Electrochemical detection of aluminium using single-use sensors

(Note: The full text of this document is currently only available in the PDF Version )

John W. Dilleen, Brian J. Birch and Barry G. D. Haggett


Abstract

An electrochemical method for the direct detection of aluminium using single-use sensors is described. The sensor consists of three electrodes (working, auxilliary and pseudo-reference) screen-printed with carbon ink. Aluminium is accumulated at the carbon electrode at –1.6 V (vs. pseudo-reference) in zinc sulfate electrolyte for 90 s prior to anodic stripping using differential-pulse voltammetry. Linear calibrations were obtained between 10 and 450 ppb using the aluminium oxidation peak. The limit of detection was 8 ppb Al3+.


References

  1. Aluminium & Health current fact sheets, European Aluminium Association, Brussels, Belgium, December, 1997 Search PubMed.
  2. Q. Cai and S. B. Khoo, Anal. Chim. Acta, 1993, 276, 99 CrossRef CAS.
  3. C. M. G. van den Berg, K. Murphy and J. P. Riley, Anal. Chim. Acta, 1986, 188, 177 CrossRef CAS.
  4. P. Akhtar, H. A. Devereaux, A. J. Downard, B. O'Sullivan and K. J. Powell, Anal. Chim. Acta, 1999, 381, 49 CrossRef CAS.
  5. J. W. Dilleen, S. D. Sprules, B. J. Birch and B. G. D. Haggett, Analyst, 1998, 123, 2905 RSC.