Jump to main content
Jump to site search

Volume 177, 2015
Previous Article Next Article

Resolution enhancement through microscopic spatiotemporal control

Author affiliations

Abstract

Operating at biologically benign conditions, multi-photon fluorescence imaging microscopy has benefitted immensely from recent developments in microscopic resolution enhancement. Fluorescence microscopy continues to be the best choice for experiments on live specimens, however, multi-photon fluorescence imaging often suffers from overlapping fluorescence of typical dyes used in microscopy, limiting its scope. This limitation has been the focus of our research where we show that by making simple modifications to the laser pulse structure, it is possible to resolve these overlapping fluorescence complications. Specifically, by using pairs of femtosecond pulses with variable delay in place of single pulse excitation, we show controlled fluorescence excitation or suppression of one of the fluorophores over the other through wave-packet interferometry. Such an effect prevails even after the fluorophore coherence timescale, which effectively results in a higher spatial resolution. Here we extend the effect of our pulse-pair technique to microscopic axial resolution experiments and show that such pairs of pulses can also ‘enhance’ axial resolution.

Back to tab navigation

Article information


Submitted
15 Sep 2014
Accepted
24 Nov 2014
First published
27 Jan 2015

Faraday Discuss., 2015,177, 203-212
Article type
Paper
Author version available

Resolution enhancement through microscopic spatiotemporal control

D. Goswami, D. Das and S. Nath Bandyopadhyay, Faraday Discuss., 2015, 177, 203
DOI: 10.1039/C4FD00177J

Social activity

Search articles by author

Spotlight

Advertisements