Jump to main content
Jump to site search

Issue 22, 2013
Previous Article Next Article

Determination of inorganic arsenic in water by a quartz crystal microbalance

Author affiliations

Abstract

A quartz crystal microbalance sensor has been developed for the determination of inorganic arsenic species in water. The gold electrode surface was modified by a self-assembled layer of dithiothreitol, and the frequency change of the modified crystal was proportional to the arsenic concentration from 0 to around 50 μg L−1, a range which spans the current US EPA maximum contaminent level of 10 μg L−1 in drinking water. As dithiothreitol is capable of reducing arsenate to arsenite, the sensor detects both species. The method was applied to the determination of arsenic in spiked rain, tap, pond and bottled water; recoveries not significantly different from 100% were obtained for a number of spike additions of less than 10 μg L−1. Arsenic was only detected in the bottled water sample, at a concentration of 8 μg L−1. This method is simple, fast, and inexpensive compared with other conventional arsenic detection methods, and has the potential to be used in the field.

Graphical abstract: Determination of inorganic arsenic in water by a quartz crystal microbalance

Back to tab navigation

Article information


Submitted
24 May 2013
Accepted
03 Oct 2013
First published
04 Oct 2013

Anal. Methods, 2013,5, 6286-6291
Article type
Communication

Determination of inorganic arsenic in water by a quartz crystal microbalance

C. Li, A. E. Başaran and J. F. Tyson, Anal. Methods, 2013, 5, 6286
DOI: 10.1039/C3AY40876K