Abstract

The requirements of high-precision amperostatic coulometry in the absolute mode are summarised, and the apparatus and instruments used described. Three constant-current sources are appraised: a 10-mA operational amplifier circuit, a commercial coulometric titrimeter and a commercial analogue computer high-current source. The first and last have shown noise and drift levels that are acceptable in high-precision work.

Article information

Article type
Paper

Analyst, 1973,98, 305-312

Precise coulometric determination of acids in cells without liquid junction. Part I. Introduction and instrumentation

E. Bishop and M. Riley, Analyst, 1973, 98, 305 DOI: 10.1039/AN9739800305

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.