K.
Van Hoecke
a,
C.
Catry
b and
F.
Vanhaecke
*a
aGhent University, Department of Analytical Chemistry, Krijgslaan 281 – S12, B-9000 Ghent, Belgium. E-mail: frank.vanhaecke@ugent.be; karen.vanhoecke@ugent.be
bJanssen Pharmaceutica NV, Turnhoutseweg 30, B-2340 Beerse, Belgium
First published on 3rd August 2012
Both new guidelines concerning elemental impurity limit concentrations and novel analytical procedures for elemental analysis of pharmaceutical substances are being adopted by the United States Pharmacopeia (USP). This paper outlines the optimization of two sample preparation procedures for organic pharmaceutical excipient products, one relying on direct dissolution in an aqueous medium and another on microwave-assisted acid digestion. Optimization particularly involved the stabilization and reliable determination of Hg, Pd and Os. Analyte concentrations and spike recoveries were determined in two organic pharmaceutical excipients using a quadrupole ICP-MS instrument equipped with a collision–reaction cell. Aqueous dissolution was achieved with 0.009 mM KBrO3 in 1% (v/v) HNO3 and 1% (v/v) HCl and microwave-assisted acid digestion was performed with aqua regia. Polyatomic interferences originally hampering the accurate determination of 51V, 52Cr, 53Cr, 55Mn, 56Fe and 75As were eliminated through introduction of He/H2 collision–reaction gas at a flow rate of 4.5–5.1 ml min−1 and applying a kinetic energy barrier of 3 V between the hexapole collision–reaction cell and the quadrupole analyzer. Limits of detection were more than a factor of ten lower than the proposed limit concentrations for parenteral administration and the impurity concentrations in the products tested never exceeded the limit concentrations defined by USP. Elemental spike recoveries ranged between 93.8 and 109.9% when standard addition was used for calibration, with relative standard deviations (RSDs) ≤12.7%. It is expected that the proposed procedures can be directly implemented for routine impurity analysis of a broad spectrum of pharmaceutical substances.
Already more than a century ago, the United States Pharmacopeia (USP) adopted a guideline for heavy metals in pharmaceutical substances, first published as Procedure 121, Time-limit Test for heavy elements1 and in its revised form currently known as USP Chapter 231, Heavy Elements Test.2 Though updated over time, the currently valid guideline describes an analytical procedure for heavy element determination that is considered obsolete and which does not reach the requirements requested for present and future impurity determination, given the available toxicological knowledge.2 The current procedure is based on a visual comparison between a sample and a lead (Pb) standard, treated identically for the precipitation of metals as sulphides in aqueous solution. It has been proven that the sulphide precipitation procedure provides inaccurate results for several important elements. Additional concerns with respect to the current procedure are, e.g., the subjective visual comparison between the sample and standard, the assumption that the sulphides formed in the sample have optical characteristics that are identical to those of the lead sulphide, the inability to identify and quantify specific heavy elements and the risk of loss of volatile elements. Furthermore, it has been experimentally confirmed that elemental recoveries are far from satisfying.3 More specifically, Lewen et al. (2004) found average recoveries <10% for Se, Sn, Sb, Ru and Hg and recoveries between 30 and 50% for Pb, As, Cd, Mo, Pd, Pt and In. Especially in view of the fact that Pb, As, Cd and Hg are among the most toxic metals, a more reliable standardized analytical procedure is urgently needed.
Recently, USP has published proposals for new general chapters concerning limit concentrations of various elemental impurities and proposed new analytical procedures to accurately determine them at the suggested limit concentrations. The respective documents,4,5 which are currently in the process of revision, are Chapters 232, Elemental Impurities – Limits, and 233, Elemental Impurities – Procedures. For Chapter 232, target elements were selected on the basis of toxicity and limit concentrations were calculated based on permitted daily exposure for a person with a body weight of 50 kg. In Table 1, the selected elements are listed together with the proposed concentration limit for a maximum daily dose of 10 g day−1 and for both parenteral or inhalational and other (oral, dermal, mucosal) administration routes. The iron and zinc limit concentrations listed in Table 1 were taken from the EMEA guideline.6 Chapter 233 contains information on the proposed sample preparation procedure and the experimental design when using inductively coupled plasma-optical emission spectrometry (ICP-OES) or inductively coupled plasma-mass spectrometry (ICP-MS) for sample analysis. Any other analytical procedure may be used as well, provided that compliance with the validation requirements listed in Chapter 233 can be proven.
Element | Oral component limit μg g−1 | Parenteral component limit μg g−1 | Internal standards | Set A | Set B | ||||
---|---|---|---|---|---|---|---|---|---|
Spike μg l−1 | Std. 1 μg l−1 | Std. 2 μg l−1 | Spike μg l−1 | Std. 1 μg l−1 | Std. 2 μg l−1 | ||||
As | 1.5 | 0.15 | Sc, Y | 1.5 | 0.75 | 3.0 | 1.5 | 0.75 | 3.0 |
Cd | 0.5 | 0.05 | Y, In | 0.5 | 0.25 | 1.0 | 0.5 | 0.25 | 1.0 |
Pb | 1 | 0.1 | Re, Tl | 1.0 | 0.5 | 2.0 | 1.0 | 0.5 | 2.0 |
Hg | 1.5 | 0.15 | Re, Tl | 0.75 | 1.5 | 3.0 | 0.75 | 1.5 | 3.0 |
V | 25 | 2.5 | Sc, Y | 7.5 | 5 | 10 | 25 | 12.5 | 50 |
Cr | 25 | 2.5 | Sc, Y | 7.5 | 5 | 10 | 25 | 12.5 | 50 |
Mn | 250 | 25 | Sc, Y | 7.5 | 5 | 10 | 50 | 25 | 100 |
Fe | 1300 | 130 | Sc, Y | 15 | 10 | 25 | 50 | 25 | 100 |
Ni | 25 | 2.5 | Li, Y | 7.5 | 5 | 10 | 25 | 12.5 | 50 |
Cu | 250 | 25 | Li, Y | 7.5 | 5 | 10 | 50 | 25 | 100 |
Zn | 1300 | 130 | Li, Y | 15 | 10 | 25 | 100 | 50 | 200 |
Mo | 25 | 2.5 | Y, In | 7.5 | 5 | 10 | 25 | 12.5 | 50 |
Ru | 10 | 1 | Y, In | 7.5 | 5 | 10 | 10 | 5 | 20 |
Rh | 10 | 1 | Y, In | 7.5 | 5 | 10 | 10 | 5 | 20 |
Pd | 10 | 1 | Y, In | 7.5 | 5 | 10 | 10 | 5 | 20 |
Os | 10 | 1 | Re, Tl | 7.5 | 5 | 10 | 10 | 5 | 20 |
Ir | 10 | 1 | Re, Tl | 7.5 | 5 | 10 | 10 | 5 | 20 |
Pt | 10 | 1 | Re, Tl | 7.5 | 5 | 10 | 10 | 5 | 20 |
Samples may be dissolved in aqueous or organic solvent or may be digested using a closed-vessel digestion procedure with concentrated acids. In order to check the appropriateness of any applied analytical protocol, elemental standards, sample solutions and spiked sample solutions should be prepared and analyzed. Accuracy and precision criteria for quantitative procedures include elemental recoveries between 70% and 150% and ≤20% relative standard deviation (RSD) on six independently prepared spiked samples.
Despite the combination of a detailed description of the analytical procedures in Chapter 233 and the large amount of freedom remaining for the analyst to develop an appropriate procedure that meets the acceptance criteria, the impurity analysis of a particular drug substance using ICP-OES and/or ICP-MS is not always straightforward. More specifically, the selection of a broadly applicable sample preparation procedure that allows quantitative recovery of all restricted elements and the development of a measurement protocol that can detect every element of interest with sufficient sensitivity, thereby avoiding important interferences, requires a systematic approach.
The present study describes the development of two sample preparation procedures, one relying on direct dissolution in an aqueous medium and one on microwave-assisted acid digestion, and the optimization of a set of instrumental parameters for adequate use of a quadrupole-based ICP-MS instrument equipped with a collision–reaction cell. Therefore, several spike recovery experiments were performed in order to finally propose two experimental protocols resulting in quantitative recovery with RSDs ≤ 15% for all USP and EMEA restricted elements.
Two organic products used as excipients in pharmaceutical products were used in this work, i.e. Neosorb 70/70 B (Roquette Frères, France) and Kollidon® 17 PF (BASF, Germany). The former product is a hydrogenated starch hydrolysate that appears as a viscous liquid, while the latter product is a white powder consisting of polyvinylpyrrolidone.
Standard mode | CCT mode | |
---|---|---|
a Optimized daily for good sensitivity (7Li+, 59Co+, 115In+, 238U+) and stability, low oxide (156CeO+/140Ce+) and doubly charged ion (138Ba++/138Ba+) formation. b Optimized for low background (78ArAr+, 80ArAr+, 51ClO+) to signal ratio and low oxide formation (156CeO+/140Ce+). | ||
Instrumental settings | ||
RF power (W) | 1200 | 1200 |
Plasma gas flow rate (l min−1) | 13 | 13 |
Auxiliary gas flow rate (l min−1) | 0.7 | 0.7 |
Nebulizer gas flow rate (l min−1) | 0.85–0.90a | 0.85–0.90 |
Sampling cone and type | Nickel, Xt | Nickel, Xt |
Skimmer cone and type | Nickel, Xt | Nickel, Xt |
Extraction lens voltage (V) | −117.6 to −160.8a | −117.6 to −160.8b |
Pole bias (V) | −4.2 to 0.1a | −17 |
Hexapole bias (V) | −3.7 to 0.4a | −20 |
CCT gas flow rate (ml min−1) | n.a. | 4.5 to 5.1b |
Data acquisition parameters | ||
Nuclides monitored | 7Li+, 60Ni+, 62Ni+, 63Cu+, 64Zn+, 65Cu+, 66Zn+, 68Zn+, 89Y+, 95Mo+, 98Mo+, 99Ru+, 101Ru+, 103Rh+, 105Pd+, 108Pd+, 111Cd+, 115In+, 187Re+, 189Os+, 191Ir+, 192Os+, 193Ir+, 194Pt+, 195Pt+, 200Hg+, 201Hg+, 202Hg+, 205Tl+, 206Pb+, 207Pb+, 208Pb+ | 45Sc+, 51V+, 52Cr+, 53Cr+, 55Mn+, 56Fe+, 75As+, 89Y+ |
Scanning mode | Peak jump | Peak jump |
Dwell time (ms) | 10 | 10 |
Acquisition points per peak | 1 | 1 |
Number of sweeps | 100 | 100 |
Number of readings | 3 | 3 |
Total acquisition time (s) | 132 | 33 |
As indicated, microwave-assisted acid digestion was performed to aid in sample dissolution, using a Milestone Microwave Labstation MLS-1200 mega (Milestone s.r.l., Italy). Each time, 0.250 g of product was digested in 2.5 ml of concentrated acid, tightly sealed in a 100 ml Teflon vessel inserted into a device-specific sarcophagus. The microwave program consisted of 1 minute heating at 250 W, 1 minute at 0 W, 4 minutes at 250 W, 4 minutes at 400 W, 4 minutes at 600 W and 5 minutes of ventilation. Before unscrewing, the vessels were taken out of the microwave unit and left to cool for several hours.
