Calibration model transfer in mid-infrared process analysis with in situ attenuated total reflectance immersion probes

Abstract

Process applications of mid-infrared (MIR) spectrometry may involve replacement of the spectrometer and/or measurement probe, which generally requires a calibration transfer method to maintain the accuracy of analysis. In this study, direct standardisation (DS), piecewise direct standardisation (PDS) and spectral space transformation (SST) were compared for analysis of ternary mixtures of acetone, ethanol and ethyl acetate. Three calibration transfer examples were considered: changing the spectrometer, multiplexing two probes to a spectrometer, and changing the diameter of the attenuated total reflectance (ATR) probe (as might be required when scaling up from lab to process analysis). In each case, DS, PDS and SST improved the accuracy of prediction for the test samples, analysed on a secondary spectrometer–probe combination, using a calibration model developed on the primary system. When the probe diameter was changed, a scaling step was incorporated into SST to compensate for the change in absorbance caused by the difference in ATR crystal size. SST had some advantages over DS and PDS: DS was sensitive to the choice of standardisation samples, and PDS required optimisation of the window size parameter (which also required an extra standardisation sample). SST only required a single parameter to be chosen: the number of principal components, which can be set equal to the number of standardisation samples when a low number of standards (n < 7) are used, which is preferred to minimise the time required to transfer the calibration model.

Graphical abstract: Calibration model transfer in mid-infrared process analysis with in situ attenuated total reflectance immersion probes

Supplementary files

Article information

Article type
Paper
Submitted
21 Jan 2022
Accepted
19 Apr 2022
First published
28 Apr 2022
This article is Open Access
Creative Commons BY license

Anal. Methods, 2022,14, 1889-1896

Calibration model transfer in mid-infrared process analysis with in situ attenuated total reflectance immersion probes

A. J. Parrott, A. C. McIntyre, M. Holden, G. Colquhoun, Z. Chen, D. Littlejohn and A. Nordon, Anal. Methods, 2022, 14, 1889 DOI: 10.1039/D2AY00116K

This article is licensed under a Creative Commons Attribution 3.0 Unported Licence. You can use material from this article in other publications without requesting further permissions from the RSC, provided that the correct acknowledgement is given.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements