Dominic
Hare
a,
Fiona
Burger
a,
Christine
Austin
a,
Fred
Fryer
b,
Rudolf
Grimm
c,
Brian
Reedy
a,
Richard A
Scolyer
d,
John F.
Thompson
d and
Philip
Doble
*a
aElemental Bio-imaging Facility, University of Technology, Department of Chemistry and Forensic Science, Sydney, Broadway, New South Wales, Australia. E-mail: philip.doble@uts.edu.au; Fax: +61 2 9514 1460; Tel: +61 2 9514 1792
bAgilent Technologies Australia, North Ryde, New South Wales, Australia
cAgilent Technologies Inc., Santa Clara, California, USA
dSydney Melanoma Unit and Melanoma Institute Australia, Royal Prince Alfred and Mater Hospitals, Sydney, New South Wales, Australia
First published on 27th November 2008
The spatial distribution of trace elements in human lymph nodes partially infiltrated by melanoma cells was determined by elemental bio-imaging. Imaging of 31P within the nodal capsule and normal lymph node tissue showed a clear demarcation of the tumour boundary, with a significant decrease in relative 31P concentration within the tumour. The location of the tumour boundary was confirmed by haematoxylin and eosin staining of serial sections and observation by light microscopy. Further enhancement of the tumour boundary was achieved by imaging the 31P/34S ratio. 31P/66Zn ratio images showed a decreasing ratio beyond the tumour boundary that extended into peritumour normal lymph node tissue.
Mortality from most solid malignancies usually occurs as a consequence of tumour metastasising from its site of origin to distant locations. Lymph nodes receive lymphatic fluid directly from the primary tumour and are the most common first sites to harbour metastases. Metastatic melanoma is present in lymph nodes in approximately 20% of melanoma patients with primary tumours ≥1 mm in thickness.4 Identification of the presence of melanoma metastases in clinically negative regional lymph nodes is usually achieved by surgical removal of the lymph node most likely to harbour metastases (i.e. the sentinel lymph node) and laborious histopathological examination of tissue sections stained both with hematoxylin and eosin (H&E) and immunohistochemically for various melanoma-associated antigens. The tumour-harbouring status of the sentinel lymph node is the most important prognostic factor in patient's presenting with early stage melanoma.5
An alternative method for detection of melanoma in lymph nodes is chemical imaging. Fourier transform infrared (FTIR) spectral imaging has been used to investigate cutaneous melanomas.6,7 Recently, K-means clustering was used to identify infrared spectral similarities in skin basal cell carcinomas (BCCs), highlighting the utility of imaging technology and chemometrics for the diagnosis of cutaneous malignancies.8
Although trace elements have long been associated with malignancy, their role and prevalence in melanoma is unknown. Phosphorus/sulfur ratios have been shown to be higher in malignant human hepatocarcinoma cell lines when compared to normal human dermal fibroblast and transformed embryo kidney cell lines.9Sulfur-based, cysteine-rich, zinc-binding metallothioneins (MTs) have shown both increased expression and down regulation in numerous types of cancers.10 Increased copper concentrations and lower iron levels were reported following the analysis of toenail clippings from 58 newly diagnosed cutaneous melanoma patients who did not show evidence of metastasis compared with 58 control samples.11 In another study, serum trace element levels in healthy controls and patients with varying stages of melanoma showed an increase in zinc levels. However, no statistical difference in serum copper levels was noted in the melanoma patients.12
Imaging mass spectrometry 13 employs matrix-assisted laser desorption/ionisation mass spectrometry (MALDI-MS) for the imaging of peptides in tissues by monitoring specific mass-to-charge (m/z) ratios. This technique has been applied to glioma biopsies and numerous other diseases.14 In a similar manner, elemental bio-imaging of tissue sections is possible by employing laser ablation-inductively coupled plasma-mass spectrometry (LA-ICP-MS).15LA-ICP-MS is capable of determining most of the elements with a 7 order magnitude linear dynamic range, with part per million detection limits. The system utilises a focused laser beam (10 to 200 μm diameter) directed onto the surface of the tissue. The sample is ablated by an energy transfer from the laser, resulting in a plume of particles swept to the ICP by a carrier gas. The spatial distribution of elements across a tissue sample can be determined by collecting mass spectra as the laser traverses the sample in two dimensions. Maps can then be constructed of both metals and non-metals essential to physiology. Recently, elemental bio-imaging has been applied to human glioblastomas,16 labelled β-amyloid-bound antibodies in Alzheimer's disease,17 implanted tumour cells18 and Cu, Zn and Pt in cis-platin dosed mice.19
The current study demonstrated the potential of elemental bio-imaging for detection of metastatic melanoma in lymph nodes in human tissues. Construction of maps presented as single elements or ratios of elements, provided clear images of the melanoma and adjacent non-tumorous tissues.
Agilent 7500ce ICP-MS | New Wave UP213 Laser Ablation | ||
---|---|---|---|
RF power | 1200 W | Wavelength | 213 nm |
Cooling gas flow rate | 15 L min−1 | Repetition frequency | 20 Hz |
Carrier gas flow rate | 1.1 L min−1 | Laser energy density | 0.3 J cm−2 |
Sample depth | 6.0 mm | Spot size | 40 µm |
QP bias | −5 V | Scan rate | 40 µm s−1 |
OctP bias | −8 V | Line spacing | 40 µm |
Scan mode | Peak hopping | Carrier gas | Ar |
Dwell time | 0.1 s per m/z | ||
Measured m/z | 13, 31, 32, 34, 56, 57, 63, 66 | ||
Extracts 1, 2 | 6.8, −126 V |
The instrument was tuned for high sensitivity and low oxide production by ablating NIST 612 glass and measuring masses 232 and 248 (Th and ThO respectively). At an RF power setting of 1200 W and sampling depth of 6 mm, ThO/Th ratios were consistently less than 0.3%. The low sample load from the ablated tissue minimised the influence of oxide-based interferences from water in the sample matrix. Measurement of 32S+/34S+ and 56Fe+/57Fe+ ratio in tissue sections confirmed the natural isotopic abundance ratios.
LA-ICP-MS data was exported in comma separated variable (CSV) file formats with a separate file created for each line of ablation in the x-axis. A Visual Basic macro was used to both normalise each data point to 13C and combine all resultant data into a single ASCII file. This file was then imported into the ITT Visual ENVI 4.2 (Research Systems Inc., Boulder, CO, USA) hyperspectral imaging software package, where final images were produced.
![]() | ||
Fig. 1 Sentinel lymph node partly replaced by metastatic melanoma (upper two thirds of the ovoid-shaped lymph node). (a) Photograph of slice before elemental imaging, (b) H&E stain, (c) 31P+ image without 13C normalisation, (d) 31P+ image, (e) 31P+/34S+ and (f) 31P+/66Zn+. |
Fig. 1(c) is a 31P+ image prior to 13C+ normalisation. There was variation in the concentration of phosphorus that corresponded to the presence of the tumour as determined by the H&E stain. Fig. 1(d) is the same image normalised to 13C+. There was significant improvement in contrast in the normalised image, particularly on the left side of the section. This normalisation process also compensated for varying thickness in the tissue slice.
There were no significant variations in concentrations of 32,34S, 56,57Fe, 63Cu and 66Zn when imaged as separate isotopes (data not shown). Fig. 1(e) clearly shows an increased 31P+/34S+ ratio in the peritumoural lymphoid tissue when compared with the metastatic tumour. Furthermore, the 31P+/34S+ ratio improved contrast between the melanoma and lymphoid cells, when compared with 31P+ images alone. Construction of 31P+/66Zn+ ratios (Fig. 1(f)) shows a slightly different perspective on the tumour boundary. The blue regions (relatively lower values) penetrated further into the peritumoural, apparently uninvolved, lymphoid tissue (determined from the H&E stain) than in the images of the 31P+ and the 31P+/34S+ ratio. These results suggest changes in biochemistry of the peritumoural lymphoid tissue that likely represents a response to the tumour.
Fig. 2 shows a 20× magnification of the tumour boundary overlaid with the corresponding 31P image. Clear evidence of metastatic melanoma was observed on the H&E stained section (the pale area in the upper right portion of the lymph node in Fig. (1b)) confirmed by immunohistochemistry. The boundary between healthy and cancerous cells correlated well with the changing intensities of 31P+, 31P+/34S+ and 31P+/66Zn+.
![]() | ||
Fig. 2 Overlaid 31P+ image and photomicrograph of lymph node showing boundary of normal and metastatic melanoma cells. |
Five further lymph nodes containing metastatic melanoma were imaged in a similar way and are shown in Fig. 3 and 4. The trends discussed above were confirmed in each of the samples. The central node in Fig. 3(a) displayed only a very small melanoma micrometastasis on the H&E stained section and elemental images clearly identified the presence of tumour cells. The heightened contrast between metastasis and non-tumourous normal lymphoid tissue in 31P+/66Zn+ images was confirmed as shown in Fig. 3 and 4.
![]() | ||
Fig. 3 Five lymph nodes, each partly involved by metastatic melanoma. (a) H&E stain, (b) 31P+ image, (c) 31P+/34S+ and (d) 31P+/66Zn+. |
![]() | ||
Fig. 4 Lymph node partly replaced by metastatic melanoma. (a) H&E stain (the tumour cells represent the pale area involving the upper right portion of the lymph node), (b) 31P+ image, (c) 31P+/34S+ and (d) 31P+/66Zn+. |