One hundred-fold concentrated spiking solutions containing all USP and EMEA restricted elements in 5% HNO3 + 1% HCl + 1% HF were freshly prepared before each analysis in order to produce spiked samples and standard solutions through addition of 0.250 ml of spike solution to 25 ml volumetric flasks. Two sets of elemental concentrations were used. Concentrations chosen initially are given in column ‘SET A’ of Table 1. For the final recovery experiments, ‘SET B’ concentrations of Table 1 were used. Except for the least toxic metals, i.e. Cu, Mn, Fe and Zn, the concentrations of spike and standards finally chosen correspond to the elemental impurity limit for parenterally administrated drug components at a component concentration of 10 g l−1. The concentrations of all USP and EMEA elements added to spiked samples and standards are given in Table 1, together with the limit concentrations in drug components for orally and parenterally administrated products.
The 0.250 ml of spike solution for elemental recovery determination was added to the appropriate samples before any sample preparation involving dissolution or digestion was performed. On the other hand, 0.250 ml of standard spike solution for preparation of external calibration or standard addition standards was added at the final step of sample preparation. Samples for direct dissolution in aqueous media were directly dissolved in the complete medium, while microwave-digested samples were quantitatively transferred to 25 ml volumetric flasks, which already contained 2.5 ml of 0.09 mM KBrO3 in 10% (v/v) HCl, and subsequently made up to volume with Milli-Q water.
In a second study, two different media were included, i.e. 0.009 mM KBrO3 in 5% HCl + 1% HNO3 and 0.009 mM KBrO3 in 1% HCl + 1% HNO3. The sample sets were prepared in duplicate and one set was analyzed the day after preparation, while the other was analyzed one day later. The concentrations used in spikes and standards are given in column ‘SET B’ in Table 1. Both external calibration and standard addition were applied.
Internal standards Li, Sc, Y, In, Re and Tl were added at a final level of 10 μg l−1. An indication of the internal standards used for each target element is given in Table 1.
The other product, Kollidon® 17 PF, was used in a recovery study performed after dissolution of 0.250 g product in 0.009 mM KBrO3 in the 1% HNO3 + 1% HCl medium in Milli-Q water.
Before transferring digested blanks and samples to volumetric flasks, 2.5 ml of 0.09 mM KBrO3 in 10% HCl was spiked into each volumetric flask. Hence, the final concentrations of KBrO3 and HCl were 0.009 mM and 1%, respectively, in case it was not added into the microwave vessels and 0.010 mM and 1.1%, respectively, in case KBrO3 in HCl solution was also added into the digestion vessels prior to sample digestion.
The Kollidon® 17 PF was digested in a HNO3/HCl (3:
1) mixture, as aqua regia could not completely digest the product with the heating program applied.
Internal standards Li, Sc, Y, In, Re and Tl were added at a final level of 10 μg l−1. An indication of the internal standards used for each element is included in Table 1.
In a second study, the same approach was used with 7 different rinsing solutions in two separate experiments. In the first experiment, the rinsing media included: 5% HCl, 10% HCl, 0.09 mM KBrO3 in 1% HCl and 0.9 mM KBrO3 in 10% HCl. Finally, an uptake and rinse experiment was performed using a 10 μg l−1 Hg standard in 1% HNO3 for uptake during 5 minutes and 0.09 mM KBrO3 in 1% HCl, 0.09 mM KBrO3 + 0.09 mM KBr in 1% HCl and 0.009 mM KBrO3 in 1% HCl solutions for rinsing. During rinsing, the rate of decrease in the 202Hg+ signal intensity was monitored for 12 minutes.
In order to stabilize mercury in samples, Hg should be dissolved in an oxidizing matrix. Standards of 0.75, 3 and 6 μg l−1 Hg were prepared in 1% HNO3, 0.09 mM KBrO3 in 1% HCl and 0.009 mM KBrO3 in 1% HCl and were left to stand for one week. Then, identical standard series were prepared and all solutions, both the fresh and the one week old ones, were measured. Mercury calibration curves in each medium were determined by measuring the standards in the order of 3, 0.75 and 6 μg l−1 Hg. In-between standards, 3 minutes washing with 0.009 mM KBrO3 in 1% HCl was applied. Calibration curves obtained on the basis of old and freshly prepared standards, respectively, were compared.
![]() | ||
Fig. 1 202Hg+ intensities measured during Hg standard solution uptake and subsequent rinsing with various solutions. |
It was expected that Hg standards prepared in a medium that could stabilize Hg would also prevent Hg accumulation in the introduction system during measurement and vice versa.
The KBrO3 solutions in 1% HCl also turned out to be able to stabilize Hg standards of 0.75–6 μg l−1 for at least one week. No difference was observed between standard series prepared in 0.09 mM and 0.009 mM KBrO3 (both in 1% HCl), respectively and between freshly prepared standard series and one week old ones. All standard series are shown in Fig. 2. Clearly, the freshly prepared 1% HNO3 solutions caused memory effects during the analysis, resulting in a non-linear calibration curve and high intensities. On the other hand, the Hg intensity in the old standards was much lower, probably because Hg volatilized out of solution. The large discrepancy between standard series prepared in KBrO3 solutions in 1% HCl and standard series prepared in 1% HNO3 illustrates the need for a medium specifically adjusted for Hg stabilization and measurement.
![]() | ||
Fig. 2 Hg standard series of 0.75, 3 and 6 μg l−1 prepared and measured twice in various media. One series was prepared one week prior to the measurements and one series on the same day of the measurements. |
Often a BrCl solution, prepared through mixing of KBr and KBrO3 solutions and addition of concentrated HCl is used in the context of Hg determination in environmental samples.7,8 It is a strong oxidizing solution that can break down organomercurials at room temperature and preserve Hg2+ in solution. As discussed above, in this study also a BrCl solution was tested at 0.46 mM, however, though efficient in resolving Hg memory effects, it was thought to be too aggressive for daily introduction into a mass spectrometer. As can be seen in Fig. 2, a more diluted KBrO3 solution was found to fulfill our needs of stabilizing inorganic mercury at the low μg l−1 concentrations, relevant for impurity determination within the context of USP and EMEA guidelines.
In subsequent recovery experiments, it was tested whether a KBrO3 solution in a diluted HCl + HNO3 mixture was suitable in the context of determination of impurity concentrations in solutions of pharmaceutical products. Correlation coefficients of calibration curves were ≥0.991 for external standards and ≥0.998 for standard addition standards. The sample sets analyzed after 24 h showed satisfying recovery results for all elements. For Pd, around 100% recovery was obtained with a relative standard deviation (RSD) <2%, both in the medium containing 5% HCl and in the medium containing 1% HCl. For Hg, the recovery was overestimated with external standardization, but a value of approximately 100% was obtained with standard addition.
The samples analyzed after 48 h showed a lower recovery for Os in the medium containing 1% HCl, while Pd recovery was still 100% after 48 h. Arsenic recovery was overestimated with external calibration, which was probably due to an enhanced ionization in the Neosorb-containing samples compared to the external standards without an organic matrix. It has been hypothesized that charge transfer reactions can occur between C+-ions and analyte element atoms. This charge transfer reactions only affect analytes with a first ionization potential (IP) lower than carbon itself, like As and Se, although other high IP elements do not show this specific effect.9–11 In general, the standard addition calibration gave more accurate recovery results compared to external calibration. Recoveries and standard deviations on average values (n = 3) determined in media containing 5% HCl and 1% HCl can be found in Table 3 and in Table 4, respectively.
Nuclide | Analyzed after 24 h | Analyzed after 48 h | ||||||
---|---|---|---|---|---|---|---|---|
Ext. std. | Std. add. | Ext. std. | Std. add. | |||||
Mean% | RSD% | Mean% | RSD% | Mean% | RSD% | Mean% | RSD% | |
51V | 101.9 | 1.0 | 102.2 | 1.0 | 97.7 | 0.8 | 99.3 | 0.7 |
52Cr | 102.6 | 1.0 | 102.2 | 1.0 | 98.5 | 0.2 | 99.3 | 0.2 |
53Cr | 103.1 | 1.2 | 103.2 | 1.2 | 99.4 | 1.0 | 99.9 | 1.0 |
55Mn | 104.8 | 1.0 | 104.8 | 1.0 | 101.0 | 1.0 | 101.5 | 1.0 |
56Fe | 107.3 | 1.6 | 105.2 | 1.8 | 102.9 | 0.3 | 101.9 | 0.3 |
60Ni | 99.9 | 1.0 | 102.0 | 1.1 | 97.9 | 0.4 | 99.2 | 0.4 |
62Ni | 100.4 | 1.3 | 101.5 | 1.3 | 97.9 | 0.7 | 99.2 | 0.7 |
63Cu | 103.7 | 0.9 | 105.5 | 1.0 | 101.4 | 0.4 | 102.1 | 0.4 |
64Zn | 109.3 | 7.3 | 103.5 | 7.8 | 109.0 | 5.7 | 98.4 | 6.4 |
65Cu | 104.3 | 1.0 | 105.4 | 1.1 | 101.4 | 0.4 | 102.5 | 0.4 |
66Zn | 109.5 | 7.0 | 103.4 | 7.5 | 103.9 | 5.8 | 94.6 | 6.4 |
68Zn | 109.9 | 7.0 | 103.1 | 7.5 | 104.0 | 5.7 | 94.5 | 6.3 |
75As | 132.5 | 7.2 | 96.5 | 11.5 | 136.1 | 1.8 | 101.2 | 1.8 |
95Mo | 99.8 | 1.4 | 100.5 | 1.4 | 99.5 | 0.5 | 99.7 | 0.4 |
98Mo | 99.6 | 1.1 | 100.0 | 1.1 | 99.5 | 0.6 | 99.4 | 0.6 |
99Ru | 104.3 | 1.2 | 101.7 | 1.3 | 100.5 | 0.5 | 100.6 | 0.5 |
101Ru | 105.1 | 1.3 | 101.8 | 1.3 | 100.6 | 0.4 | 100.9 | 0.4 |
103Rh | 99.9 | 0.8 | 100.7 | 0.8 | 100.2 | 0.5 | 101.0 | 0.5 |
105Pd | 97.6 | 0.9 | 100.9 | 0.9 | 101.1 | 0.4 | 101.4 | 0.4 |
108Pd | 96.9 | 1.0 | 100.4 | 1.0 | 101.2 | 0.3 | 101.8 | 0.3 |
111Cd | 104.6 | 1.0 | 105.7 | 1.0 | 100.5 | 1.2 | 101.9 | 1.2 |
189Os | 101.7 | 1.5 | 99.0 | 1.6 | 80.9 | 0.9 | 100.1 | 0.9 |
191Ir | 103.7 | 0.5 | 103.8 | 0.6 | 103.1 | 0.3 | 101.7 | 0.3 |
192Os | 102.2 | 1.6 | 98.9 | 1.7 | 80.3 | 1.1 | 100.4 | 1.1 |
193Ir | 103.7 | 0.3 | 103.8 | 0.4 | 106.0 | 0.5 | 102.4 | 0.5 |
194Pt | 104.0 | 0.5 | 104.6 | 0.5 | 102.4 | 0.5 | 99.3 | 0.7 |
195Pt | 104.3 | 0.4 | 104.9 | 0.4 | 102.4 | 0.4 | 101.5 | 0.4 |
200Hg | 116.5 | 1.1 | 106.4 | 1.2 | 112.2 | 3.1 | 102.9 | 3.1 |
201Hg | 115.7 | 1.3 | 106.1 | 1.4 | 112.5 | 2.3 | 102.8 | 2.3 |
202Hg | 117.7 | 1.0 | 107.7 | 1.1 | 112.9 | 2.9 | 103.0 | 2.9 |
206Pb | 93.2 | 0.9 | 104.7 | 0.5 | 106.2 | 1.5 | 101.7 | 1.8 |
207Pb | 93.7 | 0.7 | 106.1 | 0.7 | 104.1 | 1.9 | 101.9 | 1.9 |
208Pb | 94.1 | 0.4 | 105.7 | 0.3 | 103.9 | 1.7 | 102.4 | 1.8 |
Nuclide | Analyzed after 24 h | Analyzed after 48 h | ||||||
---|---|---|---|---|---|---|---|---|
Ext. std. | Std. add. | Ext. std. | Std. add. | |||||
Mean% | RSD% | Mean% | RSD% | Mean% | RSD% | Mean% | RSD% | |
51V | 100.0 | 0.7 | 100.4 | 0.7 | 97.9 | 1.6 | 98.0 | 1.6 |
52Cr | 99.8 | 0.2 | 101.4 | 0.2 | 94.8 | 2.2 | 98.8 | 2.0 |
53Cr | 100.8 | 1.6 | 101.3 | 1.6 | 93.9 | 1.5 | 99.3 | 1.4 |
55Mn | 104.0 | 0.3 | 104.4 | 0.3 | 100.2 | 1.8 | 99.8 | 1.7 |
56Fe | 102.5 | 0.6 | 103.4 | 0.6 | 101.1 | 3.0 | 102.2 | 2.9 |
60Ni | 97.7 | 0.4 | 99.1 | 0.4 | 83.9 | 0.8 | 99.8 | 0.6 |
62Ni | 97.8 | 0.9 | 99.9 | 0.9 | 83.2 | 0.3 | 99.8 | 0.2 |
63Cu | 101.8 | 0.5 | 102.7 | 0.5 | 101.8 | 0.4 | 102.7 | 0.4 |
64Zn | 98.6 | 14.6 | 104.1 | 14.4 | 106.7 | 4.3 | 107.9 | 4.2 |
65Cu | 101.9 | 0.6 | 102.7 | 0.6 | 101.5 | 0.4 | 102.4 | 0.4 |
66Zn | 98.4 | 14.7 | 103.7 | 14.4 | 107.7 | 4.1 | 107.8 | 4.1 |
68Zn | 98.8 | 15.1 | 104.1 | 14.9 | 107.8 | 4.1 | 107.7 | 4.1 |
75As | 122.7 | 4.7 | 104.8 | 5.0 | 125.7 | 1.7 | 98.5 | 1.7 |
95Mo | 99.1 | 0.7 | 99.5 | 0.6 | 98.3 | 1.0 | 99.2 | 1.0 |
98Mo | 99.7 | 0.8 | 99.2 | 0.8 | 99.7 | 0.8 | 99.3 | 0.8 |
99Ru | 125.5 | 0.6 | 99.3 | 0.6 | 165.3 | 0.4 | 100.4 | 0.4 |
101Ru | 126.1 | 0.8 | 99.7 | 0.8 | 165.3 | 0.3 | 100.4 | 0.3 |
103Rh | 100.2 | 0.6 | 100.0 | 0.6 | 101.5 | 0.4 | 100.4 | 0.4 |
105Pd | 99.2 | 0.6 | 99.3 | 0.6 | 100.8 | 0.8 | 101.5 | 0.9 |
108Pd | 99.0 | 0.6 | 99.0 | 0.6 | 100.8 | 0.8 | 101.4 | 0.8 |
111Cd | 101.3 | 1.6 | 102.2 | 1.6 | 99.3 | 0.8 | 101.2 | 0.8 |
189Os | 96.5 | 1.2 | 99.8 | 1.2 | 92.1 | 2.3 | 77.9 | 2.3 |
191Ir | 101.0 | 0.9 | 101.5 | 0.9 | 102.4 | 0.5 | 103.4 | 0.5 |
192Os | 96.3 | 1.2 | 99.6 | 1.1 | 91.6 | 2.1 | 76.8 | 2.0 |
193Ir | 101.5 | 1.0 | 102.0 | 1.0 | 102.6 | 0.6 | 104.2 | 0.6 |
194Pt | 101.5 | 0.9 | 102.6 | 0.9 | 102.3 | 0.9 | 103.1 | 0.9 |
195Pt | 101.3 | 0.9 | 102.5 | 0.9 | 101.9 | 0.8 | 103.2 | 0.8 |
200Hg | 112.7 | 0.8 | 102.6 | 0.9 | 105.1 | 1.4 | 103.3 | 1.4 |
201Hg | 114.2 | 0.7 | 105.2 | 0.7 | 103.8 | 1.1 | 102.0 | 1.1 |
202Hg | 114.0 | 1.0 | 104.0 | 1.0 | 105.2 | 1.0 | 103.4 | 1.0 |
206Pb | 102.2 | 1.3 | 103.4 | 1.3 | 96.4 | 2.4 | 106.0 | 1.9 |
207Pb | 101.9 | 1.2 | 103.4 | 1.2 | 94.8 | 2.8 | 106.0 | 2.3 |
208Pb | 102.5 | 0.9 | 102.5 | 1.0 | 95.8 | 2.1 | 105.4 | 1.8 |
The determination of volatile Pd compounds in pharmaceuticals has been the subject of a study by Al-Ammar and Northington (2011).12 In organic solvent, they could accurately determine Pd through conversion of the volatile compounds to non-volatile Pd-thioacetamide complexes. As for our experiments higher Pd concentrations were observed in blank digestions performed after spike recovery experiments, the second hypothesis of precipitation was more likely to explain the low recoveries observed (results not shown). Usually, aqua regia or different acid mixtures are used in the determination of the platinum group metals (PGMs), of which Pd is one, in a variety of samples.13 Therefore, it was expected that addition of concentrated HCl during digestion could aid in improving Pd recoveries.
However, osmium determination was most problematic. Apparently, Os was transformed into a volatile species during microwave digestion, which caused Os memory effects during measurements. Most likely, Os was oxidized to the volatile OsO4. Often, quantitative Os determination involves a procedure of oxidizing distillation and recovery of OsO4 in solution.14 Though, such a procedure is difficult to implement in a routine analysis method for large numbers of pharmaceutical substances. As Os spiked to standard solutions did not undergo the same digestion treatment, normal intensities were found for standard solutions. Interestingly, the Os intensities in standard addition standards were only around half of the intensities in external calibration standards. Consequently, Os recoveries were seriously overestimated by a factor of 5 with external standardization and by a factor of 10 with standard addition. The Os memory effects also erroneously increased 206Pb and 208Pb recoveries, due to intense 190Os16O+, 192Os14N+ and 192Os16O+ interferences. Therefore, the lead recovery calculated using the 207Pb isotope was more reliable. In conclusion, no satisfying results were obtained with concentrated HNO3 for microwave-assisted acid digestion of pharmaceutical products.
Subsequently, it was tested whether aqua regia digested samples would give better recovery results for Pd and Os. From Table 5 it can be seen that this was indeed the case. Now, Pd recoveries were 94% and Os recoveries 107%, which is well contained within the ranges outlined by USP. At the same time, the use of aqua regia did not negatively affect the recoveries of all other elements. Correlation coefficients of calibration curves were ≥0.992. Table 5 also gives determined LODs and LOQs. For Neosorb 70/70 B, not a single element listed as impurity was detected at a concentration higher than its LOQ.
Nuclide | LOD μg l−1 | LOD μg g−1 | LOQ μg l−1 | LOQ μg g−1 | Ext. std. | Std. add. | ||
---|---|---|---|---|---|---|---|---|
Recovery% | RSD% | Recovery% | RSD% | |||||
51V | 0.09 | 0.009 | 0.3 | 0.03 | 98.4 | 1.0 | 98.0 | 1.0 |
52Cr | 0.09 | 0.009 | 0.3 | 0.03 | 97.1 | 1.0 | 97.0 | 1.0 |
53Cr | 0.2 | 0.02 | 0.7 | 0.07 | 96.4 | 2.3 | 95.3 | 2.3 |
55Mn | 0.1 | 0.01 | 0.3 | 0.03 | 100.7 | 0.6 | 101.4 | 0.6 |
56Fe | 1 | 0.1 | 5 | 0.5 | 95.8 | 1.6 | 98.4 | 1.5 |
60Ni | 0.09 | 0.009 | 0.3 | 0.03 | 85.9 | 0.7 | 97.6 | 0.6 |
62Ni | 0.3 | 0.03 | 1 | 0.1 | 86.4 | 0.4 | 97.9 | 0.4 |
63Cu | 0.07 | 0.007 | 0.2 | 0.02 | 99.9 | 0.5 | 100.1 | 0.5 |
64Zn | 0.06 | 0.006 | 0.2 | 0.02 | 92.2 | 13.4 | 99.6 | 12.1 |
65Cu | 0.03 | 0.003 | 0.1 | 0.01 | 99.9 | 0.4 | 100.4 | 0.4 |
66Zn | 30 | 3 | 100 | 10 | 92.5 | 13.7 | 99.8 | 12.4 |
68Zn | 30 | 3 | 100 | 10 | 92.4 | 13.8 | 99.6 | 12.4 |
75As | 0.004 | 0.0004 | 0.01 | 0.001 | 116.1 | 1.3 | 94.9 | 1.4 |
95Mo | 0.09 | 0.009 | 0.3 | 0.03 | 99.4 | 0.4 | 97.1 | 0.4 |
98Mo | 0.08 | 0.008 | 0.3 | 0.03 | 99.0 | 0.5 | 97.3 | 0.5 |
99Ru | 0.004 | 0.0004 | 0.01 | 0.001 | 99.8 | 0.6 | 98.3 | 0.6 |
101Ru | 0.004 | 0.0004 | 0.01 | 0.001 | 99.7 | 0.4 | 98.4 | 0.4 |
103Rh | 0.001 | 0.0001 | 0.002 | 0.0002 | 100.3 | 0.5 | 99.2 | 0.5 |
105Pd | 0.1 | 0.01 | 0.5 | 0.05 | 100.4 | 12.5 | 93.8 | 12.7 |
108Pd | 0.1 | 0.01 | 0.4 | 0.04 | 100.7 | 12.3 | 94.5 | 12.5 |
111Cd | 0.003 | 0.0003 | 0.009 | 0.0009 | 95.6 | 1.3 | 99.1 | 1.3 |
189Os | 0.9 | 0.09 | 3 | 0.3 | 104.0 | 3.2 | 107.4 | 3.2 |
191Ir | 0.03 | 0.003 | 0.08 | 0.008 | 98.9 | 0.4 | 99.6 | 0.4 |
192Os | 0.9 | 0.09 | 3 | 0.3 | 103.7 | 3.3 | 107.3 | 3.3 |
193Ir | 0.03 | 0.003 | 0.09 | 0.009 | 98.6 | 0.4 | 99.2 | 0.4 |
194Pt | 0.003 | 0.0003 | 0.01 | 0.001 | 97.9 | 0.4 | 99.2 | 0.4 |
195Pt | 0.004 | 0.0004 | 0.01 | 0.001 | 97.8 | 0.3 | 99.5 | 0.3 |
200Hg | 0.02 | 0.002 | 0.08 | 0.008 | 108.9 | 1.5 | 102.7 | 1.5 |
201Hg | 0.05 | 0.005 | 0.2 | 0.02 | 109.5 | 1.3 | 101.9 | 1.2 |
202Hg | 0.03 | 0.003 | 0.09 | 0.009 | 109.5 | 1.6 | 101.5 | 1.6 |
206Pb | 0.05 | 0.005 | 0.2 | 0.02 | 96.8 | 0.5 | 101.0 | 0.5 |
207Pb | 0.05 | 0.005 | 0.2 | 0.02 | 95.9 | 0.8 | 101.2 | 0.8 |
208Pb | 0.04 | 0.004 | 0.2 | 0.02 | 96.1 | 0.6 | 100.4 | 0.6 |
It turned out that the procedure via direct dissolution gave good recoveries in this case, as can be concluded from Table 6. However, the microwave-assisted acid digestion procedure was not directly applicable to the Kollidon® 17 PF product, as incomplete digestion was observed in 2.5 ml of aqua regia. Therefore, 2.5 ml of a HNO3/HCl (3:
1) mixture was used to digest 0.250 g of the product instead. After the microwave program was finished, all samples seemed to be completely digested. However, the 50% overestimation of As recovery with external calibration, raises doubts about complete removal of the organic matrix. Probably, the samples were rather dissolved instead of digested. Unfortunately, due to the lower HCl contents, Os determination was again problematic because of important memory effects during analysis. All correlation coefficients of calibration curves obtained for standard addition were ≥0.998.
Nuclide | Impurity concentration | Direct dissolution in 0.009 mM KBrO3 in 1% HNO3 + 1% HCl | Microwave-assisted acid digestion in HNO3![]() ![]() ![]() ![]() |
|||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
Ext. std. | Std. add | Ext. std. | Std. add. | Ext. std. | Std. add | |||||||
Conc. μg l−1 | RSD% | Conc. μg l−1 | RSD% | Recovery% | RSD% | Recovery% | RSD% | Recovery% | RSD% | Recovery% | RSD% | |
51V | <LOQ | <LOQ | 99.0 | 0.9 | 97.4 | 0.9 | 99.4 | 1.4 | 99.0 | 1.4 | ||
52Cr | 0.2 | 1.3 | <LOQ | 99.0 | 0.9 | 97.6 | 0.9 | 97.7 | 1.5 | 97.0 | 1.5 | |
53Cr | <LOQ | <LOQ | 97.8 | 1.4 | 97.2 | 1.4 | 98.2 | 1.5 | 96.6 | 1.5 | ||
55Mn | <LOQ | <LOQ | 104.0 | 0.9 | 102.2 | 0.9 | 104.0 | 1.8 | 102.4 | 1.8 | ||
56Fe | <LOQ | <LOQ | 103.4 | 0.9 | 101.7 | 0.9 | 104.2 | 1.7 | 102.2 | 1.7 | ||
60Ni | 0.2 | 1.5 | <LOQ | 96.1 | 0.3 | 97.3 | 0.3 | 93.6 | 0.6 | 96.9 | 0.6 | |
62Ni | 0.3 | 0.4 | <LOQ | 97.0 | 0.2 | 97.5 | 0.2 | 93.0 | 0.4 | 96.8 | 0.4 | |
63Cu | 0.2 | 3.9 | <LOQ | 100.9 | 0.2 | 101.1 | 0.2 | 97.7 | 0.7 | 100.9 | 0.7 | |
64Zn | <LOQ | <LOQ | 96.9 | 5.8 | 102.0 | 5.9 | 98.6 | 5.4 | 94.3 | 6.3 | ||
65Cu | 0.2 | 4.3 | <LOQ | 100.5 | 0.2 | 100.7 | 0.2 | 97.6 | 0.8 | 100.9 | 0.8 | |
66Zn | <LOQ | <LOQ | 96.6 | 6.0 | 101.9 | 6.0 | 98.3 | 5.4 | 94.2 | 6.3 | ||
68Zn | <LOQ | <LOQ | 97.3 | 5.9 | 102.5 | 5.9 | 99.0 | 5.4 | 94.0 | 6.3 | ||
75As | <LOQ | 0.02 | 22.2 | 158.2 | 0.9 | 101.6 | 0.9 | 151.1 | 1.5 | 100.8 | 1.5 | |
95Mo | 0.13 | 0.74 | <LOQ | 97.4 | 1.2 | 96.6 | 1.2 | 96.3 | 0.7 | 97.0 | 0.7 | |
98Mo | 0.14 | 0.3 | <LOQ | 98.4 | 1.1 | 96.7 | 1.1 | 96.7 | 0.5 | 96.9 | 0.5 | |
99Ru | <LOQ | <LOQ | 133.3 | 0.1 | 98.4 | 0.1 | 98.2 | 0.4 | 99.9 | 0.4 | ||
101Ru | 0.03 | 0.6 | <LOQ | 133.4 | 0.2 | 98.5 | 0.2 | 98.6 | 0.4 | 99.6 | 0.4 | |
103Rh | <LOQ | <LOQ | 101.3 | 0.2 | 100.2 | 0.2 | 98.9 | 0.3 | 99.5 | 0.3 | ||
105Pd | <LOQ | <LOQ | 99.4 | 0.1 | 99.9 | 0.1 | 99.6 | 0.6 | 100.2 | 0.6 | ||
108Pd | <LOQ | <LOQ | 99.9 | 0.2 | 100.1 | 0.2 | 99.6 | 0.9 | 100.2 | 0.9 | ||
111Cd | <LOQ | <LOQ | 97.1 | 1.1 | 100.6 | 1.1 | 95.3 | 0.5 | 101.4 | 0.5 | ||
189Os | <LOQ | <LOQ | 92.6 | 1.0 | 97.5 | 0.9 | 654.9 | 12.1 | 710.3 | 13.0 | ||
191Ir | <LOQ | <LOQ | 97.7 | 0.2 | 99.9 | 0.2 | 91.2 | 0.9 | 99.9 | 0.5 | ||
