Determination of lead in drinking water by atomic-absorption spectrophotometry with electrothermal atomisation

Abstract

The use of electrothermal atomisation to determine concentrations of lead in water, although sensitive, generally suffers from suppressive interference effects that can produce large and variable negative bias. This paper describes a procedure that employs a lanthanum pre-treatment technique, which when tested with a range of 14 drinking water samples is shown to overcome such interference. The pre-treatment technique can be either impregnation of the furnace tube with lanthanum or the addition of lanthanum (as lanthanum chloride) to each sample. The procedure has a limit of detection of less than 1.0 µg l–1, a total standard deviation of less than 1.5 µg l–1, or 5% of the concentration, and a bias of less than 5 µg l–1, or 10% of the concentration over a working range of 0–100 µg l–1, for both manual and automated injection of samples.

Article information

Article type
Paper

Analyst, 1981,106, 23-31

Determination of lead in drinking water by atomic-absorption spectrophotometry with electrothermal atomisation

M. P. Bertenshaw, D. Gelsthorpe and K. C. Wheatstone, Analyst, 1981, 106, 23 DOI: 10.1039/AN9810600023

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements