Xiaole
Mao
a and
Tony Jun
Huang
*b
aThe Procter and Gamble Company, Mason, Ohio 45040, USA
bDepartment of Engineering Science and Mechanics, The Pennsylvania State University, University Park, PA 16802, USA. E-mail: junhuang@psu.edu
First published on 12th September 2012
Cellular mechanical properties have been observed to have important implications for pathogenesis and pathophysiology. These observations have led to the recent development of a unique class of biomarkers: mechanical biomarkers. Compared with the traditional biochemical-based biomarkers (e.g., antibodies), mechanical biomarkers have many advantages such as label-free, low cost, convenient maintenance, and reduced assay time. In the past few years, there has been an increasing effort to exploit cellular mechanical biomarkers in microfluidic devices. This trend makes sense because microfluidic devices often feature structures that have characteristic lengths similar to those of cells, which renders them uniquely capable of probing and utilizing mechanical biomarkers. In this Focus article, we discuss a few examples of mechanical biomarker-based microfluidic applications. We believe that these examples are just the tip of the iceberg and that the full potential of mechanical biomarkers in microfluidic-based diagnostics and therapeutics has yet to be revealed.
In recent years, efforts to utilize cellular mechanical properties or mechanical biomarkers to provide new avenues for detecting and treating diseases have increased.3 Compared with biochemical biomarkers, mechanical biomarkers are label-free—measuring cellular mechanical properties can eliminate many costly sample labeling and preparation steps and significantly reduce assay time and complexity. Since microfluidic devices4–7 often have features with characteristic lengths similar to that of cells, they are excellent platforms for probing cellular mechanical biomarkers to enable novel diagnostic and therapeutic applications. By observing the mechanical behavior of cells in these cell-sized microfluidic structures and exploiting interactions between individual cells and the microfluidic structures, many new and exciting studies and applications become possible. In this Focus article we will discuss several interesting works along this front. We expect that as we improve our understanding of the implication of cell mechanics in various types of diseases, mechanical biomarker-enabled microfluidic approaches will have a more significant impact on many biological studies and clinical applications.
Fig. 1A shows a schematic of the deformability cytometry device. It contains an array of triangle-shaped pillars. The gaps between pillars are designed to add constraints for RBCs to pass through. In a pressure-driven flow, uninfected and infected RBCs travel through the gaps at different velocities, due to their difference in deformability. Fig. 1B shows the identification of infected and uninfected RBCs in the channels, indicated by red and blue arrows, respectively. Using such a deformability-based cytometry device, the difference between the uninfected and infected RBCs can be well characterized. Fig. 1C is a two-dimensional (2D) plot of velocity vs. fluorescence intensity of RBCs tested (infected RBCs are fluorescently dyed). Clearly the infected RBCs show much lower speed travelling through the pillar array due to their reduced deformability. This device shows great promise in deformability-based, microfluidic cell analysis for diagnosis and drug development. The authors envision the use of this device for screening drug compounds that alter the rigidity of early-stage infected RBCs to help the spleen remove the infected cells or to reduce the rigidity of late-stage infected RBCs to prevent capillary blockage.
![]() | ||
Fig. 1 Principles of pillar-based deformability cytometry.8 (A) Schematic of the device design. The device contains an array of obstacles. Physical dimensions of the obstacles are shown in the inset. Each array contains 10 by 200 obstacles. (B) Optical images of infected RBCs (red arrows) and uninfected RBCs (blue arrows) in the device. (C) Velocity vs. intensity for RBCs (infected cells: gray; uninfected cells: red). |
The method described above is an excellent example of how cell deformability can be exploited by microfluidic devices to enable novel disease diagnosis and treatment methods. However in the case of malaria, the change of RBC deformability is mainly due to the uptake of rigid protist parasites. Therefore, the question is whether deformability analysis is only helpful in the diagnosis of infectious diseases or might it have a broader impact in the diagnosis of other diseases such as cancer. Recently, Gossett et al. published an interesting work that shed some light into this question.9 In this work, the authors developed a method for the clinical screening of pleural fluids for malignant cells and the characterization of stem cell differentiation states using a “hydrodynamic stretching” based deformability cytometry device. The device (Fig. 2A) combines the authors' previously introduced inertial focusing component10 with a “hydrodynamic stretching” component (Fig. 2B). The inertial focusing component ensures the alignment of cells in a single-file fashion along the same streamline, placing them at the same initial condition for better stretching uniformity before they enter the “hydrodynamic stretching” zone (Fig. 2C). The “hydrodynamic stretching” zone is located at the center of an extensional flow. The cells are stretched upon entering this region (Fig. 2D). Stretching-induced changes in cell shape are recorded using a high-speed camera, and images are analyzed to quantify the cell deformability. The “hydrodynamic stretching” based deformability cytometry device provides several advantages, e.g., high throughput, less probability of clogging (especially compared with methods using physical constraints such as pillar arrays), and flexibility for fine-tuning hydrodynamic stress. Using this approach, the authors successfully demonstrated the possibility to predict disease states of cancer and immune activities by measuring the deformability of leukocytes and malignant cells in pleural fluids, and revealed deformability as an early biomarker for pluripotent stem cell differentiation.
![]() | ||
Fig. 2 Principles of “hydrodynamic stretching” based deformability cytometry. (A) A photograph of the microfluidic deformability cytometry device. (B) A schematic of the microfluidic device including the “inertial focusing” region and the “hydrodynamic stretching” region. (C) A schematic of the deformation of a cell delivered to the center of an extensional flow via inertial focusing. (D) High-speed microscopic images showing a focused cell entering the extensional flow region. (E) Definition of the shape parameters extracted from images. (F) Density scatter plot of size and deformability measurements. Images reproduced with permission from ref. 9. |
![]() | ||
Fig. 3 The principle of deterministic lateral displacement (DLD). (A) Three fluid streams (represented by red, yellow, and blue) enter the three lanes (represented by 1, 2, and 3, from left to right, respectively) at the first obstacle row. Small particles will follow the streamlines and fall repeatedly into the same lane. (B) Big particles travel laterally as they enter the next row of obstacles. Image reproduced with permission from ref. 11. |
DLD is a very effective and highly precise method for size-based particle/cell separation. For the separation of rigid spherical particles, the outcome is fairly predictable. However, there are often complications for biological cells because their deformation and shape can affect their trajectories and hence potentially change the outcome of the separation. A recent study by Beech et al. demonstrated how shape and deformability can be utilized to develop novel methods for cell separation.12 In this study, the authors employed deformable and disc-shaped RBCs. In a DLD device, the separation of RBCs is determined by the effective diameters of cells (Fig. 4A). The effective diameters of RBCs are dependent on multiple factors including: the state of the cell (Fig. 4B), the extent of cell deformation (Fig. 4C), and the orientation of the cell (Fig. 4D). Beech et al. suggested several simple methods for changing the effective diameters of cells (such as RBCs) in DLD devices. For example, they altered the depth of the DLD device to force cells to be oriented in certain ways (Fig. 4 E, F, and G) ; they also tuned the fluid shear rates to achieve different cell deformation (Fig. 4C). Both approaches altered effective cell diameters and thus changed the outcome of cell separation. The authors tested their theory using three red blood cell types in devices with different heights and at different shear rates and achieved very promising results. This work indicates great potential in the use of tailored DLD devices to separate cells based on their mechanical properties. One could envision that once we achieve a better understanding of cellular mechanical biomarkers for various diseases, such tailored DLD devices can be invaluable in many applications that require high-throughput, label-free cell separation, e.g., separation of circulating tumor cells (CTCs).
![]() | ||
Fig. 4 DLD-based separation by cell shape and deformability.12 (A) DLD separation based on effective cell diameter. (B) Different forms of RBCs in normal and abnormal states. (C) Shear-induced deformation of RBCs affects effective cell diameter. (D) Orientation affects effective cell diameter. (E, F, G) Channel depth affects orientation, hence effective cell diameter and cell trajectory in a DLD device. |
This journal is © The Royal Society of Chemistry 2012 |