Inhibitor-treated microbial sensor for the selective determination of glutamic acid

Abstract

An amperometric sensor containing Bacillus subtilis with a high selectivity for glutamic acid has been developed. The bacteria were treated with chloromercuribenzoate and NaF, which inhibit the uptake and assimilation of glucose. The resulting electrode is characterised by a linear response to glutamic acid over the concentration range 0–550 mg l–1. In the kinetic mode the response time is 10–20 s. The sensor is stable for a period up to 14 d.

Article information

Article type
Paper

Analyst, 1987,112, 341-342

Inhibitor-treated microbial sensor for the selective determination of glutamic acid

K. Riedel and F. Scheller, Analyst, 1987, 112, 341 DOI: 10.1039/AN9871200341

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.