Advantages of CE-SDS over SDS-PAGE in mAb purity analysis

Abstract

Traditional SDS-PAGE and a newly developed CE-based method, i.e., CE-SDS, are both widely used to determine mAb purity. This work aimed to compare the two methods side-by-side in details and to establish the pros and cons of each. The sample preparation conditions were first optimized to acquire fair comparison results. When analyzing stressed mAb samples, both methods consistently showed comparable band/peak patterns. The results also suggest both methods are stability indicating. Beyond that, CE-SDS had improved advantages over SDS-PAGE in assay precision, linearity and separation resolution etc. Analysis of the same sample 21 times by utilizing SDS-PAGE introduced large standard deviation in LC and HC percentage determination. Furthermore, SDS-PAGE performed poorly in the linearity testing. The correlation coefficients from the regression were all less than 0.93 (SDS-PAGE) in contrast to those of greater than 0.97 (CE-SDS). When used to separate the non-glycosylated heavy chain (NGHC) from the heavy chain (HC), CE-SDS clearly showed much better resolution over SDS-PAGE. The CE-SDS method was further validated to assess its assay specificity, precision and robustness. For LC, HC and NGHC, the R.S.D. for repeatability, intraday and intermediate precision were determined to be <0.5% for migration time, <2.0% for corrected area percentage and <5.0% for corrected area, respectively. Finally, a small deviation in the migration time and corrected peak areas demonstrated the robustness of the CE-SDS assay when using different capillaries, analysts and sample storage conditions. All these results indicated that the newly developed CE-SDS method is superior to SDS-PAGE, and is suitable for use in areas of biochemical and biological research.

Graphical abstract: Advantages of CE-SDS over SDS-PAGE in mAb purity analysis

Article information

Article type
Paper
Submitted
29 Feb 2012
Accepted
15 Apr 2012
First published
17 Apr 2012

Anal. Methods, 2012,4, 1637-1642

Advantages of CE-SDS over SDS-PAGE in mAb purity analysis

Y. Shi, Z. Li and J. Lin, Anal. Methods, 2012, 4, 1637 DOI: 10.1039/C2AY25208B

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements