Colm P. McMahona, Gaia Rocchittaab, Pier A. Serraab, Sarah M. Kirwana, John P. Lowryb and Robert D. O'Neill*a
aUCD School of Chemistry and Chemical Biology, University College Dublin, Belfield, Dublin 4, Ireland. Fax: +353-1-7162127; Tel: +353-1-7162314E-mail: Robert.ONeill@UCD.ie
bUCD School of Biomolecular and Biomedical Sciences, University College Dublin, Belfield, Dublin 4, Ireland
First published on 9th November 2005
The apparent Michaelis constant, KM, for glutamate oxidase (GluOx) immobilised on Pt electrodes increased systematically with enzyme loading. The effect was due, at least in part, to electrostatic repulsion between neighbouring oxidase molecules and the anionic substrate, glutamate (Glu). This understanding has allowed us to increase the Glu sensitivity of GluOx-based amperometric biosensors in the linear response region (100 ± 11 nA cm−2 µM−1 at pH 7.4; SD, n = 23) by incorporating a polycation (polyethyleneimine, PEI) to counterbalance the polyanionic protein. Differences in the behaviour of glucose biosensors of a similar configuration highlight a limitation of using glucose oxidase as a model enzyme in biosensor design.
The high spatial and temporal resolution achievable with implantable amperometric biosensors has motivated the design of a number of Glu biosensor configurations incorporating the enzyme glutamate oxidase (GluOx).6,7,9–15 Based on Pt wire modified with GluOx immobilised using PPD, poly(o-phenylenediamine), the Pt/GluOx/PPD cylinder biosensor, designed specifically for detection of Glu in vivo,15 has shown promising responses for monitoring Glu in the alert brain implanted with probes of moderate sensitivity in the linear response region (∼20 nA cm−2 µM−1),16 whereas sensors of lower sensitivity (∼10 nA cm−2 µM−1) failed to detect Glu changes associated with mild behavioural stimulation.17 Thus, there is an urgent need to increase Glu biosensor sensitivity for a range of applications, including detection of small changes in ECF Glu levels. A recent study of the same Pt/GluOx/PPD design, but based on smaller Pt disks, provided a useful increase in Glu sensitivity (∼30 nA cm−2 µM−1), and showed that both geometrical configurations had good oxygen tolerance.18 Here we investigate the effects of surface enzyme loading on the Glu response of both cylinder and disk Pt/GluOx/PPD designs, a study which has led to a significant increase in linear region sensitivity (∼100 nA cm−2 µM−1) of these implantable devices.
FAD is the prosthetic group of many oxidases, including GluOx,19 and molecular oxygen the co-substrate, which re-oxidises the FADH2 to form H2O2 (reactions (1) and (2)) that can be detected amperometrically (reaction (3)). A number of sophisticated mathematical models of the behaviour of enzymes in membranes have been described.20–24 These complex analyses are often needed to understand and optimise the behaviour of thick and/or conducting layers.24,25 However, a recent study has shown that substrate diffusion is not limiting for PPD layers incorporating enzyme, due to their relatively small thickness.26 Therefore, the basic Michaelis–Menten enzyme parameters used here provide more readily accessible insights into factors affecting the responsiveness of biosensors fabricated from ultrathin (10–30 nm)27–30 insulating polymers, such as PPD.
L-glutamate + H2O + GluOx/FAD → α-ketoglutarate + NH3 + GluOx/FADH2 | (1) |
GluOx/FADH2 + O2 → GluOx/FAD + H2O2 | (2) |
H2O2 → O2 + 2H+ + 2e | (3) |
A two-substrate model is necessary to describe the kinetics of oxidase enzymes under conditions of varying concentration of both substrate and O2.31,32 When the concentration of the co-substrate is constant, however, the two-substrate equation simplifies to the one-substrate Michaelis–Menten form (eqn (4)), where the current density for the biosensor Glu response, JGlu, is a measure of the overall rate of the enzyme reaction, and Jmax is the JGlu value at enzyme saturation. Different values of Jmax, determined under the same conditions, reflect differences in the amount of active enzyme on the surface, provided the sensitivity of the electrode to H2O2 (reaction (3)) does not vary, as was the case for the PPD-modified Pt cylinders and disks used here.33–35
![]() | (4) |
The Michaelis constant, KM, is defined in terms of the rate constants for the generalised reactions (reaction (5)) describing the conversion of substrate (S) to product (P), catalysed by enzyme (E); see eqn (6). When eqn (4) is used to approximate the two-substrate case, the KM is more complex, containing co-substrate terms. KM is then the apparent Michaelis constant and phenomenologically defines the concentration of substrate that gives half the Jmax response. Thus, changes in KM are sensitive to the binding constant, k1, and have often been interpreted in terms of barriers to substrate/enzyme binding,36,37 as well as changes in oxygen demand.38 Here we present data that suggest that GluOx molecules (pI = 6.2)19 packed on an electrode surface represent an electrostatic barrier to Glu accessing the enzyme active site; this limits enzyme loading as a stand-alone strategy for increasing sensitivity of Glu biosensors in their linear response region, a design feature which is crucial for successful applications in neurochemical studies where Glu concentrations are in the low micromolar range.39,40
![]() | (5) |
![]() | (6) |
Experiments were computer controlled with data collection accomplished using either a Biodata Microlink interface or a National Instruments (NI, Austin, Texas) AT-MIO-16 data acquisition board linked to a low noise, low damping potentiostat (Biostat II, Electrochemical and Medical Systems, Newbury, UK). All electropolymerisations and calibrations were carried out amperometrically in quiescent PBS solution, pH 7.4: NaCl (BDH, AnalaR grade, 150 mM), NaH2PO4 (BDH, AnalaR grade, 40 mM) and NaOH (Sigma, 40 mM). After rinsing and a settling period at +700 mV vs. SCE in 20 mL of fresh air-saturated PBS, calibrations were performed to determine the response of the biosensors to enzyme substate and H2O2.
Non-linear regression analysis (Prism 4.02, GraphPad Software Inc.) was carried out on the current density responses for each biosensor to determine the apparent Michaelis–Menten constants, using eqn (4); see Fig. 1. The apparent Michaelis constant is symbolised by KM(Glu) for GluOx and KM(G) for GOx when required for clarity, but KM otherwise.
![]() | ||
Fig. 1 Current densities from glutamate calibrations carried out in PBS (pH 7.4) at +700 mV vs. SCE with two Pt/GluOx/PPD disk biosensors showing different active enzyme loading, Jmax. The following non-linear regression parameters were obtained for the high (![]() ![]() |
Values reported are mean ± SEM, unless stated otherwise, with n being the number of biosensors. The significance of differences in the values of parameters determined for distinct populations of electrodes was calculated using Student's two-tailed unpaired t-tests, with p < 0.05 taken to indicate statistical significance.
Because dip-evaporation is expected to lead to different amounts of deposited enzyme for each electrode, whose activity could be affected differently by the subsequent polymerisation step, it is not surprising that biosensors fabricated in this way from 125 µm diameter cylinders or disks displayed a range of Jmax values (see Fig. 1 and 2). Unexpectedly, however, there was also a strong dependence of KM on GluOx loading (Jmax) for disk-based Pt/GluOx/PPD biosensors (Fig. 1). Ideally (eqn (6)), KM values should be independent of enzyme loading, but in practice several factors related to the density of GluOx on the surface could affect the apparent Michaelis constant, including steric and electrostatic hindrance of Glu (reaction (1)) by neighbouring enzyme molecules, and limited oxygen supply (reaction (2)). Thus, for example, a slight increase in KM with enzyme loading for GOx-based glucose biosensors has been reported and discussed in terms of oxygen turnover.41 However, recent studies have shown that the Pt/GluOx/PPD disk-based Glu biosensors, which show higher Glu sensitivity than the corresponding cylinder-based devices (see Fig. 1), had a lower dependence on solution pO2,18 indicating that increased oxygen demand cannot be used to explain the increase in KM observed here.
![]() | ||
Fig. 2 Linear correlation analysis for KMvs.Jmax values obtained using eqn (4) (see Fig. 1) for Pt/GluOx/PPD biosensors based on cylinder (n = 21) and disk (n = 46) designs (slope = 34 ± 2 µM µA−1 cm2, n = 67). |
To examine further the effects of GluOx loading on KM(Glu), a regression analysis was performed on the respective KM(Glu) and Jmax values for Pt/GluOx biosensors fabricated from both disks and cylinders (Fig. 2). There was a significant correlation between these two paramaters (slope = 34 ± 2 µM µA−1 cm2, R2 = 0.85, n = 67), which fitted the data for both geometrical designs (see Fig. 2). It is clear, however, that the GluOx loading density was considerably greater on the disks than on the cylinders. This is consistent with retention of a dome of enzyme solution around the disk tip as it is removed vertically from the liquid during fabrication, as expected from surface tension considerations. Upon evaporation, the density of GluOx on the disk surface should therefore be higher than that achieved with the corresponding cylinder geometry. The magnitude of the increase in KM(Glu) from cylinders (∼0.5 mM) to disks (∼10 mM; see Fig. 2) was approximately 20-fold; this compares with only a 3-fold increase in KM(G) for Pt/GOx over a similar range of Jmax values, and therefore of oxygen demand.35 One explanation for these results is that the additional electrostatic repulsion between anionic Glu and the surface GluOx (pI = 6.2)19 may reduce access by Glu to the enzyme (decreasing k1, eqn (6)) as the GluOx loading builds up, an interaction not relevant to neutral glucose.
To determine whether the polymer contributed to the effects shown in Fig. 1 and 2, PPD-free Pt/GluOx biosensors (n = 4) were fabricated using glutaraldehyde as the immobilisation agent. Glu calibrations, performed after each of two enzyme layers was deposited in this way, showed that the second layer of GluOx resulted in an 80 ± 20% increase in Jmax and a 30 ± 10% increase in KM compared with layer 1 (p < 0.02), a trend similar to that for PPD-based electrodes. Furthermore, deposition of PPD after these calibrations led to only a small and statistically insignificant change in the KM: 1.4 ± 0.1 to 1.7 ± 0.1 mM (p > 0.13), indicating that neither the PPD polymer itself, nor diffusion through it, plays a major role in the effect of enzyme loading on KM.
Therefore, some increase in Glu sensitivity in the linear region (LR) of the calibration response (LR slope = Jmax/KM) is achievable by increasing the loading of GluOx in Pt/GluOx/PPD biosensors18 (e.g., the higher LR slopes for disk biosensors in Fig. 1 compared with previous reports for PPD-based cylinder designs12,15,16). However, much of the additional enzyme activity (Jmax) does not translate down to boosting the response at lower Glu concentrations due to an increased KM value, possibly caused by an anionic electrostatic barrier associated with enzyme crowding. The following experiments were designed to test this hypothesis.
![]() | ||
Fig. 3 Schematic representation of the immobilisation of glutamate oxidase molecules (GluOx, 140 kDa,19 polyanionic) over a molecule of polyethyleneimine (PEI, 750 kDa, polycationic). The PEI helps reduce the electrostatic barrier between neighbouring enzymes and the anionic substrate (Glu), significantly decreasing the apparent Michaelis constant (KM) for Glu and increasing Glu sensitivity in the analytically important linear response region. |
An analysis for PEI-containing biosensors, similar to that in Fig. 2, showed that PEI reduced the regression slope by a factor of ten, but had no effect on the intercept which represents KM(Glu) in the limit of zero loading of GluOx (∼0.4 mM, comparing favourably with the solution value of 0.2 mM19). Thus, PEI did not significantly affect the interaction of Glu with GluOx when the enzyme molecules were widely separated (zero loading limit). This finding provides evidence that the effect of PEI is not mediated by diffusion effects through the PEI/GluOx/PPD matrix, and is consistent with our ‘ultrathin-layer’ enzyme kinetic analysis,26 which avoids the concept of the Thiele modulus.32 It appears, therefore, that PEI specifically inhibits the repulsion between Glu and neighbouring GluOx molecules in the high loading, crowded, region (illustrated in Fig. 3), but does not modify significantly the interaction between Glu and its host enzyme molecule.
Two sets of control experiments were carried out to test the conclusion that electrostatics played a major role in the PEI-reversal of the increase in KM(Glu) caused by GluOx loading. In the first, to demonstrate that the effect was not simply due to the immobilisation of the enzyme on an organic layer as opposed to the bare metal, disk-based biosensors were prepared using a zwitterionic lipid, PEA, layer in place of the polycationic PEI. The globally neutral lipid coating had no significant effect on KM(Glu): 3.6 ± 0.7 mM, n = 11; p > 0.22 compared with PEA absent. In the second, a PEI polycationic layer was included in GOx-based biosensors. The presence of PEI in the glucose biosensors had no significant effect on the KM(G) value: 12.3 ± 0.3 mM (PEI, n = 4); 11.1 ± 0.7 mM (no PEI, n = 4); p > 0.22. These results, taken together, support the electrostatic hypothesis.
The combination of high Jmax and low KM for Pt/PEI/GluOx/PPD disk biosensors gave rise to superb sensitivity to Glu (LR slope), the exact value depending on the amount of surface PEI. When a 5% PEI solution was used in the dip-evaporation procedure, subsequent Glu calibrations gave a LR slope of 62 ± 5 nA cm−2 µM−1 (n = 12), compared with 100 ± 5 nA cm−2 µM−1 (n = 8) for 1% PEI (p < 0.001). This latter value represents a new benchmark value (Fig. 4) which has not been surpassed by other Glu biosensor designs.7,11,13,14,42,46,47
![]() | ||
Fig. 4 The linear region calibration slope (LRS) for glutamate determined using cylinder and disk biosensors fabricated by immobilising GluOx on either bare Pt or PEI-coated Pt. Data are pooled for one dip of 1–5% PEI followed by 1–4 dips of GluOx solution, except for PEI-modified disks (optimum 1% PEI, 1–2 dips GluOx; see text). Mean ± SEM with number of biosensors in brackets. |
One of the reasons for exploring the disk geometry in the design of implantable biosensors was the increased LR slope achieved through the extra loading available on disks.18 The final step in these studies was to investigate the effects of PEI on the LR sensitivity of cylinder-based Glu biosensors. As observed for the disk design, there was a substantial increase in the LR slope for biosensors fabricated from PEI-coated Pt cylinders. Surprisingly, however, there was no significant difference between the responses for cylinder biosensors produced with different amounts of PEI and GluOx. All combinations of one dip of either a 1% or 5% PEI solution followed by either two or four dips of the enzyme solution gave similar responses (nA cm−2 µM−1: 108 ± 1, n = 4; 100 ± 4, n = 11; 97 ± 4, n = 4; and 96 ± 3, n = 4). These values were therefore pooled: 100 ± 3 nA cm−2 µM−1, n = 23. In an attempt to increase the sensitivity further, a layer-by-layer approach described in the literature for the fabrication of sensors using a variety of polyelectrolytes48–51 was investigated. This was found not to be advantageous here because the deposition of polycationic PEI over the polyanionic enzyme led to a marked loss of Glu sensitivity, presumably due to the large size of PEI molecules relative to GluOx (see Fig. 3).
The LR slopes for both cylinders and disks are compared in Fig. 4. The enhanced LR slope achieved through the extra loading on PEI-free disks18 was not apparent in the presence of PEI. Thus, the polycation also enhanced the deposition of the enzyme on the vertical cylinder surface during the dip-evaporation process, and led to indistinguishably high sensitivities for the two PEI-coated geometries. The lack of a major effect of PEI on Glu sensitivity for redox hydrogel based biosensors reported previously42 may be due to the more dispersed distribution of enzyme in the hydrogel matrix. That regime coincides more with the limit of zero loading described here, where there was no effect of PEI on the KM(Glu) value.
Given the small area of these Pt disk sensors (similar to 0.5 mm cylinders made from 10 µm diameter carbon fibre),7,47 biosensors based on this configuration would provide excellent Glu sensitivity and spatial resolution in neurochemical monitoring involving small brain areas or layered structures in vivo.11,52 The enhanced sensitivity of the cylinder design will allow these larger devices to find applications in larger brain regions, such as dorsal striatum and accumbens, where the greater signal strength will be an important advantage in early in vivo investigations. The results presented here also demonstrate some limitations of using GOx as a model enzyme for developing biosensors for other enzyme systems, and provide an explanation of why multiple dip-evaporation steps increased the LR slope for biosensors incorporating GOx, but not GluOx, in a previous study.15
This journal is © The Royal Society of Chemistry 2006 |