Precise sample metering method by coordinated burst action of hydrophobic burst valves applied to dried blood spot collection

Abstract

Dried blood spot (DBS) sampling by finger-pricking has recently gained a lot of interest as an alternative sample collection method. The reduced invasiveness, requirement of lower sample volumes and suitability for long-term storage at room temperature make DBS ideal for use in home settings or low-resource environments. However, traditional protocols often suffer from biased analysis data due to variable and not exactly known blood volumes present in the samples. In this work, a novel device has been developed to split-off precisely metered volumes from a blood drop and load them on pre-cut filter paper. Hereto, hydrophobic burst valves (HBV) were developed to temporarily retain a fluid flow, configurable to burst at pressures within a range of 175–600 Pa. By combining HBVs with different burst pressures, a volume metering system was developed to allow parallel metering of multiple pre-defined sample volumes. The system was shown to be accurate and consistent for blood volumes between 5–15 μL and for hematocrit levels spanning the range of 25–70%. Finally, a point-of-care DBS sampling device was developed combining the self-powered microfluidic SIMPLE technology. To evaluate the system's practical applicability, a validation study in the context of therapeutic drug monitoring of biologicals was performed using adalimumab-spiked blood samples. Microfluidic DBS samples showed good performance compared to the traditional DBS method with improved recovery rates (86% over 62%). This innovative metering system, allowing for parallelization and integration with complex liquid manipulations, will greatly impact the field of robust sampling, sample preparation, storage and analysis at the point-of-care.

Graphical abstract: Precise sample metering method by coordinated burst action of hydrophobic burst valves applied to dried blood spot collection

Supplementary files

Article information

Article type
Paper
Submitted
12 May 2021
Accepted
07 Oct 2021
First published
08 Oct 2021

Lab Chip, 2021,21, 4445-4454

Precise sample metering method by coordinated burst action of hydrophobic burst valves applied to dried blood spot collection

D. Vloemans, L. Van Hileghem, W. Verbist, D. Thomas, F. Dal Dosso and J. Lammertyn, Lab Chip, 2021, 21, 4445 DOI: 10.1039/D1LC00422K

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements