Rapid screening of boron isotope ratios in nuclear shielding materials by LA-ICPMS – a comparison of two different instrumental setups

Christoph Kurta a, Ladina Dorta b, Florian Mittermayr c, Karl Prattes d, Bodo Hattendorf b, Detlef Günther b and Walter Goessler *a
aInstitute of Chemistry – Analytical Chemistry, University of Graz, Stremayrgasse 16, 8010 Graz, Austria. E-mail: walter.goessler@uni-graz.at
bLaboratory of Inorganic Chemistry, ETH Zürich, Wolfgang-Pauli-Strasse 10, 8093 Zürich, Switzerland
cInstitute of Technology and Testing of Building Materials, Graz University of Technology, Inffeldgasse 24, 8010 Graz, Austria
dBöhler-Edelstahl GmbH, Mariazellerstraße 16, 8605 Kapfenberg, Austria

Received 12th June 2013 , Accepted 5th November 2013

First published on 26th November 2013


Abstract

10B enriched resources are widely used in shielding materials such as boron-alloyed steels in nuclear reactors and storage containers. During production, quality control is an indispensable part of the process. In this study two methods for the rapid screening of 10B enriched special steel samples with a 10B isotopic fraction from 90 to 97% – taken from the production process – were developed and compared. We used both ns-LA-ICPMS and fs-LA-MC-ICPMS and evaluated the different instrumental setups for this application. Even though the micro-inhomogeneous boron distribution in the steel samples was determined by LA-ICPMS and EPMA no effect on the isotope ratios could be observed. Besides matrix- and isotope-ratio-matched standardization with in-house reference materials, a non-matrix- and non-isotope-ratio-matched calibration using NIST 610 glass SRM was applied successfully.


Introduction

Boron has two stable natural isotopes, 10B (19.9%) and 11B (80.1%). The neutron absorption properties of boron alloyed stainless steels depend on the content of the 10B isotope with an absorption cross-section for thermal neutrons of more than 3.8 kilobarn (10−25 m2). This isotope is an important neutron absorber used in shielding materials such as boron-alloyed steels for storage of spent nuclear fuel, reactor shielding and control, and instruments detecting neutrons.1,2

The production of shielding materials is challenging economically as well as metallurgically. During steel production, internal recycling of materials is part of the production process. Inadvertent mix-up of 10B enriched materials with natural low 10B materials tremendously increases reject costs. To avoid such mix-up during the recycling process the sampling of liquid steel and recycling materials is an integral part of process control.

Traditionally, wet chemical mineralization methods are employed to digest steel samples for the determination of the trace element content and the isotopic composition.3 Different mass spectrometric techniques like secondary ion mass spectrometry (SIMS),4 isotope dilution thermal ionization mass spectrometry (ID-TIMS),5–7 spark source mass spectrometry (SS-MS),8 glow discharge mass spectrometry (GD-MS),9 quadrupole inductively coupled plasma mass spectrometry (Q-ICPMS),9–12 single collector sector field inductively coupled plasma mass spectrometry (SF-ICPMS),13 or multi-collector inductively coupled plasma mass spectrometry (MC-ICPMS)7,14–16 have been used for the determination of boron isotope ratios, mainly in geological samples and reference materials. Unconventional approaches like electrospray mass spectrometry (ES-MS)17 or high resolution continuum source flame atomic absorption spectrometry (HR-CS-FAAS)3 have been used for assessing boron isotope ratios as well.

However, all of these methods are either work-extensive in terms of mineralization (HR-CS-FAAS, ICPMS),3,13 matrix removal (ICPMS),10,18 extraction19 and sample preparation (TIMS)20 or require rather complex instrumental setups (SIMS).21

LA-ICPMS is an attractive alternative because of its high lateral and depth resolution, low limits of detection, a minimum of sample preparation and short analysis time compared to classical sample preparation methods which are tedious and time consuming.22 Moreover, the sampling process is not limited by physical or chemical properties of the materials of interest.21 Sample preparation of steel for LA-ICPMS analysis is not critical to achieve stable signals23 but it is indispensable for EPMA and SEM.24

In 2006 Možná et al. investigated the quantification capabilities of iron-based samples by different UV-ns (ArF excimer, Nd:YAG) and femtosecond (fs) (Ti-sapphire) laser ablation systems coupled to ICPMS. Evaluation of different wavelengths and laser pulse width for the analysis of minor and trace elements using 57Fe as an internal standard has been performed. Matrix- (within metallic samples) and non-matrix-matched (using silicate glass standard reference materials) calibrations have been applied for quantification showing improved analytical results in terms of precision and accuracy using matrix matched calibration and fs-LA-ICPMS.25 In 2011 Wiltsche and Günther used fs-LA-ICPMS for the quantification of 23 metallurgical relevant elements (including boron) in unalloyed, alloyed, highly alloyed steels, and super alloys. Scanning mode ablation with large ablation spot diameters (250 μm) and calibration with matrix- and non-matrix-matched standards provided similar results agreeing within the uncertainty of the certified values.26

LA with multi-collector-ICPMS for the analysis of boron isotopes (δ11B) at <1‰ 2σ precision at the nanogram level in natural and synthetic glasses was first presented by le Roux et al. in 2004.27 Tiepolo et al. applied LA-MC-ICPMS for the δ11B characterization of geological materials in 2006.28 In 2010 Fietzke et al. proposed the use of NIST glasses for the non-matrix-matched calibration of a new in situ method using LA-MC-ICPMS (193 nm excimer laser) for the determination of stable boron isotopes in carbonates.29

In order to demonstrate the benefits and limitations of different LA-ICPMS systems, we developed a fast and simple method for the screening of boron isotope ratios in special steels for quality and production control using standard equipment (ns-LA coupled to Q-ICPMS) as well as more advanced instrumentation (fs-LA-MC-ICPMS).

Experimental

Samples

NIST SRM 612 (nominal trace element concentration 50 mg kg−1) “Trace Elements in a Glass Matrix” (Gaithersburg, MA, USA) was used for daily instrument tuning (ns-LA-ICPMS). For non-matrix-matched isotope ratio determination and tuning of fs-LA-MC-ICPMS NIST SRM 610 (nominal trace element concentration 500 mg kg−1) “Trace Elements in a Glass Matrix” (Gaithersburg, MA, USA) with a boron content of 356 ± 7 mg kg−1 (ref. 30) and an isotope ratio of 10B/11B = 0.2472 ± 0.0003 (ref. 27) was employed.

Currently no steel reference material with a certified boron isotope ratio is available on the market.31 For this reason in-house reference materials from Böhler Edelstahl GmbH, Kapfenberg, Austria were used for the experiments. Major constituents in the steel in-house reference materials S20, S35 and S97 are 19.7% Cr, 12.3% Ni, 1.8% B, 1.1% Mn, 0.02% C and a balance of Fe.3,12 Boron isotope ratios in these materials have previously been determined by Q-ICPMS, MC-ICPMS and TIMS.3,12 The isotopic ratios are presented in Table 1.

Table 1 In-house reference materials S20, S35 and S97: reference values for boron isotope ratio (mean value ± standard error in %), 10B abundance and total boron content (mean value ± standard error in fraction, %)3
S20 S35 S97
10B/11B 0.2477 ± 0.0010 0.5314 ± 0.0010 33.13 ± 0.01
Abundance of 10B (%) 19.85 ± 0.07 34.70 ± 0.09 97.07 ± 0.03
Total B (%) 1.80 ± 0.05 1.84 ± 0.05 1.79 ± 0.05


Additionally, three samples (N1, N2 and N3) with approximately 1.8% total B content and different 10B enrichment levels (>90%) were taken from the production process and analysed as unknowns.

Sample preparation

LA-ICPMS is only requiring a minimum of sample preparation (e.g. cutting and/or grinding). For wet chemical analysis of steel the sample has to be cut, reduced to chip form, digested in concentrated acids, and diluted prior to analysis – which is usually taking at least 30 minutes, compared to 5 minutes of sample preparation for LA-ICPMS. However, due to the EPMA experiments performed in this study sample preparation (60 minutes for cutting, embedding, grinding, and polishing) cannot be considered negligible any more.

Samples were taken from metal sheet plates (except in-house reference material S20, which was only available as metal chips), roughly ground and cut into pieces of approximately 3 × 6 mm in size. The samples were used for both LA-ICPMS and microprobe experiments and thus had to be embedded in epoxy resin for LA and EPMA sample holders (1 inch diameter), ground down with 600 and 1200 grid SiC powder and polished with both Buehler METADI® 3 and 1 micron diamond suspension (Lake Bluff, IL, USA) (ESI, Fig. 1).

Imaging by LA-ICPMS and electron probe microanalysis (EPMA)

In order to visualize the boron distribution in the steel samples several 2000 μm long line scans with a spot size of 100 μm, a scan speed of 10 μm s−1, a laser energy of 4.0 J cm−2 and a repetition rate of 10 Hz were performed by ns-LA-ICPMS. After a 30 second laser warm-up (laser shutter closed but laser firing) a line scan of 200 seconds was started, followed by 60 seconds of washout.

Additionally, elemental distribution images of boron were recorded by electron probe microanalysis using a JEOL JXA-8200 Superprobe (JEOL, Tokyo, Japan). The wavelength-dispersive analytical mode with 20 kV acceleration voltage and a beam current of 180 nA was used. Boron intensities were recorded at the peak position of the B Kα line at 129.29 mm on a layered dispersion element (LDEB) diffracting crystal. 1024 × 1024 point analyses and a step size of 0.5 μm yielded elemental distribution mappings of 512 × 512 μm (ESI, Fig. 2).

Instrumentation

ns-LA-Q-ICPMS. The standard equipment consisted of a 213 nm nanosecond Nd:YAG laser ablation system UP-213 (New Wave Research, Fremont, CA, USA) coupled to an Agilent 7700x quadrupole ICPMS (Agilent Technologies, Waldbronn, Germany). Helium 5.0 (Messer, Gumpoldskirchen, Austria) was used as a carrier gas for the LA system. The LA sample cell was connected to the ICPMS via Tygon® R3603 tubings (Saint-Gobain, Charny, France) with an inner diameter of 3.2 mm. Mixing of the LA aerosol with the argon carrier gas of the ICPMS was performed using a glass coaxial mixing-bulb.32 In order to reduce washout times and aerosol dispersion due to a small gas volume in the LA cell33 the dead volume of the LA cell was reduced by mounting the embedded samples in a customized PTFE sample holder. The system setup determined the length of the sample transport tubing (for ns-LA-ICPMS 2 m).
fs-LA-MC-ICPMS. The Ti-sapphire femtosecond laser with a fundamental wavelength of 795 nm and a pulse duration of ∼150 fs (Legend, Coherent Inc., Santa Clara, USA)34 was coupled to a Nu Plasma HR MC-ICPMS (Nu Instruments, Wrexham, Great Britain) operating in low resolution mode. In order to reduce washout times and aerosol dispersion due to a small gas volume in the LA cell33 the dead volume of the LA cell was reduced by mounting the embedded samples in a specific low dispersion high capacity cell with very fast washout.35 The fs-LA was connected to the MC-ICPMS via Tubclair® AL tubings (Hozelock-Tricoflex, Vitry-Le-Francois, France) with an inner diameter of 4 mm. The system setup determined the length of the sample transport tubing (fs-LA-MC-ICPMS 3 m). For boron detection the instrument was operating at 6 kV extraction voltage using the two outermost Faraday cups (L5 and H6) for measurements. Carrier gas for the LA system was helium 5.0 (PanGas, Dagmersellen, Switzerland). Mixing of the LA aerosol with the argon carrier gas of the ICPMS was again performed using a glass coaxial mixing-bulb.32

Results and discussion

Boron distribution in special steels

Mean isotope ratios for sequential ablation experiments using ns-LA and quadrupole based ICPMS analyses showed relative standard deviations (RSDs) of less than 3%. RSDs for 10B or 11B signals, however, were as high as 40% within an individual line scan or a single spot ablation. In order to obtain information about the boron distribution in the sample single line scans were performed by ns-LA-ICPMS in time resolved analysis (TRA) mode. Evidently an inhomogeneous distribution at the micro-scale for both isotopes was present in the special steel sample in-house reference material S97. Similarity for the two boron isotopes was confirmed and thus the boron distribution appeared to be homogeneous (3% RSD) with respect to isotope ratios (Fig. 1). Similar micro-inhomogeneity of the boron distribution was observed in in-house reference materials S20 and S35.
image file: c3ja50194a-f1.tif
Fig. 1 Time resolved analysis of a 2000 μm line scan (ns-LA-ICPMS, 100 μm spot size) on the polished surface of in-house reference material S97 – apparent micro-inhomogeneity in the boron distribution with similar pattern of both boron isotopes but stable boron isotope ratio.

Optimization of LA and ICPMS

ns-LA-Q-ICPMS. For optimization of signal intensity and relative standard deviation (RSD) of single spot analysis by ns-LA, laser energy, repetition rate, and spot size have been evaluated (ESI, Fig. 3–9). Due to the micro-inhomogeneous distribution of boron in the samples large spot sizes (50 μm or above) were used to achieve high intensities, allowing precise isotope ratio measurements in special steels. Because of limited sample size and the need for multi-spot analyses on small areas redeposition of ablated materials due to sputtering on the sample surface should be avoided. This side effect was inspected visually after ablation (e.g. varying laser energy) with a light microscope (ESI, Fig. 3) and a compromise between high signal intensity, low RSD and little redeposition has been selected (Table 2).
Table 2 Optimized laser ablation operating conditions for steel by ns- and fs-LA
ns-LA-ICPMS fs-LA-MC-ICPMS
LA system New Wave UP-213 Coherent Legend (795 nm)
Scan pattern Single spot Single spot
Pulse width 4 ns ∼150 fs
Energy ∼0.13 mJ ∼1.0 mJ (steel)
∼1.7 mJ (glass)
Repetition rate 10 Hz 10 Hz (guidance level)
Spot size 100 μm 200 μm
Laser warm-up – background 30 s 30 s
Laser analysis time 100 s 130 s
Laser washout 300 s 300 s
He carrier gas flow 0.95 l min−1 0.95 l min−1
Ar carrier gas flow 0.65 l min−1 0.74 l min−1


Elemental mapping using EPMA confirmed the inhomogeneous distribution of boron at the micro-scale in the samples (ESI, Fig. 2), explaining the 40% relative standard deviations during LA-ICPMS line scans or single spot ablations.

