LA-ICP-MS analysis of fluid inclusions: contamination effects challenging micro-analysis of elements close to their detection limit

Abstract

This paper presents a practical guide to an optimized analytical procedure for the reliable quantification of trace element concentrations in fluid inclusions hosted by natural minerals, using laser ablation inductively coupled plasma mass spectrometry (ICP-MS). With the improved sensitivities of modern ICP-MS instrumentation, more stringent guidelines for cleaning and contaminant reduction become decisive in extending the limits of trace element analysis, particularly for dilute fluids and smaller inclusions investigated by highly sensitive sector field ICP-MS. We have identified three sources of contamination that limit the recording and quantification of low-count signals, namely: (1) the carrier gas quality that affects the continuous instrumental background. (2) The material and cleanliness of the ablation cell, connection tubing material, and sampler and skimmer cone, all of which may cause a falsely elevated signal during ablation. (3) Contamination of the sample surface during its preparation and from the deposition of previously ablated material, including aerosols produced during extended ablation of an ICP-MS tuning material. Contamination from these sources can overlap with the host mineral and with the fluid inclusion signal. Based on extensive test experiments, we propose a workflow that allows minimization of these contamination contributions and optimizes representative sampling and quantification of fluid inclusion data. For example, Au and S in low-density vapor-dominated inclusions may be quantified down to a few ng g−1 and a few hundred μg g−1, respectively.

Graphical abstract: LA-ICP-MS analysis of fluid inclusions: contamination effects challenging micro-analysis of elements close to their detection limit

Article information

Article type
Technical Note
Submitted
17 Jan 2017
Accepted
28 Mar 2017
First published
03 Apr 2017

J. Anal. At. Spectrom., 2017,32, 1052-1063

LA-ICP-MS analysis of fluid inclusions: contamination effects challenging micro-analysis of elements close to their detection limit

K. Schlöglova, M. Wälle and C. A. Heinrich, J. Anal. At. Spectrom., 2017, 32, 1052 DOI: 10.1039/C7JA00022G

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements