Comparison of 27- and 40-MHz plasmas in inductively coupled plasma mass spectrometry

Abstract

The analytical characteristics of a 27.12- and a 40.68-MHz inductively coupled plasma (ICP) are compared as ion sources for use in inductively coupled plasma mass spectrometry. The background spectra produced by the two sources are similar; however, there is evidence of greater entrainment of air in the higher frequency discharge. Calibration graphs are linear over 6–7 orders of magnitude with both operating frequencies, although the 40.68-MHz ICP suffers a factor of 2–5 loss in sensitivity. Accordingly, detection limits are also slightly worse at the higher operating frequency, as is the precision. Interference effects are independent of operating frequency, but the extent of doubly charged, oxide and hydroxide ion formation might be affected by a more significant orifice-linked discharge with the 40.68-MHz ICP.

Article information

Article type
Paper

J. Anal. At. Spectrom., 1989,4, 749-754

Comparison of 27- and 40-MHz plasmas in inductively coupled plasma mass spectrometry

G. H. Vickers, D. A. Wilson and G. M. Hieftje, J. Anal. At. Spectrom., 1989, 4, 749 DOI: 10.1039/JA9890400749

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements