Maury E.
Howard
a and
Robert D.
Vocke, Jr.
b
aDepartment of Chemistry and Physics, Southeastern Louisiana University, SLU Box 10878, Hammond, Louisiana, 70402, USA. E-mail: mhoward@selu.edu; Fax: 985 549 5126; Tel: 985 549 5122
bChemical Science and Technology Laboratory, National Institute of Standards and Technology, 100 Bureau Drive, Mail Stop 8391 Gaithersburg, Maryland, 20899, USA. E-mail: robert.vocke@nist.gov; Fax: 301 869-0413; Tel: 301 975 4103
First published on 15th October 2004
This paper describes a robust and quantitative method to separate, purify, and assay the amount of chlorine in fossil fuels using isotope dilution mass spectrometry. The digestion/extraction process uses Carius tubes containing the fossil fuel samples in the presence of HNO3, AgNO3 and a 37Cl spike. The closed system oxidation permits complete equilibration of the sample and spike Cl. The evolved chlorine is trapped as AgCl within the Carius tube, and can then be separated and cleaned. The purified AgCl is also amenable to isotopic analysis by solid or gas source isotope ratio mass spectrometry. For isotope dilution mass spectrometry, the chlorine can be readily measured in a thermal ionization mass spectrometer in a negative ion mode as 37Cl/35Cl ratios with an approximate detection limit (LOD) of 0.3 μg Cl and a precision of 0.2%, relative. This technique may also be used to produce samples suitable for high precision measurements of chlorine isotopic variations in fossil fuels and chlorinated hydrocarbons.
Chlorine acts both as a direct pollutant and as a catalyst in chemical reactions occurring within power plant exhausts. Most of the chlorine present in coal is emitted during combustion and there is evidence of soil, air, and water contamination from chlorine4 in the regions surrounding larger thermoelectric power stations. During incineration, chlorine is also thought to play a significant role in the production and release of reactive gas mercury (RGM) to the atmosphere.5 Moreover, the presence of chlorine in combustion gases increases plant operation costs by accelerating the degradation of the stainless steel components of the exhaust and heat transfer systems.6–8 The increased utilization of the extensive and relatively untapped coal reserves in the deep, high-chlorine coal seams of the Illinois Basin have sparked concern over the potential impact of higher chlorine levels on both the environment and industry. Thus, accurate Cl assays of the feed material for power generation are necessary in order to characterize and minimize deleterious effects from both perspectives.
A recent study by the Electric Power Research Institute (EPRI)9 reported that while there are no commonly employed methods for the reliable measurement of chlorine in coal below 200 parts per million (ppm), there is a clear need for methods that can measure chlorine down to 100 ppm. A subsequent EPRI report10 identified a routine method that could potentially achieve reproducible results down to 10 ppm chlorine in coal; however, accurate standards are needed to calibrate and verify such results, especially at the lower concentration range for chlorine.
Common chlorine extraction/separation methods typically require several stages during the purification/extraction process, thereby making quantitation difficult. Recently, in an attempt to improve the accuracy and precision of chlorine measurements, isotope dilution mass spectrometry (IDMS) was coupled with pyrohydrolysis,11 a frequently used chlorine extraction method for geological samples. While an improvement over existing extraction method, this approach required a quantitative recovery of the sample/spike mix to assure isotopic equilibration. This is sometimes difficult to achieve and impossible to ensure.
In an effort to overcome these difficulties, Carius tube combustion was adapted for the extraction and isolation of chlorine. This method has already proven to be extremely effective for the extraction of sulfur and mercury from coals and other fossil fuels.1,2,12–14 By combining this technique with isotope dilution negative ion thermal ionization mass spectrometry (ID-NTIMS), accurate chlorine quantitation is readily achieved (Fig. 1). The high temperature, high-pressure and closed system nature of Carius combustion is well suited to light element extractions that might otherwise be compromised during digestion and consequent volatilization. In fact, Carius combustion has been used in organic microanalysis since the 1940s for the quantitative determination, by gravimetry, of halogens and sulfur in organic compounds and matrices.15–17 Simplicity and selectivity are its distinctive traits; extraction and separation are coupled in a single step, while the environment within the Carius tube assures equilibration of spike and sample without concomitant loss of either component.
![]() | ||
Fig. 1 Schematic view of the analytical procedure. Abbreviations are: ID (isotope dilution), IDMS (isotope dilution mass spectrometry), NTIMS (negative ion thermal ionization mass spectrometry), PTIMS (positive ion thermal ionization mass spectrometry), GS-IRMS (gas source isotope ratio mass spectrometry), GC (gas chromatography). |
![]() | ||
Fig. 2 Carius tube inside its steel casing: (a) Carius tube; (b) NaHCO3; (c) “dry ice”. |
The Carius tubes were set vertically in dry ice to freeze the sample solutions, and sealed with a gas–oxygen torch. The tubes were then placed in steel shells along with about 50 g of solid CO2 to equalize the pressure built up in the glass tubes when heated. Several grams of NaHCO3 had been previously placed in the shell to neutralize any acid that may be lost from the tube due to a pinhole leak or explosion during combustion. The samples were heated to ca. 250 °C overnight (ca. 16 h) in an oven. After cooling to room temperature, the Carius tubes were removed from the steel shells and placed behind an explosion shield. The tubes were then further cooled to ∼0 °C and the necks heated with a torch to open a small vent hole, releasing the internal pressure. The tubes were scored below the shoulder and cracked open by application of heat from a torch. The tops of the tubes were removed and the contents transferred into centrifuge tubes using a high-purity water rinse (final volume, 11–12 ml).
For the blank study, several tubes were also constructed with longer necks as part of a more aggressive cleaning study. In this study, the Carius tubes were filled with 6–7 ml of concentrated high purity HNO3 and sealed. They were then loaded into steel shells and heated to 250 °C overnight, following the procedure for digestion of samples and controls. These tubes, after cooling, were opened, covered with Parafilm® and taken to a class 10 clean room. After rinsing with high purity water, each tube was filled with high purity water and allowed to stand for several hours. All tubes were dried under laminar flow in the clean room.
A wide aperture thin lens source, modified for use with negative ions, was used for ion extraction, collimation and acceleration and is shown in Fig. 3. Pertinent modifications were as follows: (1) magnets were placed in the shield plate to deflect electrons produced at the filament, thereby preventing them from entering the lens stack where they could damage the slits; (2) six 9 V batteries, wired in series, were connected to the first lens plate to prevent the electrons from charging the lens and deflecting the ion beam.
![]() | ||
Fig. 3 Modified wide aperture thin lens source. The Cl solutions were loaded on the side filaments of the filament assembly on the extreme left side of the figure. Ions were generated thermally and accelerated to and through the slit plate on the right side of the figure. |
The Re sample and ionization filaments were out-gassed at <10−4 Pa for 5 min at 2.5 A, then for 35 min at 3.5 A in a special filament out-gasser of NIST design. All filaments were also “flash cleaned” prior to use (to eliminate alkalis) by turning them on at 3.5 A for 5 s, then off for 5 s, over a period of 1 min (also at <10−4 Pa). Two side filaments were loaded under a laminar flow hood with ca. 9–12 μg Cl (about 50% of mass per filament). The amount of solution needed to load a sample was typically 2–3 μl for the spiked samples, 4–5 μl for the unspiked samples and controls, and 9–10 μl for the blanks. Samples were dried under a 250 W infrared heat lamp as they were loaded.
The loaded and dried filaments were then assembled in a triple filament configuration with a central Re ionizing filament and put into the mass spectrometer. The source chamber was then allowed to pump down to ca. 10−4 Pa and, after filling the cold finger with liquid N2, was opened to the flight tube and detector. The source vacuum was typically in the 10−5 Pa range when the samples were being heated and ionized. Controlled heating was achieved by stepping the central filament current up by 0.2 A min−1 until a temperature of ca. 1375 °C was reached (∼2.4 A), then the current was increased by 0.1 A min−1 to ca. 1550 °C (∼3.0 A). The side filaments were then taken to approximately 0.5 A, and the center filament ramped a bit higher to 1600–1610 °C. The beam was centered and focused to give maximum intensity and the baseline was measured. The mass ratio 37/35 (37Cl/35Cl) was then measured in multiple blocks of 5 ratio sets over a period of 10–60 min. Data was generally taken over the period during which the Cl-signal reached its maximum intensity as shown in Fig. 4. The final Cl isotopic ratio precision for all sample runs was better than 0.3% (RSD).
![]() | ||
Fig. 4 Typical variations observed in the Cl-signal and the measured Cl isotope ratio during the time of sample conditioning (darker shading) and data collection (lighter shading). |
Instrument mass fractionation was corrected using a T/E (true ratio/experimental ratio) correction. This was calculated by ratioing SRM 975a’s certified 37/35 absolute abundance ratio to the average 37/35 ratio measured in the isotopic standard. This correction was computed and applied in a batch or bracketed mode to samples run between the standards used to monitor the instrument fractionation.
By cleaning the Carius tubes under combustion conditions, it was hoped that leaching might be enhanced and that would be more effective in reducing the blank. Therefore, tubes were designed with longer necks that could be sealed and cleaned under combustion conditions, then opened and resealed as described in the preceding section on cleaning procedures. The results of this study, listed in Table 1, suggest that this more aggressive method of cleaning did not reduce the blank to a significant degree. This suggests that either there must be some other source for this portion of the blank or that the possible blank reservoir represented by the interior of the tube is effectively infinite. A leaching front progresses into the glass, leaving behind a porous glass interface that communicates the blank Cl with the reacting solvents. At present, the data do not allow for a distinction between these possible sources.
![]() | ||
Fig. 5 Comparison of measured chlorine concentrations for NIST SRM 1632c, Trace Elements in Coal. |
Cl/μg g−1 | σ/μg g−1 | RSD (%) | |
---|---|---|---|
Certificate | 1139 | ± 41 | 3.6 |
ID-PTIMS | 1143.9 | ± 31.7 | 2.77 |
ID-NTIMS | 1140.0 | ± 6.2 | 0.55 |
The Carius ID-NTIMS method was also applied to three different SRM coals with varying chlorine contents: 2692b, a bituminous coal (ca. 1600 μg g−1 Cl) with 7.90 ± 0.30% ash content and 1.24 ± 0.03% moisture content; 2685b, a bituminous coal (ca. 530 μg g−1 Cl) with 15.94 ± 0.30% ash content and 2.32 ± 0.06% moisture content; 2682b, a subbituminous coal (ca. 76 μg g−1 Cl) with 6.32 ± 0.42% ash content and 12.79 ± 1.29% moisture content. The results of these analyses are shown in Table 3. The values obtained agree closely with the estimated values for the coals with higher Cl concentrations, SRM 2685b and 2692b, based on consensus values from CANSPECS 58 and CANSPECS 2000–2, respectively. However, the concentration measured in the lower-level subbituminous coal, 18.3 ± 0.6 μg g−1 (1 sd) is significantly lower than the reported CANSPECS 55 consensus value of 76 μg g−1. However, taking into account the high standard deviation of the CANSPECS 55 number, the value falls well within the statistical uncertainty.
Cl/μg g−1 | σ/μg g−1 | RSD (%) | ||
---|---|---|---|---|
a Values based on two runs of single independent samples. b Values based on two runs each of two independent samples. | ||||
2692b | CANSPECS 2000–2 | 1593 | ± 200 | 12 |
Carius/NTIMSa | 1650 | ± 24 | 1.5 | |
2685b | CANSPECS 58 | 530 | ± 161 | 30 |
Carius/NTIMSa | 540 | ± 3 | 0.6 | |
2682b | CANSPECS 55 | 76 | ± 102 | — |
Carius/NTIMSb | 18.4 | ± 0.6 | 3.3 |
The method employed by the CANSPECS 55 round robin for chlorine analysis was ASTM D 4208, Bomb Combustion Ion-Selective Electrode. ASTM reports a method reproducibility of only 77 μg g−1. Additionally, this method was one of several evaluated in an EPRI study evaluating chlorine analysis methods. EPRI stated that of the three chlorine methods they compared: “ASTM D 2631 Bomb Combustion Potentiometric Titration, ASTM D 4208 – Bomb Combustion Ion Selective Electrode, and Bomb Combustion Ion Chromatography. None of these methods could measure chlorine in coal with acceptable accuracy or precision, particularly at levels below 200 μg g−1 [equivalent to parts per million (ppm)].”9
This paper demonstrates that Carius combustion combined with NTIMS can provide robust and accurate values for chlorine in coal. It does not suffer from the complications of other commonly used extraction methods. The closed system extraction allows little chance for loss and assures equilibration of the sample and spike making ID analysis possible. Additionally, since the sample and spike chlorine are trapped as AgCl after extraction, a simple dissolution in ammonia and separation from the ash components of the sample are all that is required for sample preparation. Final analysis by NTIMS is also an advantage, in that detection limits are not really an issue. The amount of sample loaded for analysis by NTIMS may be adjusted to gain adequate intensity for an accurate ratio measurement.
Finally, additional applications such as the measurement of chlorine isotopic ratios are of a more exploratory nature, and have yet to be tried using this method. However, if Carius combustion proves robust for high-precision chlorine isotope ratio measurements, it may provide a straightforward way of assessing CAH contamination sources and the evolution of these chlorinated contaminants at a particular site.
This journal is © The Royal Society of Chemistry 2004 |