WIM SCHELLES and RENÉ VAN GRIEKEN
A dc GD mass spectrometer has been used for the analysis of ZrO2 samples with known amounts of artificial impurities. For this purpose the secondary cathode technique has been used. The research work focused both on the methodological aspects of the atomization of the ZrO2 and on some quantitative aspects, with direct analytical interest. The reproducibility was found to be better than 10% RSD in most cases; the accuracy of the ‘standardless’, raw results was within a factor of 2–3 of the known concentration.