Angular resolution dependency in 2D LA-ICP-MS mapping – the case for low-dispersion laser ablation cells

Author affiliations

Abstract

In LA-ICP-MS imaging mostly lateral resolution is considered, but this work will show that linear artifacts with an angular aspect to the line scan direction may have a different spatial resolution, depending on the blur introduced by LA sampling and aerosol washout/transfer. This implies that upon scanning an irregularly shaped artifact, the spatial resolution in different locations may vary, the severity of which will be a function of the amount of “smearing” induced by the LA-ICP-MS system, invariably related to the design of the LA cell and aerosol washout/transfer efficiency. As reported previously (DOI: 10.1021/acs.analchem.6b02052), blur can be expressed as a convolution of the input image (representing the 2D elemental distribution) with point (PSF) and line (LSF) spread functions associated with ablation crater profiles and ICP-MS profiles of single pulses, respectively. Based on empirically derived PSFs and LSFs, and using a numerical calculation approach, the continuous-scanning LA-ICP-MS output of an angular-sensitive radial resolution target (Siemens star) was simulated to visualize the angular resolution dependency associated with ablation cell characteristics.

Graphical abstract: Angular resolution dependency in 2D LA-ICP-MS mapping – the case for low-dispersion laser ablation cells

Back to tab navigation

Article information


Submitted
30 Jun 2016
Accepted
20 Jul 2016
First published
20 Jul 2016

J. Anal. At. Spectrom., 2016,31, 1998-2004
Article type
Paper

Angular resolution dependency in 2D LA-ICP-MS mapping – the case for low-dispersion laser ablation cells

J. T. van Elteren and F. Vanhaecke, J. Anal. At. Spectrom., 2016, 31, 1998
DOI: 10.1039/C6JA00234J