A rapid quantitative assay of intact paracetamol tablets by reflectance near-infrared spectroscopy

(Note: The full text of this document is currently only available in the PDF Version )

Andrew D. Trafford, Roger D. Jee, Anthony C. Moffat and Paul Graham


Abstract

Near-infrared (NIR) reflectance spectroscopy was used to determine rapidly and non-destructively the content of paracetamol in bulk batches of intact Sterwin 500 mg tablets by collecting NIR spectra in the range 1100–2500 nm and using a multiple linear regression calibration method. The developed NIR method gave results comparable to the British Pharmacopoeia 1993 UV assay procedure, the standard errors of calibration and prediction being 0.48% and 0.71% m/m, respectively. The method showed good repeatability, the standard deviation and coefficient of variation for six NIR assays on the same batch on the same day being 0.14 and 0.16% m/m, respectively, while measurements over six consecutive days gave 0.31 and 0.36% m/m, respectively. Applying the calibration to a parallel test set gave a mean bias of –0.22% and a mean accuracy of 0.45%. The developed method illustrates how the full potential of NIR can be utilised and how the ICH guidelines which recommend the validation of linearity, range, accuracy and precision for pharmaceutical registration purposes can be applied. Duplicate determinations on bulk batches could be performed in under 2 min, allowing the potential use of the method on-line for real time monitoring of a running production process.


References

  1. A. F. Zappala and A. Post, J. Pharm. Sci., 1977, 66, 292 CAS .
  2. P. Corti, E. Dreassi, G. Corbini, S. Lonardi, R. Viviani, L. Mosconi and M. Bernuzzi, Pharm. Acta Helv., 1990, 65, 28 CAS .
  3. E. Dreassi, G. Ceramelli, P. Corti, P. L. Perruccio and S. Lonardi, Analyst, 1996, 121, 219 RSC .
  4. E. Dreassi, G. Ceramelli, L. Savini and P. Corti, Analyst, 1995, 120, 319 RSC .
  5. R. A. Lodder and G. M. Hieftje, Appl. Spectrosc., 1988, 42, 518 CAS .
  6. R. Jensen, E. Peuchant, I. Castagne, A. M. Boirac and G. Roux, Spectrosc. Int. J., 1988, 6, 63 CAS .
  7. J. Gottfries, H. Depui, M. Fransson, M. Jongeneelen, M. Josefson, F. W. Langkilde and D. T. Witte, J. Pharm. Biomed. Anal., 1996, 14, 1495 CrossRef CAS .
  8. S. M. Han and P. G. Faulkner, J. Pharm. Biomed. Anal., 1996, 14, 1681 CrossRef CAS .
  9. P. Merkle and K. A. Kovar, J. Pharm. Biomed. Anal., 1997, 17, 365 CrossRef CAS .
  10. A. Eustaquio, P. Graham, R. D. Jee, A. C. Moffat and A. D. Trafford, Analyst, 1998, 123, 2303 RSC .
  11. M. Blanco, A. Eustaquio, R. D. Jee, and A. C. Moffat, Anal. Chim. Acta, in the press. Search PubMed .
  12. International Conference on Harmonisation (ICH) of Technical Requirements for Registration of Pharmaceuticals for Human Use, Validation of Analytical Procedures: Methodology, Recommended for Adoption at step 4 of the ICH Process on 6 November 1996 by the ICH Steering Committee. Internet address: http://www.mcclurenet.com Search PubMed .
  13. British Pharmacopoeia 1993, H.M. Stationery Office, London, 1993, vol. II, p. 1043 Search PubMed .
  14. R. J. Barnes, M. S. Dhanoa and S. J. Lister, Appl. Spectrosc., 1989, 43, 772 CAS .
  15. S. A. Halsey, Technical Note, NIRSystems, Silver Spring, MD, 1994 Search PubMed .
  16. W. Caulcutt and R. Boddy, Statistics for Analytical Chemists, Chapman and Hall, London, 1983, p. 70 Search PubMed .