Some figures of merit of a new double focusing inductively coupled plasma mass spectrometer

Abstract

An ICP-MS instrument with a quadrupole filter is hampered by its limited resolution, in particular polyatomic ions are a major source of interference. With a double focusing magnetic sector mass analyser, operated at high resolution, most of these interferences are eliminated. In addition, these mass analysers are characterized by a low instrumental background, yielding superior detection limits at low resolution. In this work a new type of double focusing ICP-MS instrument, the ‘Element’ from Finnigan MAT (Bremen, Germany), was tested. The instrument can be used with low resolution (mm= 300, 10% valley definition) and at high resolution settings (3000 and 7500). At low resolution, a linear response versus mass curve was observed [response of 180 × 106 counts s–1 per µg ml–1(ppm) for U and 6 × 106 counts s–1 for Be]. The sensitivity decreases by roughly a factor of 14 when going from resolution 300 to resolution 3000 and by a factor of 10 from 3000 to 7500. An instrumental detection limit at low resolution of about 8 fg ml–1(8 ppq) was measured at m/z 240. The detection limit for most elements is not determined by the instrument but by the blank level and laboratory procedures will need to be improved in order to make optimal use of the detection power of the instrument. The stability and the relative abundance of oxide and doubly charged ions are comparable to the values found in quadrupole ICP-MS. Matrix effects were compared for measurements with the ‘Element’ and with a Perkin-Elmer Elan 5000 quadrupole instrument of 115In in the presence of 500 µg ml–1 of Cs and of 2.5% of ethanol. Though in the ‘Element’ ions are accelerated over 8 kV immediately after the skimmer, matrix effects were found to be similar to those observed for the ‘Elan 5000’. The instrument was used to determine Pt in an environmental material (grass).

Article information

Article type
Paper

J. Anal. At. Spectrom., 1995,10, 569-574

Some figures of merit of a new double focusing inductively coupled plasma mass spectrometer

L. Moens, F. Vanhaecke, J. Riondato and R. Dams, J. Anal. At. Spectrom., 1995, 10, 569 DOI: 10.1039/JA9951000569

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements