Niejun Wang‡
,
Sifeng Mao‡,
Wu Liu,
Jing Wu,
Haifang Li* and
Jin-Ming Lin*
Beijing Key Laboratory of Microanalytical Methods and Instrumentation, Department of Chemistry, Tsinghua University, Beijing 100084, China. E-mail: jmlin@mail.tsinghua.edu.cn; lihaifang@mail.tsinghua.edu.cn; Fax: +86 10 62792343; Tel: +86 10 62792343
First published on 18th February 2014
In this work, we developed a chip-based continuously flowing system for online liquid–liquid extraction by using monodisperse droplets, and the droplets containing target compounds after extraction were collected on the microchip. Monodisperse droplets of extracting reagent were on-chip generated and dispersed in continuously flowing sample solution, and then liquid–liquid extraction happened when organic droplets moved forward with sample solution flow. After liquid–liquid extraction, guiding tracks were designed in our system to capture the organic droplets, which allowed the aqueous solution flowing over at the same time. Structures of the microchip were optimized to achieve generation and collection of monodisperse droplets more efficiently. To verify the feasibility of droplet based liquid–liquid extraction on the established platform, we used aqueous solution containing 10 μM fluorescein sodium as sample solution and butanol as the extracting reagent to perform the whole procedures. Fluorescein sodium was successfully extracted by the butanol droplets, and the butanol droplets were successfully collected. The results demonstrated that the developed microfluidic device was a useful tool for monodisperse droplet based liquid–liquid extraction. As all the procedures including droplet generation, extraction and collection were performed on one chip, the established platform had the potential of time-resolved monitoring.
As one of the feasible methods to purify samples on microchip, solid phase extraction (SPE) need immobilization of extracting materials in microchannels,21–23 which make the fabrication of microchips much more complicated. Meanwhile, multiple procedures including loading, washing and elution are needed during the process of SPE that impedes the real-time or online pretreatment of sample. Compare to SPE, liquid–liquid extraction based on microchip shows excellent advantages.24,25 Simple introduction of extracting reagent into microchip, which is crucial to achieve the liquid–liquid extraction, can avoid complicated procedures as SPE does and make liquid–liquid extraction well compatible to the flowing system of microfluidic device.
To implement liquid–liquid extraction in microfluidic device, both laminar flow10 and droplets26–28 can be utilized and extraction will take place near the liquid–liquid interface. Laminar of sample solution and extracting reagent is usually performed in some microfluidic systems to achieve extraction, while droplets of extracting reagent are introduced into sample solutions in other systems. When similar conditions are applied, droplet-based extraction is more efficient than the stratified one.29 Several new methods implementing liquid–liquid extraction use captured droplets28,30 which cannot be released and renewed, so these methods fail to function well in a long-time monitoring or a time-resolved detection. Therefore, developing new methods that are suitable for online time-resolved monitoring will make a great contribution to the development and application of microfluidic devices.
For the successful integration of liquid–liquid extraction onto microfluidic device, sufficient control of the fluid31 is always necessary. In order to develop some novel methods for online time-resolved monitoring, it is urgent to design ingenious methods for the control of fluid. Generation32,33 and control of droplets,34–36 two of the important fundamental subjects on microfluidic device, are widely studied and utilized. In the reported methods, T-junction32,37 and flow focusing33,38 are the most common ways to generate droplets on microchip, and these droplets can be captured or manipulated. Besides, plug-liked droplets26,27,39 and monodisperse droplets40 can also be generated on microchip. Comparing to plug-like droplets in microchannels, monodisperse droplets increase the specific surface area, which enhances the exchange of substance between droplets and continuous phase. So monodisperse droplets show better properties than the other types of droplets. But the collection and recycle of monodisperse droplets remain to be a challenge because monodisperse droplets usually float in continuous phase randomly, which limit the application of monodisperse droplets in extraction in a continuous flowing system. A suitable method to collection these monodisperse droplets will promote the application of monodisperse droplets in microfluidic devices.
In this paper, we developed a novel method for online liquid–liquid extraction on microfluidic device using monodisperse droplets and these monodisperse droplets were collected from a continuously flowing system successfully using guiding tracks. Monodisperse droplets of extracting reagent were firstly generated on microchip and dispersed in the continuously flowing sample solution to simulate the emulsification in traditional liquid–liquid extraction. At the interface between the monodisperse droplets and the sample flow, fast extraction was achieved while the droplets moved forward with the sample solution. Then these droplets were collected by guiding tracks and detected. To confirm the effectiveness of online rapid extraction in this system, fluorescent dye was chosen as model molecules because of the excellent transparency of the microchips which made fluorescence microscope to be a convenient tool to record time-resolved and spatial-resolved signals. Moreover, those collected droplets could be guided to downstream devices for further detection and analysis. As the monodisperse droplets were continuously generated and collected, the extracting reagent kept renewing as time passed, which made this method potential for online time-resolved monitoring.
According to the proposed function of this platform, three functional sections were designed to perform monodisperse droplets liquid–liquid extraction (Fig. 1): droplets generation part, liquid–liquid extraction part and droplets collection part. In droplets generation part (Fig. 1a), T-junction was designed to generate droplets, where the side channel (50 μm in height and 80 μm in width) of T-junction was vertical to the main channel and thinner than it (100 μm in height). In the meanwhile, the width of main channel decreased to a half of the original one at T-junction to increase flow rate and enhance shear force here. The extracting reagent was injected into the side channel, while the aqueous sample solution was injected into the main channel (1000 μm in width). Then monodisperse droplets would rapidly generate and disperse in the aqueous sample solution flow. In liquid–liquid extraction part (Fig. 1b), a long main channel with suitable length was designed to provide enough time for liquid–liquid extraction. In our experiments, a straight channel was chosen to simplify the experiments for confirming the liquid–liquid extraction in our system. In droplets collection part (Fig. 1c), guiding tracks (functioned as a groove with a depth of 30 μm and width of 80 μm) were designed on the top surface of the main channel. The guiding tracks, which had an included angle of 15 degree to the flow direction, covered the width direction of the main channel. At the end of the guiding tracks (Fig. 1d), a connecting channel (100 μm in width, 100 μm in length and 100 μm in height) between the main channel and the collecting channel was designed to collect the absorbing reagent. An actual image of the microchip was shown in Fig. 1e and the image of the device placed under microscope was shown in Fig. 1f.
By adjusting the flow rate of sample solution and extracting reagent, both the droplets generating rate and the droplets size could be well controlled. In our experiments, appropriate rate of droplet generation as well as the size of droplets was optimized to meet the requirements for performing successful extraction and collection of droplets. The results showed that the droplets would not be effectively trapped in the guiding track when their size was too small. However, it would be difficult for them to pass through the narrow connecting channel when the droplets size was too big. As an influence of the droplets generating speed, it was obvious that the droplets suspended in the sample flow would be lacked when the speed is too low, resulting in low extraction efficiency. In the meantime, moving velocity of the droplets in the main channel would become slower when they entered the guiding tracks. Thus, a latter droplet would catch up with the former one when the droplets generating speed is too high, which caused those two droplets to fuse to be one bigger droplet. Under this situation, it might be difficult for the bigger droplet to pass through the connecting channel and enter the collecting channel. In order to simplify the optimization process, we selected the flow rate of the sample flow in the main channel as 5 μL min−1. Then, we used different flow rates (0.01, 0.05, 0.1, 0.5 and 1.0 μL min−1) (data at high flow rates were not shown) of the extracting reagent in the side channel to investigate the influence of the flow rate of the extracting reagent on the droplets generating speed. The results showed that droplets generating speed became higher and the diameter of droplets kept unchanged (about 40 μm) when the flow rate of the extracting reagent increased from 0.01 to 0.10 μL min−1 (Fig. 2). When flow rates of extracting reagent further increased (for example, 0.5 or 1.0 μL min−1), bigger droplets (diameter was about 100 μm) would generate at a lower generating speed (as shown in droplets collection part). In the experiments with appropriate control, the droplets generated at the interchange of the main channel and the side channel could randomly enter into the sample flow at different directions, for example direction a, b and c (Fig. 2a), which detailed in Video 1.† As a result, the droplets suspended well in the sample flow well, which would raise the extraction efficiency. For further use of this microchip, speed of droplets generation and size of droplets should be adjusted according to the detailed requirement of the experiments.
Appropriate width and tilt angle of connecting channel was important for droplets collection, because they would influence the flow velocity in connecting channel. Very low flow velocity in connecting channel might lead to the failure of droplets collection, while very high flow velocity in connecting channel would allow sample solution in main channel to pass through the connecting channel and entered the collecting channel. Then a software COMSOL multiphysics was used to simulate the velocity field near the connecting channel to achieve better collection of monodisperse droplets that came from the guiding tracks. The length and width of connecting channel and its tilt angle with the main channel were optimized using this software. The results showed that the length and width of connecting channel had a greater influence on the function of connection than the tilt angle did (Fig. 3). The flow velocity in the connecting channel became lower when the length of the connecting channel obviously increased (Fig. 3a), while it became higher when the width of the connecting channel increased (Fig. 3b). For the designs with different tilt angles between the connecting channel and the main channel, only the ones with tilt angles smaller than 45 degree or bigger than 135 degree had significantly different results (Fig. 3c).
According to the simulation above, microchips with different designs of connecting channel were fabricated and tested in the experiments for droplets collection, and the results of actual experiments were consistent with the simulation results. Based on the results, connecting channel with a width of 100 μm, a length of 100 μm and a tilt angle of 150 degree was chosen for further usage in following experiments. Under the optimized conditions, we investigated droplets capture and collection using water with red dye as the sample solution and butanol as the extracting reagent. When the monodisperse droplets reached the guiding tracks, they were successfully captured and moved along the guiding tracks (Fig. 4a–d). Finally, they passed through the connecting channels, entered the collecting channel and were successfully collected (Fig. 4e and f). More details about the capture and collection of droplets were shown in Video 2.†
To confirm the feasibility of online liquid–liquid extraction on the designed microchip, we studied the extraction of fluorescein sodium in water by butanol droplets. A real sample (water containing 10 μM fluorescein sodium) was injected into the main channel at a flow rate of 5 μL min−1, while organic extracting reagent (butanol) was injected into the side channel at a flow rate of 0.5 μL min−1 in the droplets generation part. Then the monodisperse droplets of butanol with a diameter of about 100 μm would disperse in the real sample solution flow and move forward with the sample flow. After that, the butanol droplets reached the droplets collection part in about 30 s, during which time fluorescein sodium in water could be extracted into the butanol droplets. The fluorescence images observed by fluorescence microscope indicated that fluorescence intensities of the droplets became stronger while the monodisperse droplets were moving forward with the real sample solution flow (Fig. 5–c), which indicated that the fluorescein sodium was successfully extracted into the organic droplets. The fluorescent intensities of the droplets at different positions were calculated by QCapture Pro software (Version 5.1.1.14, Media Cybernetics, USA). As shown in Fig. 5d, the fluorescence intensities of the droplet increased significantly as the droplets moved away from the droplets generation part, indicating that extracting efficiency would be visibly raised when we lengthened the extraction channel which provided a longer time for extraction. Even though complete distribution equilibrium was not achieved, the current extracting efficiency after a rapid extraction in chips was enough for detection and online monitor. In the meantime, elongation of the extracting channel of this microchip was proved to be a feasible way to increase the extracting efficiency if necessary. Applying smaller droplets or lower flow rate of sample solution was also believed to be helpful for the complete distribution equilibrium. Based on these results, the length of extracting channel was set to 15 mm in this work unless special explanation was mentioned.
When these droplets of butanol reached the droplets collection part, they would enter the guiding tracks and move along the tracks as discussed before. Finally, these monodisperse droplets passed through the connecting channel and then were absorbed by absorbing reagent (butanol was used here) in collecting channel (Fig. S3b and c, ESI†). When these droplets were absorbed into the non-fluorescent organic extracting reagent, different fluorescent intensities of the mixture of absorbing reagent and organic droplets were obviously observed (Fig. S3b and c, ESI†). By adjusting the flow rates in collecting channel and the generation rate of droplets, several droplets could be absorbed and detected together to provide a stronger signal, which would enhance the detection sensitivity. However, real sample solution would flow into the connecting channel and even the collection channel, if the flow rate in the collecting channel was too low (≤0.5 μL min−1) (Fig. S3a, ESI†). Instead, absorbing reagent would flow into the connecting channel if the flow rate in the collecting channel was too high (≥4 μL min−1), which would prevent the droplets from passing through the connecting channel to the collecting channel (Fig. S3d, ESI†). Each of the two cases could result in the failure of droplets collection. The fluorescence intensities of absorbing solution flow in the collection channel were calculated by Qcapture Pro software, and the results were shown in ESI Fig. S3e.† The results showed us that the fluorescence intensities of the absorbing solution increased when the flow rates of the absorbing solution were slowed down. Thus, the flow rate of the extracting reagent was set as 1 μL min−1.
All the results indicated that online monodisperse droplets based liquid–liquid extraction of fluorescein sodium was well performed on the microfluidic chip. After droplets generation in the first part, they dispersed in the real sample flow in the main channel and extraction successfully carried out in the main channel. The organic droplets containing extracted fluorescein sodium were collected and observed by an inverted fluorescence microscope. Therefore, the established microfluidic device was believed to be a useful tool for online sample pretreatment.