Some considerations when analyzing boron with ICPMS have to be taken into account. Due to its first ionization potential of 8.3 eV (ref. 1) boron is only about 60% ionized in the plasma.36 Because of the high boron concentration in the samples (up to 1.8%) elevated blank levels for boron can be expected due to carry-over effects in a liquid sample introduction system13 and might also lead to elevated background in LA-ICPMS. Additionally an elevated carbon background can occur from the carbon content of the steel samples, from impurities in the carrier gas helium, and due to degassing of the tube transport system,37 leading to potential spectral interference on 11B+ due to tailing of 12C+.38 According to previous studies by Wang et al. optimization of the sampling depth and carrier gas flow rates is essential for high sensitivity and reduced fractionation effects in LA-ICPMS.39 Andrén et al. additionally reported that variation of the instrumental mass bias can be reduced to a minimum when optimizing the sampling depth before isotope ratio measurements by ICPMS.16 Therefore optimization parameters for ICPMS consisted of sampling depth and extraction ion lens settings. Optimization was performed aiming at highest intensities for boron and lowest boron isotope ratio RSD. Details can be found in the ESI, Fig. 10 and 11. LA-ICPMS settings were optimized daily in terms of signal intensity and oxide ratios using NIST SRM 612, containing 35 ± 3 mg kg−1 boron.30 Daily pulse/analog (P/A) detector tuning for both boron isotopes using in-house reference material S35 was performed. Additionally, a dead time calibration according to manufacturer recommendation was carried out with no significant effect on the results. Optimized operating conditions for LA and ICPMS are summarized in Tables 2 and 3.

Table 3 Optimized ICPMS operating conditions for Q- and MC-ICPMS
ICPMS system Agilent 7700x Nu plasma HR
Sampling depth 4.5 mm 1.5 mm
Extraction lens (HV 1) 6 kV
Ω lens 4.0 V
Ω bias −70 V
m/z measured 10, 11 10, 11
Detector Auto (pulse and analog counting) Faraday cups (L5, H6)
Integration time 100 ms per mass 200 ms per mass
Analysis mode Isotope ratio (IR) Time resolved analysis (TRA)
B sensitivity (100 μm spot size) ∼6500 cps per mg kg−1 B ∼1 × 107 cps per mg kg−1 B


fs-LA-MC-ICPMS. EPMA images and optimization of ns-LA have shown the need for large spot sizes for compensation of the micro-inhomogeneity in the sample. Therefore, the nominal spot size for fs-LA was set to 200 μm and only laser energy and repetition rate were adjusted to optimize the signal intensity and RSD during single spot analyses. MC-ICPMS optimization for highest signal intensity was carried out by adjustment of carrier and make-up gas flows, torch position, and lens voltages. In order to ensure that ion signals registered for fs-LA-MC-ICPMS remain in the linear dynamic range of the faraday detectors and to avoid saturation of the detectors, the laser repetition rate was adjusted while the laser energy was fixed. Several configurations were evaluated (ESI, Fig. 12 and 13) and optimum conditions were selected (Table 2). Applying the optimized laser conditions for the steel samples to the glass yielded in low signal intensities close to the threshold. Thus different laser settings were applied for steel in-house reference materials and NIST SRM 610 (Table 2).

In the MC-ICPMS tailing of 12C+ on 11B+ was not observed while a spectral overlap from 40Ar4+ on the 10B+ signal was avoided by selecting a measurement position of about 0.02 amu higher than the 10B+ centre mass, where the interference could be resolved.

LA-ICPMS settings were optimized daily for highest sensitivity using either NIST SRM 610, containing 356 ± 7 mg kg−1 boron30 or in-house reference material S97 containing 1.8% boron. Optimized settings for LA and ICPMS are summarized in Tables 2 and 3.

Data acquisition

ns-LA-Q-ICPMS. For ns-LA-ICPMS, data acquisition was performed in isotope ratio mode with Agilent ICPMS MassHunter B01.01 software. One acquisition contained 60 data points with 100 ms integration time per point for each isotope – resulting in 2.4 seconds total acquisition time per point (including settling and dead time) – and lasted 144 seconds (30 s of laser warm-up and background acquisition, 100 s of laser firing and 14 s of recorded washout). For background correction the mean value of the first ten data points was subtracted from each data point used for isotope ratio determination individually. To avoid carry over an overall flush time of 300 s was applied after each single spot analysis. Raw data were exported to spreadsheet software where 10B/11B, mass bias, 10B and 11B mean intensities, background, and relative standard deviations (RSDs) were calculated from the middle 35 data points (Fig. 2).
image file: c3ja50194a-f2.tif
Fig. 2 ns-LA-ICPMS: time resolved signals from a typical dataset (consisting of 60 data points) from a single spot experiment on in-house reference material S97 – for data evaluation 35 points (point 13–47) are extracted, for background correction the first ten points are used.

The instrument provides three acquisition modes of the detector, namely pulse, analog, and mixed mode (with automatic selection of the two modes). Analog mode extends the dynamic range of the detection system but requires an accurate calibration of the electron current vs. pulse count, which may introduce an additional uncertainty when high signal intensities are acquired in analog mode, while low intensities are registered by pulse counting (<106 cps).

For example, the 10B intensity in in-house reference material S20 (Fig. 3) was measured in analog mode for the first 18 data points and in pulse mode for the remaining 17 data points. In order to remain in one detector mode (either pulse or analog) for each isotope, data points used for calculations were filtered manually (mode extracted) – selecting the detector mode giving more accurate isotope ratio values (e.g. analog/pulse for process samples N1 and N2). Even though consequent daily calibration of “pulse to analog” (P/A) factors for both boron isotopes using in-house reference material S35 has been performed, the accuracy of the P/A factor calibration was insufficient and leads to wrong estimation of the isotope signal measured in analog mode compared to pulse mode. This caused the 10B/11B intensity ratio to shift to lower values.


image file: c3ja50194a-f3.tif
Fig. 3 Boron intensities and isotope ratio of in-house reference material S20 in a mixed mode (pulse count and analog) acquisition run (ns-LA-ICPMS) – evident change in the calculated isotope ratio when pulse counts are evaluated for the 10B isotope (data points 18 to 19).

Table 4 presents an overview about the detector modes used for different samples.

Table 4 Electron multiplier mode for measurements of NIST SRM 610, in-house reference materials S20, S35, and S97 and three process samples (N1, N2 and N3). For mixed mode experiments the selected mode is presented in the mode extracted column
Sample 10B 11B Mode extracted
NIST 610 Pulse Pulse
S20 Mixed Analog Analog (10B)
S35 Analog Analog
S97 Analog Pulse
N1 Analog Mixed Pulse (11B)
N2 Analog Mixed Pulse (11B)
N3 Analog Pulse


fs-LA-MC-ICPMS. For fs-LA-MC-ICPMS, data acquisition was performed in time-resolved analysis mode with Nu Plasma 1.4.2019 software. A typical acquisition consisted of around 1000 data points with an integration time of 200 ms per point. First the background was recorded for about 30 seconds and then data acquisition (during ablation) was performed for about 130 seconds (Fig. 4). In order to avoid carry-over an overall washout time of 300 s was maintained after each single spot analysis. Raw data were exported to spreadsheet software and background correction (individual data points corrected for average gas blank), isotope ratios, signal intensity means, standard deviations and mass bias were calculated.
image file: c3ja50194a-f4.tif
Fig. 4 fs-LA-MC-ICPMS: time resolved signals of a typical dataset (consisting of approximately 1000 data points) from a single spot experiment on in-house reference material S97 – for data evaluation, data points between 50 s and 180 s are evaluated.

Mass bias correction

The sample–standard bracketing approach provides the most applicable method to correct for mass bias in this case. Measurements in the order of standard – sample – standard – … were performed and the arithmetic mean of two mass bias values obtained for the standards was used for correction of the isotope ratio of the unknown sample measured in between using eqn (1):
 
image file: c3ja50194a-t1.tif(1)
where RS,true being the mass bias corrected isotope ratio for the sample, RS,meas the ratio measured for the sample, RB,true the reference value for the bracketing standard and RB,meas the mean ratio measured for the two bracketing standards.

In the first step the three matrix-matched in-house reference materials (Table 1) – where substantial differences in the boron isotope ratios occur (e.g. in-house reference material S20 as the standard and in-house reference material S97 as unknown) – were compared and in a second step additionally non-matrix-matched correction using NIST SRM 610 as the standard and the three steel in-house reference materials as unknowns was investigated.

All in-house reference materials (S20, S35 and S97) and NIST SRM 610 were analyzed in sequences consisting of 6 times standard analysis (e.g. S20), interspersed by 5 samples (e.g. S97) (ESI, Fig. 14).

Boron isotope ratios in reference materials by LA-ICPMS

By applying the sample–standard bracketing approach, all in-house reference materials and NIST SRM 610 were evaluated as unknowns as well as reference and their respective boron isotope ratios were determined. In this way the dependence of accuracy and precision on sample matrix and isotope ratio was evaluated. The isotope ratio reference values and results for both ns-LA-ICPMS and fs-LA-MC-ICPMS experiments are summarized in Table 5. For process control the needed accuracy and precision was defined by the reference values of the in-house reference materials given in Table 1.
Table 5 Isotope ratio measurements using both instrumental setups, applying matrix- and non-matrix-matched mass bias correction in special steel in-house reference materials S20, S35, and S97 and NIST SRM 610
Sample NIST 610 S20 S35 S97
10B/11B SD Ref. 10B/11B SD Ref. 10B/11B SD Ref. 10B/11B SD Ref.
a Standard deviation of 5 independent determinations.
Reference value 0.2472 0.0003 27 0.2477 0.0010 3 0.5314 0.0010 3 33.13 0.01 3

ns-LA-ICPMS NIST 610 S20 S35 S97
Reference sample 10B/11B SDa RSD% 10B/11B SDa RSD% 10B/11B SDa RSD% 10B/11B SDa RSD%
NIST 610 Bracketing standard 0.2525 0.0016 0.62 0.5320 0.0033 0.62 33.43 0.11 0.33
S20 0.2429 0.0027 1.10 Bracketing standard 0.5259 0.0022 0.43 33.05 0.15 0.44
S35 0.2467 0.0015 0.62 0.2499 0.0005 0.21 Bracketing standard 33.58 0.27 0.81
S97 0.2447 0.0009 0.38 0.2482 0.0010 0.40 0.5254 0.0037 0.71 Bracketing standard

fs-LA-MC-ICPMS NIST 610 S20 S35 S97
Reference sample 10B/11B SDa RSD% 10B/11B SDa RSD% 10B/11B SDa RSD% 10B/11B SDa RSD%
NIST 610 Bracketing standard 0.2477 0.0002 0.08 0.5317 0.0010 0.19 33.18 0.17 0.51
S20 0.2465 0.0002 0.08 Bracketing standard 0.5311 0.0005 0.10 33.37 0.02 0.06
S35 0.2470 0.0004 0.16 0.2479 0.0004 0.16 Bracketing standard 33.39 0.03 0.09
S97 0.2471 0.0012 0.49 0.2459 0.0001 0.04 0.5273 0.0002 0.04 Bracketing standard


ns-LA-ICPMS. Reference values and analysis results are displayed as the mean of 5 measurements and their standard deviation (SD). For all ratios, background subtraction and mass bias correction were applied. Student's t-test showed significant differences from the reference values for in-house reference materials S97 and S20 with NIST SRM 610 as the standard, in-house reference material S35 with S20 as the standard and S20 with S35 as the standard. NIST SRM 610 and in-house reference material S97 with S35 as the standard and S35 with S97 as the standard were within 99% confidence. All other results were within 95% confidence. For process control it is considered permissible to use an expanded uncertainty of 3 SD (99% confidence). With this expanded uncertainty most of the results were acceptable for process control. When using NIST SRM 610 as the bracketing standard, however, highest deviations were observed whereas results with in-house reference material S97 as the standard were found to be within 99% confidence for all samples. For in-house reference material S20 an increase in deviation from the reference value with decreasing 10B content (S97 – S35 – NIST SRM 610) was observed. The results obtained for the different standard compositions (glass vs. steel), a boron concentration (356 mg kg−1vs. 1.8%), and an isotopic composition (20–97% B (ref. 10)) suggest that in-house reference material S97 can be considered the best suited material for the determination of boron isotope ratios in special steels by ns-LA-ICPMS.

Overall, ns-LA-ICPMS was able to yield a reasonable estimate of the boron isotope ratio in the samples. The precision of the determined isotope ratios was in between 0.21 and 1.1%. Several publications over the last few years have shown that determination of boron isotope ratios by applying Q-ICPMS is possible, even though trade-offs with respect to accuracy and precision have to be made.7,10,11

fs-LA-MC-ICPMS. The isotope ratio measurements by fs-LA-MC-ICPMS yielded relative standard deviations ranging between 0.04 and 0.51% (Table 5), achieving similar precision like TIMS analyses and in most cases significantly better than with ns-LA-ICPMS. Accuracy, however, was only slightly better than with ns-LA-ICPMS, indicating that mass bias correction by standard sample bracketing was similarly effective in both cases. Nevertheless it could be shown that fs-LA-MC-ICPMS is much less affected by the composition of the standard used than ns-LA-ICPMS. Although significant differences from the reference values were detected for in-house reference material S97 and NIST SRM 610 with in-house reference material S20 as the standard, for in-house reference material S97 with S35 as the standard, and for in-house reference materials S35 and S20 with S97 as the standard all other results were found to be within 95% confidence to the reference values. It has to be highlighted that even all determinations with NIST SRM 610 as the bracketing standard yielded results within 95% confidence interval.

Boron isotope ratios in process samples

All three samples N1, N2, and N3 were analyzed against either NIST SRM 610 or one of the three in-house reference materials S20, S35, and S97 as the bracketing standard (Table 6). The relative standard deviations of the boron isotope ratios of the different samples amount to 0.22–0.59% when using ns-LA-ICPMS. For fs-LA-MC-ICPMS, RSDs between 0.04 and 0.16% were obtained, except for N3 with bracketing standard NIST SRM 610 (1.66% RSD). The elevated RSD for NIST SRM 610 most likely resulted from insufficient washout, leading to carry-over from the steel sample (N3) to the bracketing standard NIST SRM 610. However, overall precision was improved by a factor of five using fs-LA-MC-ICPMS.
Table 6 Boron isotope ratios measured in special steel samples (N1, N2 and N3) taken from the production process
Sample N1 N2 N3
10B/11B SDa RSD% 10B/11B SDa RSD% 10B/11B SDa RSD%
a Standard deviation of 5 independent determinations.
ns-LA-ICPMS
NIST 610 9.134 0.046 0.50 11.98 0.04 0.35 37.24 0.08 0.21
S20 8.994 0.036 0.40 11.90 0.08 0.65 36.69 0.11 0.30
S35 9.148 0.02 0.22 12.04 0.04 0.32 37.58 0.08 0.21
S97 9.153 0.054 0.59 12.07 0.03 0.25 37.82 0.08 0.21
 
fs-LA-MC-ICPMS
NIST 610 9.118 0.015 0.16 12.19 0.05 0.41 38.56 0.64 1.66
S20 9.210 0.004 0.04 12.20 0.05 0.41 38.09 0.03 0.08
S35 9.206 0.007 0.08 12.18 0.01 0.08 38.02 0.05 0.13
S97 9.272 0.011 0.12 12.26 0.01 0.08 37.69 0.02 0.05


