ZDENĚK WEISS
The emission intensities of five resonance atomic lines affected by self-absorption were investigated as a function of analyte concentration in a sample for a Grimm-type atomization/excitation source operated in the dc mode in argon. Based on considerations of radiative transfer within the source, and using the two-layer model and other approximations, the equationwas derived, linking the emission intensityIEM, the sputter rateqMand the analyte concentrationcEMin the matrix, whereE is a particular element in a matrix M. The RE0andβE constants resulting from fitting the experimental data to this relationship were investigated as functions of GD operating conditions. It was shown thatRE0can be regarded as the (generalized) emission yield. Dependence of theRE0andβE parameters on discharge operating conditions suggests that they could reflect more fundamental processes occurring in the discharge. Comparison ofβEfactors for different lines, however, leads to results that are not explicable within the adopted model. Methodology of quantitative GD-OES analysis was generalized to be applicable also to self-absorption lines with non-linear intensity responses.