Ekaterina
Nesterenko
a,
Oksana
Yavorska
b,
Mirek
Macka
ac,
Alexander
Yavorskyy
b and
Brett
Paull
*a
aIrish Separation Science Cluster, National Centre for Sensor Research, Dublin City University, Dublin 9, Ireland. E-mail: brett.paull@dcu.ie; Fax: +353 1 7005503; Tel: +353 1 7005060
bSchool of Chemical Sciences, Dublin City University, Dublin 9, Ireland
cAustralian Centre for Research on Separation Science, School of Chemistry, University of Tasmania, Private Bag 75, Hobart, TAS 7001, Australia
First published on 1st February 2011
Glycidyl methacrylate–ethylene dimethacrylate (GMA-co-EDMA) based monolithic porous layer open tubular (monoPLOT) columns (0.05 mm I.D., monolithic layer thickness ≈ 5 µm) have been fabricated using an automated column scanning technique, providing UV polymerisation at 365 nm. Columns were chemically modified to obtain desired diol groups on the surface, and the longitudinal homogeneity of the stationary phase was profiled using scanning capacitively coupled contactless conductivity detector (sC4D), before and after such modification. Using the automated scanning polymerisation technique, column-to-column production reproducibility, including longitudinal phase thickness, was within ±5% RSD. The prepared columns were tested to evaluate their liquid chromatographic stationary phase selectivity, efficiency and reproducibility, with a series of test protein mixtures. Under optimised gradient conditions, the separation of up to 8 proteins was demonstrated on the open tubular column (510 × 0.05 mm I.D.), with a column pressure drop of <1.5 MPa.
However, from as far back as the late 1970s there have been attempts to apply open tubular format columns to LC separations,16 although in most cases practical and instrumental limitations have resulted in only limited success being reported. These limitations include the obvious issue of low stationary phase capacity and difficulty in producing a stable longitudinally homogenous adsorbed or surface bonded stationary phase, together with the lack of sensitive small volume detectors compatible with micro-bore capillary columns, and the lack of LC instrumentation to provide effective generation of mobile phase gradients at sub-µL min−1 flow rates. Although the instrumentation issues have now largely been resolved, with many gradient capLC systems available commercially, the preparation of open tubular columns for high efficiency separations still remains a challenge, particularly surface bonded porous phases of a uniform layer thickness, which can provide sufficient retention of analytes and sample loading capacity. One of the solutions to this problem can be the use of porous polymeric stationary phases. Firstly, this type of stationary phase can provide a highly developed surface area which is relatively physically and chemically stable (especially at extreme pH values). Secondly, the use of porous polymeric stationary phases provides the possibility of varying the surface chemistry, through simple surface modification procedures, which can be carried out in situ. Up to date there have been two strategies developed for the preparation of polymeric phase open tubular columns. One requires the pre-fabrication of polymer particles, which are then bonded to the inner wall of the capillary,17 the second one is based on the in situpolymerisation of the immobilised polymeric phase as a single (monolithic) porous layer.18,19
Up to date there have been relatively few studies published on the application of OT and PLOT columns in LC, however, some impressive demonstrations of the potential of these columns for LC have been shown.19–23 For example, Yue et al.19 used a 4.2 m × 0.01 mm I.D. poly(styrene-divinylbenzene) (PS-DVB) PLOT column, combined with a 30 cm monolithic PS-DVB pre-column for solid phase extraction (SPE), for the efficient separation of peptides. Later the same research group19–22 extended the application of the reversed-phase PS-DVB PLOT column (3.2 m × 0.01 mm I.D.), in combination with a traditional strong cation-exchange column (SCX), to the two-dimensional separation of peptides. The method required only a very small volume of protein digest sample, equal to only 500 ng, and permitted the identification of over 2700 peptides, associated with 850 unique proteins. Luo et al.20 also produced a 2.5 m × 0.01 mm I.D. amino-group bonded poly(vinylbenzyl chloride-divinylbenzene) PLOT column, which was applied to the analysis of N-linked glycans, using a hydrophilic interaction chromatographic approach (HILIC). Rogeberg et al.23 also recently successfully demonstrated the potential of PS-DVB PLOT columns for the efficient separation of intact proteins.
In the work presented herein, relatively short, 0.30 m × 0.05 mm I.D. monolithic porous layer open tubular (monoPLOT) columns were prepared by in situ surface layer co-polymerisation of glycidyl methacrylate (GMA) and ethylene dimethacrylate (EDMA), followed by further stationary phase chemical modification for chromatographic application. Significantly, the longitudinal homogeneity of the obtained polymeric stationary phase layer could be characterised non-destructively using scanning capacitively coupled contactless conductivity detector (sC4D), both before and after any chemical modification steps. The fabricated monoPLOT columns were evaluated for the separation of test protein mixtures under low pressure conditions, using simple mobile phase gradients. The effects of gradient profile, mobile phase pH and composition, and column temperature were studied, with optimised conditions allowing the complete separation of the test protein mixture on the short monoPLOT column, with column pressure drops of below 2 MPa.
![]() | ||
Fig. 1 Schematic diagram of the automated scanning photo-polymerisation technique developed for the preparation of open tubular monolithic columns. |
Following polymerisation, residual monomers and solvents were washed off the monoPLOT column with methanol at a flow rate of 5 µL min−1, and the column dried under a flow of nitrogen for 15–20 min. Scanning electron microscopy (SEM) was used to characterise the open tubular monolith at multiple positions along the length of the capillaries to confirm structural homogeneity and measure monolithic thickness to the capillary wall. All samples were gold-sputtered with a 60 nm layer prior to imaging in order to minimise charging and improve the image contrast.
For evaluation and comparison of column permeability, a fully monolithic (non-PLOT) GMA-co-EDMA capillary column (270 × 0.05 mm I.D.) was prepared. Here, the fused silica capillary was pre-treated and silanised as described above, and the monomer mixture used consisted of 24 wt% glycidyl methacrylate, 16 wt% ethylene dimethacrylate, 30 wt% cyclohexanol, 30 wt% 1-decanol and 1 wt% AIBN (in respect to monomers). Polymerisation was performed for 20 hours at 60 °C. After completion of polymerisation, the prepared monolith was washed with ACN overnight at a flow rate of 1 µL min−1.
![]() | ||
Scheme 1 Modification of surface epoxy groups to polar diol groups using dilute sulfuric acid solution at 60 °C. |
![]() | ||
Fig. 2 Typical cross-sectional SEM images of the GMA-co-EDMA monolithic porous layer open tubular column (50 µm I.D.). |
![]() | ||
Fig. 3 SEM image of the GMA-co-EDMA monolith-capillary wall interface, for a 100 µm I.D., monolithic porous layer open tubular column. |
Although the use of SEM provides an accurate means to measure the thickness of the layer at each specified location, it is not a suitable method for the study of complete stationary phase homogeneity, as it is obviously destructive and rather time consuming. However, sC4D has been previously demonstrated as an effective non-invasive technique for this very purpose.24–27 In this application, sC4D can also be utilised for comparison of replicate columns and evaluation of the surface modification procedure. The overall sC4D response for the capillary-housed monoPLOT columns developed here is the sum of signal contributions originating from the residual silanol groups on the surface of the silica housing, from the monolith layer itself and attached functional groups on the surface of the polymer, and of the solution phase within the capillary. For the unmodified monoPLOT phase, any significant longitudinal variation observed is then a combination of variation in the capillary housing itself (assumed to be negligible), and from the thickness and density of the monolithic phase, which indirectly effects the volume of solution phase within the detection zone, and is reflected within the detector response. As here the monolithic phase itself was converted from a non-polar to polar phase, a low conducting solution phase, in this case simply dionised water, allowed a single profile of the columns before and after modification to confirm both physical and chemical homogeneity of the porous layer phase (see Fig. 4). Hence, the presence or absence of physical gaps in the porous layer, or significant thickness changes, or variation in the concentration of charged groups on the surface of the stationary phase, could be confirmed using this simple technique.
![]() | ||
Fig. 4 Scanning C4D of two individually prepared 50 µm I.D. monoPLOT columns, before and after modification to provide surface diol groups. |
Two individually prepared 270 mm long GMA-co-EDMA monoPLOT columns were scanned, recording the signal every 1 mm (Fig. 4, bottom trace). After conversion of the two columns to a surface diol functionality, the procedure was repeated (Fig. 4, top trace). The recorded profiles confirmed that the stationary phase layer exhibited a homogeneous structure, as the profiles for non-modified columns were practically flat without significant deviations, and the same pattern was observed after modification. Also, significantly, it can be seen from comparing the scans of the two columns that column-to-column fabrication reproducibility, including surface modification, was excellent, with less than ±5% variation between the two replicate columns, both before and after modification.
The same approach was applied to the monoPLOT columns produced within this study. The 270 × 0.05 mm I.D. monoPLOT column provided a backpressure of 1.2 MPa at a flow rate of 0.2 µL min−1 using water as the mobile phase, and the calculated permeability found to be 7.2 × 10−12 m2, while for the bare 0.05 mm I.D. fused silica capillary of the same length this value was found to be approximately 4.5 times higher. The lower permeability and higher pressure drop for the monoPLOT column are certainly due to the presence of porous layer, which reduces the internal diameter of the open tube. Compared to the 0.01 mm I.D. PLOT column,19 the permeability of the PLOT column studied in the current work was 5.5 times higher, which was obviously due to the larger diameter of the open tube.
The permeability of the monoPLOT column was also compared to that of a fully monolithic GMA-co-EDMA capillary column (non-PLOT), prepared in a 270 × 0.05 mm I.D. fused silica capillary. For this column the backpressure was 6 MPa at a flow rate of 1.3 µL min−1 for ACN. The calculated permeability was found to be 1.9 × 10−14 m2, which is 360 times lower than for the monoPLOT column.
In the work of Karger and co-workers,19–22 and later by Rogeberg et al.,23PS-DVB PLOT columns based upon 10 µm I.D. fused silica capillaries, of up to 5 m in length, with approximately 0.5 to 1 µm thick porous layers were investigated. The long PLOT columns were used in gradient LC applications, so no data on column efficiency for intact proteins or digest peptides were presented from either group. However, under such conditions, despite the long analysis times, peak capacities were impressive for open tubular LC.
In the work presented herein, a similar ratio of capillary I.D. to porous layer thickness was achieved (10:
1), although the resultant shorter wider bore columns exhibited a much lower column backpressure. However, a 10-fold reduction in capillary column length compared to the above studies would significantly reduce peak capacities for such separations, although very low pressure separations of simpler mixtures of large biomolecules and intact proteins were achieved.
Initial chromatographic performance was investigated using a test mixture of 8 proteins, namely ribonuclease, insulin, cytochrome C, horseradish peroxidase, carbonic anhydrase, enolase, alcohol dehydrogenase and phosphorylase. A simple 20 min ACN–water mobile phase gradient (constant 0.1% TFA) was applied from 0 to 90% ACN, with a column temperature of 20 °C to evaluate relative retention of each of the above analytes on the monoPLOT column. The column backpressure over the flow rate range of 0.1 to 1.5 µL min−1 was also recorded under both starting (100% water) and final gradient conditions (90% ACN). The flow rate was increased at increments of 100 nL min−1 and the backpressure was allowed to stabilise at each flow rate value. Fig. 5(a) shows that backpressure linearly increases with the increase in flow rate under both conditions, indicating a stable porous layer. The highest flow rate applied was 1.5 µL min−1, giving the column backpressure of only 6 MPa (excluding extra-column system backpressure).
![]() | ||
Fig. 5 (a) Column backpressure dependence on mobile phase flow rate. (b) Overlayed chromatograms of 8 proteins on the monoPLOT column. Mobile phase A: 0% ACN, 0.1% TFA; B: 90% ACN, 0.1% TFA. Gradient elution: 0.0–20.0 min from 100% A to 100% B; 20.0–20.1 min from 100% B to 100% A; 20.1–60 min 100% A. Column: diol functionalised GMA-co-EDMA monoPLOT, 270 × 0.05 mm I.D. UV detection at 214 nm. Temp. = 25 °C. |
The above mobile phase gradient was applied to the separation of the test protein mixture at each of the above flow rates. Fig. 5(b) shows the initial overlayed chromatograms of the resultant separations for each of the 8 proteins.
The relationship between the temperature and gradient retention factor (kg) is illustrated in the Van't Hoff plot shown in Fig. 6 constructed using retention data from the repeat injection of the mixed protein standard at increasing temperature. Gradient retention factor kg can be determined as:29
kg = (tr/tm) − 1 | (1) |
![]() | ||
Fig. 7 The effect of mobile phase pH and concentration of TFA on the separation of proteins: 1—ribonuclease, 2—insulin, 3—cytochrome C, 4—horseradish peroxidase, 5—carbonic anhydrase, 6—enolase, 7—alcohol dehydrogenase and 8—phosphorylase. Column: diol functionalised monoPLOT (510 × 0.05 mm I.D.) Gradient conditions as in Fig. 4. TFA concentrations (a) 0.01%, pH 3.4, (b) 0.05%, pH 2.8, (c) 0.2%, pH 1.9. Flow rate = 0.4 µL min−1. UV detection at 210 nm (blank baseline subtracted). |
Footnote |
† Electronic supplementary information (ESI) available. See DOI: 10.1039/c0ay00649a |
This journal is © The Royal Society of Chemistry 2011 |