Issue 4, 1988

Comparative study of the electrochemical behaviour of sepiolite and hectorite modified carbon paste electrodes in the determination of dinocap

Abstract

A comparative study of the electrochemical behaviour of sepiolite and hectorite modified carbon paste electrodes in the determination of dinocap has been carried out. Dinocap was adsorbed on the electrodes in an open circuit (pH 4.0) and was determined by differential-pulse voltammetry at pH 5.3 in 0.5 M KNO3. The calibration graph obtained is linear for 2–14 ng ml–1 of dinocap using the sepiolite modified electrode and for 18–72 ng ml–1 of dinocap using the hectorite modified electrode.

Article information

Article type
Paper

Analyst, 1988,113, 621-623

Comparative study of the electrochemical behaviour of sepiolite and hectorite modified carbon paste electrodes in the determination of dinocap

L. Hernandez, P. Hernandez, E. Lorenzo and Z. S. Ferrera, Analyst, 1988, 113, 621 DOI: 10.1039/AN9881300621

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.