Ultrasonic nebulization with an infrared heated pre-evaporation tube for sample introduction in ICP-OES: application to geological and environmental samples

Abstract

An improved sample introduction system was developed for inductively coupled plasma optical emission spectrometry (ICP-OES). The heater/condenser (HC) of a commercial ultrasonic nebulizer (USN) system was replaced by a pre-evaporation tube (PET) and a sheathing device. The aerosol exiting the USN was heated to 400 °C by infrared (IR) heating. Multivariate optimizations were conducted under robust conditions to find operating conditions improving the detection limit. Under optimum conditions and compared to conventional pneumatic nebulization, 10–25 fold improvement in sensitivity and detection limit was obtained for a range of elements using USN-PET(IR), including elements that are removed in the HC (such as Hg and B) of USN-HC. Compared to USN-PET where the PET is heated with heating tape, and to conventional USN-HC, USN-PET(IR) provided a two-fold improvement in sensitivity and detection limit. The improvement was more significant for ionic emission lines than atomic emission lines. Plasma robustness, measured by the Mg II/Mg I intensity ratio, also increased significantly, enabling the accurate multi-element analysis of geological and environmental samples using an external calibration, with an Ar emission line for internal standardization. Moreover, compared to an identical USN-PET set-up operated with the same aerosol carrier gas flow rate, but with heating tape, USN-PET(IR) significantly reduced the washout time during the analysis of geological materials, making it essentially independent of the element (with the possible exception of Hg) and analyte concentration (at least up to 400 mg L−1), which suggests that it now mostly arises from the dead volume of this sample introduction system.

Graphical abstract: Ultrasonic nebulization with an infrared heated pre-evaporation tube for sample introduction in ICP-OES: application to geological and environmental samples

Article information

Article type
Paper
Submitted
07 Feb 2012
Accepted
25 Apr 2012
First published
26 Apr 2012

J. Anal. At. Spectrom., 2012,27, 1254-1263

Ultrasonic nebulization with an infrared heated pre-evaporation tube for sample introduction in ICP-OES: application to geological and environmental samples

A. Asfaw, W. R. MacFarlane and D. Beauchemin, J. Anal. At. Spectrom., 2012, 27, 1254 DOI: 10.1039/C2JA30041A

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements