Duangjai Nacapricha*a, Saowapha Muangkaewa, Nuanlaor Ratanawimarnwonga, Juwadee Shiowatanaa and Kate Grudpanc
aDepartment of Chemistry, Faculty of Science, Mahidol University, Bangkok, 10400, Thailand.. E-mail: scdnc@mahidol.ac.th
bThe Institute of Science and Technology for Research and
Development, Mahidol University, Salaya, Nakorn Pathom 73170, Thailand
cDepartment of Chemistry, Faculty of Science, Chiang Mai University, Chiang Mai, 50200, Thailand
First published on 1st December 2000
This paper describes the use of flow injection (FI) techniques for the determination of iodine in urine, based on the catalytic effect of iodide in the redox reaction between Ce(IV) and As(III). The proposed procedures minimize errors in the conventional batch method arising from the reading of absorbance at a fixed time after addition of Ce(IV) reagent. Two FI systems, for the continuous and stopped modes of operation were assembled. In the continuous-FI system, a thermostated bath was used to increase the sensitivity. However this is not necessary for the stopped-FI system. The two systems are comparable in terms of sensitivity, sample throughput and detection limit. The continuous-FI and the stopped-FI exhibited detection limits (3σ) of 2.3 and 3 μg I l−1 respectively. Both systems have equal sample throughputs of 35 samples h−1. Calibration plots for both techniques are linear. The FI procedures provide very short analysis times compared to the batch procedure. Using the linear regression test, there is no significant difference between the results from the four methods, i.e., continuous-FI, stopped-FI, conventional method and ICP-MS. The proposed methods are readily applicable for automation and can be an alternative to the conventional procedure for the survey of the iodine deficiency disorder. A condition for sample digestion is also proposed to reduce the amount of chloric acid required for complete digestion. Kinetic information of the reaction can also be obtained from the stopped flow mode.
For the quantitation of free iodide, different modes of high performance liquid chromatography (HPLC) such as ion2 and ion-pair reversed-phase3–5 were proposed with either electrochemical detection3–5 or post-column reaction.2 Yabu et al. have proposed a method to measure iodide content in urine using an iodide-selective electrode.6
Total content of iodine can be determined by using a hyphenated technique such as inductively coupled plasma mass spectrometry (ICP-MS). Usually dilution of sample with water7 or other types of reagent such as nitric acid7,8 and ammonia9 is adequate for this technique. Variations in the signals of the detection mass, are corrected by addition of some internal standards, e.g., europium,8 rhodium or indium.7 An ICP-MS method for determination of urinary iodine using isotope dilution with iodine-129 was presented by Haldimann et al.9
Surrounded by other methods, the most common method used in laboratories dealing with monitoring of IDD status is a spectrometric method with catalytic effect of iodide on a redox reaction between Ce(IV) and As(III).1,10 The method is more preferable than other methods because it is more practical as a survey tool. Sandell and Kolthoff first described the use of this reaction for determination of trace iodide catalyst in 1934.11 Rodriguez and Pardue proposed some possible mechanisms based on their experimental data achieved in 1969.12
Mushtakova et al.13 proposed a Sandell and Kolthoff method for determination of free iodide in urine. Their method requires no urine digestion step prior to spectrometric detection of Ce(IV). However extensive investigation of interfering effects from potential species in urine is yet required for this method.
Description of the classical method was provided by Dunn et al.1,10 The procedure is usually carried out batchwise. Interferences to the redox reaction in urine must be eliminated prior to the step of spectrometric measurement. Digestion with chloric acid is more often used than the alkaline ashing method.14 For chloric acid digestion, iodine species are oxidised to iodate ions after digestion. Unpublished results15 apparently indicated that chloride ion added in the arsenious solution helps in conversion of iodate to iodide within less than a minute. The results have also demonstrated that the disappearance of Ce(IV) is a first-order process under the conditions of the conventional method.
After chloric acid digestion, the sample is added with As(III) solution which contains sodium chloride. Finally after an accurate volume of Ce(IV) solution is added to this solution the reading of absorbance (405–420 nm) must be taken at a fixed time for all the samples and the standards. Hence technician error can easily occur in the conventional batch method. Automation of the procedure has been reported through the use of an autoanalyser, which is based on the air-segmented flow (ASF) technique.4,16–19 The use of flow injection (FI) techniques offers various advantages.20 Nevertheless, there has been no report of a FI system to use in conjunction with this type of sample preparation.
In 1996, Yaping et al.21 proposed a FI method for the determination of iodine in urine. The method employed the Sandell and Kolthoff reaction but with the use of a different type of reagent (H2SO4 + KMnO4 + K2Cr2O7) for on-line digestion. With this mixture of oxidant, and with the detection wavelength at 480 nm, spectral interference is at risk especially from KMnO4. Another disadvantage of the method is that the method unnecessarily used a toxic brucine as a reagent stream.
In this article, two methods developed for the determination of urinary iodine based on flow injection techniques are described. The first method is proposed for use in the normal continuous mode. The system can be modified to operate under a stopped-FI mode.20 Both methods are readily applicable for automation to reduce technician error, which often arises in the convention batch method.
![]() | ||
Fig. 1 The manifold of the continuous-FI system used for determination of iodine in urine. |
The system of the continuous-FI shown in Fig. 1 was modified to operate under the stopped-FI mode. In the stopped mode, the glass reaction coil was replaced by a coil made of Tygon tube (100 cm). The water bath was not used in this system. The pumping rate for each channel was 1 ml min−1.
Urine (4.00 ml) or working iodine standard was pipetted into a test tube, which contained 4.8 ml of 28% m/v chloric acid. After a 1 h digestion in a heating block at 105–110 °C, the solution was transferred into a volumetric flask and was made up to 10.00 ml with water. The reagent blank was prepared in a similar way using deionized distilled water. There is also enough volume of liquid for replicate injections in the FI methods.
To avoid any course of error, calibration of samples was always made against the set of iodine standards digested in the same run. The working range of standards was the same for the batch and the flow methods, which ranged from 20 to 200 μg I ml−1.
![]() | ||
Fig. 2 Profiles of iodine standards obtained from the (a) continuous-FI system and (b) stopped-FI system. |
The studies were made using the continuous-FI system shown in Fig. 1. The sixteen chemical species were separately added into 1 ml aliquots of 1000 μg I l−1 potassium iodate solution in a volumetric flask. These solution mixtures were made up to the mark with water to 10.00 ml. After the digestion, the signal obtained from these solutions were compared to the signal achieved from a pure standard of potassium iodate (100 μg I l−1). A species was considered to interfere if its presence resulted in a signal alteration of greater than ±10%.
In the analysis, 10-fold dilution with 0.16 M nitric acid was made on the samples. In, added as indium nitrate, was used as the internal standard to correct for non-spectral interference and for signal instability.
With the current volume ratio, concentrations of iodine after digestion lay within the appropriate range of standards for the FI methods as well as the batch method. There was also enough volume of liquid for replicate injections in the FI methods.
Optimization of the flow injection method was carried out using the set up shown in Fig. 1 in continuous mode. The concentration of Ce(IV) was varied over the range of 6.0 × 10−3 to 2.0 × 10−2 M when the concentration of As(III) was fixed at 0.1 M. Replicate injections of 100 μg I l−1 solution were made. The results have shown that the size of signal increased to approximately 0.2 absorbance value as the concentration of Ce(IV) was increased up to 1.5 × 10−2 M. Above this concentration, the signal stayed fairly constant. The concentration of Ce(IV) at 8.0 × 10−3 M was chosen, because at this concentration acceptable sensitivity was obtained. Under this condition the molar ratio between Ce(IV) and As(III) was 1∶12.5 (the pumping rates of these two reagent streams were equal).
Chantore et al.22 have found that raising the temperature of the reaction zone can increase the sensitivity. For the continuous-FI system, the effect of temperature on sensitivity was studied at two bath temperatures. Linear calibrations were obtained from both temperatures, e.g., ΔA = 0.0010[I] + 0.049, r2 = 0.998 (26 °C) and ΔA = 0.0016[I] + 0.133, r2 = 0.997 (43 °C). The intercepts represent signals of the reagent blank. The results have shown that the sensitivity was better achieved at 43 °C than at room temperature (26 °C), in terms of slope and intercept of calibration.23 Temperatures greater than 43 °C although giving higher sensitivity can decrease the solubility of air and sometimes cause air bubbles inside the glass coil. Thus for the continuous-FI, the bath temperature was fixed at 43 °C.
The set up condition of the stopped mode was similar to the condition illustrated in Fig. 1 except that the flow rate could be made faster to 3 ml min−1. The results in Table 1 show that prolonging of the stopped time increased the sensitivity. For this work, a stopped time of 1 min was sufficient to provide reasonable absorbance readings over the working range of 50 to 200 μg I l−1.
Stop interval/s | Slope (× 10−4) | Intercept | r2 | Sample throughput/ sample h−1 |
---|---|---|---|---|
0 | 3.0 | 0.019 | 0.999 | 88 |
20 | 6.0 | 0.039 | 0.999 | 59 |
40 | 9.0 | 0.067 | 0.999 | 44 |
60 | 12 | 0.095 | 0.999 | 35 |
120 | 17 | 0.188 | 0.996 | 22 |
180 | 21 | 0.265 | 0.996 | 16 |
Due to some differences such as configuration, the operating procedure, and under the conditions studied, the flow rates, the sensitivities given by the two modes, are incomparable. Nevertheless, it was observed that the sensitivity of the continuous system is greater than that obtained from the condition of the stopped mode. For example, the regression equations are ΔA = 0.0016[I] + 0.133, r2 = 0.989 (for the continuous-FI, 43 °C) and ΔA = 0.0011[I] + 0.091, r2 = 0.996 (for the stopped-FI, 26 °C).
In the stopped-FI mode, it is expected that the size of signal will be enlarged if the time spent by the reaction zone is prolonged. The results in Table 1 demonstrate that increasing the interval length resulted in larger size of signals including the signal of the reagent blank (intercept). Also the results have indicated that the slope became greater with time. Thus, prolonging of the stopped time increases the sensitivity of the stopped mode.
Besides the sensitivity, sample throughput is often a parameter used for consideration. This parameter is an indication of speed of analysis. The throughputs as determined by stopped interval are summarized in Table 1. Although the stopped time of 180 s delivered the highest sensitivity, the analysis was time consuming. For this work, the stopped interval of 1 min was chosen as it resulted in a reasonable sensitivity and acceptable throughput of sample (35 samples h−1). This sample throughput is approximately equal to that given by the continuous-FI system.
Foreign species | Added as | Concentration level tested/M | Signal alteration (%) |
---|---|---|---|
a Three-fold of the typical levels in urine. b These levels of foreign ions were previously tested on a continuous-FI system proposed by Yaping et al.21 | |||
Cl− | NaCl | 1.5a | +8.1 |
SO42− | Na2SO4 | 0.2a | +5.2 |
HCO3− | NaHCO3 | 0.06a | +0.9 |
(NH4)2HPO4 | (NH4)2HPO4 | 0.144a | −4.8 |
Ca2+ | CaCl2 | 0.084a | 0.0 |
Glucose | C6H12O6 | 2.34 ma | +0.9 |
Urea | CO(NH2)2 | 9.9 ma | +2.2 |
K+ | KCl | 9.9 mb | +0.9 |
Mg2+ | MgCl2 | 20.6 mb | +1.8 |
SCN− | KSCN | 0.3 mb | +4.0 |
Fe3+ | FeCl3 | 0.36 mb | +7.3 |
F− | NaF | 0.53 mb | −0.8 |
Mn2+ | MnSO4 | 36.4 μb | −0.9 |
Oxalic acid | C2H6O6 | 4.0 mb | 0.0 |
Uric acid | C5H4O3N4 | 11.9 mb | −4.0 |
Ascorbic acid | C6H8O6 | 0.57 mb | +0.9 |
Urine sample | Iodine content/ μg I l−1 | Added/ μg I l−1 | Found/ μg I l−1 | Recovery (%) |
---|---|---|---|---|
a Determination of samples A to F and G to N were respectively carried out using the continuous mode (n = 2) and the stopped mode (n = 3). | ||||
A | 31 ± 2 | 100.00 | 131 ± 0 | 100 |
B | 78 ± 2 | 100.00 | 178 ± 0 | 100 |
C | 83 ± 2 | 100.00 | 167 ± 1.5 | 84 |
D | 106 ± 5 | 100.00 | 206 ± 3 | 100 |
E | 147 ± 0 | 100.00 | 256 ± 3 | 109 |
F | 330 ± 2 | 100.00 | 440 ± 6 | 111 |
G | 15 ± 1 | 50.00 | 67 ± 0.6 | 110 |
H | 33 ± 1.5 | 50.00 | 79 ± 0.6 | 85 |
I | 51 ± 0.6 | 50.00 | 100 ± 1 | 98 |
J | 52 ± 0.6 | 50.00 | 105 ± 0.6 | 107 |
K | 53 ± 0.6 | 50.00 | 103 ± 0.6 | 99 |
L | 81 ± 2.1 | 50.00 | 126 ± 2.5 | 93 |
M | 86 ± 0.6 | 50.00 | 133 ± 2.7 | 96 |
N | 94 ± 3.5 | 50.00 | 146 ± 0.6 | 101 |
The calibration plot as suggested for use in the continuous-FI system was the plot between the size of signal (ΔA) versus iodine concentration. Under the conditions used, the calibration was linear which means that within 100 s the size of signal was directly proportional to the rate of reaction. Similar to the continuous-FI method, linear calibration was also obtained from the stopped-FI method. It is thus an advantage of the proposed methods.
Twenty samples of urine were analysed for iodine contents using the conventional method and the continuous flow injection method. The two procedural approaches gave good agreement in the results. The plot between the mean values of both techniques showed the absence of any analytical bias. The Pearson coefficient of correlation (n = 24) was found to be 0.952 and a reasonable linear regression line was obtained (y = 0.938x + 4.63).
Precision of the measurement between the batch and the flow method was compared. Ten replicates of injection (100 μg I l−1) were made on the continuous-FI set up. Both methods contributed satisfactorily low values of RSD (1.6% and 0.4% for the batch and the flow methods respectively).
