Liquid chromatographic behaviour of chelates of vanadium(V), copper(II), cobalt(III) and chromium(III) with 2-(3,5-dibromo-2-pyridylazo)diethylaminophenol
Abstract
The reversed-phase liquid chromatographic behaviour of the chelates of VV, CuII, CoIII and CrIII with 2-(3,5-dibromo-2-pyridylazo)diethylaminophenol(3,5-diBr-PADAP) was studied using a C18 column, and a method for the simultaneous determination of VV, CuII, CoIII and CrIII was developed. A mobile phase of methanol-acetonitrile-water (84 + 12 + 4) containing 0.11 mmol dm–3 cetyltrimethylammonium bromide and 0.1 mmol dm–3 NaOAc–HCOOH(pH 3.5) was found to be the best. The chelates were detected at 590 nm. The calibration graphs were linear for 0.20–6.40 µg ml–1 of VV, 0.40–12.80 µg ml–1 of CuII, 0.08–2.56 µg ml–1 of CoIII and 0.40–6.40 µg ml–1 of CrIII. The detection limits were 0.2, 0.4, 0.04 and 0.2 ng for VV, CuII, CoIII and CrIII, respectively. The method was applied to the determination of VV, CuII, CoIII and CrIII in electroplating waste water samples.