Issue 7, 1988

H-point standard additions method. Part 1. Fundamentals and application to analytical spectroscopy

Abstract

A modification of the standard additions method called the “H-point standard additions method” is proposed in order to obtain an unbiased analyte concentration when both analyte and interferents are present in a sample. It also permits the determination of an interferent known to be present. The method uses the analytical signal data at two selected wavelengths, giving two straight lines that have a common point with coordinates H (–CH, AH), where –CH is the unknown analyte concentration and AH the analytical signal due to an interferent. Examples of the application of the proposed method are given.

Article information

Article type
Paper

Analyst, 1988,113, 1011-1016

H-point standard additions method. Part 1. Fundamentals and application to analytical spectroscopy

F. B. Reig and P. C. Falcó, Analyst, 1988, 113, 1011 DOI: 10.1039/AN9881301011

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.