192Os | <LOQ | <LOQ | 92.5 | 0.9 | 97.4 | 0.9 | 674.4 | 12.3 | 731.3 | 13.3 | ||
193Ir | <LOQ | 0.02 | 6.8 | 98.1 | 0.3 | 99.9 | 0.3 | 91.1 | 1.0 | 99.4 | 0.6 | |
194Pt | <LOQ | 0.05 | 3.3 | 95.8 | 0.2 | 100.7 | 0.2 | 88.6 | 1.1 | 99.6 | 0.4 | |
195Pt | <LOQ | <LOQ | 96.1 | 0.1 | 100.4 | 0.1 | 88.7 | 1.2 | 99.5 | 0.4 | ||
200Hg | <LOQ | <LOQ | 110.5 | 1.3 | 106.7 | 1.3 | 100.8 | 1.4 | 107.6 | 0.5 | ||
201Hg | <LOQ | <LOQ | 109.4 | 0.8 | 106.6 | 0.8 | 100.7 | 3.4 | 106.9 | 2.3 | ||
202Hg | <LOQ | <LOQ | 111.1 | 1.4 | 106.8 | 1.4 | 103.4 | 1.9 | 109.9 | 1.0 | ||
206Pb | <LOQ | <LOQ | 98.7 | 1.2 | 99.7 | 1.2 | 102.8 | 1.5 | 109.3 | 2.0 | ||
207Pb | <LOQ | <LOQ | 99.4 | 0.6 | 100.3 | 0.6 | 94.3 | 1.5 | 100.3 | 1.2 | ||
208Pb | <LOQ | <LOQ | 99.2 | 1.0 | 99.8 | 1.0 | 100.9 | 1.4 | 106.8 | 1.9 |
It is clear that Os impurity determination will be difficult in products that do not dissolve in water and cannot be digested using aqua regia. As a result, solubility or digestion issues can limit the applicability of the proposed analytical procedures.
![]() | ||
Fig. 3 Overview of limits of quantification obtained via microwave-assisted acid digestion with aqua regia and subsequent addition of KBrO3 to a final concentration of 0.009 mM, in comparison to the impurity limit concentrations outlined by USP, all expressed as μg g−1. |
In order to stabilize Hg solutions and to prevent memory effects in the sample introduction system, 0.009 mM KBrO3 + 1% HCl needs to be added to all solutions. Samples soluble in an aqueous medium can be directly dissolved in 0.009 mM KBrO3 + 1% HCl + 1% HNO3 in Milli-Q water, whereas for samples subjected to microwave-assisted acid digestion destruction in 2.5 ml of aqua regia, tenfold dilution with Milli-Q water containing 0.009 mM KBrO3 in 1% HCl is recommended. This medium was able to maintain Pd into solution and to prevent Os from causing memory effects in the sample introduction system, whereas other media were not.
Overall, elemental recoveries obtained for one excipient product, Neosorb 70/70 B, ranged between 85.9% and 126.1% when external calibration was used and between 93.8 and 107.4% when standard addition was applied. Relative standard deviations on mean recoveries (n = 3) were ≥0.2% and ≤12.7%. Due to the high LOD, RSDs for Zn were up to 15%.
A second excipient product, Kollidon® 17 PF, resulted in elemental recoveries from 88.6% to 158.2% with external calibration and from 94.0% to 109.9% when standard addition was used for calibration. Relative standard deviations ranged between 0.1 and 6.3%. As the product was not completely digested in aqua regia, a modified acid mixture consisting of HNO3:
HCl (3
:
1) was used for digestion. However, with the latter acid mixture, memory effects observed for Os rendered reliable determination of this element impossible.
With external calibration, the organic sample matrix caused a systematic overestimation of the As recovery, up to 158%. In general, more accurate results and more narrow recovery ranges were obtained with standard addition. However, with exception of As, the recoveries obtained with external calibration were still in agreement with the USP-defined allowable recovery range of 70–150%.
We suggest that the sample preparation and analysis procedures described here are broadly applicable to a wide range of pharmaceutical products. Additional attention might be necessary for Os determination in products that are not soluble in an aqueous medium and which cannot be digested in aqua regia.
Footnote |
† Electronic supplementary information (ESI) available: See DOI: 10.1039/c2ja30128h |
This journal is © The Royal Society of Chemistry 2012 |