Assessment of a procedure to determine trace and major elements in atmospheric aerosol

Giuseppa Toscano a, Andrea Gambaro ab, Gabriele Capodaglio *ab, Warren R. L. Cairns b and Paolo Cescon b
aDept. Environmental Sciences, University Ca' Foscari, Dorsoduro, 2137, Venice, Italy. E-mail: capoda@unive.it; Fax: +39 041 2348549
bInst. for the Dynamics of Environmental Processes-CNR, Dorsoduro, 2137, Venice, Italy

Received 20th March 2008 , Accepted 13th October 2008

First published on 19th November 2008


Abstract

The determination of trace elements in atmospheric particulate is affected by a number of problems that arise from some critical points such as the blank of the filters, sample heterogeneity and pre-analytical treatments. In the framework of a monitoring campaign conducted in the Venice Lagoon the analytical methodology for the determination of 20 trace elements (Al, As, Ca, Cd, Co, Cr, Cs, Cu, Fe, K, Li, Mg, Mn, Na, Ni, Pb, Rb, Sr, V, Zn) in atmospheric particulate samples by inductively coupled plasma quadrupole mass spectroscopy (ICP-QMS) has been optimized taking into account the individual critical points. Tests were carried out to estimate the blank contributions, and minimize the detection limit (LOD), measurements were also carried out to evaluate the accuracy and the repeatability. To obtain a complete dissolution of aerosol dust material and good recoveries of the elements, the acid mixture and the microwave assisted digestion program were optimized. The blank contributions from membrane filter manipulation and transportation prior to exposure were tested for the slotted and back filters by placing them on the sampling device for some minutes without air flowing to obtain field blanks (FBs). The contribution to the blank values of passive deposition and by contact with the samplers (quoted as campaign blanks, CBs) was measured by exposing the membranes throughout the sampling session (fifteen days) without any air flow. Instrumental ICP-QMS parameters were optimized and calibration curve intervals were selected on the basis of the necessity of simultaneous determination of the elements present at different levels of concentration. The limits of detection for each elements and the investigated method were suitable to determine the 20 elements reported above in the atmospheric aerosol fractionated in 6 classes ranging between 10 to 0.49 µm. It allows the determination of trace elements in aerosol in a large range of concentrations that can be observed in areas characterized by remarkable variability and regions with different levels of contamination.


Introduction

In the last few years much attention has been given to the evaluation of the elemental content of airborne particulate matter due their detrimental effects on human health. Many epidemiological studies have revealed that the degree of adverse respiratory effects depends on the physical and chemical properties of atmospheric aerosol.1–3 The particulate matter with an aerodynamic diameter <10 µm (PM10) constitutes the inhalable fraction of aerosol while the particulate matter with an aerodynamic diameter <2.5 µm (PM2.5) could have more serious toxic effects as it constitutes the respirable fraction of the aerosol. Under this framework, European Commission (EC) legislation requires that member states monitor the PM10 and the lead concentration in atmosphere.4 Furthermore the EC has recently proposed the monitoring of other toxic elements such as arsenic, cadmium and nickel in atmospheric particulate.5

Since the atmospheric particulate samples are often constituted of minute quantities of dust (fractions of milligrams) collected onto large cellulose filters, as such total sample digestion is widely used for trace element analysis. Various methods using microwave-assisted digestion with different acid mixtures and different heating programs are described in the literature.6–9 Yet in the total digestion of aerosol samples, elements contained as impurities in the cellulose material as well as the large amounts of chemicals required could give high blank levels and matrix effects during analysis. Therefore a careful quality control and optimisation of the analytical procedure is indispensable.

The objectives of the present study were to develop a suitable methodology for sampling and for the quantitative determination of trace elements in atmospheric particulate with an accurate blank control, adequate detection limits, recoveries, accuracy and repeatability. The analytical methodology was then used to study the PM10 composition in relation to the content of Al, As, Ca, Cd, Co, Cr, Cs, Cu, Fe, K, Li, Mg, Mn, Na, Ni, Pb, Rb, Sr, V and Zn in the aerosol of some sites of the Venice area with different anthropic influences.

Because of the significant differences between the local sources, we expect a large variability for elements at the trace level; therefore the analytical methodology will be used to detect an extended interval of concentrations.

Materials and methods

Size-fractionated PM10 aerosol samples were collected by brushless high volume samplers (Tisch Environmental Inc., Village of Cleves, Ohio, USA) equipped with a six-stage cascade impactor separating particles in the following dimensional aerodynamic defined classes: 10–7.2 mm, 7.2–3 mm, 3–1.5 mm, 1.5–0.95 mm, 0.95–0.49 mm, <0.49 mm. Slotted cellulose filters (14 × 15 cm) were used to collect the first five fractions and a cellulose filter paper (Whatman grade 41, 20 × 25 cm) was used as a back filter at the end of impactor outlet to inertially collect particles of size <0.49 mm, the operating flow was 1.13 m3 min−1.

Aerosol sampling was carried out over four campaigns from March 2002 to June 2003 at four sites in the Venice area. Sampling sites, as well wind direction and speed were chosen in order to evaluate the composition of the inorganic pollutants in areas affected by different sources (Fig 1). Marine and “long range” sources were monitored at site 2.


Sampling sites in the Venice Lagoon area.
Fig. 1 Sampling sites in the Venice Lagoon area.

(“P. Sabbioni” station, lat. 45°25′21.8″ N, long. 12°26′12.2″ E), located on the Northern inlet of the lagoon on the light tower when winds were blowing from the south-east. Aerosol from urban and industrial sources were collected at site 1 (“Moranzani” station, lat. 45°25′38.5″ N, long. 12°12′47.6″ E) located south of the Mestre urban area and the Porto Marghera industrial area at site 4 (“Tessera” station, lat. 45°29′37.9″ N, long. 12°19′27.2″ E), located close the Venice Airport when winds were blowing from north-east. Aerosol less affected by local sources was sampled at site 3 located on Monte Grande in the Euganei Hills, a wooded area at 460 m above sea level and 50 Km from Venice (“Teolo” station, lat. 45°21′43.0″ N long. 11°40′22.4″ E), when wind blew from the north-east. For all the sites aerosol samples were taken when winds blew at a speed greater than 1 m s−1. The sampling time was variable as a function of metereological characteristics, therefore the gathered volume of each sample was changeable, it ranged between 500 m3 to 12000 m3.

Filter and vessels cleaning

All the analytical procedures (cleaning of filters and all plastic items, filter weighing, sample treatment, preparation of standard solutions) were performed under a laminar flow clean area in a class 100 clean chemistry laboratory. Ultrapure water was obtained from a purification system (PURELAB ULTRA from Elga Labwater/VWS - UK Ltd) and suprapure grade reagents (HCl and HNO3 Merck) were used for all cleaning procedures.

Preliminary analysis carried out on untreated filters showed an extremely variable trace metal content, e.g. the content (SD) in µg/filter of Al, Ca Cd, Cu, V and Zn were 8(3), 23(21), 0.071(0.004), 0.20(0.04), 2.3(0.1) and 2.0(1.8) respectively. Therefore the filters before use were washed by HCl suprapure grade diluted 1:100 by ultrapure water and rinsed with ultrapure water, the procedure was repeated then the membranes were dried at room temperature in a class 100 laminar flow area. The procedure reduced the content for some elements of about one order of magnitude and averall increase significantly the repeatability of blank measurement. All filters were conditioned for 24 h at a temperature of 20° ± 5 °C and a relative humidity of 30 ± 5% and were then weighed by a balance with 0.01 mg precision (CP225D Sartorious). Filter weighings were repeated at least three times to assess aerosol masses ≥ 5 mg with a standard deviation of 5–10%. The slotted filters were cut into 4 equal sections and the backup filters were cut into 6 equal sections, using a polyethylene cutter. The aerosol samples were then closed in washed petri dishes inside plastic bags and stored at –20 °C until analysis. Acid cleaned low density polyethylene (LPDE) bottles were used to store the digested samples. Blank filters was washed, dried and weighed following the same analytical protocol used for the sample filters.

Sample treatment and procedure

The sample digestions were carried out in a microwave oven (Milestone model Ethos 1600) provided with a ten places rotor (HPR 1000/10S) and 100 ml TMF (tretrafluoromethoxil) vessels. Reagents used for sample digestion were ultra pure grade (Ultrapure HCl, HNO3 and HF supplied from Romil Ltd Cambridge).

Before sample digestion the TMF vessels were filled with 10 ml supra pure 67–69% HNO3 and cleaned with a three step digestion programme (step 1: 5 min at 300 W; step 2: 5 min at 0 W; step 3: 5 min at 400 W). After digestion the solutions were diluted to a final volume of 25 ml with ultrapure water in acid cleaned LPDE bottles for the determination of trace elements by ICP-QMS. A reagent blank, composed of the same amounts of chemicals used in the sample digestion, was included in each sample digestion batch to monitor the cleanliness of microwave digestion and sample handling processes during digestion.

Instrumentation and quantification

Al, As, Ca, Cd, Co, Cr, Cs, Cu, Fe, K, Li, Mg, Mn, Na, Ni, Pb, Rb, Sr, V, and Zn were determined by inductively coupled quadrupole mass spectroscopy (Agilent model 7500) equipped with a double pass spray chamber in PFA fitted with the Agilent V-groove nebulizer and a torch with a ceramic injector tube, resistant to hydrofluoric acid.

Prior to use the ICP-QMS spectrometer was tuned to obtain CeO+/Ce+ ratio <0.5% and Ce2+/Ce+ ratio <2%10 with the sample introduction system in place. The carrier gas flow rate was optimised to maximize the sensitivity for 7Li, 89Y, and 205Tl and minimize the CeO+ and Ce2+ signal. The best signal to noise ratio was found to be between a flow rate of 1.0 and 1.2 L min−1. Since the determined elements are present in aerosol samples at very different concentrations, differentiated integration times (see Table 1) were used to optimise the instrumental sensitivities for each element.

Table 1 Instrumental conditions and measurement parameters for the Agilent 7500 inductively coupled plasma quadrupole mass spectrometer
RF power 1450 W
Spray chamber temperature (S/C) 2 °C
Carrier gas flow rates 1.1 L min−1
Ion sampling depth Optimised to obtain minimum interference signal intensity (8 mm)
Selected isotopes 7Li, 23Na, 24Mg, 27Al, 39K, 44 Ca, 51V, 53Cr 55Mn, 57Fe, 59Co, 60Ni, 63Cu, 66Zn, 75As, 85Rb, 88Sr, 111Cd, 133Cs, 208Pb
Sample uptake rate 0.5 ml/min
Washing time 3 min
Take up time 45 s
Integration time per point 1 s for Cs, Li, Rb; 0.5 s for Cd, Co; 0.2 s for As, Cr, Cu, Mn, Ni, Sr, V; 0.1 s for Al, Ca, Fe, K, Mg, Na, Pb, Zn
Number acquisition points per peak 3


Quantification was performed using matrix-matched calibration curves prepared by using multi-element standard solutions for As, Cd, Co, Cr, Cs, Cu, Li, Mn, Ni, Pb, Rb, Sr, V, Zn, and by single element standard solutions for Al, Ca, Fe, K, Mg and Na. Every analysis cycle calibration curve was prepared using one blank sample where 11 additions were carried out to cover the entire concentration range of elements in the aerosol samples. The following range of concentrations were used: 0.005–200 µg L−1 for As, Cd, Co, Cr, Cs, Cu, Li, Mn, Ni, Pb, Rb, Sr, V, Zn, and 100–1000 µg L−1 for Al, Ca, Fe, K, Mg and Na. A 100 µg L−1 Os and Ru solution was added on line using a T-piece and was used as an internal standard during calibration and sample analysis. Multi-element standard solution of metals (10 mg L−1) were supplied from Ultra Scientific (North Kingstown, RI, USA). Single-element standard solutions were supplied from SPEX CertiPrep. INC. Metuchen (Os and Ru, 10 mg L−1) and from Ultra Scientific (Al, Ca, Fe, K, Mg, Na 1000 mg L−1). For each sampling campaign the blank concentration was measured and used to correct the concentration measured in the dissolved filters.

Results and discussion

Blanks and detection limits

Several blanks were collected under particular conditions to better evaluate the contributions from handling, exposure and treatment of the filter and to establish baseline concentrations, which were then subtracted from the concentrations of trace elements in the exposed filters.

The contribution to the blank values from filter handling during aerosol sampling was evaluated by keeping the filters in the samplers at the sampling station for a few minutes without any air flowing and then transferring them to a plastic bag without any further exposure. These blanks, defined field blanks (FBs), were then treated and analysed following the procedure applied to the samples.

Since sampling of the aerosol was carried out only when the wind blew in particular conditions the samplers remained inactive for long time periods. To estimate the contribution to the blank values by passive deposition, the filters were kept at the sampling stations for fifteen days without any air flow. These blanks, defined campaign blanks (CBs), were then treated and analysed as the aerosol samples. In order to emphasize the effect of handling during aerosol sampling, measurements of blank concentration of some elements in filters washed and maintained always inside the class 100 clean area were carried out (filter blank).

The amounts of trace metals found in cellulose slotted filters used as filter blanks, field blanks and campaign blanks are presented in Table 2. The values found in FBs and CBs ranged from 0.0006 µg/filter of Cs to 27 µg/filter of K for the campaign blanks and from 0.0006 µg/filter of Cs to 11 µg/filter of Na for the field blanks. One anomalous result was obtained for arsenic, the concentration in the field blank was about five time higher than the campaign blank. The concentrations of elements measured in filter blank were for larger part of element determined significantly lower than FBs and CBs; therefore is evident that handling the membrane for exposition, transport and settling they on the samplers may contribute significantly at the blank concentration. In spite of the long contact of the CBs with the sampler no positive trend was observed with respect to FBs, therefore the contribution to the trace elements blank from passive deposition and contact with the sampling system is considered as negligible. Since FBs are easier to collect, they were used to quantify the “limit of detection” (LOD), which was calculated as 3 times the standard deviation values of the blank content, an average air volume of 4449 m3 was used to calculate the LOD in term of ng m−3.11 The FBs average blank values for back filters are presented in Table 3. The amounts found ranged from 0.002 µg/filter for Cs to 24 µg/filter for Na.

Table 2 Mean amounts of elements (µg/filter) and standard deviation (SD) measured in campaign blanks (CBs) and field blanks (FBs) of slotted filters for the whole processes of sampling, treatment and analysis
Element Campaign blanks Field blanks Filter blank
Meana SDb Meana SDb Meana SDb
a Mean. b Standard deviation values calculated by 7 repeats for campaign blank, 12 repeats for field blank and 5 repeats for the filter blank.
Al 3 1 2.0 0.7 0.07 0.04
As 0.09 0.01 0.47 0.07
Ca 8 2 2.6 0.7
Cd 0.006 0.0007 0.002 0.001
Co 0.0045 0.0005 0.005 0.002
Cr 0.23 0.04 0.6 0.2
Cs 0.0006 0.0001 0.0006 0.0005
Cu 0.07 0.01 0.05 0.03 0.04 0.03
Fe 5.1 0.3 3.1 0.8
K 27 3 2.9 0.7
Li 0.004 0.001 0.013 0.003
Mg 0.8 0.2 0.5 0.1
Mn 0.04 0.01 0.05 0.02 0.007 0.005
Na 15 2 11 3
Ni 0.05 0.01 0.11 0.05
Pb 0.021 0.003 0.022 0.006 0.005 0.004
Rb 0.0047 0.0007 0.004 0.002
Sr 0.024 0.009 0.011 0.008
V 0.016 0.008 0.07 0.02 0.01 0.008
Zn 1.6 0.5 0.4 0.1 0.24 0.12


Table 3 Mean amounts of elements (µg/filter), standard deviation (SD), limit of detection (LOD ng m−3) estimated by back filter field blanks
Element Field blanks
Meana SD b LODc
a Mean. b Standard deviation values calculated by 7 independent part of filters at least. c LOD in ng m−3 was calculated as three times the blank standard deviation and using an average air volume of 4449 m3.
Al 3 1 0.7
As 0.65 0.06 0.04
Ca 4.5 0.8 0.5
Cd 0.004 0.002 0.002
Co 0.005 0.001 0.001
Cr 0.7 0.1 0.1
Cs 0.002 0.001 0.001
Cu 0.09 0.03 0.02
Fe 3.9 0.5 0.3
K 4.1 0.5 0.3
Li 0.017 0.003 0.002
Mg 0.9 0.3 0.2
Mn 0.05 0.01 0.008
Na 24 1 1
Ni 0.09 0.03 0.02
Pb 0.04 0.02 0.01
Rb 0.007 0.002 0.001
Sr 0.015 0.008 0.006
V 0.09 0.02 0.02
Zn 1.0 0.4 0.3


Slotted filters, which have a lower content of cellulose than back filters, show lower blank values (Table 2 and Table 3) for the almost all elements so different blank corrections must be used for the first five fractions (slotted filters) and for the final fraction (size <0.49 mm) (back filters).

Comparison with limits of detection previously published shows that our results are comparable with values reported in the literature. The LODs reported by Yang et al.7 ranged from 0.015 ng m−3 for Cd to 46 ng m−3 for Ca using microwave digestion and ICP-MS analysis. Fernández Álvarez8 reported LOD values ranging between 0.025 ng m−3 for As to 0.94 ng m−3 for Pb using microwave digestion and inductively coupled plasma atomic emission spectroscopy analysis (ICP-AES). Detection limits found by A. A. Karanasiou et al.6 range from 0.0004 ng m−3 for Cd to 4.3 ng m−3 for Al using microwave digestion and electrothermal atomic absorption spectroscopy analysis (ETAAS).

Accuracy and repeatability

Several microwave digestion programs with different combinations of power and time settings and different reagent mixtures were tried to achieve complete dissolution of the particulate matter. The best recoveries were obtained by the following step program settings of microwave oven: step 1: 10 min at 250 W; step 2: 1 min at 0 W; step 3: 20 min at 400 W; step 4: 2 min at 0W; step 5: 20 min at 500 W, step 6: 2 min at 0 W; step 7: 3 min at 650 W and two acid mixtures (a) and (b):

(a) 1.5 ml ultra pure water, 5 ml HNO3 and 1 ml HF (Romil Ultrapure acids);

(b) 1.5 ml ultra pure water, 5 ml HNO3, 1 ml HF and 0.5 ml HCl (Romil Ultrapure acids).

In order to validate the microwave digestion methods, portions of 1/6 of a blank back filter were loaded with about 10 mg of Standard Urban Dust Reference Material (NIST, SRM-1648). The reagents were added and the filter and reference material were digested following the selected program settings, the solution was diluted until a final volume of 25 ml by ultrapure water. The procedure was repeated at least five time using different portions of the filter for each of the reagent mixtures listed above as (a) and (b). As, Cd, Cr, Cu, Mn, Ni, V concentrations were quantified using matrix-matched calibration curves without any dilution of the digested NIST sample. Al, Fe, K, Na, Pb and Zn concentrations were quantified after a 1:100 dilution of the digested reference material sample using external calibration curves. Experimental values and certified values are compared in Table 4. For most of the elements satisfactory agreement between the measured and certified values was found using both the reagent mixtures. Recoveries of the elements content, expressed as a percentage of the NIST certified values, ranged from 79% for Cr to 120% for Zn in samples digested with the H2O, HNO3 and HF mixture and from 84% for V to 105% for K in sample digested with the H2O, HNO3, HF and HCl mixture. Since no significant difference was found for the two reagent mixtures tested, sample digestion without HCl were preferred to minimize the interference from 40Ar35Cl+ in the determination of As.

Table 4 Elemental contents (µg g−1) and percent of recovery obtained for the certified urban particulate (NIST 1648), dissolved by microwave-assisted digestion with two different acid mixturesd,e
Element Certified valuesb Found valuesc,d Recovery, % Found valuec,e Recovery %
a Content in weight percentage (w/w %). b Average ± 95% tolerance limits. c Average (standard deviation). d Digestion acid mixture: 1.5 ml H2O, 5 ml HNO3, 0.5 ml HCl, 1 ml HF (Romil Ultrapure acids). e Digestion acid mixture: 1.5 ml H2O, 5 ml HNO3, 1 ml HF (Romil Ultrapure acids).
Al a 3.42 ± 0.11 3.2 (0.1) 94 3.12 (0.05) 91
Fe a 3.91 ± 0.10 3.6 (0.1) 92 4.0 (0.4) 102
K a 1.05 ± 0.01 1.1 (0.2) 105 0.84 (0.04) 80
Na a 0.425 ± 0.002 0.40 (0.01) 94 0.41 (0.01) 96
Pba 0.655 ± 0.008 0.59 (0.01) 90 0.63 (0.01) 96
Zn a 0.476 ± 0.014 0.48 (0.02) 101 0.57 (0.02) 120
As 115 ± 10 101 (3) 88 105 (3) 91
Cd 75 ± 7 66 (1) 88 67 (2) 89
Cr 403 ± 12 370 (16) 92 320 (29) 79
Cu 609 ± 27 550 (11) 90 560 (15) 92
Mn 786 ± 17 780 (10) 99 760 (36) 97
Ni 82 ± 3 78 (2) 95 68 (3) 83
V 127 ± 7 107 (3) 84 110 (4) 87


Dividing the filters into smal sections was necessary to avoid the overloading microwave vessels due to the large weight of the cellulose filters (from 1.4 g for slotted filters to 4 g for back-filters). The back filters were divided into six equal sections and slotted filers were divided into four equal sections. Since only one section was used to assess the sample concentration, the uniformity of atmospheric particulate distribution on the filters was tested by digestion and analysis of different sections of the same aerosol sample. The results obtained for digestion and analysis of four sections of a randomly chosen aerosol sample collected on a back filter are reported in Table 5. The relative standard deviations of element concentrations measured for most elements was less than 10% showing that the sampled atmospheric particulate can be considered uniformly distributed on the filters and that a section of the filter is representative of the entire sample.

Table 5 Results (ng m−3), mean values and standard deviation (SD) of elemental content in tests of uniformity of aerosol samples. Digestion and analysis were performed on four sections of a sample collected on a back filtera
Element Results ng m−3 in repeatability tests Mean SD
a u.d.l.: under detection limit.
Al 31.4, 31.1, 31.4, 32.3 31.5 0.5
As u.d.l. u.d.l. u.d.l.
Ca 14.6, 14.9, 16.3, 16.3 15.5 0.9
Cd 0.18, 0.22, 0.23, 0.20 0.21 0.02
Co 0.0095, 0.0093, 0.0103, 0.0095 0.0096 0.0004
Cr 0.16, 0.19, 0.19, 0.25 0.20 0.04
Cs 0.0107, 0.0112, 0.0119, 0.0103 0.0110 0.0007
Cu 0.62, 0.68, 0.74, 0.77 0.70 0.07
Fe 26, 28, 31, 30 29 2
K 52, 61, 62, 55 58 5
Li 0.0269, 0.0264, 0.0272, 0.0272 0.0270 0.0004
Mg 8.1, 8.0, 8.6, 8.5 8.3 0.3
Mn 0.93, 1.02, 1.07, 1.00 1.01 0.06
Na 17.6, 18.4, 17.7, 16.7 17.6 0.7
Ni 0.79, 0.97, 0.98, 0.89 0.91 0.09
Pb 3.0, 3.7, 3.8, 3.3 3.5 0.4
Rb 0.14, 0.16, 0.17, 0.15 0.16 0.01
Sr 0.119, 0.127, 0.137, 0.132 0.129 0.008
V 1.6, 1.9, 2.0, 1.7 1.8 0.2
Zn 6.3, 7.3, 7.2, 6.9 6.9 0.4


Determination of trace elements in the atmospheric aerosol of the Venice Lagoon

In order to test the methodology to characterize aerosols with remarkable different composition, the procedure was applied to determine major and trace element concentrations in the PM10 and to define the distribution between fine (aerodynamic diameter <3 µm) and coarse particles (aerodynamic diameter <10 µm and >3 µm) in the atmosphere aerosols collected in areas of the Venice region characterized by airborn particulate of different origin.

Mean values of PM10 and of element concentrations during the entire sampling period are summarized in Table 6. Similar PM10 mean mass concentrations at sites 1, 2 and 4 were obtained during the four sampling campaigns (39.3 µg m−3, 21.0 µg m−3 and 34.2 µg m−3 respectively), while, a higher median value (56,5 µg m−3) was obtained at station 3 (rural area), that was ascribed to occasional fires occuring during the summer season in the neighbouring wood. The results are consistent with those reported for other urban-industrial sites in Europe. Röösli et al.12 found in Basel (Central Europe) a mean mass concentration of 28.3 µg m−3, Viana et al.13 measured at Ghent in west Europe a mass concentration of 28.6 µg m−3 during winter 2005 and at Barcelona in the Mediterranean a mass concentration of 34.0 µg m−3 during winter 2004. Higher values (78–89 µg m−3) are found by Voutsa et al.14 at Thelassoniki in Greece during the summer 1997–1998 and by Weckwerth15 (54 µg m−3) in the atmosphere of Cologne (Germany).

Table 6 Mean particle content (µg m−3) and trace element concentration (ng m−3) in PM10, coarse and fine factions of particulate airborne collected in marine, rural, urban and industrial areas of Venice during four campaigns from March 2002 to June 2003
  Site 1 (industrial-urban area) Site 2 (marine area) Site 3 (rural area) Site 4 (urban area)
PM10 Coarse Fine PM10 Coarse Fine PM10 Coarse Fine PM10 Coarse Fine
a Mean concentration in µg m−3.
Particulate 39.3 8.98 30.3 21.0 2.67 18.3 56.5 7.28 49.2 34.2 5.91 28.3
Al a 0.468 0.180 0.289 0.117 0.046 0.071 0.221 0.111 0.109 0.085 0.039 0.046
As 4.2 0.25 4.0 0.40 0.046 0.35 0.81 0.046 0.76 5.9 0.23 5.7
Caa 0.88 0.44 0.44 0.135 0.097 0.038 0.21 0.121 0.086 0.27 0.155 0.118
Cd 2.5 0.2 2.3 0.42 0.012 0.41 0.81 0.22 0.59 2.9 0.18 2.7
Co 0.38 0.101 0.28 0.056 0.0176 0.039 0.167 0.035 0.133 0.091 0.028 0.064
Cr 2.3 0.47 1.9 0.45 0.13 0.32 1.2 0.13 1.1 1.2 0.19 1.0
Cs 0.029 0.0077 0.021 0.0078 0.0012 0.0066 0.0099 0.0018 0.0081 0.017 0.0031 0.014
Cu 14 5.0 9.0 3.4 0.94 2.5 9.1 3.7 5.5 8.7 2.6 6.1
Fe a 0.48 0.20 0.28 0.130 0.046 0.084 0.21 0.083 0.124 0.21 0.074 0.136
K a 0.31 0.071 0.24 0.13 0.025 0.106 0.18 0.034 0.15 0.25 0.029 0.22
Li 0.508 0.153 0.355 0.078 0.029 0.048 0.118 0.043 0.075 0.102 0.033 0.069
Mg 259 124 135 80 51 29 111 63 49 91 49 42
Mn 24 9.9 13.7 4.8 1.2 3.6 9.2 2.76 6.4 7.1 1.9 5.2
Na a 0.36 0.151 0.206 0.36 0.25 0.108 0.21 0.090 0.120 0.198 0.097 0.101
Ni 17 2.9 14 3.3 0.49 2.8 3.1 0.38 2.7 2.2 0.36 1.9
Pb 29 4.9 24 11 0.85 10 17 1.9 15 20 1.7 18
Rb 0.98 0.27 0.72 0.42 0.076 0.35 0.63 0.15 0.49 0.71 0.098 0.61
Sr 2.6 1.2 1.41 0.84 0.46 0.39 1.37 0.56 0.81 0.89 0.43 0.46
V 14 2.2 12 7.3 0.88 6.4 4.5 0.26 4.3 2.2 0.25 1.9
Zn 108 30 79 31 3.4 28 58 10.8 47 54 7.5 47


Measured elemental concentrations differ by several orders of magnitude from one metal to another. In relation to their crustal abundance, the highest concentrations are observed for Al and Fe at the µg m−3 level, whereas the lowest concentrations, in agreement with their very low crustal abundance, are observed for elements such as Cs, at the sub-ng m−3 level.16

The ratio between highest and the lowest concentration detected for trace elements cover an interval ranging between 3 to 15. Comparison with relevant data from other urban sites shows that the trace element concentrations observed in this study are consistent and close to the ones detected in the aerosol of other cities of the central and northern Europe. Concentrations of As, Cd, Ni, Pb and Zn in the PM10 are comparable with those measured in Basel during 97–98, (1 ng m−3 for As, 1 ng m−3 for Cd, 9 ng m−3 for Ni, 55 ng m−3 for Pb, 84 ng m−3 for Zn),12 in Edinburgh during 1999–2000 (0.4 ng m−3 for As, 0.34 ng m−3 for Cd, 3.4 ng m−3 for Ni, 14.1 ng m−3 for Pb, 1.1 ng m−3 for V, 13.3 ng m−3 for Zn)17 and in Budapest during spring 2002 (0.99 ng m−3 for As, 27 ng m−3 for Pb, 2.8 V ng m−3, 90 ng m−3 for Zn).18

The particle content and trace element concentration in the fine and coarse fractions are reported in Table 6. The element distribution between fine and coarse is consistent with sources; fine particle concentrations account for 87% (site 2), 87% (site 3), 77% (site 1) and 83% (site 4) of the PM10 particle mass. All the elements that are prevalently originating from anthropogenic sources such as automotive traffic, oil combustion or industrial processes (As, Cd, Cr, Ni, Pb, V, Zn) are essentially associated to the fine particle fraction (70–95%). On the other hand, elements such as Al and Fe that are mainly from natural sources (soil dust) were almost equally distributed in fine and coarse fractions.

The highest PM10 metal concentration for almost all of the elements was found at station 1 downwind of the large industrial zone of Porto Marghera where the main activities are petrochemical production and refining, iron and steel industries with incineration and thermal plants (fed by fossil fuels). Moreover the site is affected by the vehicular and urban emissions of Mestre.

The concentrations of all the elements, except Na, in the samples collected at site 2, are lower than those measured at St.1 by a factor of 2–4.

Conclusions

The proposed method of aerosol sampling, sample microwave digestion and inductively coupled mass spectroscopic analysis allows accurate determination of a large number of elements at trace levels in minute amounts of aerosol samples. The microwave digestion procedure with a simple acid mixture, containing only HNO3 and HF, was validated using the certified reference material NIST SRM 1648 (Urban Particulate Matter). The results show that it is an efficient method for the determination of trace elements in airborne particles with element recoveries always higher than 80% for all the analysed elements. The contributions of different processes to the blank were evaluated to identify possible sources of sample contamination. The results show that the blank contributions from filter handling and sampling procedures are critical and must be controlled for a correct evaluation of the analytical results. The contribution is significant from the filter handling, transport and settling on the sampling system, while the contributions to the blank values by passive deposition is negligible therefore blank contributions can be evaluated by placing filters on the sampling device at every sampling station for some minutes without any air flow. The measured LODs for the method allow the determination of trace elements in aerosol in the range of concentrations observed in the four areas characterized from aerosol of different origin if the sampled air-volume is 1000 m3 or higher.

Acknowledgements

This study has been supported by the Consortium for Coordination of Research Activities concerning the Venice Lagoon System (CORILA) within the project “Role of aerosol and secondary pollution in the chemical contamination of the lagoon of Venice”. We thank the Ente Zona Marghera and ARPAV for sampling support and V. Zampieri for technical support.

References

  1. N. Englert, Toxicol. Lett., 2004, 149, 235–242 CrossRef CAS.
  2. C. A. III Pope, Toxicology, 1996, 111, 149–155 CrossRef.
  3. J. Schwartz, D. W. Dockery and L. M. Neas, J. Air Waste Manage. Assoc., 1996, 46, 927–939 CAS.
  4. Common Position (EC) No. 57/98, 98/C360/04, Official Journal of European Communications C360/99 23 November 1998.
  5. Commission of the European Communities, COM (2003) 423 final, 16 July 2003.
  6. A. A. Karanasiou, N. S. Thomaidis, K. Eleftheriadis and P. A. Siskos, Talanta, 2005, 65, 1196–1202 CrossRef CAS.
  7. K. X. Yang, K. Swami and L. Husain, Spectrochim. Acta B, 2002, 57, 73–84 CrossRef.
  8. F. Fernández Álvarez, M. Ternero Rodríguez, A. J. Fernández Espinosa and A. Gutiérrez Dabán, Anal. Chim. Acta, 2004, 524, 33–40 CrossRef CAS.
  9. N. J. Pekney and C. I. Davidson, Anal. Chim. Acta, 2005, 540, 269–277 CrossRef CAS.
  10. E. McCurdy and D. Potter, Spectroscopy Europe, 2001, 13/3, 14–20 CAS.
  11. IUPAC, Nomenclature, symbols, units and their usage in spectrochemical analysis II, Spectrochim. Acta B, 1978, 33, 241–245 Search PubMed.
  12. M. Röösli, G. Theis, N. Künzli, J. Staehelin, P. Mathys, L. Oglesby, M. Camenzind and C. Braun-Fahrländer, Atmos. Environ., 2001, 35, 3701–3713 CrossRef CAS.
  13. M. Viana, W. Maenhaut, X. Chi, X. Querol and A. Alastuey, Atmos. Environ., 2007, 41, 315–326 CrossRef CAS.
  14. D. Voutsa, C. Samara, T. H. Kouimtzis and K. Ochsenkühn, Atmos. Environ., 2002, 36, 4453–4462 CrossRef CAS.
  15. G. Weckwerth, Atmos. Environ, 2001, 35, 5525–5536 CrossRef CAS.
  16. K. H. Wedepohl, Geochim. Cosmochim. Acta, 1995, 59, 1217–1232 CrossRef CAS.
  17. M. R. Heal, L. R. Hibbs, R. M. Agius and I. J. Beverland, Atmos. Environ., 2005, 39, 1417–1430 CrossRef CAS.
  18. I. Salma and W. Maenhaut, Environ. Pollut., 2006, 143, 479–488 CrossRef.

This journal is © The Royal Society of Chemistry 2009
Click here to see how this site uses Cookies. View our privacy policy here.