Effect of experimental design on assay calibration

Abstract

Criteria are developed for the maximisation of the amount of information available from an analytical calibration graph. For linear calibration graphs, the methods involve replicate determinations of standards at two points, which are chosen so that they occur at the extremes of the calibration range. The increase in precision that can be obtained by using an optimised design can be considerable, particularly when the variance in the assay errors is not constant. The methods described can be extended readily to non-linear calibration graphs.

Article information

Article type
Paper

Analyst, 1981,106, 1249-1254

Effect of experimental design on assay calibration

L. Aarons, Analyst, 1981, 106, 1249 DOI: 10.1039/AN9810601249

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.