Measuring 0.01‰ to 0.1‰ isotopic variations by MC-ICPMS—testing limits for the first time with Pb δ-iCRMs

Abstract

A blind comparison on Pb-isotope δ-scale measurements by MC-ICPMS of 0.01‰ to 0.1‰ level was organised, involving five laboratories. Test samples were obtained from the series of candidate ERM-3810 δ-isotopic Certified Reference Materials (δ-iCRMs), and comprise four pairs of a material with ∼natural Pb-isotopic composition (‘delta zero’ or ‘δ-0’) and the same natural Pb progressively enriched in 207Pb (with δ207Pb values certified to ∼0.1% relative uncertainty, k = 2). Participants were free to apply the measurement strategy of their choice. A result was considered ‘acceptable’ only when, simultaneously, there was agreement within stated uncertainties with the corresponding reference value and the relative uncertainty stated by the participant was < 100%. This study illustrates the high degree of difficulty inherent to these δ-scale measurements by ‘routine’ MC-ICPMS methodologies (in this case, three participants reported 55% of their results which were deemed accurate, and the other two reported none). The closer to unity the isotope ratio value the better the results became (‘acceptable’ results mostly for δ7/6‰ and δ7/8‰ measurements). This first experiment of its kind demonstrates that Pb δ-scale isotopic measurements by MC-ICPMS can be reliably carried out down to 0.05‰ levels (two participants delivered accurate results above this threshold systematically for δ7/6‰, δ7/8‰ and δ7/4‰). Below this limit, at ∼0.01‰ and ∼ 0.03‰ levels, results are no longer consistent or reproducible and appear to be susceptible to a number of effects introducing error (such as short term changes in mass discrimination) which are either not well understood, or not controlled and/or not corrected for at a sufficiently low level of uncertainty. These results also suggest that ‘routine’ methods for absolute (calibrated) Pb-isotope ratio determination by MC-ICPMS produce relative combined uncertainties on results which are unlikely to be better than 0.05‰ (k = 2).

Graphical abstract: Measuring 0.01‰ to 0.1‰ isotopic variations by MC-ICPMS—testing limits for the first time with Pb δ-iCRMs

Article information

Article type
Paper
Submitted
28 Nov 2008
Accepted
06 Feb 2009
First published
03 Mar 2009

J. Anal. At. Spectrom., 2009,24, 407-412

Measuring 0.01‰ to 0.1‰ isotopic variations by MC-ICPMS—testing limits for the first time with Pb δ-iCRMs

C. R. Quétel, E. Ponzevera, I. Rodushkin, A. Gerdes, R. Williams and J. Woodhead, J. Anal. At. Spectrom., 2009, 24, 407 DOI: 10.1039/B821403B

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements