aFaculty of Science and Engineering, School of Science and the Environment, Division of Chemistry and Environmental Science, Manchester Metropolitan University, Chester Street, Manchester M1 5GD, UK. E-mail: o.sutcliffe@mmu.ac.uk; c.banks@mmu.ac.uk; Fax: +44-(0)-161-247-6831; Tel: +44-(0)-161-247-1531, +44-(0)-161-247-1196
bSchool of Mechanical Engineering, University of Leeds, Leeds LS2 9JT, UK
First published on 27th July 2015
The global increase in the production and abuse of cathinone-derived New Psychoactive Substances (NPSs) has developed the requirement for rapid, selective and sensitive protocols for their separation and detection. Electrochemical sensing of these compounds has been demonstrated to be an effective method for the in-field detection of these substances, either in their pure form or in the presence of common adulterants, however, the technique is limited in its ability to discriminate between structurally related cathinone-derivatives (for example: (±)-4′-methylmethcathinone (4-MMC, 2a) and (±)-4′-methyl-N-ethylmethcathinone (4-MEC, 2b) when they are both present in a mixture. In this paper we demonstrate, for the first time, the combination of HPLC-UV with amperometric detection (HPLC-AD) for the qualitative and quantitative analysis of 4-MMC and 4-MEC using either a commercially available impinging jet (LC-FC-A) or custom-made iCell channel (LC-FC-B) flow-cell system incorporating embedded graphite screen-printed macroelectrodes. The protocol offers a cost-effective, reproducible and reliable sensor platform for the simultaneous HPLC-UV and amperometric detection of the target analytes. The two systems have similar limits of detection, in terms of amperometric detection [LC-FC-A: 14.66 μg mL−1 (2a) and 9.35 μg mL−1 (2b); LC-FC-B: 57.92 μg mL−1 (2a) and 26.91 μg mL−1 (2b)], to the previously reported oxidative electrochemical protocol [39.8 μg mL−1 (2a) and 84.2 μg mL−1 (2b)], for two synthetic cathinones, prevalent on the recreational drugs market. Though not as sensitive as standard HPLC-UV detection, both flow cells show a good agreement, between the quantitative electroanalytical data, thereby making them suitable for the detection and quantification of 4-MMC and 4-MEC, either in their pure form or within complex mixtures. Additionally, the simultaneous HPLC-UV and amperometric detection protocol detailed herein shows a marked improvement and advantage over previously reported electroanalytical methods, which were either unable to selectively discriminate between structurally related synthetic cathinones (e.g. 4-MMC and 4-MEC) or utilised harmful and restrictive materials in their design.
Electrochemistry is an advantageous analytical tool that is adaptable to an in-the-field device, in light of its portability, and can exhibit sensitivity and selectivity toward many target analytes.5,8–14 Our previous work on the development of robust electrochemical methods for the sensing of the synthetic cathinones, mephedrone (2a) and 4-MEC (2b), either in their pure form5 (LOD = 39.8–84.2 μg mL−1), using electroanalytical oxidation, or in the presence of common adulterants15 (i.e. products containing synthetic cathinones in combination with caffeine or benzocaine) (LOD = 11.6–11.8 μg mL−1), using direct electrochemical reduction, has the potential to be rapid, simple and cost-effective on-the-spot analytical screening tools with graphite screen-printed macroelectrodes (GSPEs). Krishnaiah et al. have also reported the electrochemical reduction of mephedrone (2a, LOD = 2.2 × 10−3 μg mL−1) using a dropping mercury electrode (DME).16 Though sensitive, the use of DME for in-field sensors is restrictive, as mercury is widely considered harmful and its use is banned in numerous countries.17–20
Though our initial work has indicated the application of GSPEs to the electrochemical detection of synthetic cathinones5 and demonstrated an excellent agreement between our electroanalytical protocol and that of high performance-liquid chromatography (HPLC) for street samples,15 the ability to simultaneously detect and quantify structurally related cathinones (for example: 4-MMC and 4-MEC within a single sample) eluded us due to co-incident voltammetric waves for the target analytes.5 The application of electrochemical detection (ED) in HPLC has been used for a range of areas including toxicology, therapeutic drug monitoring, drug metabolism and pharmacokinetics,21–23 however, the application of the technique towards the analysis of synthetic cathinones remains unexplored. Consequently in this paper, for the first time, the direct combination of HPLC with electrochemical detection for the qualitative and quantitative analysis of synthetic cathinones (4-MMC and 4-MEC) is reported using both a commercially available impinging jet flow cell (Dropsens, FC-A, Fig. 1a and b) and a custom-made iCell channel flow-cell (University of Leeds, FC-B, Fig. 1c and d)24 incorporating embedded GSPE macroelectrodes. GSPEs offer a cost-effective, reproducible and reliable sensor platform for the amperometric detection (AD) of the target analytes and the validated technique, high performance liquid chromatography-amperometric detection (HPLC-AD), has been shown to be suitable for the routine detection and quantification of the two synthetic cathinones either in their pure form, in the presence of common adulterants (e.g. caffeine) or simultaneously within blended street samples of the evolved “legal high” product, NRG-2.
The two flow cells used in this study were obtained from Metrohm UK, Runcorn, UK (impinging jet flow cell; product code: DRP-FLWCL-TEF-71306; 3.3 × 6.0 × 3.3 cm, flow chamber volume = 8 μL; denoted as FC-A, Fig. 1a and b) or the University of Leeds, UK (iCell channel flow cell; 4.5 × 4.5 × 4.0 cm, flow-chamber volume = 120 μL; denoted as FC-B, Fig. 1c and d). The iCell (FC-B) was fabricated as previously reported.24 Graphite screen-printed macroelectrodes (GSPEs) with a working electrode (3 mm diameter) were fabricated in-house with appropriate stencil designs using a DEK 248 screen-printing machine (DEK, Weymouth, UK).13 For the fabrication of the screen-printed sensors, firstly, a carbon-graphite ink formulation (Gwent Electronic Materials Ltd, UK; product code: C2000802P2) was screen-printed onto a polyester (Autostat, 250 μm thickness) flexible film (denoted throughout as GSPEs). This layer was cured in a fan oven (60 °C/30 min) and an Ag/AgCl reference electrode incorporated by screen-printing Ag/AgCl paste (Gwent Electronic Materials Ltd, UK; product code: C2040308D2) onto the polyester substrate. Finally, a dielectric paste (Gwent Electronic Materials Ltd, UK; product code: D2070423D5) was then printed onto the polyester substrate to cover the connections. After curing (60 °C/30 min) the screen-printed electrodes are ready to be used. Note that a new GSPE was utilized for each experiment performed, including during the “street sample” analysis study.
Khreit et al. have reported the application of HPLC and LC-MS techniques for the analysis of NRG-2 products using an ACE 3 C18 column (150 mm × 4.6 mm i.d., particle size: 3 μm) in combination with a mobile phase consisting of methanol:
10 mM ammonium formate (46
:
54% v/v).3 The validated HPLC method (which can detect 4-MMC (2a), 4-MEC (2b) and caffeine at levels of 0.02 μg mL−1) was further developed by Smith et al. to screen for these analytes in the presence of other synthetic cathinones and benzocaine based on new intelligence received from law enforcement agencies.15 A gradient elution program was employed to ensure both optimal detection of the analytes and a rapid analysis time. As gradient elution can affect the performance of electrochemical detectors,25 due to changes in the composition of the electrolyte/eluent employed, the original isocratic method of Khreit et al. was adapted to screen for caffeine, 4-MMC and 4-MEC, simultaneously via UV and amperometric detection by employing an mobile phase with a reduced percentage of organic modifier (30% v/v methanol) in combination with 10 mM ammonium acetate buffer containing a suitable electrolyte (100 mM KCl). The pH of the eluent was adjusted to 4.3 both to ensure the cathinones (2a: pKa = 8.692; 2b: pKa = 8.883) were fully ionised and, as the electrochemical responses of (2a) and (2b) have been shown to be sensitive to pH, to optimise their detection amperometrically.14,15
The two amperometric detectors used in this study were either of impinging jet flow cell (FC-A, Fig. 1a and b)25,26 or iCell channel flow cell (FC-B, Fig. 1c and d)24 design. The flow cells accommodated the GSPEs without any need for further modification. The optimum configuration (Fig. 1e) of the HPLC-AD system required the amperometric detector to be connected after the UV detector, via PTFE tubing (230 × 1.6 mm, i.d. 0.3 mm, internal volume: 16.25 μL). This configuration minimised system back-pressure and thereby reduced the leakages (from the flow-cells) observed when the amperometric detector precedes the UV detector. To distinguish the HPLC-AD system employing the impinging jet (FC-A) from the iCell channel (FC-B) flow-cells the two systems were denoted LC-FC-A and LC-FC-B respectively. Based on the previous reported validated HPLC-UV methods,2,3,5,15 employed in the separation of caffeine, (2a) and (2b) an ACE 3 C18 column was selected and the extra-column volumes associated with the system (e.g. connective tubing and/or flow cell internal volumes) reduced to minimise both eddy- and longitudinal-diffusional processes respectively – thereby optimising the efficiency of a chromatographic resolution between components within a mixture and ensuring the accuracy in their quantification. The anodic over-potential for 4-MMC (100 μg mL−1) in the mobile phase was determined using cyclic voltammetry (data not shown) with the peak maxima found to occur at +1.1 E V−1. Using the peak maxima, in conjunction with the optimised instrumental configuration, the potential required to achieve the optimal detector response (for 2a) was determined, for both LC-FC-A and LC-FC-B, by measuring the amperometric response (peak current, μA) as a function of anodic potential (E V−1), over the range +1.1 to +1.4 E V−1. The maximum response (0.25 μA ± 2.09%, n = 3) was observed, for (2a), at +1.4 E V−1 and this potential, which was also shown to be optimal for caffeine (1.36 μA ± 1.77%, n = 3) and (2b) (0.17 μA ± 3.15%, n = 3) was used herein for the detection of the target analytes. Due to the variation in internal chamber volumes of the two flow cells (FC-A = 8 μL vs. FC-B = 120 μL) a solution of (2a) (150 μg mL−1) was injected (n = 10) at different flow rates (0.8–1 mL min−1) and the amperometric response measured to determine the optimal linear velocity required for maximum amperometric response for each system. The system employing the impinging jet flow cell (LC-FC-A) gave the best response (+0.47 μA ± 6.58%, n = 10) at 0.8 mL min−1 with higher linear velocities giving a decreased response (circa 19%) with concomitant increase in back-pressures – due to the nature of the impinging jet design. The corresponding system employing the iCell channel flow cell (LC-FC-B) gave, under similar conditions, the best response (+0.028 μA ± 4.94%, n = 10) at a flow rate of 1.0 mL min−1. Using the optimised parameters (see Experimental section) the standard mixture (500 μg mL−1) of caffeine, (2a) and (2b) was rapidly separated on both systems, employing a reverse-phase column (with UV detection), eluting at 5.5 (or 4.3) (caffeine), 9.4 (or 7.5) (4-MMC) and 11.7 (or 9.3) minutes (4-MEC) at 0.8 (or 1) mL min−1 respectively (see Fig. 2a and c), exhibiting baseline resolution (Rs > 2) with slight peak fronting (asymmetry factor, As ∼ 0.53–0.64) in each case. The amperometric response (Fig. 2b and d) corresponding to this mixture shows a slight delay of 1.22 (LC-FC-A) and 0.98 s (LC-FC-B) respectively due to variation in flow rates and the connecting PTFE tubing between the HPLC-UV and AD detectors.
System suitability tests (resolution, column efficiency (N), height of a theoretical plate (H) and asymmetry factor) were used to verify that the system was performing adequately to ensure confidence in the analytical method and the results obtained. The developed method, for system LC-FC-A, shows that all of the standard system suitability parameters are within acceptable limits. The HPLC-UV validation parameters, for the LC-FC-A system, are summarized in Table 1.
System (detection) | LC-FC-A (HPLC-UV) | LC-FC-B (HPLC-UV) | ||||
---|---|---|---|---|---|---|
Flow rate | 0.8 mL min−1 (t0 = 2.01 min)a | 1 mL min−1 (t0 = 1.57 min)a | ||||
Analyte | Caffeine | 4-MMC (2a) | 4-MEC (2b) | Caffeine | 4-MMC (2a) | 4-MEC (2b) |
a Determined from the retention time of a solution of uracil (10 μg mL−1) eluting from the column. b Relative retention time (with respect to 4-MMC, 2a). c Relative response factor (with respect to 4-MMC, 2a). d N expressed in plates per m. e Limit of detection (based on the standard deviation of the response and the slope). f Limit of quantification (based on the standard deviation of the response and the slope). g y = 28.005x + 17.842. h y = 42.457x − 59.662. i y = 40.176x − 72.103. j y = 22.325x + 31.399. k y = 33.8x − 16.925. l y = 32.083x − 34.811. | ||||||
t R (min) | 5.5 | 9.4 | 11.7 | 4.3 | 7.5 | 9.3 |
RRTb | 0.56 | 1 | 1.24 | 0.57 | 1 | 1.24 |
RRFc | 0.8 | 1 | 1.1 | 0.8 | 1 | 1.1 |
Capacity factor (k′) | 1.7 | 3.7 | 4.8 | 1.7 | 3.7 | 4.9 |
N (plates) | 10![]() ![]() |
13![]() ![]() |
13![]() ![]() |
10![]() ![]() |
12![]() ![]() |
13![]() ![]() |
H (m) | 1.40 × 10−5 | 1.15 × 10−5 | 1.11 × 10−5 | 1.47 × 10−5 | 1.17 × 10−5 | 1.15 × 10−5 |
Resolution (Rs) | — | 14.3 | 5.9 | — | 14.2 | 5.98 |
Asymmetry factor (As) | 0.59 | 0.54 | 0.53 | 0.64 | 0.58 | 0.56 |
LODe (μg mL−1) | 2.03 | 2.50 | 2.99 | 1.79 | 1.95 | 2.41 |
LOQf (μg mL−1) | 6.14 | 7.58 | 9.05 | 5.43 | 5.90 | 7.29 |
Co-efficient of regression | 0.999g | 0.999h | 0.999i | 0.999j | 0.999k | 0.999l |
Precision (%RSD, n = 6) | ||||||
50 μg mL−1 | 0.06 | 0.06 | 0.05 | 0.03 | 0.06 | 0.03 |
100 μg mL−1 | 0.02 | 0.01 | 0.03 | 0.02 | 0.01 | 0.04 |
200 μg mL−1 | 0.04 | 0.02 | 0.03 | 0.03 | 0.03 | 0.03 |
300 μg mL−1 | 0.03 | 0.01 | 0.03 | 0.02 | 0.01 | 0.01 |
400 μg mL−1 | 0.06 | 0.05 | 0.05 | 0.01 | 0.02 | 0.06 |
500 μg mL−1 | 0.02 | 0.04 | 0.04 | 0.02 | 0.16 | 0.16 |
Validation of the LC-FC-A system, in terms of amperometric detection, was carried out using the calibration standards (50–500.0 μg mL−1) employed in the UV-detection validation study (vide supra) and indicated that (2a), (2b) and caffeine again demonstrated a linear response (r2 = 0.99) with good repeatability (%RSD = 0.32–1.00%; n = 6). The limits of detection for the analytes were determined to be within the range of 9.35–14.66 μg mL−1 and, though these are approximately 5× higher than UV-detection, agree with the previously reported levels (11.6–11.8 μg mL−1) reported by Smith et al.15 The limits of quantification were determined, from the standard deviation of the response and the slope, to be 37.06 (caffeine), 44.42 (4-MMC) and 28.33 μg mL−1 (4-MEC) respectively. The AD validation parameters, for the LC-FC-A system, are summarized in Table 2.
System (detection) | LC-FC-A (AD) | LC-FC-B (AD) | ||||
---|---|---|---|---|---|---|
Flow rate | 0.8 mL min−1 | 1 mL min−1 | ||||
Analyte | Caffeine | 4-MMC (2a) | 4-MEC (2b) | Caffeine | 4-MMC (2a) | 4-MEC (2b) |
n.d. = not determined.a Relative retention time (with respect to 4-MMC, 2a).b limit of detection (based on the standard deviation of the response and the slope).c Limit of quantification (based on the standard deviation of the response and the slope).d y = 0.0105x + 0.2039.e y = 0.0025x − 0.0211.f y = 0.0011x + 0.0082.g y = 0.0013x + 0.0563.h y = 0.0003x + 0.0053.i y = 0.00009x + 0.026. | ||||||
t R (min) | 5.52 | 9.42 | 11.72 | 4.32 | 7.52 | 9.32 |
RRTa (min) | 0.59 | 1 | 1.24 | 0.57 | 1 | 1.24 |
LODb,e (μg mL−1) | 12.23 | 14.66 | 9.35 | 23.38 | 57.92 | 26.91 |
LOQc,f (μg mL−1) | 37.06 | 44.42 | 28.33 | 70.86 | 175.51 | 81.54 |
Co-efficient of regression | 0.995d | 0.993e | 0.997f | 0.992g | 0.953h | 0.990i |
Precision (%RSD, n = 6) | ||||||
50 μg mL−1 | 0.58 | 0.55 | 0.74 | n.d. | n.d. | n.d. |
100 μg mL−1 | 0.32 | 0.87 | 0.81 | n.d. | n.d. | n.d. |
200 μg mL−1 | 0.53 | 0.91 | 1.00 | 0.07 | 0.19 | 0.74 |
300 μg mL−1 | 0.53 | 0.81 | 0.80 | 0.32 | 0.45 | 0.68 |
400 μg mL−1 | 0.71 | 0.91 | 1.00 | 0.15 | 0.55 | 0.45 |
500 μg mL−1 | 0.57 | 0.87 | 0.48 | 0.10 | 0.87 | 0.38 |
The corresponding liquid chromatography-amperometric detection system, [LC-FC-B], employing the iCell channel flow cell (FC-B) was also validated in terms of UV-detection (Table 1) after increasing the flow rate to 1 mL min−1 to ensure a satisfactory elution time of the three target analytes (Fig. 2c). As the HPLC-UV detection system was identical to that employed with the impinging jet flow cell, the repeatability, specificity, linear response, limits of detection, limits of quantification and the system suitability tests for the three analytes, showed no significant differences over the 50–500.0 μg mL−1 range to the system employing the impinging jet flow cell (LC-FC-A).
Interestingly, in terms of the amperometric detection, the modified system incorporating the iCell channel flow cell (LC-FC-B) demonstrated better repeatability (RSD = 0.07–0.87%; n = 6), than LC-FC-A, for the three analytes, however, the linear response was significantly reduced (r2 = 0.95–0.99) over the 200.0–500.0 μg mL−1 range. The limits of detection for these components were confirmed as being in the range of 23.38–57.92 μg mL−1 and the limits of quantification were established to be 70.86 (caffeine), 175.51 (2a) and 81.54 μg mL−1 (2b) respectively, which less sensitive than those obtained for impinging jet flow-cell. It is rationalised that in the case of the iCell channel flow cell, the larger chamber volume (120 μL), is increasing sample dispersion, diluting the analytes, and thereby reducing the sensitivity of the GSPE sensor platform via mass transfer/diffusion to the electrode surface.27–29 The amperometric validation parameters, for the modified system (LC-FC-B), are summarized in Table 2.
LC-MS (n = 3) | LC-FC-A (HPLC-UV) (n = 3) | LC-FC-B (HPLC-UV) (n = 3) | |
---|---|---|---|
a Sample analysed during this study using the method reported by Khreit et al.3 | |||
NRG-2-A | t R = 5.34 min [m/z = 192.2 [M + H]+, 4-MEC]3 | t R = 11.7 min [24.03% w/w ± 0.03, 4-MEC] | t R = 9.3 min [24.01% w/w ± 0.05, 4-MEC] |
NRG-2-B | t R = 4.48 min [m/z = 178.1 [M + H]+, 4-MMC]3 | t R = 9.4 min [49.24% w/w ± 0.03, 4-MMC] | t R = 7.5 min [48.18% w/w ± 0.02, 4-MMC] |
NRG-2-C | t R = 2.57 min [major, m/z = 195.1 [M + H]+, caffeine]; 5.34 min [minor, m/z = 192.2 [M + H]+, 4-MEC]13 | t R = 5.5 min [major, 76.19% w/w ± 0.22, caffeine]; 11.7 min [minor, 23.58% w/w ± 0.49, 4-MEC] | t R = 4.3 min [major, 74.83% w/w ± 0.16, caffeine]; 9.3 min [minor, 25.81% w/w ± 0.23, 4-MEC] |
NRG-2-D | t R = 2.57 min [major, m/z = 195.1 [M + H]+, caffeine]; 4.48 min [minor, m/z = 178.1 [M + H]+, 4-MMC]13 | t R = 5.5 min [major, 83.04% w/w ± 0.03, caffeine]; 9.4 min [minor, 15.64% w/w ± 0.45, 4-MMC] | t R = 4.3 min [major, 82.93% w/w ± 0.35, caffeine]; 7.5 min [minor, 16.58% w/w ± 1.13, 4-MMC] |
NRG-2-E | t R = 2.57 min [m/z = 195.1 [M + H]+, caffeine; 4.48 min [m/z = 178.1 [M + H]+, 4-MMC]; 5.34 min [m/z = 192.2 [M + H]+, 4-MEC]a | t R = 5.5 min [36.55% w/w ± 0.08, caffeine; 9.4 min [15.64% w/w ± 0.46, 4-MMC]; 11.7 min [24.03% w/w ± 0.03, 4-MEC] | t R = 4.3 min [34.09% w/w ± 0.77, caffeine; 7.5 min [16.71% w/w ± 0.05, 4-MMC]; 9.3 min [25.84% w/w ± 0.01, 4-MEC] |
With substantial evidence supporting an electroanalytical oxidation approach for detecting various substituted cathinones in street samples the viability of the proposed protocol was tested. The NRG-2 samples were reanalysed (in triplicate) using the validated LC-AD method, using both flow cells, at a concentration of 500 μg mL−1. The HPLC-UV results (Table 3), obtained using the system employing the impinging jet flow cell (LC-FC-A), confirmed that two of the samples contained only synthetic cathinones (NRG-2-A: 24.03 ± 0.03% w/w 4-MEC and NRG-2-B: 49.24 ± 0.03% w/w 4-MMC); two of the samples contained predominantly caffeine (circa 80% w/w) in combination with 4-MMC or 4-MEC (NRG-2-C: 76.19 ± 0.22% w/w caffeine, 23.58 ± 0.49% w/w 4-MEC and NRG-2-D: 83.04 ± 0.03% w/w caffeine, 15.64 ± 0.45% w/w 4-MMC) and one sample contained a complex mixture of the three analytes (NRG-2-E: 36.55 ± 0.08% w/w caffeine, 15.64 ± 0.46% w/w 4-MMC, 24.03 ± 0.03% w/w 4-MEC). These observations are in agreement with the information reported by Khreit et al., Brandt et al. and Smith et al. who noted that many second-generation “legal high” products contained increased levels of commonly used diluents and adulterants.3,4,15
The qualitative results, obtained from the amperometric detector (LC-FC-A), also confirmed the constitution of the five NRG-2 samples and comparison of two methods (HPLC-UV vs. AD, Table 4) indicated that in samples containing caffeine (NRG-2-C, NRG-2-D and NRG-2-E) the two methods were comparable in terms of their ability to quantify the levels of caffeine present (NRG-2-C, HPLC-UV: 76.19 ± 0.22% w/w vs. AD: 78.26 ± 0.68% w/w; NRG-2-D, HPLC-UV: 83.04 ± 0.03% w/w vs. AD: 80.54 ± 2.06% w/w; NRG-2-E, HPLC-UV: 36.55 ± 0.08% w/w vs. AD: 42.22 ± 1.43% w/w).
System | LC-FC-A | LC-FC-B | ||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
Flow rate | 0.8 mL min−1 | 1 mL min−1 | ||||||||||
Detection | HPLC-UV (% w/w) (n = 3) | Amperometric (AD) (% w/w) (n = 3) | HPLC-UV (% w/w) (n = 3) | Amperometric (AD) (% w/w) (n = 3) | ||||||||
Sample | Caffeine | 4-MMC (2a) | 4-MEC (2b) | Caffeine | 4-MMC (2a) | 4-MEC (2b) | Caffeine | 4-MMC (2a) | 4-MEC (2b) | Caffeine | 4-MMC (2a) | 4-MEC (2b) |
n.d. = not detected. | ||||||||||||
t R (min) | 5.5 | 9.4 | 11.7 | 5.52 | 9.42 | 11.72 | 4.3 | 7.5 | 9.3 | 4.32 | 7.52 | 9.32 |
NRG-2-A | n.d. | n.d. | 24.03(±0.03) | n.d. | n.d. | 54.39 (±1.24) | n.d. | n.d. | 24.01 (±0.05) | n.d. | n.d. | 65.07 (±1.21) |
NRG-2-B | n.d. | 49.24 (±0.03) | n.d. | n.d. | 60.80 (±0.57) | n.d. | n.d. | 48.18 (±0.02) | n.d. | n.d. | 75.28 (±1.71) | n.d. |
NRG-2-C | 76.19 (±0.22) | n.d. | 23.58 (±0.49) | 78.26 (±0.68) | n.d. | 20.69 (±1.72) | 74.83 (±0.16) | n.d. | 25.81 (±0.23) | 80.35 (±0.99) | n.d. | 18.77 (±2.45) |
NRG-2-D | 83.04 (±0.03) | 15.64 (±0.45) | n.d. | 80.54 (±2.06) | 18.95 (±2.96) | n.d. | 82.93 (±0.35) | 16.58 (±1.13) | n.d. | 85.38 (±0.48) | 8.82 (±2.21) | n.d. |
NRG-2-E | 36.55 (±0.08) | 15.64 (±0.46) | 24.03 (±0.02) | 42.22 (±1.43) | 8.56 (±3.30) | 54.40 (±1.19) | 34.09 (±0.77) | 16.71 (±0.05) | 25.84 (±0.01) | 36.42 (±1.14) | 27.53 (±0.32) | 44.01 (±1.59) |
Samples containing only 4-MMC and 4-MEC (NRG-2-A and NRG-2-B) showed a significant over estimation of the quantities of the synthetic cathinones present in comparison to the HPLC-UV detection (NRG-2-A, HPLC-UV: 24.03 ± 0.03% w/w 4-MEC vs. AD: 54.39 ± 1.24% w/w 4-MEC; NRG-2-B, HPLC-UV: 49.24 ± 0.03% w/w 4-MMC vs. AD: 60.80 ± 0.57% w/w 4-MMC) and though a new GSPE was utilised during each sample analysis, the loss in analytical performance, in terms of the inconsistency, maybe due to adsorption of the synthetic cathinones onto the surface of the GSPE during the timescale of the analysis.
Yao et al. has recently reported that purine bases have the ability to adsorb competitively onto the surface of carbon electrodes.30 As caffeine, which is structurally similar to purine, is present in the NRG-2-C and NRG-2-D samples, it is feasible that it may compete, with the synthetic cathinones, for sites on the surface of the GSPE. This may explain the increase observed in the levels of caffeine, and more consistent quantification of 4-MMC and 4-MEC, present when HPLC-UV and amperometric detection, of these samples, are contrasted (NRG-2-C, HPLC-UV: 76.19 ± 0.22% w/w caffeine, 23.58 ± 0.22% w/w 4-MEC vs. AD: 78.26 ± 0.68% w/w caffeine, 20.69 ± 1.72% w/w 4-MEC; NRG-2-D, HPLC-UV: 83.04 ± 0.03% w/w caffeine, 15.64 ± 0.45% w/w 4-MMC vs. AD: 80.54 ± 2.06% w/w caffeine 18.95 ± 2.96% w/w 4-MEC). Though a good correlation between the HPLC-UV and amperometric methods was observed for the fifth sample, NRG-2-E, in terms of their ability to quantify the levels of caffeine. Interestingly, by contrasting the two detection methods, the levels of 4-MEC (2b) were overestimated and 4-MMC (2a) was underestimated amperometrically in the case of NRG-2-E (NRG-2-E, HPLC-UV: 36.55 ± 0.08% w/w caffeine, 15.64 ± 0.46% w/w 4-MMC, 24.03 ± 0.02% w/w 4-MEC vs. AD: 42.22 ± 1.43% w/w caffeine, 8.56 ± 3.30% w/w 4-MMC, 54.40 ± 1.19% w/w 4-MEC). These observations are difficult to rationalise as simple analyte adsorption on to the GSPE surface as the efficient chromatographic separation of the target analytes, before their detection, should intrinsically contribute to avoid competitive adsorption. Additionally within the system, there may not be enough time for one analyte (or other adsorbates) to occupy all the adsorption sites on the electrode in a flowing system, and the solution flowing may effectively make the analytes desorbed before detection of other analytes.
The NRG-2 samples were also analysed using the modified liquid chromatography-amperometric detection system, [LC-FC-B] at a concentration of 500 μg mL−1. The HPLC-UV results, obtained using the system employing the iCell channel flow cell (LC-FC-B), showed no significant differences and were consistent with those observed on the LC-FC-A (Table 3) utilising the impinging jet flow cell. The amperometric detection results follow a similar trend to those observed with LC-FC-A, however, in the case the sample containing caffeine, (2a) and (2b) (NRG-2-E), the results show an over estimation of the synthetic cathinones. This variation in the results may again be due to the adsorption of the analytes onto the surface of the GSPE. However, as the iCell channel flow cell (FC-B) has a larger chamber volume, the effect of reduced mass transfer/diffusion to the electrode surface, due to sample dispersion may also be a factor in reducing the sensitivity of the GSPE sensor platform.
In summary, though our LC-AD system has lower sensitivity than simple HPLC-UV,5 this work demonstrates an improvement over our previous work, which indicated that there was no electrochemical selectivity of the electrochemical detection of 4-MMC and 4-MEC. Efficient chromatographic separation of these analytes, before their detection, allows us to now rapidly separate, discriminate between and quantify, two structurally related cathinones within a complex street sample mixtures (Table 4) indicating that the proposed HPLC-AD protocol can be considered suitable for the detection and quantification of the two synthetic cathinones either in their pure form, in the presence of common adulterants (e.g. caffeine) or simultaneously within blended street samples of the evolved “legal high” product, NRG-2. We concede that the observed amperometric limits of detection (for the electrochemical oxidation of 4-MMC and 4-MEC) reported herein are lower to the values reported in our previous work5 (2a: 39.8 μg mL−1 and 2b: 84.2 μg mL−1), however, this is sufficient for use in the field opposed to the values reported by Krishnaiah et al.16 who utilised a dropping mercury electrode (DME) which is not suitable for use in the field and banned in many countries. Notwithstanding the loss in analytical performance, when compared to HPLC-UV detection, this proof-of-concept study is still adequate for quantifying the synthetic cathinones present within seized samples and work to (i) ascertain the physical processes at the electrode surface; (ii) optimise of the shape of the flow cell to yield greater sensitivity and (iii) employ microfluidics to develop a miniaturised detection system which can be employed in the field testing of new psychoactive substances is currently underway.