Analytical validation using a gas mixing system for the determination of gaseous formaldehyde

Abstract

Formaldehyde levels in the atmosphere are a concern in the indoor and outdoor air and many methods for determining this compound have been developed. The use of 2,4-dinitrophenylhydrazine (DNPH) for reaction with formaldehyde, catalyzed by acid, forming a hydrazone derivative in cartridges is considered the standard method for analyzing formaldehyde compounds in the air. However, formaldehyde is quantified using an analytical curve, created by diluting liquid standards of the formaldehyde-DNPH product. The analysis aims to quantify the gas phase formaldehyde, and it may be subject to experimental biases from the differences in the matrix of the sample (gas) and calibration standard (liquid). The objective of this work was to build an analytical curve in the gaseous phase using a synthetic air/formaldehyde mixing system (SFMS) and sampling with SPE-DNPH-tubes, comparing with the analytical curve in the liquid phase adopted by the Environmental Protection Agency (EPA). Parameters of linearity, sensitivity, limit of detection (LOD), limit of quantification (LOQ), precision and accuracy (recovery) were determined from the analytical curve in the gaseous phase. The best recovery in DNPH-tubes was obtained using the range of 400–1600 mL min−1 of flow rates in the gaseous phase. The sampling and reaction/elution of formaldehyde using DNPH-tubes presented adequate linearity and a similar sensitivity in the liquid analytical curve. Considering the LOD and LOQ in the gaseous phase, the values in nanograms are higher than those in the liquid phase. This study suggests that the quantification of formaldehyde in ambient air may be subject to bias due to differences in derivatization reaction efficiency. However, the results prove the efficiency of formaldehyde recovery from the atmosphere and the validity of the use of this DNPH-tube method.

Graphical abstract: Analytical validation using a gas mixing system for the determination of gaseous formaldehyde

Article information

Article type
Paper
Submitted
15 Jul 2020
Accepted
28 Sep 2020
First published
29 Sep 2020

Anal. Methods, 2020,12, 5247-5256

Analytical validation using a gas mixing system for the determination of gaseous formaldehyde

M. D. O. Souza, B. Sánchez, M. Fuentes, J. Gilaranz and M. C. Canela, Anal. Methods, 2020, 12, 5247 DOI: 10.1039/D0AY01363C

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements