Ying
Zhou
ab,
Patrick J.
Parsons
*ac,
Kenneth M.
Aldous
ac,
Paul
Brockman
d and
Walter
Slavin
e
aWadsworth Center, New York State Department of Health, P.O. Box 509, Albany, NY 12201-0509, USA
bDepartment of Chemistry, The University at Albany, Washington Avenue, Albany, NY 12222, USA
cDepartment of Environmental Health and Toxicology, School of Public Health, The University at Albany, PO Box 509, Albany, NY 12201-0509, USA
dExeter Analytical Inc., 7 Doris Drive, North Chelmsford, MA 01863, USA
eBonaire Technologies, Box 1089, Ridgefield, CT 06877, USA
First published on 13th December 2000
Several novel W-filament designs are investigated and compared as atomization devices for blood Pb measurements by atomic absorption spectrometry. Two designs (short and long wire filaments) are based on rigidly-wound W-filaments from Osram, while a third (etched filament) is based on lithographic imaging and photoetching technology. The filaments are positioned in an aluminum atomization cell that is mounted in the optical path of a compact AAS instrument. This instrument is designed specifically for blood lead testing using W-filament technology with a Pb hollow cathode lamp and incorporates a self-reversal background correction system. Whole blood is diluted 1 + 4 with a phosphate modifier, Triton X-100 and nitric acid, and 15 µL are deposited on the filament using a micropipette. A simple heating program dries, pyrolyzes and atomizes the sample. Integrated absorbance measurements are used for improved precision and detection limits. Matrix-matched Pb standards are necessary for blood lead calibration. Of the three filaments studied, the long wire filament proved most suitable, with a characteristic mass (m0) of 200 pg, a method detection limit (3σ) of1–2 µg dL−1 (0.05–0.10 µmol L−1), and a lifetime of the order of 60–70 firings. Method accuracy is verified using blood lead reference materials from the National Institute of Standards and Technology (SRM 955a,b), the Centers for Disease Control (BLLRS) and the New York State Department of Health (RMs 028, 040, and 043). For the long wire filament, blood lead results below 40 µg dL−1 were mostly within ±1 µg dL−1 of certified values, and within ±10% above 40 µg dL−1; within-run precision was almost always better than ±10%. Additional validation is reported using proficiency test materials and human blood specimens. All blood lead results were within the acceptable limits established by regulatory authorities in the US.
Currently, the principal technology for blood lead measurement is electrothermal atomization atomic absorption spectrometry (ETAAS) using a graphite furnace (GFAAS) with platform atomization. Its high sensitivity, accuracy and low detection limits makes it suitable for screening purposes and clinical diagnosis of lead poisoning. However, its relatively large size, high cost and electrical power requirements limits its use to centralized clinical laboratories. In the US, such laboratories are regulated under the Clinical Laboratories Improvement Amendment of 1988 (CLIA '88)3 and must participate successfully in a blood lead proficiency testing (PT) program approved by the US Department of Health and Human Services.
Since 1993, the CDC has encouraged the development of new, innovative technologies for blood lead measurements suitable for use in point-of-care testing. The desirable characteristics of such technologies are: precision of <±1 µg dL−1 to support a detection limit of 1–2 µg dL−1; accuracy performance within CLIA '88 requirements (±4 µg dL−1 for blood lead levels of less than 40 µg dL−1; ±10% for blood lead levels of above 40 µg dL−1); minimum specimen volume of around 200 µL to enable capillary blood lead measurements; and portable, low cost per test, easy-to-operate, and analysis time of less than 5 min per specimen.
Electrochemical technologies such as anodic stripping voltammetry (ASV) are well established for blood lead measurements.4 Recent improvements in electrochemical methods have led to new approaches for blood lead measurement based on disposable microarray electrodes5,6 and screen-printed sensors.7,8 Portable, hand-held instrumentation for blood lead testing based on ASV with disposable microarray electrodes are now being marketed commercially. However, electrochemical methods have the disadvantage of requiring Pb to be decomplexed from proteins and other macromolecules for reaction at the working electrode.
Others have proposed using a capacitively coupled microwave plasma (CCMP) coupled with atomic emission spectrometry (AES) for blood lead determinations.9,10 In one approach, a 2 µL aliquot of whole blood is directly deposited on a tungsten filament, which becomes the plasma-supporting electrode. A microwave-induced plasma subsequently dries, pyrolyzes and atomizes the sample leading to Pb atomic emission signals.
The use of tungsten filaments as electrothermal atomization devices in AAS began as early as 1972 with the pioneering work of Williams and Piepmeier,11 who employed a wound filament purged with Ar gas. Berndt and Schaldach12 later proposed using inexpensive commercially available tungsten coils (150 W, 15 V) enclosed in a glass cell and mounted on the optical bench of a commercial AAS instrument in place of a graphite furnace or burner head. This technology was later transferred to other laboratories for further application development.13 Several groups have described the use of the tungsten filament atomizer for trace element determinations in biological materials,13 ground water and tap water.14 Other applications include barium in water,15 Al in hemodialysis fluid16 and cobalt in animal feces.17
The application of tungsten filament atomization to blood lead measurements has been independently proposed by several groups.18,19 In our early work, lead was atomized directly from diluted blood using a 150 W tungsten filament operated with a current controlled 12 V power supply. Further refinements and additional details were later published20 simultaneously with a report from another group21 who described using W-filament AAS to measure Pb in blood samples following protein precipitation to overcome matrix interferences. Other workers have also explored W-filament AAS for blood lead measurements, with some proposing pretreatment of blood with either microwave digestion22,23 or preconcentration and extraction into methyl isobutyl ketone (MIBK) to overcome matrix interferences.24
In this paper, we report advances in the measurement of blood lead using tungsten filaments of various designs mounted on a compact AAS instrument designed specifically for blood lead measurements using W-filament atomization, and operated with a self-reversal background correction scheme.
![]() | ||
Fig. 1 Schematic drawings of the Osram HLX short wire filament, the P-290 etched filament and the Osram H-1431 long wire filament. All dimensions are in millimeters. |
![]() | ||
Fig. 2 Exploded view (not to scale) of the atomization cell used in these studies. |
A second filament (etched filament) was manufactured from a 0.18 mm thick tungsten sheet (Rembar, Inc., Dobbs Ferry, NY, USA) using photoetching technology performed by Buckbee-Mears, Inc. (St. Paul, MN, USA). The design is first developed into a mask that is used to transfer the coil images to the tungsten sheet using photo lithography. The filaments are then separated from the tungsten sheet using a chemical etching process. We explored two styles of etched filaments before settling on the final design (P-290) that was 22 mm in length, of which 5.6 mm comprised the middle segment that could accommodate up to 20 µL of sample (Fig. 1). The etched filament incorporated a novel three-segment arrangement, designed to improve the isothermality of the middle section of the filament on which the sample is deposited by isolating it from the cooler end segments. Initially, etched filaments were installed by Scientific Instrument Services (Ringoes, NJ, USA) on nickel posts mounted on a stainless-steel backplate. Later, we spot welded them in the laboratory, but with less precision than attainable through SIS, such that each new filament required minor optimization adjustments with respect to the light beam. The backplate was then installed in the atomization cell (Fig. 2).
A third filament (H-1431 long wire filament) was obtained unmounted from Osram, Inc., (Waldoboro, ME, USA). It was similar in design to the short wire filament, but was longer by a factor of about three (13 mm) and so could easily accommodate up to 50 µL of sample (Fig. 1). These long wire filaments were spot welded in-house to nickel posts and installed in the atomization cell as described above.
![]() | ||
Fig. 3 Optical layout of the PBA-10 instrument; dashed lines indicate the optical path. |
For self-reversal background correction, the hollow cathode lamp is pulsed at 100 Hz such that uncorrected AA measurements are taken during a 1.0 ms low current (9 mA) pulse, followed by a 0.1 ms high current (167 mA) pulse during which the background absorption (BKG) signals are read. The lamp then remains off for 8.9 ms. The complete timing sequence is given in Fig. 4. Data acquisition is operated at 100 Hz to synchronize with the pulsed hollow cathode lamp. Integrated absorbance measurements are used throughout these studies.
![]() | ||
Fig. 4 Timing scheme of the pulsed hollow cathode lamp operating in self-reversal mode (not to scale). During each 10 ms cycle, uncorrected AA measurements are taken at low current pulse stage and BKG measurements taken at high current pulse stage. |
A compact power-controlled internal 12 V dc supply was initially used to drive the etched filaments. Later it became necessary to use two different power supplies to drive the two wire filaments. We used a current-controlled 12 V dc external power supply to drive the short wire filaments through the drying, pyrolysis, atomization and cleaning stages. The long wire filaments were driven through drying, pyrolysis and atomization using the internal 12 V dc power supply, however, a 28 V dc external power supply was required to clean the filament at 75 W. A compact power controller was used to regulate the power delivered to the filament via a series of software setpoints. The setpoints are arbitrarily related to actual power, which is calculated empirically. Setpoint–power relationships were established for the three filament–power supply arrangements, and optimized power settings are given in Table 1. For the measurements in this paper, the instrument was interfaced to a personal computer, running software to control the heating program and data acquisition.
Programstep | Power/W | Hold time/s | Gas flow rate/L min−1 | ||
---|---|---|---|---|---|
HLX(Short) | P-290(Etched) | H-1431(Long) | |||
Dry | 0.8 | 0.8 | 0.8 | 200 | 1.0 |
Pyrolysis | 2.6 | 2.8 | 2.6 | 30 | 1.0 |
Cool down | 0 | 0 | 0 | 10 | 1.0 |
Atomization | 38 | 40 | 45 | 3 | 1.0 |
Clean | 65 | 73 | 75 | 60 | 1.0 |
For validation purposes, we used a Perkin-Elmer Model 4100ZL AAS instrument equipped with a transversely heated graphite furnace and a longitudinal Zeeman background correction system, and a well established method for blood lead.25
All reference materials, quality control and human blood specimens were diluted 1 + 4 with a modifier. The volume of diluted sample deposited on the filament was generally 15 µL, equivalent to 3 µL of whole blood except where indicated. A manual double-action micropipette (Finnpipette, Helsinki, Finland) was used to deposit samples on the filament. An alignment jig mounted on top of the atomization cell was necessary to guide the micropipette tip, to ensure reproducible sample deposition onto the coil (Fig. 5). The alignment jig, which was machined from an aluminum block, was designed with an inverted internal taper to minimize sample cross contamination, and a thread to provide fine adjustment of the tip to coil distance. This device was particularly important for depositing samples on the long wire filament.
![]() | ||
Fig. 5 Schematic drawing of the alignment jig. |
![]() | ||
Fig. 6 Pyrolysis and atomization curves of the long wire filament using a NYS blood lead reference material (42.8 µg dL−1). Bars indicate the range of the measurements (n = 3). |
![]() | ||
Fig. 7 The effect of purge gas flow rate on sensitivity using an etched filament. A NYS blood lead reference material (42.8 µg dL−1) was used for this study. Bars indicate the range of the measurements (n = 3). |
In contrast, our previous experience with peak height measurements using a commercial AAS optical bench20 showed improving sensitivity when flow rate was varied over roughly the same range. Similarly, Luccas et al16 reported that the aluminum peak height sensitivity almost doubled as the flow rate was increased from 0.4 to 1.4 L min−1. Use of integrated absorbance measurements here is almost certainly responsible for avoiding the susceptibility to purge gas flow rate observed with peak height.
![]() | ||
Fig. 8 Typical atomization profiles for the (a) HLX short wire filament, (b) P-290 etched filament and (c) H-1431 long wire filament using a NYS blood lead reference material (42.8 µg dL−1). |
![]() | ||
Fig. 9 Calibration curves obtained from matrix matched standards (•) and aqueous standards + modifier (o) using the (a) HLX short wire filament, (b) P-290 etched filament and (c) H-1431 long wire filament. Bars indicate the range of the three measurements. Characteristic masses (m0) and r2 statistics are labeled next to the calibration curves. |
![]() | ||
Fig. 10 Linear dynamic range ( 0–500 µg L−1) of the W-filament AAS method using the long wire filament, m0 = 304 pg. |
Filament | PE 3110 AAS | EAI Prototype PBA-10 | |||||
---|---|---|---|---|---|---|---|
Optimized | Method detection limit/ | Optimized | Method detection limit/ | Linear range/µg L−1 | |||
m 0/pg | µg dL−1 | µmol dL−1 | m 0/pg | µg dL−1 | µmol dL−1 | ||
a Data based on peak height measurements (Ap), reported in ref. 20. | |||||||
HLX (Short) | 26a | 2–3a | 0.10–0.15a | 400 | 4–5 | 0.19–0.24 | — |
P-290 (Etched) | 100 | 2–3 | 0.10–0.15 | 200 | 2–3 | 0.10–0.15 | — |
H-1431 (Long) | — | — | — | 200 | 1–2 | 0.05–0.10 | 0–500 |
Reference materials | Certified valuea/µg dL−1 | HLX (Short) | P-290 (Etched) | H-1431 (Long) | ||||||
---|---|---|---|---|---|---|---|---|---|---|
Value foundb/µg dL−1 | RSD(%) | Bias/µg dL−1 | Value foundb/µg dL−1 | RSD(%) | Bias/µg dL−1 | Value foundb/µg dL−1 | RSD(%) | Bias/µg dL−1 | ||
a Definition of uncertainty in certified values varies among the RM producers, see text for details. b Uncertainty for measured values are ±SD (n = 5). | ||||||||||
NIST SRM 955a-1 | 5.01 ± 0.09 | 3.9 ± 1.7 | 44 | −1.1 | 4.7 ± 0.9 | 19 | −0.3 | 4.0 ± 0.3 | 8 | −1.0 |
NIST SRM 955a-2 | 13.53 ± 0.13 | 11.2 ± 0.6 | 5 | −2.3 | 10.4 ± 3.1 | 30 | −3.1 | 13.3 ± 0.7 | 5 | −0.2 |
NIST SRM 955a-3 | 30.63 ± 0.32 | 27.1 ± 1.2 | 4 | −3.5 | 30.9 ± 3.3 | 11 | 0.3 | 30.3 ± 0.5 | 2 | −0.3 |
NIST SRM 955a-4 | 54.43 ± 0.38 | 52.3 ± 3.1 | 6 | −2.1 | 56.5 ± 0.7 | 1 | 2.1 | 51.7 ± 1.1 | 2 | −2.7 |
NYS RM Lot 028 | 6.7 ± 1.2 | 6.6 ± 0.7 | 11 | −0.1 | 6.5 ± 1.1 | 17 | −0.2 | 6.4 ± 0.7 | 11 | −0.3 |
NYS RM Lot 043 | 17.7 ± 0.5 | 18.7 ± 3.5 | 19 | 0.9 | 18.5 ± 0.2 | 1 | 0.7 | 17.1 ± 1.2 | 7 | −0.7 |
NYS RM Lot 040 | 42.8 ± 0.9 | 43.9 ± 0.8 | 2 | 1.1 | 43.5 ± 1.6 | 4 | 0.7 | 44.4 ± 0.2 | 0 | 1.6 |
NIST SRM 955b-1 | 4.04 ± 0.15 | — | — | — | — | — | — | 4.5 ± 0.6 | 12 | 0.5 |
NIST SRM 955b-2 | 10.30 ± 0.10 | — | — | — | — | — | — | 9.5 ± 0.3 | 3 | −0.8 |
NIST SRM 955b-3 | 20.59 ± 0.21 | — | — | — | — | — | — | 19.8 ± 0.1 | 1 | −0.8 |
NIST SRM 955b-4 | 39.36 ± 0.36 | — | — | — | — | — | — | 41.1 ± 1.2 | 3 | 1.8 |
CDC BLLRS 1198 | 45.67 ± 0.12 | — | — | — | — | — | — | 45.2 ± 1.3 | 13 | −1.1 |
CDC BLLRS 2398 | 18.49 ± 0.04 | — | — | — | — | — | — | 15.5 ± 1.9 | 12 | −3.0 |
CDC BLLRS 1299 | 5.58 ± 0.27 | — | — | — | — | — | — | 4.5 ± 0.6 | 3 | −0.5 |
Blood lead PT samples from NYS and the WSLH were also analyzed with the long wire filament and are shown in Fig. 11 as a scatterplot [Fig. 11(a)] and a difference plot [Fig. 11(b)]. These PT samples were only used as reference materials; target values were obtained from the routine PT practice. Again all reported values were within PT program acceptability limits (dashed lines) as defined above under CLIA '88.
![]() | ||
Fig. 11 Analysis of blood lead PT materials from New York State and from the WSLH using the PBA-10 instrument with the long wire filament: (a) a scatter plot of W-filament AAS values versus the target values; and (b) a difference plot of the W-filament AAS data versus the target values. Dashed lines indicate the PT program acceptability limits as defined under CLIA '88. |
For clinical validation purposes, we compared data obtained from the long wire filament with the blood lead values from a well established furnace AAS method25 for 23 human specimens. Results are shown in Fig. 12 as a scatterplot of W-filament AAS data versus furnace AAS. Agreement between the two techniques was within the ±4 µg dL−1 limits as defined above and shown as dashed lines in Fig. 12.
![]() | ||
Fig. 12 Scatterplot of 23 human blood lead values obtained from W-filament atomization using the long wire filament versus values from a well established GFAAS method.25 Dashed lines show the US regulatory limits for accuracy under CLIA '88. |
By contrast, the rigidly-wound Osram wire coils (both the short and long designs) showed better resistance to warping, perhaps because they are better able to tolerate the stresses produced under intense heating. Warping develops later in the coil lifetime compared to the etched filament. Typically Osram wire coil warping begins to form after 30–40 firings, but is not as severe as with the etched filament. Between the two Osram coils, the long wire filament is more flexible and has a lifetime of 60–70 firings compared to 40–50 firings for the short wire filament. As the long wire coil ages, the atomization peak height decreases while peak width increases, but the integrated absorbance remains virtually unchanged (Fig. 13). As with furnace AAS,29 this is another good reason to use integrated absorbance measurements rather than peak height in W-filament AAS.
![]() | ||
Fig. 13 Comparison of an atomization profile from a NYS blood lead reference material (42.8 µg dL−1) on (a) a new long wire filament and (b) the same filament after 40 firings. |
In the EAI PBA-10 instrument, we operated the data acquisition system at 100 Hz to synchronize signal measurements with the pulsing frequency of the hollow cathode lamp. Faster data acquisition, while possible, might be incompatible with the limitations for pulsing the hollow cathode lamp to achieve self reversal correction. At 100 Hz, reliable integrated absorbance measurements are feasible. Peak area measurements also improve precision and thus the method detection limit for blood lead. For example, on the long wire filament the precision for peak height measurements at a blood Pb concentration of 6.7 µg dL−1 is only 13.8% but, using integrated absorbance measurements, this improves by almost a half to 8.4%. The estimated 3σ method detection limit based on peak height is approximately 3 µg dL−1 (n = 10) while, for integrated absorbance measurements, it is roughly 2 µg dL−1.
Most previous workers12–17,21,23 used continuum background correction systems installed in commercial AAS instruments. In fact, our earlier report20 also relied on a D2 arc background correction system. But continuum correction is not really practical for a portable instrument because it requires an extra lamp and creates additional cost. One group22,24 has used a "near-line" background correction method for a portable blood lead instrument based on W-filament atomization AAS. Near-line background correction was first proposed 40 years ago30 for use with flame atomization AAS, and its application to Pb in biological fluids was described more than 30 years ago.31 This approach assumes that background absorption at both the resonance and non-resonance lines is exactly the same. However, for electrothermal atomization of complex matrices such as blood, where a structured background often appears, this is rarely true and can lead to serious errors. In the near-line method proposed by Sanford et al.,22 microwave-assisted digestion with concentrated nitric acid is required for blood lead measurements, and, in a later near-line method described by Salido et al.,24 it is necessary to separate Pb from blood matrix by chelation and extraction into MIBK solvent, prior to deposition on the coil. Other workers have proposed methods that require sample pretreatment such as protein precipitation21 and microwave digestion23 for blood lead measurement using W-filament AAS.
The self-reversal background correction system is simple, effective and requires no additional instrumental components. Self reversal is based on spectral line broadening that occurs within a hollow cathode lamp that is operated with a very high current, e.g., 100–300 mA.32 Atomic absorption measured during the high current pulse is greatly reduced but background absorption remains constant. Correction is accomplished by pulsing the hollow cathode lamp and taking the difference in absorbance measurements at low and high currents. Under a self-reversal correction scheme, sensitivity for Pb is reduced by as much as 50–60% and this is an important limitation. Thus, improvements in precision are critical for reaching the required detection limits.
Footnote |
† Use of trade names is for informational purposes only and does not imply an endorsement by the New York State Department of Health. |
This journal is © The Royal Society of Chemistry 2001 |