effect in the range of concentrations from 5 to 700 pg g−1. The contribution from the uncertainty on the certified value was found to become dominant at concentrations above 20 pg g−1. A better precision can be achieved when using the literature data obtained for NIST SRM 981 by double spike techniques24,25,28,29 instead of the NIST certified values, but it could be argued that the confidence in these data cannot be as high as in the certified values. At Pb concentrations below 20 pg g−1, the uncertainty associated to the Pb isotope ratios can be reduced by up to 40% by increasing the instrument sensitivity. Additionally, lowering the blank level could result in reducing the uncertainty by more than 20% for samples with Pb concentrations below 10 pg g−1.
The analytical procedure required 20 g of snow and about 10 minutes for each isotope ratio measurement. The Pb concentration was estimated by comparison of the Pb intensity observed for the sample to that of the closest standards (NIST SRM 981 at 10 ng mL−1) and it ranged from 5 to 707 pg g−1 (0.1 to 14.1 ng of Pb), with a median value of 23 pg g−1 (0.5 ng of Pb). The estimated Pb amount, the ratios obtained and their uncertainties (95% confidence interval) are reported in Table 2.
Date | Pb (ng) | 207Pb/206Pb | U | 208Pb/206Pb | U | 208Pb/207Pb | U |
---|---|---|---|---|---|---|---|
2006 | |||||||
Dec-05 | 2.5 | 0.8650 | 0.0002 | 2.1048 | 0.0006 | 2.4333 | 0.0008 |
Jan-06 | 0.2 | 0.8630 | 0.0010 | 2.0984 | 0.0016 | 2.4317 | 0.0014 |
Feb-06 | — | — | — | — | — | — | — |
Mar-06 | 0.5 | 0.8685 | 0.0002 | 2.1057 | 0.0005 | 2.4244 | 0.0004 |
Apr-06 | 0.5 | 0.8747 | 0.0002 | 2.1160 | 0.0005 | 2.4190 | 0.0004 |
May-06 | 0.8 | 0.8700 | 0.0002 | 2.1082 | 0.0005 | 2.4233 | 0.0004 |
Jun-06 | 0.5 | 0.8691 | 0.0002 | 2.1070 | 0.0005 | 2.4242 | 0.0004 |
Jul-06 | 0.4 | 0.8629 | 0.0002 | 2.1020 | 0.0008 | 2.4360 | 0.0007 |
Aug-06 | 0.2 | 0.8455 | 0.0011 | 2.0727 | 0.0016 | 2.4513 | 0.0017 |
Sep-06 | 0.3 | 0.8542 | 0.0004 | 2.0914 | 0.0007 | 2.4484 | 0.0006 |
Oct-06 | 1.1 | 0.8445 | 0.0002 | 2.0692 | 0.0005 | 2.4503 | 0.0005 |
Nov-06 | 0.1 | 0.8469 | 0.0009 | 2.0774 | 0.0014 | 2.4529 | 0.0013 |
Dec-06 | — | — | — | — | — | — | — |
![]() |
|||||||
2010 | |||||||
Jan-10 | 0.2 | 0.8616 | 0.0020 | 2.0971 | 0.0035 | 2.4340 | 0.0025 |
Feb-10 | 2.0 | 0.8661 | 0.0002 | 2.1046 | 0.0004 | 2.4299 | 0.0004 |
Mar-10 | 1.4 | 0.8626 | 0.0002 | 2.0979 | 0.0004 | 2.4321 | 0.0003 |
Apr-10 | 1.0 | 0.8650 | 0.0002 | 2.1041 | 0.0004 | 2.4324 | 0.0003 |
May-10 | 14 | 0.8701 | 0.0002 | 2.1108 | 0.0006 | 2.4259 | 0.0007 |
Jun-10 | 0.4 | 0.8596 | 0.0002 | 2.0939 | 0.0005 | 2.4358 | 0.0005 |
Jul-10 | 0.6 | 0.8596 | 0.0003 | 2.0945 | 0.0006 | 2.4367 | 0.0007 |
Aug-10 | 0.2 | 0.8606 | 0.0004 | 2.0969 | 0.0008 | 2.4364 | 0.0008 |
Sep-10 | 1.3 | 0.8640 | 0.0005 | 2.1006 | 0.0008 | 2.4312 | 0.0012 |
Oct-10 | 0.3 | 0.8591 | 0.0004 | 2.0956 | 0.0007 | 2.4393 | 0.0007 |
Nov-10 | 0.1 | 0.8499 | 0.0005 | 2.0799 | 0.0014 | 2.4473 | 0.0019 |
Dec-10 | 0.4 | 0.8608 | 0.0002 | 2.0974 | 0.0006 | 2.4365 | 0.0006 |
At 5 pg g−1 Pb concentration, the uncertainty on the 207Pb/206Pb ratio (n = 2) was assessed to be ∼0.08%. This result is almost a factor of 5 better than that calculated for the data reported by Vallelonga et al.1 (0.39%, n = 7) at 10 pg g−1 of Pb, using ID-TIMS and 10 g of snow.
In Fig. 3, the data are presented in a three-isotope plot. The Pb isotope ratios plotted on a straight line and are in agreement with data reported in literature.32 Particularly, data lie between the natural and the anthropogenic end-members reported by Van de Velde et al.,26 and the actual values are similar to those measured at the Hercules Névé for the period 1986–1994. The same authors pointed out that the natural end-member is the result of a mixing of volcanic and crustal inputs, whereas the anthropogenic end-member was governed by both South American and Australian industrial emissions.
![]() | ||
Fig. 3 Three-isotopes plot showing the results obtained for the Antarctic snow samples collected at Dome C during the 2006 (![]() ![]() ![]() |
Also Pb isotope ratios measured in the coastal area of East Antarctica33 are included in Fig. 3. These values are closer to the anthropogenic end-member than those measured on the Plateau, possibly indicating a larger anthropogenic input for the coastal area.
The time-series obtained for the years 2006 and 2010 are reported in Fig. 4. The first year showed a strong seasonal variation of the 208Pb/207Pb ratio, characterised by lower values from March to July 2006, and higher values from both December 2005 to January 2006, and from September to November 2006. Particularly, the ratios measured during spring showed values similar to those of the pelagic sediments and South Sandwich volcanic rock,26 thus suggesting a predominance of crustal Pb for this period. This result partly contradicts that reported by Burn-Nunes et al.,34 who found a maximum of both anthropogenic and mineral dust occurring during spring and autumn in the late 1880s. On the other hand, Planchon et al.32 reported evidence for close-to-radiogenic values for the 206Pb/207Pb ratio occurring in spring/summer during maxima of dust deposition, likely originating from South America.
![]() | ||
Fig. 4 Annual variation of the 208Pb/207Pb ratio for both the periods considered. Error bars represent 95% confidence intervals. |
On the contrary, a clear seasonal trend was not apparent in the Pb isotope ratios recorded for the 2010 samples, showing just minor fluctuations around the average value (208Pb/207Pb = 2.435 ± 0.005). The absence of a regular seasonal pattern was also reported by other authors, suggesting that the transport of aerosols and their deposition on the Antarctic ice sheet are complex and may vary from year to year.26,35 To improve the understanding of this variability and to highlight possible correlation with meteorological patterns, the analysis of more years and the use of additional chemical markers is clearly required.
The small sample consumption allowed us to combine the method with a simple analyte pre-concentration procedure consisting of freeze-drying of 20 g samples of snow, thus providing accurate ratios at Pb concentration levels down to 5 pg g−1. The mean concentration of the blanks (0.5 ± 0.3 pg g−1) was sufficiently low to obtain reliable Pb isotope ratios in this range of concentrations. Particularly, at 5 pg g−1, the uncertainty associated to the blank subtraction accounted for about 38% of the total uncertainty budget, thus still not representing the major contribution.
The analytical protocol developed was applied to the measurement of Pb isotope ratios in recent snow samples collected at Dome C (Antarctic Plateau) during the 2006 and 2010 campaigns. These are the first data of Pb isotope ratios measured in the Antarctic snow using MC-ICP-MS. The relative uncertainty of the ratios measured at 5 pg g−1 turned out to be competitive with that reported in literature for ID-TIMS. Moreover, the data provided some preliminary insights into the complex pattern of the transport of Pb towards Antarctica and its deposition on the ice sheet.
This journal is © The Royal Society of Chemistry 2015 |