Rapid method to confirm suitability of laboratory LC systems

Gregory K. Webster *a, Peter D. Redfern a and Sean J. Orlowicz b
aGlobal Analytical R&D, Abbott Laboratories, Abbott Park, IL 60046, USA. E-mail: gregory.webster@abbott.com; Tel: +1 847 935-1295
b411 Madrid Avenue, Torrance, CA 90501, USA. E-mail: SeanO@phenomenex.com; Tel: +1 310 212-0555

Received 7th September 2009 , Accepted 27th November 2009

First published on 16th December 2009


Abstract

A simple method for the routine assessment of system performance for liquid chromatographs is presented. The method is a novel use of a commercially available test mix and simple gradient profile to challenge system precision and linearity in order to assess the chromatographic suitability of the system for GxP applications.


A. Introduction

One of the many challenges analysts face in regulated industries is maintaining the qualification (“suitability”) of their instrumentation. In the pharmaceutical industry, the “readiness” of an instrument is documented in its instrument qualification, operational qualification and performance qualification, known as IQ/OQ/PQ.1 These qualifications are detailed tests verifying that the instrumentation being used is meeting manufacturer specifications. It is important to note that IQ/OQ/PQ criteria should never exceed manufacturer specifications.2 If a company desires tighter specifications, either an alternate vendor should be chosen or the company should work with the vendor(s) to ensure such specifications are within the original design qualification for that instrument.

For chromatographic instrumentation, the IQ/OQ and PQ are initially established at the time of installation. Depending on the usage, the PQ (or equivalent) is re-performed on a scheduled basis. In the interim, pharmaceutical laboratories use method or system suitability criteria to establish the validity of analytical runs.3 Typically, this includes (1) a series of standard injections to establish suitable precision, (2) resolution, tailing and/or plate number determinations to establish suitable chromatographic performance and (3) injection of a practical detection limit or quantification limit standard to establish appropriate sensitivity.

In our laboratory, we have found using the suitability of the various methods we have running at any one time is not a good enough diagnostic tool for LC performance. Several older instruments were running with periodic failures in meeting suitability criteria. This is a significant issue for it entails not only a regulatory issue in quality audits, but incurs significant costs in terms of repeating assays, performing investigations and completing paperwork.

To address this issue, a quick instrumentation-independent suitability method was developed for use as a routine diagnostic tool to investigate instrument performance. The criteria for such a method was (1) be “simple” so that analysts would actually use it and (2) be efficient so minimal cost in analyst and instrument time would be incurred. The final criterion was that the method must be able to detect performance issues so these issues could be addressed prior to the instruments being released for routine GxP testing. An added bonus is that the sequence only takes a few hours or can run overnight. The method presented here meets proposed criteria for UV and fluorescence detection using a commercially available test mix.

B. Reagents

 1. WaterLC Grade or equivalent.

2. Acetonitrile, LC Grade or equivalent.

3. Test Mix, Reversed Phase #2 Test Mix, Phenomenex (Part # ALO-3045). This is a standard test mixture containing Uracil (void volume marker), Acetophenone, Benzene, Toluene and Naphthalene in Acetonitrile.

C. Equipment parameters and conditions

 1. An HPLC system with gradient elution capability, in-line mobile phase degasser, auto-sampler with 2–25 μL injection volume capability and either an ultraviolet absorbance detector, a fluorescence detector or both in series.

2. A data acquisition system.

3. Phenomenex Gemini C18 HPLC column, 100 mm × 4.6 mm, 3 μm (Phenomenex, Part # 00D-4439-E0). (See Tables 1 and 2)

Table 1 Method
1 UV Wavelength 254 nm
F Wavelength Ex: 284 nm
Em: 324 nm
2 Flow Rate: 1 mL min−1
3 Injection Volume 2–25 μL
4 Column Temperature: 30 °C
5 Mobile Phase A: Water
B: Acetonitrile


Table 2 Suitability Gradient Profile
Time (min.) %A % B
0.00 80 20
2.00 80 20
20.00 20 80
22.00 20 80
22.10 80 20
Re-equilibrate 10 min.
Run Time: 22 min


D. Solution preparations

Preparation of test solution

 Prepare a 1[thin space (1/6-em)]:[thin space (1/6-em)]100 dilution of the Phenomenex Reversed

Phase #2 Test Mixture using Acetonitrile.

Preparation of needle rinse (if necessary)

 Prepare an 80[thin space (1/6-em)]:[thin space (1/6-em)]20 mixture of Acetonitrile/Water.

E. System preparation

 1. Purge injector.

2. Reprime all lines used.

NOTE: Priming of each line should require a minimum of about 20 mL of mobile phase to be drawn to waste as most degassers retain a volume of roughly 15 mL. Adjust this volume as necessary based on the specification of the instrument in question.

3. Allow the system to pump at the initial gradient conditions for at least 30 min, prior to initiating any test runs, to allow the column to equilibrate.

F. System suitability and analysis

 1. Inject Acetonitrile. This injection is intended to complete the equilibration of the system prior to performing the analysis.

2. Make a second injection of Acetonitrile (Blank) and confirm that there are no peaks, other than void volume disturbances, visible in the resulting chromatogram.

3. Make six 10 μL injections of the Test Solution, calculate the %RSD of the peak areas for the six injections. Typical Retention Times are shown in Table 3 and a typical Chromatogram can be seen in Fig. 1.

Table 3 Representative Retention Times for the Test Solution
Analyte RT/min RRT
Uracil 1.1 (unretained) 0.12
Acetophenone 8.9 1.00
Benzene 12.3 1.38
Toluene 14.7 1.65
Naphthalene 16.4 1.84



Representative Chromatograms of the Test Solution. (A) Uracil, (B) Acetophenone, (C) Benzene, (D) Toluene, (D) Naphthalene. Retention time.
Fig. 1 Representative Chromatograms of the Test Solution. (A) Uracil, (B) Acetophenone, (C) Benzene, (D) Toluene, (D) Naphthalene. Retention time.

Requirement: The %RSD should be NMT 1.0.

4. Inject Acetonitrile (blank).

Requirement: There should be no significant carryover of any of the peaks present in the Test Solution.

5. Make a 2, 5 and 8 μL injection of the Test Solution. These are the first 3 injections used to determine injector linearity.

6. Make three 10 μL injections of the Test Solution, for use as a bracketing standard.

7. Make a 12, 15, 20 and 25 μL injection of the Test Solution. These are the final 4 injections used to determine injector linearity.

8. Make three 10 μL injections of the Test Solution, for use as a bracketing standard.

9. Perform a linearity calculation using the data from steps 5 and 7. Plot injection volume vs. peak area and calculate the correlation coefficient.

Requirement: The correlation coefficient should be NLT 0.9950.

G. Results and discussion

The procedure is designed so that the system is first challenged to meet system precision requirements of USP.3

After injection of a standard, it is important that the autosampler does not exhibit carryover greater than the practical qualification limit for the LC. This value is typically NMT 0.05% of the nominal method standard. This is where many suitability checks stop. However, systems in use need to be challenged with more than simple precision and carryover.

The procedure moves to challenge not only the linearity of the detector, but the proportionality of the injection system. While it is not critical that an injection value be exactly 10 μL, it is important that the 8 μL and 12 μL injections are 80% and 120% of the nominal 10 uL value respectively. Current LCs on the market today easily post a correlation coefficient of NLT 0.999 from 2–25 uL. For single standard methods, injector precision is more critical than injector accuracy. However, knowing the range of proportionality for each LC injector is an important diagnostic tool as well as documentation of the qualified range in which the instrument can be used in regulated methods. It works as a diagnostic for the detector and autosampler.

The recovery of each standard from the bracketed standard is a better diagnostic of the acceptable injector range than linearity. Each injected volume should be accurate to 98–102% recovery of the theoretical recovery.

In Table 4, the values for a LC system suspected of under performing are presented. Note the high RSD for the precision injections and the poor accuracy of the injections in the linearity run. It should also be noted that correlation is a poor diagnostic tool. Many labs would consider R = 0.99 good enough to illustrate an injector is performing well. Here, the injector is not suitable for GxP chromatography.

Table 4 Suitability Results from an underperforming LC
Uracil Acetophenone Benzene Toluene Naphthalene
Rt 1.166 8.249 11.527 14.035 15.706
Rt RSD 0.303 0.411 0.141 0.126 0.189
Std 5.881 253.955 111.047 140.118 178.658
Area RSD 2.864 6.706 6.475 6.592 6.593

Uracil Acetophenone Benzene Toluene Naphthalene
Area Area Recovery (%) Area Recovery (%) Area Recovery (%) Area Recovery (%)
uL 2 37.915 74.6 16.461 74.1 20.839 74.4 26.584 74.4
5 118.217 93.1 51.639 93.0 65.173 93.0 82.936 92.8
8 124.067 61.1 54.274 61.1 68.433 61.0 87.191 61.0
10 241.912 95.3 105.963 95.4 133.587 95.3 170.329 95.3
12 299.343 98.2 130.712 98.1 164.610 97.9 209.991 97.9
15 379.953 99.7 166.005 99.7 209.651 99.8 266.993 99.6
20 512.010 100.8 222.902 100.4 281.141 100.3 358.673 100.4
25 619.105 97.5 269.649 97.1 340.619 97.2 433.644 97.1
k′ 0.2 7.3 10.6 13 14.7
T 1.2 1.1 1 1 1
N 2932 36439 54630 88684 128520
r 0.99224 0.99223 0.99226 0.99220


The LC instrument from Table 4 was subsequently moved offline for maintenance by the vendor. The vendor determined the autosampler needed to be replaced. After the vendor was finished, the method was again run on this system. The results are listed in Table 5. A significant improvement in precision, accuracy of injection and linearity of response is found. This is the performance expected of GxP instrumentation and with our suitability method, we have documentation that our system is performing at this level.

Table 5 Suitability Results from an acceptable LC
Uracil Acetophenone Benzene Toluene Naphthalene
Rt 1.153 8.307 11.599 14.076 15.76
Rt RSD 0 0.34 0.195 0.226 0.206
Std 7.840 351.144 163.360 199.767 251.101
Area RSD 0.384 0.151 0.077 0.001 0.002

Uracil Acetophenone Benzene Toluene Naphthalene
Area Area Recovery (%) Area Recovery (%) Area Recovery (%) Area Recovery (%)
uL 2 71.382 101.6 32.590 99.7 40.445 101.2 51.124 101.8
5 178.301 101.6 82.815 101.4 101.447 101.6 127.555 101.6
8 275.660 98.1 128.230 98.1 156.978 98.2 197.619 98.4
10 351.519 100.1 163.448 100.1 199.765 100.0 251.104 100.0
12 413.062 98.0 192.760 98.3 235.777 98.4 295.765 98.2
15 520.729 98.9 242.474 99.0 296.888 99.1 373.074 99.1
20 689.016 98.1 321.501 98.4 393.271 98.4 493.770 98.3
25 855.746 97.5 399.462 97.8 489.939 98.1 614.859 97.9
k′ 0.2 7.3 10.6 13 14.7
T 1.2 1.1 1 1 1
N 2932 36439 54630 88684 128520
r 0.99993 0.99994 0.99996 0.99995


Another novel aspect of this method is its ability to challenge the performance of the pumps and gradient proportioning ability by monitoring the resolution of the analyte peaks. Under current evaluation is whether we can use resolution to monitor the pumps and gradient proportioning value.

Fluorescence detection

In the course of this investigation, the need to qualify a fluorescence detector arose. Knowing the test mix contained a fluorophore (naphthalene), the challenge of suitability method was also used with florescence detection. Listed in Table 6, the method illustrated that the fluorescence detector used in series with the UV detector was working suitably for GxP studies.
Table 6 Suitability Results from an acceptable fluorescence detector
Naphthalene (UV) Naphthalene (F)
Rt 16.182 16.247
Rt RSD 0.073 0.077
Std 238.847 23.396
Area RSD 0.562 0.716

Naphthalene (UV) Naphthalene
Area Recovery (%) Area Recovery (%)
uL 2 48.240 101.0 4.742 101.3
5 118.802 99.5 11.661 99.7
8 191.061 100.0 18.711 100.0
12 287.179 100.2 28.034 99.0
15 360.211 100.5 34.991 99.7
20 477.044 99.9 46.236 98.8
25 595.051 99.7 57.697 98.6
r 0.99998 0.99998


H. Conclusion

The method proposed here is an effective diagnostic tool for routine suitability and diagnostic testing of liquid chromatographs. The method can be used for maintenance or for spot checking LCs such as prior to use for lengthy projects where small changes in performance are not acceptable. In addition, the method can serve as a problem solving tool for validation and CAPA (corrective action/preventative action) investigations.

Notes and references

  1. P. Bedson and M. Sargent, Accredit. Qual. Assur., 1996, 1, 265–74 CrossRef.
  2. C. Burgess, D. C. Jones and R. D. McDowall, Analyst, 1998, 123, 1879–86 RSC.
  3. Chromatography: USP General Chapter <621>, US Pharmacopeia 32, 2009, United States Pharmacopeia: Rockville, Maryland 20852-1790, USA Search PubMed.

This journal is © The Royal Society of Chemistry 2010
Click here to see how this site uses Cookies. View our privacy policy here.