Urea pH electrodes: characterisation and optimisation for plasma measurements

Abstract

A potentiometric urea enzyme electrode is reported where consumption of H+ by the urease-catalysed degradation of urea is followed at a glass pH electrode. The response of the electrode in both weak buffers and diluted plasma is described, and the special conditions needed for plasma assay are outlined. Response times were 1–3.5 min. The use of 0.002 M phosphate assay buffer gave linear calibration graphs up to 5 × 10–4M urea. The response to plasma diluted 1 + 49 or 1 + 999 initially proved unreliable, but corresponded to that in aqueous standards when allowance was made for background pH variation using an auxiliary pH electrode.

Article information

Article type
Paper

Analyst, 1986,111, 875-878

Urea pH electrodes: characterisation and optimisation for plasma measurements

P. Vadgama, Analyst, 1986, 111, 875 DOI: 10.1039/AN9861100875

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.