Dong Sung
Kim
ab,
Se Hwan
Lee
c,
Chong H.
Ahn
c,
Jae Y.
Lee
d and
Tai Hun
Kwon
*ab
aDepartment of Mechanical Engineering, Pohang University of Science and Technology (POSTECH), San 31 Hyoja-dong Nam-gu, Pohang, Kyungbuk 790-784, Korea. E-mail: thkwon@postech.ac.kr; Fax: 82 54 2795899; Tel: 82 54 2792175
bResearch Center for Micro Mechanical Systems, Pohang University of Science and Technology (POSTECH), San 31 Hyoja-dong Nam-gu, Pohang, Kyungbuk 790-784, Korea
cMicroSystems and BioMEMS Laboratory, Department of Electrical & Computer Engineering and Computer Science, University of Cincinnati, P.O. Box 210030, Cincinnati, OH 45221-0030, USA
dCollege of Medicine, Ohio State University, Columbus, OH 45221, USA
First published on 4th April 2006
Blood typing is the most important test for both transfusion recipients and blood donors. In this paper, a low cost disposable blood typing integrated microfluidic biochip has been designed, fabricated and characterized. In the biochip, flow splitting microchannels, chaotic micromixers, reaction microchambers and detection microfilters are fully integrated. The loaded sample blood can be divided by 2 or 4 equal volumes through the flow splitting microchannel so that one can perform 2 or 4 blood agglutination tests in parallel. For the purpose of obtaining efficient reaction of agglutinogens on red blood cells (RBCs) and agglutinins in serum, we incorporated a serpentine laminating micromixer into the biochip, which combines two chaotic mixing mechanisms of splitting/recombination and chaotic advection. Relatively large area reaction microchambers were also introduced for the sake of keeping the mixture of the sample blood and serum during the reaction time before filtering. The gradually decreasing multi-step detection microfilters were designed in order to effectively filter the reacted agglutinated RBCs, which show the corresponding blood group. To achieve the cost-effectiveness of the microfluidic biochip for disposability, the biochip was realized by the microinjection moulding of COC (cyclic olefin copolymer) and thermal bonding of two injection moulded COC substrates in mass production with a total fabrication time of less than 20 min. Mould inserts of the biochip for the microinjection moulding were fabricated by SU-8 photolithography and the subsequent nickel electroplating process. Human blood groups of A, B and AB have been successfully determined with the naked eye, with 3 µl of the whole sample bloods, by means of the fabricated biochip within 3 min.
Among many blood typing methods, a plate method,1,2 a gel card method3,4 and an automated blood bank test system5,6 have been regarded as the most representative blood typing methods. The plate method is relatively simple but tedious due to its whole manual process. Furthermore, it may result in undesirable mistyping of blood group of a sample blood due to poor reliability and user performance. The gel card method is regarded as a semi-automatic and relatively accurate test method. However, the test kit is expensive and it requires a practised hand and extra equipment, such as a centrifuge. The blood bank system offers fully automated blood typing of numerous bloods. This system, however, is very expensive and huge so that only major medical centers or blood banks can maintain it. Recently, Lee et al.7 presented a miniaturized blood typing system fabricated by the micro-stereolithography technology, which assembled flow splitting channels, chaotic micromixers8 and reaction chambers. They used 20 µl of a sample blood per test line to determine the blood group and showed that an efficient micromixer should be required to get reliable agglutination reaction in microchannels. However, their system has the limitations of disposability and portability due to the high fabrication cost and requirement of extra detection equipment to observe agglutinated red cells in the reaction chambers.7
Therefore, the previous blood typing methods including the plate, gel card, automated system and miniaturized system explained above have a limitation for emergency blood typing due to the requirement of extra-large scale equipment and much testing time for blood sedimentation. In other words, they are not suitable for point-of-care monitoring and treatment, especially for the emergency case. Furthermore, the previous methods generally require a blood volume of more than 20 µl.5 If the repeated collection of blood from patients is required, withdrawing blood becomes a critical problem, in particular for infants though the test blood volume is relatively small.
In this regard, we propose a low cost, lab-on-a-chip for blood typing, integrated microfluidic biochip of compact size, which requires a very small sample blood volume of the order of 1 µl. In the following sections we introduce the design and operation process of the microfluidic biochip and components integrated in the biochip: flow splitting microchannels, chaotic micromixers,9 reaction microchambers and detection microfilters. As the cost-effective mass fabrication method of the biochip, a microinjection moulding process with electroplated nickel mould inserts and a subsequent thermal bonding process are applied. Then, we present successful blood typing results of blood groups of A, B and AB by means of the fabricated biochip.
In general, both red cell typing and serum typing are carried out for the accurate determination of blood group in order to compare the agglutination results of the other test. Red cell typing could have seven test channels, which contain sample RBCs with standard Anti-A, Anti-B, Anti-AB, Anti-A1, Anti-H, Anti-D and negative control, to determine blood groups of A, B, AB, O, Rh (D) and also weak-A. Anti-A1 and Anti-H are used for determining weak blood sub-groups and negative control is required to check the self-agglutination of the sample blood. For serum typing, three test channels, which contain sample serum with standard red cell-A, cell-B and cell-O, could be applied to confirm the test result of the red cell typing.
![]() | ||
Fig. 1 Schematic diagram of one blood typing unit of a microfluidic biochip in which flow splitting microchannels, chaotic micromixers, reaction microchambers and detection microfilters are fully integrated. The reaction microchamber can store about 3 µl of a mixture of blood and serum. The agglutinated red blood cells formed by antigen–antibody reaction between red blood cells and test serum can be effectively observed by the efficient detection microfilters (insets are enlarged views of the microfilter): (a) top view; and (b) three-dimensional view. |
A blood typing process by means of the biochip in this study runs as follows. The constant volume (about 1–1.5 µl per test line) of a sample blood (RBCs) and the standard sera are loaded into the corresponding inlets, as indicated in Fig. 1(a) (process step (i) in Fig. 1(b)). The loaded sample blood and standard sera are injected into the biochip through the flow splitting microchannels via a syringe pump, for example. In this case, the stream of the sample blood splits into and is guided to 2 or 4 separate microchannels (process step (ii)). Both the divided blood stream and the corresponding serum stream are introduced into each connected chaotic micromixer, SLM, to be efficiently mixed with each other in order to induce the effective agglutination of RBCs with the corresponding antibodies (process step (iii)). The mixtures of blood and serum streams are then led to each reaction microchamber. The reaction microchamber keeps the mixture during the appropriate reaction time about 1–3 min for agglutination of RBCs. It should be mentioned that, in the present study, we stopped the syringe pump at this process step to eliminate the driving pressure applied to the biochip, thereby keeping the mixtures inside the reaction microchambers during the reaction time (process step (iv)). The reacted (or non-reacted) mixture of blood RBCs and serum passes through each detection microfilter and finally flows out from each outlet (process steps (v) and (vi)). The large size reacted agglutinated cells are effectively filtered inside each detection microfilter, whereas non-reacted RBCs easily pass the detection microfilters. Therefore, one can observe the agglutination results by the naked eye without any extra large-scale equipment such as a microscope, and thus one can easily determine the blood group of the test blood sample by means of the biochip alone.
It might be noted that one can perform 2 or 4 blood agglutination tests in parallel with the present design. For the case of 2 blood agglutination tests (e.g., with Anti-A and Anti-B, that is a simple ABO blood typing), one can utilize two of four test lines in the one blood typing unit in Fig. 1 by opening two inlets of sera and the corresponding two outlets (of course, also opening blood inlet). For the 4 blood agglutination tests (e.g., with Anti-A, Anti-B, Anti-D and Anti-A1), all the four test lines are utilized by opening all inlets and outlets. Finally, it might be mentioned that when there exists any need to design multiple flow splitting microchannels having odd splitting channels or different splitting volumes, one can utilize the microchannel design equations suggested in ref. 10 to achieve that.
The SLM has a great advantage of mass producible simple geometry while maintaining the higher level of chaotic mixing over the wide range of Re. Furthermore, with the SLM, one could effectively reduce the chances of non-specific adsorption of proteins (antibodies in this study) to the channel surface by decreasing contact area between the surface of the microchannel and stream.
Two sample RBCs and standard serum streams injected from the connected flow splitting microchannels are efficiently mixed up through the SLM. Each mixture of RBCs and serum is then led to the subsequent reaction microchamber. For the detailed mixing mechanism and characterization of the SLM, one may refer to ref. 9.
Fig. 1 schematically shows the reaction microchamber between the SLM and the detection microfilter. The width and total thickness of the reaction chamber was designed to become 2500 µm and 120 µm (two layers), respectively. As was mentioned before, the reaction microchamber is patterned both top and bottom substrates for the purpose of increasing the storage volume capacity of the chamber, so that the reaction microchamber can contain about 3.3 µl of mixture of blood and serum.
After the prescribed reaction time, reacted (or non-reacted) mixtures of blood and serum are injected into the connected detection microfilters for observation of agglutination.
Insets of Fig. 1 schematically show the detection microfilters designed in this study. Various trapezoidal planar shaped micropillars that make the filters are periodically arranged at the regular interval of 500 µm in the downchannel direction. The trapezoidal shape of the micropillars makes a wide inlet and a narrow outlet of the filter gaps. This wide inlet guides the injected agglutinated or non-agglutinated RBCs to be easily flowed into the narrow outlet so that it enhances the efficiency of the microfilter. The detection microfilter was designed to have five gradually decreasing outlet filter gaps, i.e., the minimum distances between the adjacent micropillars in the transverse direction, along the downstream direction: 200 µm (2 columns), 150 µm (2 columns), 100 µm (6 columns), 80 µm (10 columns) and 50 µm (10 columns) (see electronic supplementary information (ESI) for detailed dimensions of the microfilter†). It might be mentioned that we determined the minimum size of the micropillar features and the filter gaps as 50 µm, taking into account processing conditions of photolithography and microinjection moulding processes, which are the fabrication method of the present disposable microfluidic biochip. With the present design of the microfilter, normal RBCs can easily pass the filter gaps without any interference, whereas the large size agglutinated RBCs of which the size is greater than 50 µm could not pass the filter gaps and thereby are filtered during the flow through the microfilter.
Fig. 2 illustrates the schematic diagram of the whole fabrication process of the blood typing integrated microfluidic biochip. The fabrication process of the mould inserts is as follows (Fig. 2(a)–2(d)). First of all, two chrome photomasks corresponding to the top and bottom layers of the biochip were fabricated based on the design shown in Fig. 1. Two nickel substrates, of which the diameter and thickness are 3 inch and 1.5 mm, respectively, were prepared as the mould inserts for the microinjection moulding process. The nickel disks were cleaned and then dried in a 120 °C oven for about 2 h before the spin coating process.
![]() | ||
Fig. 2 Fabrication process of the blood typing integrated microfluidic biochip: (a) UV photolithography, (b) SU-8 and OmniCoat™ layers patterned on the nickel substrates as masters for electroplating, (c) electroplated nickel in the masters, (d) nickel mould inserts after removing SU-8 and OmniCoat™ layers, (e) injection moulding of COC substrates and (f) final microfluidic biochip after thermal bonding. |
In order to improve the adhesion between the nickel surface and the thick negative photoresist of SU-8, first an adhesion promoter, OmniCoat™ (Microchem Corp., MA, USA), was spin coated over the nickel substrates at a speed of 3000 rpm for 30 s and baked on the hot plate at 200 °C for 1 min. The SU-8 2075 (Microchem Corp., MA, USA) was spin coated on the baked thin OmniCoat™ layer at a speed of 950 rpm for 45 s. After spin coating, a soft baking process was carefully performed at 65 °C for 10 min and at 95 °C for 1 h and 45 min. Then, UV light was exposed to the soft baked SU-8 and OmniCoat™ layers on the nickel substrates with a total dose of 250 mJ cm−2, as schematically shown in Fig. 2(a). After the UV exposure, a post exposure baking (PEB) process was carried out at 65 °C for 10 min and at 95 °C for 10 min. Then, both SU-8 and OmniCoat™ layers were developed by immersing the nickel substrates into both a SU-8 developer and an OmniCoat™ developer (Microposit MF 319) to obtain masters for the electroplating process as depicted in Fig. 2(b).
As a final step of the mould insert fabrication, the electroplating process was carried out in a nickel electroplating bath with the prepared electroplating masters at a current density of 10 mA cm−2 for 6 h (Fig. 2(c)). Finally, two nickel mould inserts, which contain 60 µm thick electroplated nickel microstructures on the nickel substrates, of the microfluidic biochip for the microinjection moulding were achieved after stripping the residual SU-8 2075 and OmniCoat™ layers by immersing them in a stripper (Remover PG), as schematically shown in Fig. 2(d). It should be noted here that in order to realize the precise alignment of two polymer layers of the biochip during the bonding stage, the cavities of pairs of pins and holes, the complementary self-alignment microstructures, were patterned on two nickel mould inserts as indicated in Fig. 2(d).12 One may refer to ref. 12 for the detailed process of this self-assembly technique.
Figs. 3 and 4 show the photographs and representative titled SEM (scanning electron microscope) pictures of the two fabricated nickel mould inserts. It might be noted that one can clearly observe the minimal surface roughness of the electroplated nickel positive relief on the nickel substrate from the SEM pictures of Fig. 4.
![]() | ||
Fig. 3 Mould inserts of the blood typing integrated microfluidic biochip for the microinjection moulding process, which contain 60 µm thick electroplated nickel positive relief on the nickel substrate: (a) top side; and (b) bottom side. |
![]() | ||
Fig. 4 Tilted SEM pictures of the electroplated nickel positive relief (cavities) in the fabricated mould inserts: (a) flow splitting microchannel; (b) F-shape mixing unit of the SLM; and (c) detection microfilter of 80 µm gap. |
![]() | ||
Fig. 5 Tilted SEM pictures of the injection moulded microstructures in the COC substrates: (a) flow splitting microchannel; (b) F-shape mixing unit of the SLM; and (c) detection microfilter with 80 µm gap. |
Processing condition | Value |
---|---|
Injection melt temperature | 290 °C |
Mould temperature | 110 °C |
Injection pressure | 300 bar |
Injection filling time | 0.25 s |
Packing pressure | 300 bar |
Packing time | 2 s |
Cooling time | 30 s |
The microfluidic biochip was finally made by thermal bonding of the two injection moulded COC substrates by means of a hot embossing machine after drilling holes (0.8 mm diameter drill bit) for fluidic interconnection for inlets and outlets on the top side COC substrate, as depicted in Fig. 2(f). The thermal bonding process was carried out at a heating temperature of 123 °C and an applied pressure of 1.94 MPa for a duration time of 10 min. It should be noted here that thermal bonding of the present reaction microchamber requires special attention to prevent the collapse of the microchamber due to its large area. We could successfully form the reaction microchamber without the collapse via a holed pressure equalizing plate (HPEP) which includes holes of the same size and shape as the reaction microchambers.15 One may refer to the detailed bonding technique including the fabrication of the HPEP in ref. 15. During the bonding process, two replicated COC substrates were precisely aligned via the self-alignment microstructures, as indicated in Fig. 2(f). The final size of the individual biochip is 56 × 44 × 3.8 mm3. It should be noted that the total fabrication cycle time of each biochip (from the microinjection moulding to the thermal bonding) was just less than 20 min (1 min for the injection moulding, 3 min for the post processing of mechanical cutting and drilling of injection moulded COC substrates and 11 min for the thermal bonding). Fig. 6 shows the finally fabricated integrated microfluidic biochip with two blood typing units in which flow splitting microchannels, SLM's, reaction microchambers and detection microfilters are fully assembled.
![]() | ||
Fig. 6 Blood typing microfluidic biochip in which flow splitting microchannels, SLMs, reaction microchambers and detection microfilters are fully integrated. Red dyed water was injected into one blood typing unit of the biochip to enhance the observation. |
![]() | ||
Fig. 7 Blood typing experimental setup with the developed microfluidic biochip: (a) schematic diagram of the configuration; and (b) microfluidic biochip installed in the blood typing experimental setup. Human whole blood and standard sera (Anti-A and Anti-B) were already loaded. |
(1) First, three types of sample human blood (blood groups of A, B and AB) were withdrawn from three healthy young donors. (It might be mentioned that about 1 ml of sample blood was withdrawn from each donor by means of a disposable syringe.) The sample whole bloods were mixed with the heparin in a 6∶1 volume ratio to prevent coagulation.
(2) The test volume of 3 µl of each whole blood was loaded into the blood inlet reservoir of each microfluidic biochip and also each 3 µl of Anti-A and Anti-B (Murine IgM monoclonal, Serologicals Ltd.) was loaded into serum inlet reservoirs, as indicated in Fig. 7.
(3) The loaded sample whole blood and the standard sera were injected into the biochip by means of two syringe pumps with the help of saline as a working fluid through the tubing until the blood and sera flow into the reaction microchambers through the flow splitting microchannels and SLMs.
(4) The mixture of the whole blood and sera was kept for about 2 min as the reaction time of agglutination.
(5) The saline was then injected into the biochip only through the blood inlet (i.e., using only one syringe pump for blood injection) until the reacted or non-reacted mixtures of blood and sera pass through the detection microfilters and then flow out from the outlets.
(6) We observed the detection microfilters, whether agglutinated RBCs were filtered or not, with the naked eye and then we finally determined the blood group of the three sample bloods based on the filtering results.
In this study, we have also carried out the plate-type blood typing in order to confirm the blood typing results from the biochip. In the plate-type blood typing, each 20 µl of the sample whole blood (i.e., a total of 40 µl of the whole blood used) was first loaded at two test wells. Each 20 µl of Anti-A and Anti-B was then loaded at the two test wells and they were manually mixed with the already loaded blood. After 3 min of agglutination time, we observed the agglutination results.
Fig. 8 shows the successful blood typing results of the three sample bloods. Insets in Fig. 8 show microscope views of the successfully filtered agglutinated RBCs and air entrapments in the detection microfilters (and the reaction microchambers) after flushing out. It should be noted that one can easily observe the filtered agglutinated RBCs in the detection microfilters by the naked eye from the blood typing results in Fig. 8. Therefore, one is able to simply and precisely determine the blood group of the sample bloods with the blood typing integrated microfluidic biochip developed in this study. We have carried out more than three blood typing tests per each blood sample and confirmed all the successful identification of agglutination of blood samples. Also, we have found that the blood typing results of the biochip were in exact accordance with the results of the plate-type blood typing method performed in this study (see electronic supplementary information† for blood typing results of the plate-type method).
![]() | ||
Fig. 8 ABO blood typing results of the three sample bloods by means of the blood typing integrated microfluidic biochip developed in this study: (a) blood group A; (b) blood group B; and (c) blood group AB. Insets in the figures are enlarged microscope views of the filtered agglutinated red cells. |
It may be mentioned that the size of the agglutinated red cells of blood group A is generally so large that they could not pass the relatively narrow microchannel in front of the microfilter, thereby frequently remaining in the reaction microchamber, as shown in Fig. 8 (a) and (c). This might be explained by the high reaction activity of the agglutinogen of blood group A with the corresponding agglutinin.
To achieve cost effectiveness of the biochip, we have manufactured it by the microinjection moulding of COC (cyclic olefin copolymer) and thermal bonding of two injection moulded COC substrates with a total fabrication time of less than 20 min. The mould inserts for the microinjection moulding process were fabricated by SU-8 photolithography and subsequent nickel electroplating.
For the feasibility test of the developed biochip, we have demonstrated successful ABO blood typing results with the fabricated biochip. Three kinds of human blood groups, A, B and AB, were successfully determined with 3 µl of the whole blood within a test time of 3 min by means of the fabricated biochip.
Based on the results of the present work, we want to simply mention the advantages of the present biochip in terms of cost, rapidity, easiness and reliability, although it might be too early to discuss this last item. The present biochip can reduce the reagent cost since the biochip requires a smaller test volume (about 3 µl) than the previous methods (more than 20 µl). The fabrication cost of the biochip may be also less than the gel card, automated system and miniaturized system, since we applied a mass fabrication method to the fabrication of the biochip. The microfluidic channel of the biochip increases surface to volume ratio so that it could enhance the reaction of agglutinogens on RBCs and agglutinins in serum, thereby resulting in the fast testing time. Because of the compact size of the biochip, it is easy to handle. The multi-step microfilter of the biochip increases chances to filter out agglutinated RBCs and, therefore, it can increase the reliability of the biochip. Also, one can preserve the blood typing test result performed with the biochip.
This integrated microfluidic biochip developed in this study may form the basis of development of microfluidic devices for automated diagnosis, suitable for the level of hospitals in function and capacity. In addition to that, the proposed biochip could be applied to clinical examinations based on agglutination reactions in antigen–antibody reactions, not to mention the blood typing.
Footnote |
† Electronic supplementary information (ESI) available: supplementary figures. See DOI: 10.1039/b516495h |
This journal is © The Royal Society of Chemistry 2006 |