Yan
Li
a,
Brett
Paull
a,
Marius N.
Müller‡
b and
Pavel N.
Nesterenko
*a
aAustralian Centre for Research on Separation Sciences (ACROSS), School of Physical Sciences, University of Tasmania, Hobart, Tasmania, Australia. E-mail: Pavel.Nesterenko@utas.edu.au
bInstitute for Marine and Antarctic Studies (IMAS), University of Tasmania, Private Bag 129, Hobart, Tasmania 7001, Australia
First published on 13th October 2014
A new high performance chelation ion chromatography method for the simultaneous determination of trace magnesium and strontium in various calcium carbonate samples was developed. Separations were performed on a monolithic silica column (Chromolith Si, 100 × 4.6 mm I.D.) chemically modified with hydroxyethyliminodiacetic acid functional groups. At a flow rate of 1.0 mL min−1, an eluent containing 80 mM NaCl and 20 mM picolinic acid at pH 5.30 was found to provide complete separation of Ni2+, Cu2+, Mg2+, Cd2+, Sr2+ and Ca2+ in 15 minutes, in matrices with 20000 fold excess of Ca2+. Two post-column reagents, o-cresolphthalein complexone (o-CPC) and ZnEDTA-PAR, were compared for post-column reaction based photometric detection at 570 and 490 nm, respectively. This method provides sensitive detection of Mg2+ (LOD 20 μg L−1 and 5 μg L−1 for o-CPC and ZnEDTA-PAR, respectively) and Sr2+ (LOD 200 μg L−1 and 39 μg L−1 for o-CPC and ZnEDTA-PAR, respectively). Using o-CPC, the linear range was from 0.5 to 24 mg L−1 for Mg2+ and from 1.0 to 32 mg L−1 for Sr2+. For ZnEDTA-PAR, the linear range was from 0.5 to 4 mg L−1 for Mg2+ and 0.1 to 32 mg L−1 for Sr2+. The method was applied to the analysis of a variety of calcium carbonate samples, including laboratory reagents, limestone NIST certified reference material, and the calcite based shells of marine microorganisms. The accuracy of the method was confirmed using inductively coupled plasma mass spectrometry.
Other methods have also been reported for the determination of traces of magnesium or strontium in calcite, such as flame photometry,10 neutron activation analysis (NAA)11 and atomic absorption spectroscopy (AAS).12,13 However, these methods also have their limitations. For example, the linear range of both atomic emission and absorption spectroscopy is rather narrow,13,14 and due to the high ionic strength and complexity of samples, analysis generally requires the removal of the matrix and/or dilution.
Thus, alternative accurate and practical methods for the determination of Mg and Sr in calcium carbonate based matrices are in significant demand. Ion chromatography (IC) has obvious potential for such samples,15 and has been widely used for the determination of Mg and Sr in various complex sample matrices, including seawater,16 plant-tissues17 and ceramic superconductors.18 However standard IC, using traditional cation-exchange columns, suffers considerably when applied to samples of high ionic strength or those containing very diverse ion ratios, such as magnesium in concentrated calcium carbonate. To overcome these issues, high performance chelation ion chromatography (HPCIC) was developed, and specifically presented as a solution to the determination of alkaline earth and transition metal ions in such complex samples. The separation mechanism of HPCIC includes the formation of kinetically labile complexes between chelating groups and the sample metal ions, so that their retention and separation are not affected by the ionic strength of the samples.19,20 Generally, chelating ligands are conjugate bases, resulting in their strong affinity to hydrogen ions. As a result of this affinity, pH can be used to control the separation selectivity and time.
The determination of alkaline earth metals in complex samples is one of the most effective applications of HPCIC. As far back as 1994, Jones et al.21 published the determination of Ba2+ and Sr2+ in calcium-containing matrices using HPCIC. However, this chromatographic system could not separate Mg2+ from the huge amount of Ca2+. In addition, since the technique was in its infancy at that time (and column limitations), the whole separation took an excessive 50 minutes. Recent studies reported much faster (5–15 min) determination of Be2+ in stream sediments,22 and Mg2+, Ca2+ and Sr2+ in seawater6,23,24 and in saturated brines for the chlor-alkali industry.19
The current work presents the results of determination of trace magnesium (and strontium) in calcium carbonate based matrices, such as certified argillaceous limestone, using HPCIC. The HPCIC methodology developed is based on hydroxyethyliminodiacetic acid (HEIDA) bonded monolithic silica columns and optimised post-column reaction (PCR) detection. The applicability of the method to the analysis of calcareous skeletons of marine microorganisms, such as coccolithophores (in particular Emiliania huxleyi), is also demonstrated, with its potential as a new method for inorganic magnesium determination in coccolithic calcite discussed.
A certified reference sample (1d Limestone (Argillaceous)) was purchased from the National Institute of Standards and Technology, NIST, USA. Full details of its composition can be found at the web-page http://www.nist.gov/srm.
The coccolithophore Emiliania huxleyi (isolated from the Southern Ocean, strain EHSO 5.14) was cultured in 0.20 μm filtered natural sea water, with a salinity of 35. Macro- and micro-nutrients were added in excess to ensure the supply of all necessary nutrients for growth. The medium was continuously bubbled with sterile air supplying carbon dioxide. Cultures received a constant daily illumination of 150 μmol quanta m−2 s−1 and were harvested at a density of 3000
000 cells per mL. Sample pellets were produced by centrifugation and subsequently dried and stored at 60 °C till further processing. After rinsing with water, drying in air and weighing the sample (40 mg) was suspended in 9 mL of deionised water. Titration with Cu2+ was carried out in suspension to replace the labile Mg (see the ESI†), and then the suspended sample was dissolved by addition of nitric acid (0.1 mol L−1) until the pH of the solution reached 2.0 to ensure complete dissolution of the calcite matrix.
Generally, the affinity to alkali metal cations is very low for most of the iminodiacetic acid (IDA) type chelating columns. Hence high alkali metal concentrations originating from the marine microorganism samples should not affect the resultant separation.19 Additionally, with non-complexing eluents, alkaline earth metals such as Mg2+, Ca2+ and Sr2+ are generally less retained on such chelating phases than the majority of transition metal ions, including Cu2+.23,25,26 Therefore herein, particular attention to the eluent conditions was required to pre-elute residual titrated Cu2+ ions.
Further control of selectivity can be obtained through the inclusion of complexing ligands within the eluents, to compete with the complexing ligand on the surface of the stationary phase.30 Recent work6 using HPCIC for the direct determination of Sr2+ in seawater demonstrated the separation of Mg2+, Sr2+ and Ca2+ using an eluent containing glycolic acid (complexing agent) and sodium chloride (ionic strength regulator). However, for the current application, greater resolution of the above metals was required, due to a diverse concentration ratio, and the additional resolution of the alkaline earth metal ions from Cu2+.
In 2008,30 Jones and Nesterenko investigated the effects of different complexing agents as eluent additives. As Cu2+ exhibits very strong retention on all IDA based columns, including HEIDA, the use of complexing agents is generally required. Dipicolinic acid and picolinic acid have been proven to provide appropriate affinity to Cu2+, compared to other complexing reagents,30,31 and so were applied in this application. The addition of picolinic acid to the eluent has little effect on the separation selectivity of Sr2+, Ca2+ and Mg2+, as the stability of complexes of these metals with picolinic acid is lower than that of complexes formed with the surface bonded HEIDA group. Alternatively, the β values of Sr2+, Ca2+ and Mg2+ with dipicolinic acid are higher than that with the surface bonded HEIDA group, and resultant selectivity reflects this.
Using a 0.15 mM dipicolinic acid eluent, with 6 mM HNO3 as a starting point, a poor separation of Mg2+ and Ca2+ was achieved, whilst Cu2+ failed to elute (Table 1). Obviously, a higher concentration of dipicolinic acid was required to elute Cu2+, however, as expected from the high β of Mg2+ and Ca2+ with dipicolinic acid, increased eluent concentrations of the acid reduced both the retention and resolution of Mg2+ and Ca2+, and was therefore unsuitable. A change of complexing agent to picolinic acid and an increase in concentration (4 mM) provided the elution of Cu2+, whilst the separation selectivity (α) between Mg2+ and Ca2+ remained relatively unaffected (Table 1).
Metal ions | Complexing additives | ||
---|---|---|---|
0.15 mM dipicolinic acid | 2 mM picolinic acid | 4 mM picolinic acid | |
a Note: “No peak” means strongly retained. | |||
Mg2+ | 4.69 | 4.56 | 4.57 |
Ca2+ | 5.13 | 4.99 | 5.00 |
Cu2+ | No peak | 13.72 | 6.39 |
α (Mg2+ and Ca2+) | 1.15 | 1.16 | 1.16 |
With the addition of sodium chloride (0.25 M) and a slight increase of pH (5.0), Mg2+, Ca2+ and Cu2+ could be separated with a retention order of Mg2+ < Ca2+ < Cu2+. However, the elution of Cu2+ after Ca2+ is not applicable to the analysis of the samples with massive excess of Ca2+. Therefore, a significant increase of picolinic acid concentration (20 mM) in the eluent was used to further reduce the retention of Cu2+, whilst a simultaneous reduction in eluent sodium chloride concentration (from 0.25 to 0.1 M) provided slightly increased retention of both Mg2+ and Ca2+. Under these conditions the desired elution order of Cu2+ < Mg2+ < (Cd2+, potential internal standard cation) < Sr2+ < Ca2+ was achieved. The optimisation was completed by attenuation of the eluent pH (5.0) and column temperature (40 °C) and resulted in the separation of five metals of interest in less than 10 minutes (see Fig. 1).
These conditions provided not only excellent resolution for minor Mg2+ and excessive Ca2+ peaks, but also a reasonable separation of Sr2+ and Ca2+ (αSr2+/Ca2+ = 1.15). However, for the latter pair the quantitative determination of strontium still could be a problem in the analysis of real samples (Fig. 2(a)).
However, according to work,6 the retention of Ca2+ on HEIDA functionalised silica is more sensitive to the changes in eluent pH as compared with Mg2+ and Sr2+. So, a slight increase in eluent pH to 5.3 improved dramatically the separation selectivity of Sr2+ and Ca2+ (αSr2+/Ca2+ = 1.30), while the retention of magnesium increased insignificantly (Fig. 2(b)).
Under the above final conditions the retention of a range of divalent and trivalent cations (Ni2+, Mn2+, Fe2+/3+, Zn2+, Co2+, Cu2+, Mg2+, Cd2+, Sr2+, Ca2+, and La3+) was also checked to ensure the absence of the possible interference for the determination of magnesium and strontium (see Table 2).
Metal ions | Ni2+ | Mn2+ | Fe2+/3+ | Zn2+ | Co2+ | Cu2+ | Mg2+ | Cd2+ | Sr2+ | Ca2+ | La3+ |
---|---|---|---|---|---|---|---|---|---|---|---|
a Note: “No peak” means strongly retained. | |||||||||||
Time (min) | 2.18 | 2.17 | 1.71/2.18 | 2.26 | 2.87 | 2.93 | 3.57 | 5.51 | 7.12 | 10.19 | No peak |
Herein, for quantitative analysis, the linear range test for alkaline earth metals was checked for both ZnEDTA-PAR (at pH 10.65) and o-CPC (at pH 11.0), with photometric detection at 570 and 490 nm, respectively. The flow rate of the eluent was 1.0 mL min−1 and PCR reagent was delivered at a flow rate of 0.44 mL min−1. With o-CPC, the linearity was confirmed over the concentration range of 1 to 29 mg L−1 for Mg2+, 1 to 32 mg L−1 for Sr2+, n = 6, with R2 values = 1.00 and 0.999, respectively. For ZnEDTA-PAR, the linear range was found to be only 0.05–4 mg L−1 for Mg2+ and 0.1 to 32 mg L−1 for Sr2+, (n = 6, R2 = 0.999 for both metals).
The limits of detection (LODs) for Mg2+ and Sr2+ with each PCR reagent were calculated using the signal to noise criteria S/N = 3. These were found to be 20 μg L−1 and 200 μg L−1 for o-CPC, and 5 μg L−1 and 39 μg L−1 for ZnEDTA-PAR, for Mg2+ and Sr2+, respectively.
In the following experiments ZnEDTA-PAR was used because of higher sensitivity for Mg2+ and Sr2+and the possibility of detection of various transition metal ions.
To evaluate the suitability of the developed method to the determination of Mg2+ and Sr2+ traces in the samples with high matrix calcium, three laboratory grade CaCO3 samples from different producers were also analysed. Fig. 4(a) and (b) show the chromatograms resulting from the injection of a 10 mg mL−1 solution prepared from CaCO3 obtained from AJAX and BDH, respectively. In both cases the sample concentrations were high enough to quantitate both Mg2+ and Sr2+ without loss of resolution between the peak of Sr2+ and the massive peak of Ca2+. However, a diluted solution (1 mg mL−1) was prepared for the analysis of the CaCO3 sample obtained from Strem Chemicals, Inc. because this contained a significant impurity of Mg2+ (see Fig. 4(c)), which otherwise will be at concentration beyond the linearity range. In the latter case, both the peaks for Mg2+ and Sr2+ impurities can be clearly identified. Using a standard addition method, the concentration of Mg2+ in this sample was found to be 5.05 ± 0.04 mg g−1, with RSD = 0.76% (n = 5). This again matched well with the results of HR-ICP-MS analysis, with which the concentration of Mg2+ was found to be 5.10 mg g−1, which corresponds to less than 1% difference between methods.
Finally, the developed method was applied to the determination of non-labile inorganic magnesium and the Sr/Ca ratio in coccolithic calcite skeletons (Emiliania huxleyi). For this purpose the sample of calcite was pre-titrated using Cu2+ to replace organically bound or physically adsorbed magnesium in coccolithic skeletons. Organic magnesium is mainly presented as chlorophyll and can be substituted in its molecules by copper. The content of inorganic or calcite entrapped magnesium was obtained by subtraction of the concentration of organic magnesium measured in the filtrate of the sample titrated with Cu2+ from total concentration of magnesium in coccolith. The latter concentration was measured for the sample completely dissolved in nitric acid. The chromatogram of the filtrate of the sample titrated with Cu2+ is shown in Fig. 5. The ratio of non-labile inorganic Mg to Ca in coccolithic calcite skeletons was determined to be 5.39 ± 0.23 mmol mol−1, with RSD = 4.3% (n = 5). This value is in good agreement with the literature data,2 where the Mg/Ca ratio varied in the range of 0.066–98.6 mmol mol−1.
Using the HPCIC method, the ratio of Sr to Ca in coccolithic calcite skeletons was found to be 2.79 ± 0.22 mmol mol−1, with RSD = 7.8% at n = 7. This is in good agreement with 2.73 ± 0.22 mmol mol−1 reported in the literature3 for this sample, when ICP-MS was used for the analysis.