Determination of aluminum in beverages by automated non-segmented continuous flow analysis with fluorescent detection of the lumogallion complex

(Note: The full text of this document is currently only available in the PDF Version )

Richard L. Hoch


Abstract

A method for the automated determination of aluminum in a variety of beverages is described. The method utilizes lumogallion as a complexing agent in a buffer solution. The system is very similar to flow-injection analysis (FIA), however, the tubing id is larger than that typically used in FIA. Therefore, the system is best described as non-segmented continuous flow analysis using fluorescence spectroscopy detection. The method is extremely simple, requiring virtually no sample preparation and only one reagent. The instrument detection limit for aluminum is 0.012 µg ml–1 and calibration is linear to 3 µg ml–1. Results from a variety of beverage matrices are discussed and compared with the frequently used 8-hydroxyquinolone method utilizing a chloroform extraction and fluorescence spectroscopy detection.


References

  1. E. Goon, J. E. Petley, W. H. McMullen and S. E. Wiberley, Anal. Chem., 1953, 25, 608 CrossRef CAS.
  2. F. Carrillo, C. Pérez and C. Cámara, Anal. Chim. Acta, 1992, 262, 91 CrossRef CAS.
  3. H. He, H. Lee, S. Li, A. Hsieh and K. Siow, J. Chromatogr. Sci., 1997, 35, 333 CAS.
  4. S. H. Sutheimer and S. E. Cabaniss, Anal. Chem., 1995, 67, 2342 CrossRef CAS.
  5. J. Quintela, M. Gallego and M. Valcárcel, Analyst, 1993, 118, 1199 RSC.
  6. F. S. Rojas, E. C. Alcaraz and J. M. Cano Pavón, Analyst, 1994, 119, 1221 RSC.
  7. C. Jiang, B. Tang, C. Wang and X. Zhang, Analyst, 1996, 121, 317 RSC.
  8. M. J. Ahmed and J. Hossan, Talanta, 1995, 42, 1135 CrossRef CAS.
  9. S. H. Sutheimer and S. E. Cabaniss, Anal. Chim. Acta, 1995, 303, 211 CrossRef CAS.
Click here to see how this site uses Cookies. View our privacy policy here.