Irrespective of the bracketing standard used, isotope ratios determined by ns-LA-ICPMS were lower than those determined by fs-LA-MC-ICPMS (Table 6), which could result from different detector modes (pulse/analog) and an insufficient accuracy of the P/A factor calibration. ANOVA analysis revealed significant differences (95% confidence interval) between ns-LA-ICPMS and fs-LA-MC-ICPMS results for all samples except for process sample N1 with NIST SRM 610 as the bracketing standard. When expanding the confidence interval to 99% also N1 and N3 with in-house reference material S97 as the standard were not showing significant deviations. Since no additional analytical data of these process samples were available, results from fs-LA-MC-ICPMS with in-house reference material S97 as the bracketing standard were assumed to be closest to the true values.

Conclusions

In the present study we evaluated different LA-ICPMS setups for the determination of boron isotope ratios in special steel samples. The micro-inhomogeneous boron distribution in the steel samples was apparent from EPMA analyses and also appears in the LA-ICPMS transient signals but had no significant effect on the isotope ratios when large ablation craters were employed. ns-LA-ICPMS has some limitations concerning precision (by about a factor of five less than fs-LA-MC-ICPMS) and accuracy, especially when the isotopic composition and concentration of standard and sample are substantially different. A major limitation in this case is caused by the detection system. The sequential signal acquisition of quadrupole based ICPMS is inherently affected by slight variations of the aerosol density inside the ICP, caused by the pulsed laser ablation event. Additionally, the signal detection using the electron multiplier can cause significant bias in the isotope ratio when the isotopes are detected in different modes (pulse counting vs. analog detection). Thus the selection of the right detector mode is of great importance for ns-LA-ICPMS. Nevertheless, with some special considerations (P/A factor calibration, background correction, extended washout times, and the use of matrix- and isotope-ratio-matched standards), this approach can be useful for the rapid screening of steel samples. The costs for ns-laser ablation and Q-ICPMS equipment are affordable and thus this technique could find or has already found its way into the routine laboratory of steel manufacturers. For high precision measurements however fs-LA-MC-ICPMS is the method of choice. Even though the use of NIST SRM 610 as the bracketing standard is possible, results are improved by the use of matrix- and isotope-ratio-matched standards. For this application fs-LA-MC-ICPMS is comparable to TIMS in terms of accuracy and precision (Table 5). When appropriate standards are available, this technique is even superior because sample preparation can be reduced to a minimum and shorter measurement times are required. Due to the relatively long washout times (5 minutes per analysis) the total measurement times however were substantially extended. An optimization of the sample introduction system especially using a shorter transport tubing should still improve this situation. In this study, an extensive sample preparation procedure has been applied to make the samples suitable for EPMA. For routine application by LA-ICPMS alone, however, the benefit of almost no sample preparation can be sustained and would make the technique interesting for high-throughput analysis.

References

  1. Handbook of Chemistry and Physics, ed. D. R. Lide, CRC Press, Boca Raton, 2002 Search PubMed.
  2. H. Möllinger, Themen zur Chemie des Bors, Dr. Alfred Hüthig Verlag, Heidelberg, 1976 Search PubMed.
  3. H. Wiltsche, K. Prattes, M. Zischka and G. Knapp, Spectrochim. Acta, Part B, 2009, 64, 341–346 CrossRef PubMed.
  4. I. Marchetti, L. Menichetti, C. Kusmic, L. A. De las Heras, P. Salvadori, R. Fuoco, F. Belloni, A. L'Abbate and M. Betti, Spectrochim. Acta, Part B, 2009, 64, 911–920 CrossRef PubMed.
  5. K. G. Heumann and H. Zeininger, Int. J. Mass Spectrom. Ion Processes, 1985, 67, 237–252 CrossRef CAS.
  6. K. G. Heumann, S. Eisenhut, S. Gallus, E. H. Hebeda, R. Nusko and A. Vengosh, Analyst, 1995, 120, 1291–1299 RSC.
  7. J. Aggarwal, F. Böhm, G. Foster, S. Halas, B. Hönisch, S.-Y. Jiang, J. Kosler, A. Liba, I. Rodushkin, T. Sheehan, J. J.-S. Shen, S. Tonarini, Q. Xie, C.-F. You, Z.-Q. Zhao and E. Zuleger, J. Anal. At. Spectrom., 2009, 24, 825–831 RSC.
  8. R. A. Lukaszew, J. G. Marrero, R. F. Cretella and C. J. Noutary, Analyst, 1990, 115, 915–917 RSC.
  9. F. Chartier and M. Tabarant, J. Anal. At. Spectrom., 1997, 12, 1187–1193 RSC.
  10. D. C. Gregoire, Anal. Chem., 1987, 59, 2479–2484 CrossRef CAS.
  11. A. Al-Ammar, E. Reitzernová and R. M. Barnes, Spectrochim. Acta, Part B, 2000, 55, 1861–1867 CrossRef.
  12. K. Prattes, G. Hochoertler and W. Wegscheider, in Proceedings of the Chemists' Conference, Corus Research, Development & Technology, 2000, vol. 51, pp. 12–14 Search PubMed.
  13. J. Vogl, M. Rosner and W. Pritzkow, J. Anal. At. Spectrom., 2011, 26, 861–869 RSC.
  14. H.-E. Gäbler and A. Bahr, Chem. Geol., 1999, 156, 323–330 CrossRef.
  15. J. K. Aggarwal, K. Mezger, E. Pernicka and A. Meixner, Int. J. Mass Spectrom., 2004, 232, 259–263 CrossRef CAS PubMed.
  16. H. Andrén, I. Rodushkin, A. Stenberg, D. Malinovsky and D. C. Baxter, J. Anal. At. Spectrom., 2004, 19, 1217–1224 RSC.
  17. M. C. B. Moraes, J. G. A. Brito Neto and C. L. Do Lago, J. Anal. At. Spectrom., 2001, 16, 1259–1265 RSC.
  18. A. G. Coedo, T. Dorado, B. J. Fernandez and F. J. Alguacil, Anal. Chem., 1996, 68, 991–996 CrossRef CAS.
  19. K. E. Burke, Analyst, 1972, 97, 19–28 RSC.
  20. D. J. Scott and G. Gauthier, Chem. Geol., 1996, 131, 127–142 CrossRef CAS.
  21. J. Pisonero, B. Fernández and D. Günther, J. Anal. At. Spectrom., 2009, 24, 1145–1160 RSC.
  22. J. S. Becker, Spectrochim. Acta, Part B, 2002, 57, 1805–1820 CrossRef.
  23. A. Bengtson, M. Granfors, I. Gustavsson, M. Sedlakova, A. Gómez Coedo, I. Padilla, H. Dillen, K. Van den Bergh, D. Günther and C. Latkoczy, High spatial resolution analysis of steel samples using laser ablation ICP-MS, 2006, EUR 21995 Search PubMed.
  24. S. J. B. Reed, Electron Microprobe Analysis and Scanning Electron Microscopy in Geology, Cambrige University Press, Cambridge, 2005 Search PubMed.
  25. V. Možná, J. Pisonero, M. Holá, V. Kanický and D. Günther, J. Anal. At. Spectrom., 2006, 21, 1194–1201 RSC.
  26. H. Wiltsche and D. Günther, Anal. Bioanal. Chem., 2011, 399, 2167–2174 CrossRef CAS PubMed.
  27. P. J. le Roux, S. B. Shirey, L. Benton, E. H. Hauri and T. D. Mock, Chem. Geol., 2004, 203, 123–138 CrossRef CAS PubMed.
  28. M. Tiepolo, C. Bouman, R. Vannucci and J. Schwieters, Appl. Geochem., 2006, 21, 788–801 CrossRef CAS PubMed.
  29. J. Fietzke, A. Heinemann, I. Taubner, F. Bohm, J. Erez and A. Eisenhauer, J. Anal. At. Spectrom., 2010, 25, 1953–1957 RSC.
  30. N. J. G. Pearce, W. T. Perkins, J. A. Westgate, M. P. Gorton, S. E. Jackson, C. R. Neal and S. P. Chenery, Geostand. Newsl., 1997, 21, 115–144 CrossRef CAS.
  31. J. Vogl and W. Pritzkow, J. Anal. At. Spectrom., 2010, 25, 923–932 RSC.
  32. D. Bleiner and D. Günther, J. Anal. At. Spectrom., 2001, 16, 449–456 RSC.
  33. W. Müller, M. Shelley, P. Miller and S. Broude, J. Anal. At. Spectrom., 2009, 24, 209–214 RSC.
  34. J. Koch, M. Wälle, J. Pisonero and D. Gunther, J. Anal. At. Spectrom., 2006, 21, 932–940 RSC.
  35. M. B. Fricker, D. Kutscher, B. Aeschlimann, J. Frommer, R. Dietiker, J. Bettmer and D. Günther, Int. J. Mass Spectrom., 2011, 307, 39–45 CrossRef CAS PubMed.
  36. K. E. Jarvis, A. L. Gray and R. S. Houk, Handbook of inductively coupled plasma mass spectrometry, Chapman and Hall, New York, 1992 Search PubMed.
  37. R. Kovacs and D. Günther, J. Anal. At. Spectrom., 2008, 23, 1247–1252 RSC.
  38. J. A. Moreton and H. T. Delves, J. Anal. At. Spectrom., 1999, 14, 1545–1556 RSC.
  39. Z. Wang, B. Hattendorf and D. Günther, J. Am. Soc. Mass Spectrom., 2006, 17, 641–651 CrossRef CAS PubMed.

Footnote

Electronic supplementary information (ESI) available: Optimization of LA and ICPMS parameters. See DOI: 10.1039/c3ja50194a

This journal is © The Royal Society of Chemistry 2014