DOI:
10.1039/B006484J
(Paper)
Analyst, 2001,
126, 33-36
Fluorimetric determination of isatin in human urine and serum by
liquid chromatography postcolumn photoirradiation
Received 8th August 2000, Accepted 18th October 2000
First published on 23rd November 2001
Abstract
For the fluorimetric determination of isatin in human urine and
serum, HPLC-postcolumn photoirradiation using a mobile phase has been developed.
Isatin in the urine or serum sample was separated on a Capcell Pak C1 column
(250 × 4.6 mm id). The mobile phase consisted of 70 mmol l−1
phosphate buffer (pH 7.2)–tetrahydrofuran (85 + 15% v/v) containing
5 mmol l−1 hydrogen peroxide, which was irradiated with germicidal
light to induce fluorescence (λex 302 nm, λem
418 nm). The addition of tetrahydrofuran to the mobile phase led to the peaks
showing good separation as well as increased sensitivity. The calibration
graph for isatin was linear over the range of 0.16–10.7 ng. The pretreatment
of the acidified urine or serum samples consisted of diluting steps or deproteinizing
steps using perchloric acid, respectively. The mean recovery of isatin from
urine and serum was greater than 94%.
Introduction
Isatin (indole-2,3-dione, IST) has been used in analytical chemistry as
a reagent for proline or tryptophan.1,2
In 1988, IST was discovered in human and rat urine by Glover et al.3 The increased IST concentration indicated the condition
of neurological patients including stress, anxiety and epilepsy.4–6
Although the metabolic pathway of IST has not been fully clarified, speculations
of IST biosynthesis are reported to be linked to the peroxidation of tryptophan7 and the effectiveness of the intestinal flora.8,9 A number of procedures have already been described
for the measurement of IST in biological samples such as gas chromatography-mass
fragmentometry (GC-MS)3,10 and HPLC with
a spectrometric detector.11,12 GC-MS
is useful in determining the existence of interesting substances; however,
the need to maintain purity makes it unsuitable for routine work. Although
HPLC is useful in analyzing a biological sample, its spectrophotometric detection
is liable to lack sensitivity and/or selectivity. It requires a tedious pretreatment.
For example, in the assay of Manabe et al.,11
the urine sample was treated by a liquid–liquid and solid-phase extraction,
followed by preparative HPLC. As reported here, HPLC based on postcolumn photoirradiation
was established with a Capcell Pak C1 column and with a mobile phase containing
tetrahydrofuran and hydrogen peroxide. This method allowed easy pretreatment
by dilution or deproteinizing and efficient measurements of IST in human urine
and serum.Experimental
Chemicals and pooled serum
IST was obtained from Sigma (St. Louis, MO, USA) and was further purified
by recrystallization from benzene. The standard IST sample was diluted using
0.1 mol l−1 hydrochloric acid and stored at −20 °C.
All other chemicals were purchased from Wako (Osaka, Japan). Tetrahydrofuran
(THF) was used as a solution free from stabilizer and was stored at 4 °C
until the mobile phase was prepared. A freeze dried pooled serum (Consera®)
was obtained from Nissui Seiyaku (Tokyo, Japan).Chromatographic system
The chromatographic system consisted of a high pressure pump (Model SSP
DM-3U, Sanuki Kogyo, Tokyo, Japan), a valve injector (Model E1E005, Shimamura,
Tokyo, Japan) fitted with a 200 µl loop, an analytical column, a photoreactor
(Model S-3900, Soma Optics, Tokyo, Japan), a fluorescence monitor (Model RF-530,
Shimadzu, Kyoto, Japan) with a 12 µl flow cell, and a Shimadzu Chromatopac
C-R3A (Shimadzu, Kyoto, Japan) recorder–integrator. In this system the
analytical column (250 × 4.6 mm id) was packed with Capcell Pak C1 (particle
size 5 µm, Type SG-120, Shiseido, Tokyo, Japan). The S-3900 photoreactor
consisted of a stainless steel photoreflector, an electric fan and two tubes
of GL-4 germicidal light (4 W electric power; Nippon Denki, Tokyo, Japan).
The photoirradiation was carried out in ETFE (a copolymer of ethylene and
tetrafluoroethylene: Tefzel®, GL Sciences, Tokyo, Japan) tubing
(0.25 mm id × 1.59 mm od, 4 m long) which was wound around each germicidal
light tube. Following the photoirradiation coil, the back-pressure tubing
was composed of PEEK (polyether ether ketone, GL Sciences, Tokyo, Japan) tubing
(0.13 mm id × 0.5 m long) and PTFE tubing (0.25 mm id × 2 m long)
was connected. The fluorescence of the effluent solution was measured with
excitation and emission wavelengths of 302 nm and 418 nm, respectively. Chromatography
was performed at ambient temperature.Separation conditions
The mobile phase, which consisted of 70 mmol l−1 phosphate
buffer (pH 7.2) containing 5 mmol l−1 hydrogen peroxide and
15% (v/v) tetrahydrofuran, was delivered at a flow rate of 0.56 ml min−1.Analytical procedure for human urine and serum
IST was diluted with hydrochloric acid to maintain its stability and to
solubilize any sediment in the biological sample. The human urine of healthy
subjects (ordinary diet) was collected over a 24 h period in the vessel containing
10 ml of hydrochloric acid per liter of urine. The urine samples were stored
at −20 °C until analysis.One milliliter of 5 mol l−1 hydrochloric acid was added
to the urine (1 ml) and then heated for 10 min in boiling water. After the
urine sample was cooled in an ice–water bath, the sample was diluted
50-fold with distilled water. An aliquot (20 µl) of the sample was injected
into the previously described system. For the serum samples, 200 µl
of 5 mol l−1 hydrochloric acid was added to the serum (200 µl)
and then heated for 10 min in boiling water followed by cooling in an ice–water
bath. The mixture was centrifuged at 9600g for 1 min.
One hundred microliters of 1.5 mol l−1 perchloric acid
were added to 200 µl of the heated serum in a polypropylene tube (1.5
ml). The mixture was then vortex-mixed. The entire mixture was added to 100 µl
of 1.5 mol l−1 potassium chloride. After being vortex-mixed
and centrifuged for 1 min, a 50 µl portion of the supernatant fluid
was injected into the chromatograph. The recovery test was carried out to
observe the influence of the biological component on the fluorophore. The
spiked urine or serum was prepared as follows. One milliliter of IST standard
solution (854 ng ml−1) was added to 0.5 ml of the heated
urine in the glass measuring flask which was filled to 25 ml with distilled
water. Then, 1.5 mol l−1 perchloric acid solution was added
to the serum containing the IST of 49.7 ng ml−1 concentration
and deproteinized by the method described above. The peak heights obtained
for the spiked sample were compared with those of the standard sample.
Results and discussion
Fluorescence reaction upon irradiation
Fig. 1 shows the excitation (a) and emission
spectra (b) for the IST obtained using the LC system. The excitation (λex)
and emission (λem) maxima were at 302, and 392
and 418 nm, respectively. IST is converted to anthranilic acid (2-aminobenzoic
acid) in the presence of sodium hydroxide and hydrogen peroxide.13
In order to produce a fluorophore, the products were developed with the system
described except for photoirradiation and two peaks were confirmed. The retention
time of main and minor products were 7.6 min (λex
312 nm, λem 388 nm) and 9.0 min (λex
302 nm, λem 418 nm), respectively. The characteristics
of the main product agreed with those of anthranilic acid. Although the minor
product is still unknown, the spectrum of IST is obtained with two fluorophores. |
| Fig. 1 Fluorescence
excitation (a) and emission (b) spectra of the HPLC eluate of IST. a1
and b1, 12.8 µg ml−1; a2 and
b2, sample blank for a1 and b1, respectively. | |
Optimization of postcolumn photoirradiation
The phosphate buffer for the maximum fluorescence intensity in the mobile
phase was observed at pH 6.8–8.0. pH 7.2 was adopted based on the peak
shape described in Optimization of chromatographic conditions (below) and
the stability of the mobile phase. Table 1
shows the effects of organic solvents on the fluorescence intensity. The usual
organic modifiers, such as methanol and acetonitrile, did not provide adequate
sensitivity. If THF was added after the photoirradiation, it was not effective.
The best range of concentrations in the mobile phase for the photoirradiation
reaction was 15–30% v/v. For the separation of IST in biological samples,
the concentration of 15% v/v THF was adopted. The optimum concentration of
hydrogen peroxide were found to be 3–8 mmol l−1 for
the 4.0 m long ETFE tubing while 5 mmol l−1 was adopted for
additional experiments. The addition of hydrogen peroxide enhanced the fluorescence
intensity by about 20%.
Table 1 Effect of organic modifier on the fluorescence
intensity
Organic modifier (20% v/v) | Relative fluorescence intensity (%) |
---|
None | 2.6 |
Methanol | 6.5 |
Ethanol | 8.8 |
Propan-1-ol | 6.0 |
Propan-2-ol | 5.9 |
Acetonitrile | 12.9 |
Tetrahydrofuran | 100 |
Optimization of chromatographic conditions
When the mobile phase which contained the 15% (v/v) tetrahydrofuran in
70 mmol l−1 phosphate buffer (pH 7.2) was used, the IST retention
times with Capcell Pak C1, C8, C18 and CN (size 150 × 4.6 mm id) were
10, 8.6, 8.6 and 8.8 min, respectively. The Capcell Pak C1 column provided
the best separation of IST, anthranilic acid (degradation product of IST)
and coexisting components in the urine sample. The size of the adapted C1
column was 250 × 4.6 mm id to ensure separation with injection of a
large volume of samples and with the serum sample. The IST retention time
was not significantly influenced by the pH of the phosphate buffer in the
eluent. The pH affected the peak shape of IST. Fig.
2 shows the effect of the pH of the phosphate buffer on the peak
shapes. The range of pH 4.5–6.5 gave a broad and fronting peak shape,
which almost became symmetrical at pH 7.2. The hydrolysis of IST has been
reported by Casey et al.,14i.e.,
above pH 6 IST undergoes cleavage of the amide bond and forms an open ring
structure, while at pH 4 there is an approximately equal ratio of the closed-
and opened-ring forms. |
| Fig. 2 Effect
of pH on the peak shapes of isatin. (a) pH 5.9, (b) pH 6.5 and (c) pH 7.2
in 70 mmol l−1 phosphate buffer. | |
The fronting peak of IST appears during the acidified mobile phase.12 As the IST of the open-ring form showed a symmetrical
peak, the phosphate buffer in the eluent was kept at pH 7.2. In addition,
several compounds such as tryptophan, melatonin, kynurenine, kynurenic acid,
anthranilic acid (retention time 7.6 min), nicotinic acid, tyrosine, phenylalanine,
histidine, pyridoxine and thiamine did not interfere with the analysis although
they were eluted faster than the IST peak.
Calibration and precision
The measured relative fluorescence intensity was linear with the amount
of IST over the range of 0.16–10.7 ng (1.1–72.7 pmol). The correlation
coefficient was 0.998, the slope was 26.8 and the intercept was −0.8
using a least squares regression analysis. The parameters y and x
are the relative fluorescence intensity in % and the amount of IST in ng,
respectively. The relative standard deviation of the standard IST was 1.9%
at 0.62 ng (n = 8) and 1.2% at 4.0 ng (n = 8) . The detection
limit (signal to noise = 3) was determined to be 0.11 ng (0.75 pmol).Selectivity and recovery
Fig. 3 shows chromatograms of authentic
IST and urine samples. The retention time of IST was 16 min. The IST peak
was observed in chromatograms (a), (b) and (c) at the corresponding retention
times, whereas a chromatogram (d) of the unirradiated sample showed no peak
at that position. Furthermore, the spectrum of the peak of interest was similar
to that of IST (Fig. 4). These comparisons
of the chromatograms demonstrate the specificity of the present method. The
recovery (±s) of IST added to urine (854 ng per 25 ml of urine)
was 96.3 ± 4.7% (n = 7). In the urine of six normal subjects,
the range and mean ±s of urinary IST was 184–1513 µg
l−1 and 735.4 ± 491.5 µg l−1,
respectively. This value in human urine was similar to the 708.4 ±
766.9 µg l−1 reported by Manabe et al.11Fig. 5 compares
the chromatograms of irradiated and non-irradiated samples in deproteinized
serum. |
| Fig. 3 Chromatograms
obtained with a standard and the urine samples. Irradiation with UV light:
(a) standard of IST (0.68 ng); (b) urine sample; (c) urine sample spiked with
IST;. (d) the same sample as in (b), but without irradiation. Analytical procedure
and conditions as described in the Experimental section. Injection volume
was 20 µl. Applied concentration of IST was 34.1 ng ml−1
of diluted urine. | |
 |
| Fig. 4 Fluorescence
excitation and emission spectra of the HPLC eluate of (a) standard peak (20.8
ng ml−1) and (b) component of IST peak in urine sample. Spectra
determined by Shimadzu Spectrofluorometer RF-5000. | |
 |
| Fig. 5 Chromatograms
obtained with a standard and the serum samples. Irradiation with UV light:
(a) standard of IST (0.62 ng); (b) serum sample; (c) serum sample spiked with
IST; (d) the same sample as in (b), but without irradiation. Analytical procedure
and conditions were as described in the Experimental section. Injection volume
was 50 µl. Applied amount of IST was 49.7 ng in deproteinized serum. | |
The assay of the serum was sufficient using 200 µl and is possible
at 1/25 the volume compared with UV detection. The spectra of the IST peak
was similar to that shown in Fig. 4. The recovery
(±s) of IST added to serum (49.7 ng ml−1
in perchloric acid) was 94.3 ± 4.4% (n = 6). The mean ±s
of serum IST was 54.7 ± 4.2 ng ml−1 (n =
6). For reference, the value in human plasma was 134.2 ± 120.6 ng ml−1
as reported by Manabe et al.11 The
variation in the IST level is presumed to correspond to the difference in
dietary tryptophan content and might be influenced by intestinal flora.
This method is both sensitive and sufficiently specific to estimate IST
in human urine and serum and should be useful in biochemical and clinical
studies.
Acknowledgment
The authors acknowledge the lending
of the photoreactor (S-3900) from Soma Optics.References
- F.
N. Boctor, Anal. Biochem., 1971, 43, 66 CrossRef CAS.
- C. I. Trigoso, N. Ibanez and J. C. Stockert, J. Histochem.
Cytochem., 1993, 41, 1557 Search PubMed.
- V. Glover, J. M. Halket, P. T. Watkins, A. Clow, B.L. Goodwin and M. Sandler, J. Neurochem., 1988, 51, 656 CAS.
- V. Glover, S. K. Bhattacharya and M. Sandler, Indian
J. Exp. Biol., 1991, 29, 1 Search PubMed.
- D. Hota and S. B. Acharya, Indian J. Exp. Biol., 1994, 32, 710 Search PubMed.
- N. Hamaue, Yakugaku Zasshi, 2000, 120, 352 Search PubMed.
- S. Ghosal, S. K. Bhattacharya, A.V. Muruganandam and K.S. Satyan, Biogenic Amines, 1997, 13, 91 Search PubMed.
- A. H. Jackson, R. T. Jenkins, M. Grinstein, A.-M. Ferramola de Sancovich and H. A. Sancovich, Clin. Chim.
Acta, 1988, 172, 245 CrossRef CAS.
- M. Sandler, A. Przyborowska, J. Halket, P. Watkins, V. Glover and M. E. Coates, J. Neurochem., 1991, 57, 1074 CAS.
- J. M. Halket, P. J. Watkins, A. Przyborowska, B. L. Goodwin, A. Clow, V. Glover and M. Sandler, J.
Chromatogr., 1991, 562, 279 CrossRef CAS.
- S. Manabe, Q. Gao, J. Yuan, T. Takahashi and A. Ueki, J. Chromatogr.
B, 1997, 691, 197 CrossRef CAS.
- N. Hamaue, N. Yamazaki, M. Minami, T. Endo, M. Hirahuji, Y. Monma and H. Togashi, Gen. Pharmac., 1998, 30, 387 Search PubMed.
- W. J. Houlihan, in The Chemistry of Heterocyclic Compounds, Indoles,
eds. A. Weissberger and E. C. Taylor, Wiley, New
York, 1972, pt. 1, ch. 1, pp. 157–158. Search PubMed.
- L. A. Casey, R. Galt and M. I. Page, J.
Chem. Soc., Perkin. Trans., 1993, 2, 23 Search PubMed.
|
This journal is © The Royal Society of Chemistry 2001 |
Click here to see how this site uses Cookies. View our privacy policy here.