Although the batch method did not give a prominently high value of RSD, the results could be worse in the hands of different operators. The RSD results have shown that the continuous flow mode has allowed for a better control of timing than the batch method.
Repeatability of the stopped FI system was also measured and the RSD was found to be 1.2%. This RSD value was obtained when the pump was turned on and off manually. The precision index could have been improved if the system had been fully automated.
Excluding the step of digestion, the two systems of FI have given the analysis time for a sample of approximately 1 min 40 s. The analysis time is therefore much less than the time required in the batch method (25 to 35 min per sample).
Iodine content/μg I l−1 ± s | ||||
---|---|---|---|---|
Sample | Continuous-FI | Stopped-FI | Batch | ICP-MS |
1 | 5.44 ± 22 | 11.2 ± 15 | 18.9 ± 8.1 | 12.1 ± 15 |
2 | 13.3 ± 6.7 | 14.5 ± 1.5 | 17.4 ± 3.6 | 16.9 ± 4.2 |
3 | 15.5 ± 11 | 13.3 ± 13 | 14.6 ± 19 | 15.5 ± 1.7 |
4 | 31.2 ± 1.1 | 27.1 ± 4.9 | 28.8 ± 8.4 | 38.4 ± 17 |
5 | 34.0 ± 6.1 | 37.1 ± 4.0 | 32.5 ± 3.9 | 42.3 ± 2.2 |
6 | 37.7 ± 2.7 | 35.6 ± 2.9 | 39.5 ± 5.3 | 74.5 ± 0.61 |
7 | 43.4 ± 0.7 | 49.9 ± 3.5 | 49.8 ± 4.7 | 55.3 ± 2.5 |
8 | 57.1 ± 4.4 | 52.5 ± 0.8 | 59.6 ± 3.7 | 73.4 ± 0.19 |
9 | 78.8 ± 1.3 | 74.1 ± 2.6 | 84.3 ± 7.3 | 87.2 ± 11 |
10 | 79.0 ± 2.6 | 83.9 ± 2.3 | 76.6 ± 3.4 | 97.8 ± 1.7 |
11 | 82.5 ± 1.3 | 84.9 ± 1.4 | 81.8 ± 7.5 | 105 ± 1.2 |
Applying linear regression,24 comparison was made for each method pair via the plot between the results achieved from the two methods. Parameters of the regression line such as intercept (a), slope (b) and correlation coefficient (r2) were then obtained.
The correlation coefficients (r2) shown in Table 5 indicate reasonably good correlation (r2 >0.850). The confidence limits suggest that the slope and the intercept do not differ significantly from the ideal values. Thus, there is no evidence for systematic differences between the results obtained from the four techniques, i.e., continuous-FI, stopped-FI, batch and ICP-MS. This agreement of the results shows that the two modes of flow injection can be an alternative to the conventional batch method.
Method pair (x,y) | b ± tsb | a ± tsa | r2 |
---|---|---|---|
Batch, continuous-FI | 1.04 ± 0.27 | −4.45 ± 11.96 | 0.968 |
Batch, stopped-FI | 1.03 ± 0.17 | −4.15 ± 8.97 | 0.959 |
Stopped-FI, continuous-FI | 0.981 ± 0.11 | 0.860 ± 5.74 | 0.982 |
ICP-MS, batch | 0.732 ± 0.23 | 4.87 ± 15.18 | 0.894 |
ICP-MS, continuous-FI | 0.784 ± 0.27 | 0.797 ± 17.6 | 0.923 |
ICP-MS, stopped-FI | 0.789 ± 0.27 | 1.09 ± 17.7 | 0.915 |
![]() | ||
Fig. 3 Kinetics obtained from the stopped-FI system, showing the disappearance of Ce(IV) at different concentrations of iodine standard in sample zones. Full lines represent the exponential fittings. |
It was found that the calibration plot is linear for both the FI systems. The signal measured at a fixed time is therefore directly proportional to the rate of reaction.
The sensitivity and sample throughput given by the two systems was not much different. The systems give a satisfactorily low value of detection limit (≤3 μg I l−1), although results from the determination of samples having contents below 20 μg I l−1(severe IDD) can be erratic. However the systems are still suitable for screening these types of samples from the samples above the margin of severe IDD status. If necessary the samples containing this very low level of iodine can be re-determined in the stopped mode by increasing the stopped time.
For this work, the condition for urine digestion was developed by modifying the former condition1,10 to reduce the amount of chloric acid. The proposed condition improves the effect of interferents and provides complete digestion.
Besides, the stopped-FI system can be used to study kinetic profiles of the reaction. Some results have shown that the kinetics is first-order in Ce(IV) concentration under the conditions used.
This journal is © The Royal Society of Chemistry 2001 |