Abstract

GC × GC investigations are well known to generate a substantial amount of information-rich and structurally complex data, requiring advanced data processing strategies like chemometrics. Many workflows are available for data handling and processing, such as the peak-table and pixel-based approaches. The goal of this work is to present a solution based on method development to solve the missing pixel problem that may be encountered in experiments performed with GC and GC × GC coupled to the Fourier transform orbital ion trap (FT-Orbitrap) mass analyzer. Data input is vital for pixel-based chemometric analyses, as some post-processing solutions may lead to significant loss of chemical information in the data set. Hence, a key requisite is that the chemical information is consistently indexed in the data arrays for proper pixel-based data handling and analysis. In this study, we carefully evaluated the ion management parameters to preserve the intrinsic structure and information of the data arrays of the GC × GC-FT-Orbitrap for future pixel-oriented chemometric analysis. The most acceptable conditions yielded acquisition rates up to 42.6 spectra per s, while a routine setting of 24.7 Hz was successfully employed in analyses of different petroleum fractions, producing both consistent tensor sizes and acceptable peak reconstructions. A data acquisition rate of 24.7 spectra per s and a mass resolving power of 15 000 allowed the resolution of a mass split of only 0.004 Da – which is an interesting configuration for challenging applications in petroleomics. Using such advanced settings, the missing pixel problem was reduced from up to 30% to much less than 0.04% of the data array dimension. Thus, the proposed configuration can be employed in studies that require pixel-oriented multivariate data analysis.

Graphical abstract: Advanced tuning of the ion management parameters in GC × GC-HRMS using a Fourier transform Orbitrap mass analyzer for pixel-based data handling and multivariate analysis

Supplementary files

Article information

Article type
Technical Note
Submitted
23 Feb 2022
Accepted
01 Apr 2022
First published
04 Apr 2022

Anal. Methods, 2022,14, 1646-1654

Advanced tuning of the ion management parameters in GC × GC-HRMS using a Fourier transform Orbitrap mass analyzer for pixel-based data handling and multivariate analysis

A. Moreira de Oliveira, C. A. Teixeira and L. W. Hantao, Anal. Methods, 2022, 14, 1646 DOI: 10.1039/D2AY00314G

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements