Issue 8, 2000

Abstract

A home-made, conventional glass pneumatic, concentric nebulizer has been used in radially viewed inductively coupled plasma atomic emission spectrometry for low sample consumption uptake rates, i.e., 50–200 µl min−1, by applying desolvation of the aerosol before it has entered the plasma. The desolvation system comprised a heated spray chamber and a condenser. The condenser temperature was 5 °C and the cooling water was fed by a thermostatted water bath. The operating conditions, such as heating aerosol temperature, nebulizer gas flow rate and viewing height were established to achieve optimum signal-to-background ratios (S/B). Over the range of measured liquid flow rates, the highest S/B values of the low-energy atomic lines were obtained at a heating aerosol temperature of 80 °C, while the best temperature was 120 °C for the high-energy atomic lines and most of the ionic lines. At a liquid uptake rate of 102 µl min−1 and a heating aerosol temperature of 120 °C, detection limits were reached which were similar to, or up to 10 times lower than, those which were obtained using the nebulizer in a common way, i.e., at a liquid flow rate of 1 ml min−1 and without pre-desolvation. Acid interferences of nitric and sulfuric acid, both in concentrations of 1.4 and 4.0 mol l−1, were studied. The interference seemed to be most serious for the low-energy atomic lines. To reduce the acid effect of the two nitric acid concentrations to within 10%, at liquid flow rates ranging from 50 to 100 µl min−1, heating aerosol temperatures of 160 and 200 °C were needed for the measured ionic and low-energy atomic lines, respectively. The same interference reduction could even be achieved for the ionic lines of elements in a sulfuric acid solution of 4.0 mol l−1, provided the liquid uptake rate is not higher than 50 µl min−1.

Article information

Article type
Paper
Submitted
25 Apr 2000
Accepted
13 Jun 2000
First published
05 Jul 2000

J. Anal. At. Spectrom., 2000,15, 959-966

Efficient use of a conventional pneumatic concentric nebulizer in ICP-AES at low liquid uptake rates by applying a desolvation system: determination of detection limits and degrees of acid interferences

J. W. Elgersma, D. Thanh Thuy and R. P. Groenestein, J. Anal. At. Spectrom., 2000, 15, 959 DOI: 10.1039/B003250F

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.