Abstract

A method is proposed for conducting a coulometric titration in a continuous flow of solution containing the analyte and the reagents for titrant generation where the titrant is generated in rectangular impulses of definite duration and different current magnitude. Coulometric titration was applied to redox titration reaction with FeII obtained electrochemically from FeIII solution and solutions of K2Cr2O7, KMnO4, Ce(SO4)2 and NH4VO3 were measured in the concentration range 1 × 10–4–1 × 10–2 equiv. 1–1, with relative standard deviations from 6.4 to 1.6%.

Article information

Article type
Paper

Analyst, 1994,119, 1835-1838

Pulse coulometric titration in continuous flow

A. D. Dakashev and V. T. Dimitrova, Analyst, 1994, 119, 1835 DOI: 10.1039/AN9941901835

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.